NO-CODING BLOOD GLUCOSE MONITORING SYSTEM, MODELS TD-4251, TD-4252, TD-4253 AND TD-4254

K092099 · Taidoc Technology Corporation · NBW · Dec 10, 2009 · Clinical Chemistry

Device Facts

Record IDK092099
Device NameNO-CODING BLOOD GLUCOSE MONITORING SYSTEM, MODELS TD-4251, TD-4252, TD-4253 AND TD-4254
ApplicantTaidoc Technology Corporation
Product CodeNBW · Clinical Chemistry
Decision DateDec 10, 2009
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Intended Use

FORA TD-4251 / U-RIGHT TD-4252/ U-RIGHT TD-4253/U-RIGHT TD-4254 No-Coding Blood Glucose Monitoring System, model no. TD-4251A/ TD-4252C/ TD-4253A/ TD-4254A are intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and the following alternative sites: the palm, the forearm, the upper-arm, the calf and the thigh. It is intended for use by healthcare professionals and people with diabetes mellitus at home as an aid in monitoring the effectiveness of diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates. The alternative site testing in the FORA TD-4251 / U-RIGHT TD-4252/ U-RIGHT TD-4253/U-RIGHT TD-4254 No-Coding Blood Glucose Monitoring System can be used only during steady-state blood glucose conditions.

Device Story

System consists of glucose meter, test strips, control solution, lancing device, and lancets. Uses electrochemical biosensor technology with glucose oxidase to measure glucose in fresh capillary whole blood. Operated by healthcare professionals or patients at home. Provides quantitative glucose concentration output to assist in monitoring diabetes control programs. Modifications from predicate include software updates, physical appearance changes, and associated labeling revisions.

Clinical Evidence

Bench testing only. System accuracy performance compared to predicate device. Mechanical resistance testing performed per EN ISO 15197. Software verification and validation confirmed performance, safety, and effectiveness equivalent to predicate.

Technological Characteristics

Electrochemical biosensor using glucose oxidase. Components: glucose meter, test strips, control solution, lancing device. Software-controlled. Meets EN ISO 15197 standards for mechanical resistance.

