Healgen® URS Test Strips

K251800 · Healgen Scientific,, LLC · JMT · Dec 23, 2025 · Clinical Chemistry

Device Facts

Record IDK251800
Device NameHealgen® URS Test Strips
ApplicantHealgen Scientific,, LLC
Product CodeJMT · Clinical Chemistry
Decision DateDec 23, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1510
Device ClassClass 1

Intended Use

The Healgen® URS Test Strips are for the qualitative detection of leukocytes (white blood cells) and nitrite in urine as an aid in the screening of a urinary tract infection (UTI). It is intended for over-the-counter home use only.

Device Story

Healgen® URS Test Strips are in vitro diagnostic devices for qualitative detection of leukocytes and nitrites in human urine. Device consists of two color-reactive pads on a strip. User dips strip in urine; color change occurs after 2 minutes. User visually compares pad colors against a provided color chart. Leukocyte detection uses hydrolysis of indoxyl ester derivative via leukocyte esterase, reacting with diazonium salt to produce pink/purple color. Nitrite detection uses conversion of nitrate to nitrite via p-arsanilic acid, forming a diazonium compound that couples with 1,2,3,4-tetrahydrobenzo(h)quinoline to produce a pink color. Intended for over-the-counter home use by patients to screen for UTI. Results provide immediate visual feedback to assist in clinical decision-making regarding potential infection.

Clinical Evidence

Bench testing included sensitivity (LOD: 10 cells/μL leukocytes, 0.05 mg/dL nitrite), precision (100% agreement across 45 replicates), and analytical specificity (no interference from 35 substances, though high glucose/ascorbic acid may affect results). Method comparison study (n=234) against K111999 showed high agreement (exact match >95% for most levels). Lay-user study (n=234) confirmed that non-professionals could correctly perform the test and interpret results with high agreement compared to healthcare professionals. Flesch-Kincaid analysis confirmed package insert readability (grade level <8).

Technological Characteristics

The device consists of reagent-impregnated pads on a strip. Nitrite detection uses p-arsanilic acid and 1,2,3,4-tetrahydrobenzo(h)quinoline. Leukocyte detection uses an indoxyl ester derivative and a diazonium salt. The device is a qualitative/semi-quantitative visual test with a 2-minute read time. It is stable for 24 months at 2-30°C and has an open-pouch stability of 2 hours. No electronic components, software, or connectivity features are present.

Indications for Use

Indicated for individuals with symptoms of a urinary tract infection (UTI) for the qualitative detection of leukocytes and nitrite in urine to aid in screening.

