The Teco Diagnostics UTI Test Strips are intended for the qualitative detection of Nitrite and Leukocytes in urine as an aid in the screening of urinary tract infection (UTI). It is intended for over-the-counter home use only.
Device Story
Device consists of firm plastic strips with two dry reagent pads; pads contain bibulous material saturated with chemically active substances. User dips strip into urine sample; chemical reactions produce color changes on pads. Nitrite test: Gram-negative bacteria convert nitrate to nitrite; nitrite reacts with p-arsanilic acid to form diazonium compound, which couples with 1,2,3,4-tetrahydrobenzo(h) quinolin to produce pink color. Leukocyte test: Leukocyte esterase catalyzes hydrolysis of indoxyl ester derivative; liberated indoxyl reacts with diazonium salt to produce beige-pink to purple color. Results obtained by visual comparison against color chart provided on foil pouch. Used by lay persons in home setting for UTI screening. Output provides qualitative indication of infection presence; assists users in identifying need for professional medical consultation.
Clinical Evidence
Clinical study conducted at point-of-care sites and with lay persons comparing Teco UTI Test Strips to predicate Multistix 10 SG. Study evaluated clinical accuracy and ability of inexperienced lay users to obtain results comparable to professionals. Results indicated lay users achieved performance comparable to professional users.
Technological Characteristics
Plastic test strips with two dry reagent pads; bibulous material saturated with chemical reagents. Nitrite detection via p-arsanilic acid/1,2,3,4-tetrahydrobenzo(h) quinolin coupling. Leukocyte detection via indoxyl ester hydrolysis/diazonium salt reaction. Visual readout via color chart. Storage 15-30°C. No electronic components or software.
Indications for Use
Indicated for qualitative detection of Nitrite and Leukocytes in human urine to aid in screening for urinary tract infections (UTI). Intended for over-the-counter home use by lay persons.
Regulatory Classification
Identification
A nitrite (nonquantitative) test system is a device intended to identify nitrite in urine. Nitrite identification is used in the diagnosis and treatment of uninary tract infection of bacterial origin.
Predicate Devices
Multistix 10 SG Reagent Strips for Urinalysis (K905396)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows a black hexagon with the letters 'TC' in white inside. The letters are in a simple, sans-serif font and are stacked on top of each other. The hexagon is solid black, providing a strong contrast to the white letters. The overall design is simple and modern.
# FEB 1 7 2011
## 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR §807.92 .-
The Assigned 510(k) number is k100024
## Submitter's Identification:
Teco Diagnostics, Inc. 1268 N. Lakeview Avenue Anaheim, CA 92807
Tel: 714-463-1111 Fax: 714-463-1169
Date Prepared: December 18, 2009
## Contact Person:
Dr. KC Chen
## Proprietary Name of the Device:
Teco Diagnostics UTI Test Strips
## Common Name:
Urine Reagent Strips (URS)
#### Regulation Section and Classification:
21 CFR § 862.1510 Nitrite (Non-Quantitative) Test System Leukocyte Peroxidase Test 21 CFR § 864.7675
Class I: Urinary Leukocytes, Nitrite
## Product Code:
Test, Urine Leukocyte LJX Nitrite (urinary, non-quantitative) test system JMT
## Medical Specialty:
Clinical Chemistry
{2}------------------------------------------------
Image /page/1/Picture/1 description: The image shows a logo with the letters 'TC' in white, set against a black hexagon. The letters are bold and sans-serif, with the 'T' slightly taller than the 'C'. The hexagon provides a solid, geometric background that makes the letters stand out.
# Predicate Device:
Multistix 10 SG Reagent Strips for Urinalysis, K905396 Bayer Corporation, marketed by Bayer Corporation, located at Elkhart, IN 46515, USA.
# Device Description:
Teco Diagnostics UTI Test Strips provides tests for the qualitative determination of leukocytes and nitrite in urine. The product is a firm plastic strip to which is affixed two separate dry reagent pad areas that is used for urinalysis. The reagent pad areas are bibulous material saturated with chemically active substances. All UTI Test Strips are carefully packaged along with a desiccant in a sealed, foil pouch. The color chart for reading the strips is on the outside of the foil pouch. A package insert is packaged along with the foil pouch into a box. The package insert contains all the necessary product information.
Results of each test are based on the color produced from the reaction of each reagent pad area once the Urine Reagent Strip comes into contact with a urine sample. Each parameter is color coded accordingly as described in the color chart. Results with Teco Diagnostics UTI Test Strips can be obtained in clinically meaningful units directly by comparison with the color chart. The color blocks represent nominal values will vary around the nominal values.
## Intended Use:
The Teco Diagnostics UTI Test Strips are intended for the qualitative detection of Nitrite and Leukocytes in urine as an aid in the screening of urinary tract infection (UTI). It is intended for over-the-counter home use only.
## Test Principles:
Nitrite: This test depends on the conversion of nitrite by the action of Gram-negative bacteria in the urine. The nitrite reacts with p-arsanilic acid to from a diazonium compound in an acid medium. The diazonium compound in turn couples with 1,2,3,4-tetrahydrobenzo(h) quinolin to produce a pink color.
Leukocyte: This test is based on the action of esterase present in leukocytes, which catalyzes the hydrolysis of an indoxyl ester derivative. The indoxyl ester liberated reacts with a diazonium salt to produce a beige-pink to purple color.
## Substantial Equivalence:
The Teco Diagnostics UTI Test Strips for Leukocyte and Nitrite are substantially equivalent to the Bayer Multistix 10-SG Reagent Strips for Urinalysis (K905396).
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image shows a logo with the letters 'TC' in white, set against a black hexagon. The letters are bold and sans-serif, with the 'T' slightly overlapping the 'C'. The hexagon is oriented with one point facing downwards, creating a visually stable and recognizable mark. A thin black line is visible above the hexagon, possibly indicating the top edge of a document or page.
