Safecare Urinary Tract Infection Test

K242767 · Safecare Biotech (Hangzhou) Co., Ltd. · JMT · Jan 10, 2025 · Clinical Chemistry

Device Facts

Record IDK242767
Device NameSafecare Urinary Tract Infection Test
ApplicantSafecare Biotech (Hangzhou) Co., Ltd.
Product CodeJMT · Clinical Chemistry
Decision DateJan 10, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1510
Device ClassClass 1

Indications for Use

The Safecare Urinary Tract Infection Test is for the qualitative detection of Leukocytes (LEU,white blood cells) and nitrite (NIT) in urine as an aid in the screening of a Urinary Tract infection (UTI). It is intended for over-the-counter home use.

Device Story

The Safecare Urinary Tract Infection Test is an in vitro diagnostic device for home screening of UTIs. It consists of a test strip with two color pads: one for leukocyte esterase and one for nitrite. The user collects a midstream urine sample, exposes the strip to the urine, and visually compares the resulting color change on the pads against a provided color chart after 2 minutes. The leukocyte pad detects leukocyte esterase via hydrolysis of an indoxyl ester derivative, which reacts with a diazonium salt to produce a pink-to-purple color. The nitrite pad detects nitrite via conversion of nitrate to a diazonium compound in an acid medium, which couples with 1,2,3,4-tetrahydrobenzo(h)quinoline to produce a pink color. The device provides qualitative results to aid in UTI screening. Healthcare providers use these results in conjunction with clinical symptoms to guide further diagnostic or treatment decisions. The device benefits patients by providing rapid, accessible, and convenient preliminary screening for UTI symptoms in a home setting.

Clinical Evidence

Bench testing only. Analytical performance included precision/reproducibility (n=45 per level), interference studies, and limit of detection (LOD) verification. Method comparison study conducted with 154 lay users comparing self-test results to professional predicate testing; demonstrated high agreement (Leukocyte exact match 88.89-100%; Nitrite 100%). No clinical diagnostic sensitivity/specificity studies performed.

Technological Characteristics

The device is a visual-read reagent strip. Leukocyte detection uses indoxyl ester hydrolysis; nitrite detection uses the Griess reaction (diazonium coupling). Traceable to NIST SRM 8040 (nitrite). Stability is 24 months at 4-30 °C; open-pouch stability is 1 hour. The device is a standalone, non-electronic, disposable diagnostic test.

Indications for Use

Indicated for individuals with symptoms of a urinary tract infection (UTI) for the qualitative screening of leukocytes and nitrite in urine. Intended for over-the-counter home use.