Regulatory Classification

Identification

A nitrite (nonquantitative) test system is a device intended to identify nitrite in urine. Nitrite identification is used in the diagnosis and treatment of uninary tract infection of bacterial origin.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION December 23, 2025 Healgen Scientific LLC % Jenny Xia Director LSI International Inc 504 East Diamond Ave., Suite H Gaithersburg, MD 20877 Re: K251800 Trade/Device Name: Healgen® URS Test Strips Regulation Number: 21 CFR 864.1510 Regulation Name: Nitrite (Non-Quantitative) Test System Regulatory Class: Class I (meets the limitations of exemptions in 862.9(c)(9)) Product Code: JMT, LJX Dated: December 4, 2025 Received: December 4, 2025 Dear Jenny Xia: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food &amp; Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K251800 - Jenny Xia Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K251800 - Jenny Xia Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, PAULA V. CAPOSINO -S Paula Caposino, Ph.D. Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K251800 | | | Device Name Healgen® URS Test Strips | | | Indications for Use (Describe) The Healgen® URS Test Strips are for the qualitative detection of leukocytes (white blood cells) and nitrite in urine as an aid in the screening of a urinary tract infection (UTI). It is intended for over-the-counter home use only. | | | Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} 510(k) SUMMARY K251800 1. Date: December 7, 2025 2. Submitter: Healgen Scientific LLC 5213 Maple Street Bellaire, TX77401 3. Contact person: Jenny Xia LSI International Inc. 504 East Diamond Ave., Suite H Gaithersburg, MD 20878 Telephone: 301-525-6856 Fax: 301-916-6213 Email: jxia@lsi-consulting.org 4. Device Name: Healgen® URS Test Strips Classification: Class I (meets the limitations of exemptions in 862.9(c)(9)) | Product Code | CFR # | Panel | | --- | --- | --- | | LJX | 21 CFR § 864.7675, Leukocyte Peroxidase Test System | Hematology | | JMT | 21 CFR § 862.1510, Nitrite Test System | Clinical Chemistry | 5. Predicate Devices: K231045 Healgen URS Test Strips 6. Intended Use: The Healgen® URS Test Strips are for the qualitative detection of leukocytes (white blood cells) and nitrite in urine as an aid in the screening of a urinary tract infection (UTI). It is intended for over-the-counter home use only. 7. Device Description: The Healgen® URS Test Strips are in vitro diagnostic test devices for the qualitative detection of leukocytes and nitrites in urine. The device is composed of two color pads aligned on a strip. One pad is employed for testing leukocytes and the other for nitrites by visually reading the color change of the applicable pad and comparing with the corresponding blocks on a color chart. {5} 2 # 8. Substantial Equivalence Information | Item | Device | Predicate – K231045 | | --- | --- | --- | | Indication(s) for use | The Healgen® URS Test Strips are for the qualitative detection of leukocytes (white blood cells) and nitrite in urine as an aid in the screening of a urinary tract infection (UTI). It is intended for over-the-counter home use only. | The Healgen URS Test Strips are for the qualitative detection of Leukocytes (white blood cells) and nitrite in urine as an aid in the screening of a Urinary Tract infection (UTI). It is intended for over-the-counter home use only. | | Nitrite test methodology | By conversion of nitrate to nitrite using the action of p-arsanilic acid to form a diazonium compound in an acid medium. This compound then couples with 1, 2, 3, 4-tetrahydrobenzo(h) quinoline to produce a pink color. | Same | | Leukocyte test methodology | By hydrolysis of an indoxyl ester derivative through the action of leukocyte esterase. The liberated indoxyl ester reacts with a diazonium salt to produce a colored compound (pink to purple). | Same | | Specimen Type | Human urine | Same | | Testing parameters | Leukocytes and Nitrite | Same | | Intended Users | For over-the-counter use | Same | | Test Read Time | 2 minutes for both leukocytes and nitrite | Same | | Detection Method | Visual | Same | | Test Format | Dip or Midstream | Midstream | # 9. Test Principle The Healgen® URS Test Strips measure the color developed in 2 reaction zones (leukocytes and nitrite pads) on the test strips following application of a urine sample. The developed colors are then compared {6} to calibration colors