## 1268 N. Lakeview Ave., Anaheim, CA 92807, U.S.A. Phone: (714) 463-1111 Toll Free: 1-800-222-9880 Fax: (714) 463-1169 www.tecodiag.com
# Comparison of Characteristics with Predicate:
| | Teco Diagnostics UTI Test Strips | Bayer Multistix 10 SG Reagent<br>Strips |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | For the qualitative detection of Nitrite<br>and Leukocyte in urine as an aid in the<br>screening of urinary tract infection (UTI) | For qualitative detection of Glucose,<br>Bilirubin, Ketone, pH, Blood (Occult),<br>Specific Gravity, Protein,<br>Urobilinogen, Leukocyte and Nitrite in<br>urine to assist in diagnosis in kidney<br>function, liver function, metabolic<br>disorders and urinary tract infection. |
| Intended<br>Users | Lay persons, over-the-counter use | For professional use in point-of-care<br>urine testing |
| Specimen | Human Urine | Same |
| Materials<br>Provided | Plastic test strips affixed with reagent<br>pads | Same |
| Storage | 15 - 30°C | Same |
| Nitrite<br>Test Time | 60 seconds | Same |
| Leukocyte<br>Test Time | 2 minutes | Same |
| Nitrite<br>Parameter<br>Methodology | This test depends on the conversion of<br>nitrate to nitrite by the action of Gram-<br>negative bacteria in the urine. The<br>nitrite reacts with p-arsanilic acid to<br>from a diazonium compound in an acid<br>medium. The diazonium compound in<br>turn couples with 1,2,3,4-<br>tetrahydrobenzo(h) quinolin to produce<br>a pink color. | Same |
| Leukocyte<br>Parameter<br>Methodology | This test is based on the action of<br>esterase present in leukocytes, which<br>catalyzes the hydrolysis of an indoxyl<br>ester derivative. The indoxyl ester<br>liberated reacts with a diazonium salt to<br>produce a beige-pink to purple color. | Same |
The devices differ in their intended users. UTI Test Strips are intended for lay users over the counter, whereas Multistix 10 SG is intended for professional use in point-of-care clinical sites. The possible errors associated with lay user testing were investigated in a clinical, comparison study where both lay users and professional users tested the strips on the same sample. The study demonstrated that the strips were able to be read correctly and easily among lay users.
The predicate device, Multistix 10 SG, also tests eight other analytes, including Glucose, Bilirubin, Ketone, pH value, Blood (Occult), Specific Gravity, Protein and Urobilinogen, intended in the assisting the diagnosis in the areas of kidney function, liver function and metabolic disorders. However, these analytes are not used to aid in the diagnosis of urinary tract infection and the safety and effectiveness of Teco Diagnostics' UTI Test Strips are not affected by the drop of these test analytes.
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a logo with the letters "TC" in a bold, sans-serif font. The letters are stacked, with "T" above "C". The logo is contained within a black hexagon. The letters "TC" are white, creating a contrast with the black background of the hexagon.
#### Summary of Clinical Tests Performed:
The clinical studies were conducted at Point of Care sites and with lay persons using the Teco Diagnostics UTI Test Strips versus Multistix 10 SG. Clinical data were presented evaluating clinical accuracy of results. Clinical study results indicate that the inexperienced lay users were able to obtain comparable testing data compared to those obtained by the professionals when using the Teco Diagnostics UTI Test Strips and the legally marketed Bayer Multistix 10 SG Reagent Strips for Urinalysis (K905396).
#### Summary of Laboratory Tests Performed:
The Laboratory studies were conducted in house and included sensitivity studies, reproducibility study, interference studies, stress study, and stability studies. The obtained laboratory data indicate that Teco Diagnostics UTI Test Strips run well and met all required performance characteristics.
#### Conclusion:
The performance characteristics of the Teco Diagnostics UTI Test Strips were verified by sensitivity study, reproducibility study, interference studies, and temperature stress study. Testing results indicate that Teco Diagnostics UTI Test Strips can perform satisfactorily when used according to the "Test Procedure" directions on the package insert.
The laboratory testing results and clinical studies demonstrated a substantial equivalency on performance between the Teco Diagnostics UTI Test Strips and a legally marketed product, Bayer Multistix 10 SG Reagent Strips for Urinalysis (K905396), with similar product characteristics and intended use of aid in diagnosis in urinary tract infection.
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized caduceus-like symbol with three abstract shapes resembling birds or leaves. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993
Image /page/4/Picture/4 description: The image shows a date, which is February 17, 2011. The month is abbreviated as FEB. The day is 17, and the year is 2011. The date is written in a bold, sans-serif font.
TECO Diagnostics c/o Dr. KC Chen 1268 North Lakeview Avenue Anaheim, CA 92807
Re: k100024
> Trade Name: TECO Diagnostics UTI Test Strips Regulation Number: 21 CFR 862.1510 Regulation Name: Nitrite (Non-Quantitative) Test System. Regulatory Class: Class I: (meets the limitations of exemptions in 21 CFR 862,9(c)(9) Product Codes: JMT, LJX Dated: January 12, 2011 Received: February 02, 2011
Dear Dr. Chen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice. labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{6}------------------------------------------------
#### Page 2 -
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Viro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-fire no (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
C.C.
Courtney Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number (If Known): k100024
Device Name: Teco Diagnostics UTI Test Strips
Indications for Use:
The Teco Diagnostics UTI Test Strips are intended for the qualitative detection of Nitrite and Leukocytes in urine as an aid in the screening of urinary tract infection (UTI). It is intended for over-the-counter home use only.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use__ > (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Director Sign Off
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) k/00024
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.