Regulatory Classification

Identification

A nitrite (nonquantitative) test system is a device intended to identify nitrite in urine. Nitrite identification is used in the diagnosis and treatment of uninary tract infection of bacterial origin.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY ## I Background Information: A 510(k) Number K242767 B Applicant Safecare Biotech (Hangzhou) Co., Ltd. C Proprietary and Established Names Safecare Urinary Tract Infection Test D Regulatory Information | Product Code(s) | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | JMT | Class I, meets the limitations of exemptions in 21 CFR 862.9 (c)(9) | 21 CFR 862.1510 - Nitrite (Nonquantitative) Test System | CH - Clinical Chemistry | | LJX | Class I, meets the limitations of exemptions in 21 CFR 864.9 (c)(9) | 21 CFR 864.7675 - Leukocyte Peroxidase Test | HE - Hematology | ## II Submission/Device Overview: A Purpose for Submission: New device B Measurand: Urinary nitrite and leukocytes Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 www.fda.gov {1} C Type of Test: Qualitative and semi-quantitative urinalysis III Intended Use/Indications for Use: A Intended Use(s): See Indications for Use below. B Indication(s) for Use: The Safecare Urinary Tract Infection Test is for the qualitative detection of Leukocytes (LEU, white blood cells) and nitrite (NIT) in urine as an aid in the screening of a Urinary Tract infection (UTI). It is intended for over-the-counter home use. C Special Conditions for Use Statement(s): OTC - Over The Counter D Special Instrument Requirements: Not Applicable. IV Device/System Characteristics: A Device Description: The Safecare Urinary Tract Infection Test is an in vitro diagnostic test device for qualitative detection of leukocytes and nitrite in urine. The device is composed of two-color pads aligned on a strip. One pad is employed for testing leukocyte and the other for nitrite by visually reading the color change of the pad and comparing to the corresponding blocks on a color chart. B Principle of Operation: The Safecare Urinary Tract Infection Test measures the color developed in two reaction zones (leukocytes and nitrite pads) on the test strips following application of a urine sample. The developed colors are then compared to calibration colors on the color chart. This leukocyte test is based on the action of leukocyte esterase, which cleaves a derivatized pyrrolidine amino acid ester substrate to liberate derivatized hydroxy pyrazole. This compound then reacts with a diazonium salt to produce a colored compound (pink to purple). This nitrite test uses the conversion of nitrate to nitrite by the action of bacteria in the urine. Urinary nitrite reacts with p-arsanilic acid to form a diazonium compound in an acidic medium. The compound then couples with 1,2,3,4-tetrahydrobenzo[h]quinolin-3-ol to produce a pink color. K242767 - Page 2 of 9 {2} V Substantial Equivalence Information: A Predicate Device Name(s): Healgen URS Test Strips B Predicate 510(k) Number(s): K231045 C Comparison with Predicate(s): | Device & Predicate Device(s): | K242767 | K231045 | | --- | --- | --- | | Device Trade Name | Safecare Urinary Tract Infection Test | Healgen URS Test Strips | | General Device Characteristic Similarities | | | | Intended Use/Indications For Use | For the qualitative detection of leukocytes and nitrite in urine. | Same | | Nitrite test methodology | By conversion of nitrate to nitrite using the action of p-arsanilic acid to form a diazonium compound in an acid medium. This compound then couples with 1, 2, 3, 4 - tetrahydrobenzo(h) quinoline to produce a pink color. | Same | | Leukocyte test methodology | By hydrolysis of an indoxyl ester derivative through the action of leukocyte esterase. The liberated indoxyl ester reacts with a diazonium salt to produce a colored compound (pink to purple). | Same | | Specimen type | Human urine | Same | | Conditions for use | Over-the-counter | Same | | General Device Characteristic Differences | | | | Storage condition | 4 to 30 °C | 2 to 30 °C | | Limit of Detection (LOD) | Leukocyte esterase: 0.0018 mg/mL | Leukocyte: 15 cells/μL | K242767 - Page 3 of 9 {3} K242767 - Page 4 of 9 | Device & Predicate Device(s): | K242767 | K231045 | | --- | --- | --- | | | Nitrite: 0.05 mg/dL LOD defined as the concentration of analyte that produces positive results at least 95% of the time | Nitrite: 0.05 mg/dL LOD defined as the concentration of analyte that produces positive results approximately 50% of the time | ## VI Standards/Guidance Documents Referenced: None were referenced. ## VII Performance Characteristics (if/when applicable): ### 1. Analytical Performance: #### a. Precision/Reproducibility: A precision study was performed with the Safecare Urinary Tract Infection Test at three (3) test sites with two (2) operators at each site. Samples were fresh urine samples collected and pooled on the same day that the study was conducted, then spiked with leukocyte esterase and nitrite stock solutions to