located on the Healgen color chart card and the result for each pad is determined based on the minimum color distance between the developed colors and calibration colors. The leukocytes test is using the hydrolysis of an indoxyl ester derivative through the action of leukocyte esterase. The liberated indoxyl ester reacts with a diazonium salt to produce a colored compound (pink to purple). The Nitrite test uses the conversion of nitrate to nitrite by the action of p-arsanilic acid to form a diazonium compound in an acid medium. This compound then couples with 1, 2, 3, 4-tetrahydrobenzo $(h)$ quinoline to produce a pink color. # 10. Performance Characteristics # 1. Assay Cut-off A sensitivity study was performed to evaluate the lower limits of detection for each analyte on the Healgen® URS Test Strips. Urine samples were spiked to known concentrations of each analyte. These samples were then diluted to the lowest positive concentrations that are indicated on the color chart. Each sample was tested in duplicate with three device lots by five different operators. | Leukocyte Concentration (cells/μL) | Negative | Positive | Limit of Detection | | --- | --- | --- | --- | | 15 | 0 | 30 | 100% | | 12 | 6 | 24 | 80% | | 10 | 14 | 16 | 53% | | 5 | 26 | 4 | 13% | | 3 | 30 | 0 | 0 | | Nitrite Concentration (mg/dL) | Negative | Positive | Limit of Detection | | --- | --- | --- | --- | | 0.08 | 0 | 30 | 100% | | 0.07 | 2 | 28 | 93% | | 0.06 | 7 | 23 | 77% | | 0.05 | 15 | 15 | 50% | | 0.04 | 30 | 0 | 0 | {7} LOD values of 10 cells/μL and 0.05 mg/dL are verified for Leukocyte and nitrite respectively. ## 2. Precision/Reproducibility The precision study on the Healgen® URS Test Strips was performed at three testing sites with one operator at each site. The evaluation included three replicate assays over five days. A total of forty-five assays results on each concentration were obtained. All sample concentrations were masked. Three lots of the device were used with each level of control. The obtained results are listed in the following tables. Leukocyte | Concentration (cells/μL) | Expected Range | N | % Agreement with Expected Results | | --- | --- | --- | --- | | 0 | Neg. | 45 | 100 | | 15 | 15 | 45 | 100 | | 70 | 70 | 45 | 100 | | 125 | 125 | 45 | 100 | | 500 | 500 | 45 | 100 | Nitrite | Concentration (mg/dL) | Expected Range | N | % Agreement with Expected Results | | --- | --- | --- | --- | | 0 | Neg. | 45 | 100 | | 0.08 | Positive | 45 | 100 | | 0.1 | Positive | 45 | 100 | ## 3. Analytical specificity Potentially interfering substances were added to negative urine or urine with different analytes concentrations. These samples were tested with three lots of the Healgen® URS Test Strips by three different operators. The following substances showed no interference with the tests at the specified concentrations. High glucose levels (≥1000 mg/dL) and high ascorbic acid (≥150 mg/dL) may decrease leukocyte readings. High ascorbic acid (≥150 mg/dL) may cause a false negative nitrite reading. | Substances | Testing Concentration | | --- | --- | | Albumin | 1000 mg/dL | | Ascorbic Acid | 100 mg/dL | | Hemoglobin | 100 mg/dL | {8} | Citric Acid | 50 mg/dL | | --- | --- | | Bilirubin | 10 mg/dL | | Ciprofloxacin | 1 mg/dL | | Creatine | 8 mg/dL | | Acetoacetate Acid | 1 mmol/L | | Ammonium Chloride | 400 mg/dL | | Calcium Chloride | 50 mg/dL | | Creatinine | 800 mg/dL | | Galactose | 80 mg/dL | | Glucose | 500 mg/dL | | Glycine | 1000 mg/dL | | KCL | 1500 mg/dL | | NaCl | 2800 mg/dL | | Oxalic Acid | 70 mg/dL | | Sodium Acetate | 1200 mg/dL | | Sodium Bicarbonate | 1500 mg/dL | | Sodium Nitrate | 10 mg/dL | | Sodium Nitrite | 10 mg/dL | | Sodium Phosphate | 1000 mg/dL | | Sulfamethoxazole | 40 mg/dL | | Urobilinogen | 3.0 mg/dL | | Urea | 4000 mg/dL | | Riboflavin | 50 mg/dL | | Theophylline | 100 mg/dL | | Phenolphthalein | 120 mg/dL | | Phenazopyridine | 30 mg/dL | | Glutathione | 200mg/dL | | Hypochlorite | 1mg/dL | | Hydrochloric Acid | 1mg/dL | | Peroxide | 0.1mg/dL | | Atropine | 30mg/dL | | Fructose | 5000 mg/dL | | Lactose | 5000 mg/dL | | Ketone | 200 mg/dL | | Mesna | 50mg/dL | To investigate the effect of urine specific gravity and urine pH, urine samples, with 1.005 to 1.035 specific gravity or urine samples with pH 5 to 8.5 were tested at different leukocyte and nitrite concentrations. The test results show that there is no effect of both pH and specific gravity on leukocyte and nitrite