K031052 · Access Bio Incorporate · DHA · Jun 13, 2003 · Clinical Chemistry
Device Facts
Record ID
K031052
Device Name
CARESTART HCG ONE-STEP PREGNANCY TEST
Applicant
Access Bio Incorporate
Product Code
DHA · Clinical Chemistry
Decision Date
Jun 13, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1155
Device Class
Class 2
Indications for Use
CareStanTM hCG One Step Pregnancy Test is intended for the qualitative detection of human chorionic gonadotropin (hCG) in urine for the early detection of pregnancy. The test is intended for use by health care professionals.
Device Story
CareStart hCG One-Step Pregnancy Test is a lateral flow immunoassay for qualitative detection of human chorionic gonadotropin (hCG) in urine. Device used by healthcare professionals in clinical settings for early pregnancy detection. Principle of operation involves immunochromatographic capture of hCG hormone in urine sample, resulting in visual color change on test strip. Output is qualitative (positive/negative) visual result interpreted by clinician. Provides rapid diagnostic information to confirm pregnancy status.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Lateral flow immunochromatographic assay; qualitative visual detection of hCG in urine; manual test strip format.
Indications for Use
Indicated for qualitative detection of hCG in urine for early pregnancy detection in patients. Intended for use by healthcare professionals.
Regulatory Classification
Identification
A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.
Related Devices
K031048 — CARESTART HCG COMBO ONE-STEP PREGNANCY TEST AND CARESTART PLUS HCG COMBO ONE-STEP PREGNANCY TEST · Access Bio Incorporate · Jun 16, 2003
K103574 — PREGNANCY ONE STEP RAPID TEST · Lehnus & Associates Consulting · Jul 28, 2011
K233624 — HCG One Step Pregnancy Test Strip OTC, HCG One Step Pregnancy Test Strip Rx, HCG One Step Pregnancy Test Midstream OTC, HCG One Step Pregnancy Test Midstream Rx · Hangzhou Aichek Medical Technology Co., Ltd. · Apr 12, 2024
K984076 — INSTANT-VIEW PREGNANCY URINE TEST · Alfa Scientific Designs, Inc. · Feb 16, 1999
K981892 — QUICK & SURE ONESTEP HOME PREGNANCY TEST · Abacus Diagnostics · Aug 18, 1998
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three heads facing to the right. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
## JUL 1 1 2003
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Mr. Young Ho Choi Director Access Bio, Inc. 675 U.S. Highway 1 North Brunswick, NJ 08902
Re: k031052 Trade/Device Name: CareStart™ hCG One-Step Pregnancy Test CareStart Plus - hCG One-Step Pregnancy Test Regulation Number: 21 CFR 862.1155 Regulation Name: Human chorionic gonadotropin (HCG) test system Regulatory Class: Class II Product Code: JHI Dated: March 31, 2003 Received: April 16, 2003
Dear Mr. Young Ho Choi:
This SE letter corrects the earlier letter stamp dated June 13, 2003. The name of the device had the word Combo in the wording for the name. It has now been removed from the Device Name.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{1}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Steven Butman
Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## INDICATIONS FOR USE STATEMENT
Ver/ 3 - 4/24/96
Applicant: Access Bio Inc.
510(k) Number (if known): New Application
Device Name: CareStart™ hCG One-Step Pregnancy Test CareStart Plus- hCG One-Step Pregnancy Test
Indications For Use:
CareStanTM hCG One Step Pregnancy Test is intended for the qualitative detection of human chorionic gonadotropin (hCG) in urine for the early detection of pregnancy. The test is intended for use by health care professionals.
Tan Coggm
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K031052
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use ( -
or
Over-The-Counter Use(
(Per 21 CFR 801.109)
(Optional Format 1-2-96)
4
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.