obtain the concentration levels (1-8) listed in the table below. The evaluation included three (3) replicate assays over five (5) days. A total of forty-five (45) assay results on each of eight levels of test sample were obtained. Three lots of the device were used with each level of control. All sample concentrations were masked. Results are listed in the following table: | Level | Analyte | Concentration | Expected Value | N | % Agreement with Expected results | | --- | --- | --- | --- | --- | --- | | Level 1 | Leukocyte | 0 mg/mL | Negative | 45 | 100 | | | Nitrite | 0 mg/dL | Negative | 45 | 100 | | Level 2 | Leukocyte | 0.0018 mg/mL | Trace | 45 | 100 | | | Nitrite | 0 mg/dL | Negative | 45 | 100 | | Level 3 | Leukocyte | 0.0084 mg/mL | Small (+) | 45 | 100 | | | Nitrite | 0 mg/dL | Negative | 45 | 100 | | Level 4 | Leukocyte | 0.015 mg/mL | Moderate (++) | 45 | 100 | | | Nitrite | 0 mg/dL | Negative | 45 | 100 | | Level 5 | Leukocyte | 0.06 mg/mL | Large (+++) | 45 | 100 | | | Nitrite | 0 mg/dL | Negative | 45 | 100 | | Level 6 | Leukocyte | 0 mg/mL | Negative | 45 | 100 | | | Nitrite | 0.1 mg/dL | Positive | 45 | 100 | | Level 7 | Leukocyte | 0 mg/mL | Negative | 45 | 100 | {4} K242767 - Page 5 of 9 | Level | Analyte | Concentration | Expected Value | N | % Agreement with Expected results | | --- | --- | --- | --- | --- | --- | | | Nitrite | 0.3 mg/dL | Positive | 45 | 100 | | Level 8 | Leukocyte | 0.0084 mg/mL | Small (+) | 45 | 100 | | | Nitrite | 0.1 mg/dL | Positive | 45 | 100 | b. Linearity: Not Applicable. c. Analytical Specificity/Interference: Potentially interfering substances were added into negative urine or urine with different leukocyte esterase and nitrite concentrations. These samples were tested with three lots of the Safecare Urinary Tract Infection Test by three different operators (one operator per lot). The following substances showed no interference with the test at the specified concentrations: | Substances | Concentration (mg/dL) | | --- | --- | | Albumin | 1000 | | Ammonium Chloride | 400 | | Ascorbic Acid | 100 | | Bilirubin | 10 | | Ciprofloxacin | 1 | | Creatinine | 600 | | Fructose | 18 | | Galactose | 15 | | Glucose | 500 | | Glycine | 900 | | Hemoglobin | 100 | | Lactose | 29 | | Oxalic Acid | 1 | | Phenazopyridine | 30 | | Phenolphthalein | 4 | | Potassium Chloride | 1200 | | Riboflavin | 50 | | Sodium Nitrate | 10 | | Sodium Nitrite* | 10 | | Sodium Phosphate | 1000 | | Sulfamethoxazole | 40 | | Theophylline | 4 | | Urea | 4000 | * This interferent is tested for Leukocyte results only. High levels of glucose (≥1000 mg/dL) or high ascorbic acid (≥150 mg/dL) may decrease leukocyte readings. High ascorbic acid (≥150 mg/dL) may cause a false negative nitrite reading. {5} To investigate the effect of urine specific gravity and urine pH, urine samples with specific gravity ranging from 1.000 to 1.035 and urine samples with pH ranging from pH of 5 to 9 were tested at different leukocyte esterase and nitrite concentrations. The test results show that pH > 8.0 may cause false positive leukocyte readings, and specific gravity ≥ 1.035 may cause false negative leukocyte readings. Neither pH nor specific gravity affects nitrite testing. ## Sample Carryover A sample carryover study was performed. Three (3) lots of test strips were used for testing. The study was performed by either dipping the test strip into the sample or simulating urination with the sample and then holding the test strip vertically upwards or downwards for 10 seconds to allow the sample to flow from one pad to the other pad (i.e., run-over of high-concentration leukocyte esterase samples from the leukocyte pad to the nitrite pad and run-over of high-concentration nitrite samples from the nitrite pad to the leukocyte pad). Testing was also repeated in the other direction. The study demonstrated that carryover (run-over) does not impact the test results. ## d. Assay Reportable Range: The results of the analytical studies support the following measurement ranges: | Analyte | Measurement range | | --- | --- | | Leukocyte | qualitative: Negative, Trace, +, ++, +++ | | Esterase | semi-quantitative: Negative, 0.0018, 0.0084, 0.015, 0.06 (mg/mL) | | Nitrite | qualitative: Negative, Positive | ## e. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods): ### Traceability The nitrite test is traceable to NIST SRM 8040, and the leukocyte pad is traceable to a commercially available solution. ### Stability The sponsor provided information to support that the device is stable for 15 months when stored at 4-30°C based on real-time stability studies. The sponsor instructs the user to use the test strip immediately after the foil pouch is opened, but no longer than 1 hour after opening, based on the results of an open-pouch stability study. ## f. Detection Limit: See Assay Cut-off section below. ## g. Assay Cut-Off: A cut-off study was performed to evaluate the lower limits of detection for each analyte on the Safecare Urinary Tract Infection Test. Urine samples were