test results. {9} # 4. Stability The device is stable at $2 - 30^{\circ}\mathrm{C}$ for 24 months based on real-time stability studies. # 5. Method Comparison and Lay-user Studies A total of 234 clinical samples were obtained from 234 subjects with UTI symptoms. These samples were randomized and tested by seven operators using three lots of the candidate device and one lot of the Healgen 10 Reagent Strips for Urinalysis device (K111999). The obtained results are shown below. # Comparison Results between Candidate Device and Healgen 10 Reagent Strips for Urinalysis (K111999) Device by Health Professionals Leukocytes | Healgen 10 Reagent Strips for Urinalysis Healgen® URS Test Strips | 500 (3+) | 125 (2+) | 70 (1+) | 15 (±) | Negative | Total | | --- | --- | --- | --- | --- | --- | --- | | 500 (3+) | 13 | 0 | 0 | 0 | 0 | 13 | | 125 (2+) | 0 | 43 | 1 | 0 | 0 | 44 | | 70 (1+) | 0 | 2 | 55 | 1 | 0 | 58 | | 15(±) | 0 | 0 | 0 | 35 | 1 | 36 | | Negative | 0 | 0 | 0 | 0 | 83 | 83 | | Total | 13 | 45 | 56 | 36 | 84 | 234 | | % Agreement (Exact Match) | 100 | 95.6 | 98.2 | 97.2 | 98.8 | | | % Agreement (+/- Color Block) | 100 | 100 | 100 | 100 | 100 | | Nitrite | Healgen 10 Reagent Strips for Urinalysis Healgen® URS Test Strips | Positive | Negative | Overall | | --- | --- | --- | --- | | Positive | 100 | 0 | 100 | | Negative | 0 | 134 | 134 | | Total | 100 | 134 | 234 | {10} 7 | % Agreement (Exact Match) | 100 | 100 | | | --- | --- | --- | --- | | % Agreement (+/- Color Block) | 100 | 100 | | ## Lay-user Study Three sites were selected to perform the lay-user studies. 234 lay users with UTI symptoms were recruited to test their own urine sample using the Healgen® URS Test Strips. Laypersons performed one test with the Healgen® URS Test Strips according to the product insert and then collected a sample of their urine for comparison testing by healthcare professionals. The results obtained by the lay users compared to the results obtained by the healthcare professionals are summarized below: ## Comparison Results between Lay-user and Health Professionals Using Healgen® URS Test Strips ### Leukocytes | Results by Health Professionals Layperson Results | 500 (3+) | 125 (2+) | 70 (1+) | 15 (±) | Negative | Total | | --- | --- | --- | --- | --- | --- | --- | | 500 (3+) | 13 | 0 | 0 | 0 | 0 | 13 | | 125 (2+) | 1 | 44 | 0 | 0 | 0 | 45 | | 70 (1+) | 0 | 0 | 55 | 1 | 0 | 56 | | 15(±) | 0 | 0 | 2 | 33 | 1 | 36 | | Negative | 0 | 0 | 0 | 0 | 84 | 84 | | Total | 14 | 44 | 57 | 34 | 85 | 234 | | % Agreement (Exact Match) | 92.9 | 100 | 96.5 | 97.1 | 98.8 | | | % Agreement (+/- Color Block) | 100 | 100 | 100 | 100 | 100 | | ### Nitrite | Results by Health Professionals Layperson Results | Positive | Negative | Overall | | --- | --- | --- | --- | | Positive | 100 | 0 | 100 | {11} Comparison Results between Lay-user and Health Professionals Using the Healgen 10 Reagent Strips for Urinalysis Device (K111999) Leukocytes | Results by Health Professionals Layperson Results | 500 (3+) | 125 (2+) | 70 (1+) | 15(±) | Negative | Total | | --- | --- | --- | --- | --- | --- | --- | | Large (3+) | 13 | 0 | 0 | 0 | 0 | 13 | | Moderate (2+) | 1 | 42 | 1 | 0 | 0 | 44 | | Small (1+) | 0 | 2 | 54 | 2 | 0 | 58 | | Trace | 0 | 0 | 2 | 32 | 2 | 36 | | Negative | 0 | 0 | 0 | 0 | 83 | 83 | | Total | 14 | 44 | 57 | 34 | 85 | 234 | | % Agreement (Exact Match) | 92.9 | 95.5 | 94.7 | 94.1 | 97.6 | | | % Agreement (+/- Color Block) | 100 | 100 | 100 | 100 | 100 | | Nitrite | Results by Health Professionals Layperson Results | Positive | Negative | Overall | | --- | --- | --- | --- | | Positive | 100 | 0 | 100 | | Negative | 0 | 134 | 134 | | Total | 100 | 134 | 234 | | % Agreement (Exact Match) | 100 | 100 | | | % Agreement (+/- Color Block) | 100 | 100 | | {12} Lay-users were also given surveys on the ease of understanding the package insert instructions. All lay users indicated that the device instructions can be easily followed. A Flesch-Kincaid reading analysis was performed on the package insert and the score revealed a reading grade level of less than 8. 6. Clinical Studies Not applicable. 11. Conclusion Based on the test principle and performance characteristics of the device including LOD, precision, interference, method comparison and lay-user studies of the devices, it's concluded that the Healgen® URS Test Strips are substantially equivalent to the predicate. 9
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...