Indications for Use

Indicated for use as a blood glucose monitoring system for the quantitative measurement of glucose in capillary whole blood.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # Section 10. 510(k) Summary ### 510(K) SUMMARY This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92. The Assigned 510(k) number is: _______________________________________________________________________________________________________________________________________________ - 1. Submitter's Identification: TaiDoc Technology Corporation 6F, No.127, Wugong 2nd Rd., Wugu Township, Taipei County, 248, Taiwan Correspondence: Debra Liang Regulatory Affairs Specialist Tel: +886-2-6625-8188 #1198 Fax: +886-2-6625-0288 Email: debra.liang@taidoc.com.tw Date of submission: - 2. Device name: Proprietary name: FORA TD-4251 / U-RIGHT TD-4252/ U-RIGHT TD-4253/U-RIGHT TD-4254 No-Coding Blood Glucose Monitoring System, model no. TD-4251A/ TD-4252C/ TD-4253A/ TD-4254A Regulatory information: - Regulation section: 21 CFR 862.1345 Glucose Test System A. - B. Classification: Class II CGA, Glucose Oxidase, Glucose C. Product Code: NBW, System, Test, Blood Glucose, Over The Counter Chemistry (75) D. Panel: - 3. Intended Use: FORA TD-4251 / U-RIGHT TD-4252/ U-RIGHT TD-4253/U-RIGHT TD-4254 10-1 DEC 1 0 2009 {1}------------------------------------------------ No-Coding Blood Glucose Monitoring System, model no. TD-4251A/ TD-4252C/ TD-4253A/ TD-4254A are intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and the following alternative sites: the palm, the forearm, the upper-arm, the calf and the thigh. It is intended for use by healthcare professionals and people with diabetes mellitus at home as an aid in monitoring the effectiveness of diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates. - 4. Device Description: FORA TD-4251 / U-RIGHT TD-4252/ U-RIGHT TD-4253/U-RIGHT TD-4254 No-Coding Blood Glucose Monitoring System consists of Glucose meter, Test Strips, Two levels of Control solution, A commercially available (510(k) cleared) lancing device and sterile Lancets, User Manual, and Storage Pack. - 5. Substantial Equivalance Information: Predicate device name: A. FORA TD-4251/ URIGHT TD-4252 / URIGHT TD-4254 Blood Glucose Monitoring System Predicate K number: K082482 B. Comparison with predicate: C. The modified FORA TD-4251 / U-RIGHT TD-4252/ U-RIGHT TD-4253/U-RIGHT TD-4254 No-Coding Blood Glucose Monitoring System, model no. TD-4251A/ TD-4252C/ TD-4253A/ TD-4254A have the following similarities to the predicate device: - same operating principle, ■ - same fundamental scientific technology, ■ - incorporate the same basic circuit design, ■ - 미 incorporate the same materials, - same shelf life - packaged using the same materials, and ■ - Manufactured by the same process. {2}------------------------------------------------ The modifications encompass: - A modification in the software of the glucose meter I - Modification in the physical appearance ■ - Labeling change due to the software modification . ### 6. Test Principle: The detection and measurement of glucose in blood is by an electrochemical biosensor technology using glucose oxidase. - 7. Performance Characteristics: The FORA TD-4251 / U-RIGHT TD-4252/ U-RIGHT TD-4253/U-RIGHT TD-4254 No-Coding Blood Glucose Monitoring Systems have the same performance characteristics as the predicate device. A comparison of system accuracy performance demonstrated that the FORA TD-4251 //U-RIGHT TD-4252/ U-RIGHT TD-4253/U-RIGHT TD-4254 No-Coding Blood Glucose Monitoring System, model no. TD-4251A/ TD-4252C/ TD-4253A/ TD-4254A and the currently marketed FORA TD-4251/ URIGHT TD-4252 / URIGHT TD-4254 Blood Glucose Monitoring System are substantially equivalent. A mechanical resistance test showed that the model TD-4253 meet the requirement of EN ISO 15197. Software verification and validation testing confirmed that the performance, safety and effectiveness of the FORA TD-4251 / U-RIGHT TD-4252/ U-RIGHT TD-4253/U-RIGHT TD-4254 No-Coding Blood Glucose Monitoring System, model no. TD-4251A/ TD-4252C/ TD-4253A/ TD-4254A are equivalent to the predicate device. - 8. Conclusion: Based on the information provided in this submission, the FORA TD-4251 / U-RIGHT TD-4252/ U-RIGHT TD-4253/U-RIGHT TD-4254 No-Coding Blood Glucose Monitoring System, model no. TD-4251A/ TD-4252C/ TD-4253A/ TD-4254A are substantially equivalent to the predicate FORA TD-4251/ URIGHT TD-4252 / URIGHT TD-4254 Blood Glucose Monitoring System. 10-3 {3}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-0609 Silver Spring, MD 20993-0002 TaiDoc Technology Corporation c/o Debra Liang Regulatory Affairs Specialist 6F, No.127, Wugong 2nd Rd., Wugu Township Taipei County, Taiwan DEC 1 0 2009 #### k092099 Re: Trade/Device Name: FORA TD-4251/U-RIGHT TD-4252/ U-RIGHT TD-4253/ U-RIGHT TD-4254 No-Coding Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Code: NBW, CGA Dated: November 11, 2009 Received: November 12, 2009 Dear Ms. Liang We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. {4}------------------------------------------------ ### Page 2 If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or ( 301 ) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. CA Courtney C. Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ### Section 7 Indications for Use ## Indications for Use ### K092099 510(k) Number: Device Name: FORA TD-4251 / U-RIGHT TD-4252/ U-RIGHT TD-4253/U-RIGHT TD-4254 No-Coding Blood Glucose Monitoring System, model no. TD-4251A/ TD-4252C/ TD-4253A/ TD-4254A Indications for Use: FORA TD-4251 / U-RIGHT TD-4252/ U-RIGHT TD-4253/U-RIGHT TD-4254 No-Coding Blood Glucose Monitoring System, model no. TD-4251A/ TD-4252C/ TD-4253A/ TD-4254A are intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and the following alternative sites: the palm, the forearm, the upper-arm, the calf and the thigh. It is intended for use by healthcare professionals and people with diabetes mellitus at home as an aid in monitoring the effectiveness of diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates. The alternative site testing in the FORA TD-4251 / U-RIGHT TD-4252/ U-RIGHT TD-4253/U-RIGHT TD-4254 No-Coding Blood Glucose Monitoring System can be used only during steady-state blood glucose conditions. Prescription Use (21 CFR Part 801 Subpart D) And/Or Over the Counter Use X (21 CFR Part 801 Subpart C) - (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD) Division Sign-off Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 8092090 510(k) Page I of 7-1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...