spiked to known K242767 - Page 6 of 9 {6} concentrations of each analyte. Each sample was tested in three (3) device lots by three (3) different operators in a total of 30 replicates per sample. The sponsor defined the lowest concentration with ≥ 95% test of results positive as the limit of detection. The results support the claim that the sensitivity of the leukocyte test is 0.0018 mg/mL leukocyte esterase, and the sensitivity of the nitrite test is 0.05 mg/dL. Leukocyte: | Leukocyte Esterase Concentration (mg/mL) | Results | | Positive (%) | | --- | --- | --- | --- | | | Negative | Positive (±) | | | 0.00036 | 30 | 0 | 0 | | 0.0006 | 27 | 3 | 10 | | 0.0012 | 15 | 15 | 50 | | 0.0018 | 0 | 30 | 100 | | 0.0036 | 0 | 30 | 100 | Nitrite: | Concentration (mg/dL) | Results | | Positive (%) | | --- | --- | --- | --- | | | Negative | Positive | | | 0 | 30 | 0 | 0 | | 0.04 | 7 | 23 | 76.7 | | 0.05 | 0 | 30 | 100 | | 0.06 | 0 | 30 | 100 | | 0.08 | 0 | 30 | 100 | | 0.1 | 0 | 30 | 100 | The sponsor provided studies to support the recommended wetting time (1-2 seconds), the recommended reading time (2 minutes), and provided a lighting study supporting that the Safecare Urinary Tract Infection test can be correctly read under fluorescent, incandescent, and natural lighting conditions. ## 2. Comparison Studies: a. Method Comparison with Predicate Device: Three (3) sites were selected to perform the lay-user method comparison studies. One hundred and fifty-four (154) lay users with UTI symptoms were recruited to test their own urine sample using the Safecare Urinary Tract Infection Test. Laypersons performed one test with the Safecare Urinary Tract Infection Test according to the product insert and then collected a sample of their urine for comparison testing by professionals using the predicate device. The results obtained by the lay users using the Safecare Urinary Tract Infection Test compared to the results obtained by the professionals using the predicate device when used following its product insert are summarized below. Results from 154 subjects testing their own urine samples using the midstream collection method for leukocytes: K242767 - Page 7 of 9 {7} | LEU | Professional test results | | | | | Total | | | --- | --- | --- | --- | --- | --- | --- | --- | | | | +++ | ++ | + | Trace | | Negative | | Lay person test results (candidate device) | +++ | 9 | 1 | 0 | 0 | 0 | 10 | | | ++ | 1 | 20 | 1 | 0 | 0 | 22 | | | + | 0 | 1 | 31 | 2 | 0 | 34 | | | Trace | 0 | 0 | 2 | 16 | 0 | 18 | | | Negative | 0 | 0 | 0 | 0 | 70 | 70 | | Total | | 10 | 22 | 34 | 18 | 70 | 154 | | % Agreement (Exact Match) | | 90.00 | 90.91 | 91.18 | 88.89 | 100.00 | | | % Agreement (+/- Color Block) | | 100 | 100 | 100 | 100 | 100 | | Results from 154 subjects testing their own urine samples using the midstream collection method for nitrite: | NIT | Professional test results | | Total | | | --- | --- | --- | --- | --- | | | | Positive | | Negative | | Lay person test results | Positive | 60 | 0 | 60 | | | Negative | 0 | 94 | 94 | | Total | | 60 | 94 | 154 | | % Agreement (Exact Match) | | 100 | 100 | | Lay users were also given surveys on the ease of understanding the package insert instructions. All lay users indicated that the device instructions were clear, simple, very clear, or very simple. The sponsor conducted a Flesch-Kincaid reading analysis on the package insert and the score revealed a reading grade level of less than 7. b. Matrix Comparison: Not Applicable. This device is intended to be used for urine samples using the midstream collection method. ## 3. Clinical Studies: a. Clinical Sensitivity: Not Applicable. b. Clinical Specificity: Not Applicable. c. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable): Not Applicable. K242767 - Page 8 of 9 {8} K242767 - Page 9 of 9 4. Clinical Cut-Off: Not Applicable. 5. Expected Values/Reference Range: The following information is provided in the labeling: | TEST STRIP RESULTS | | RECOMMENDATIONS | | --- | --- | --- | | Leukocytes (LEU) | Nitrite (NIT) | | | Negative | Negative | Results suggest no sign of a UTI. If you have symptoms, consult your physician; there are some cases of UTI that the test may not detect. | | Trace | Negative | Results suggest additional testing is required. Test again the following day with your first morning urine using a new strip. If you still have symptoms or if you get another trace LEU, consult your physician. | | Positive +, ++, +++ | Negative | Results suggest a sign of a UTI. Repeat the test next time you urinate using a new test strip. Make sure to wash the genital area first. If LEU is still positive, consult your physician. | | Negative or Trace | Positive | Results suggest a sign of a UTI. Consult your physician immediately. | | Positive +, ++, +++ | Positive | Results suggest a sign of a UTI. Consult your physician immediately. | VIII Proposed Labeling: The labeling supports the finding of substantial equivalence for this device. IX Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...