The Pregnancy One Step Rapid Test Strip is an in vitro diagnostic visual qualitative immunochromatographic assay designed for the rapid determination of human chorionic gonadotropin (hCG) in urine to aid in the early detection of pregnancy. For professional use only. The Pregnancy One Step Rapid Test Cassette is an in vitro diagnostic visual qualitative immunochromatographic assay designed for the rapid determination of human chorionic gonadotropin (hCG) in urine to aid in the early detection of pregnancy. For professional use only. The Pregnancy One Step Rapid Midstream Test is an in vitro diagnostic visual qualitative immunochromatographic assay designed for the rapid determination of human chorionic gonadotropin (hCG) in urine to aid in the early detection of pregnancy. For OTC self-testing use.
Device Story
Rapid sandwich immunoassay for qualitative hCG detection in human urine; utilizes membrane pre-coated with monoclonal anti-hCG antibodies (test line) and anti-mouse antibodies (control line). Urine sample reacts with mouse anti-hCG antibody-colloidal gold conjugate; mixture migrates via capillary action. Positive result indicated by red test band; control band confirms proper flow/volume. Available in strip, cassette, and midstream formats. Strip/cassette used by healthcare professionals in clinical settings; midstream format used by lay users for OTC self-testing. Results interpreted visually by user/clinician to aid pregnancy diagnosis. Benefits include rapid, early detection of pregnancy.
Clinical Evidence
No clinical data provided; substantial equivalence established via bench testing of analytical performance.
Technological Characteristics
Qualitative sandwich immunochromatographic assay. Membrane pre-coated with monoclonal anti-hCG antibodies and anti-mouse antibodies. Dye conjugate: mouse anti-hCG antibody-colloidal gold. Formats: strip, cassette, midstream. Storage: 2-30°C. Traceable to WHO 4th International Standard for hCG. No electronic components or software.
Indications for Use
Indicated for women of childbearing age, including those peri-menopausal or suspected of pregnancy, for the qualitative detection of hCG in urine to aid in early pregnancy detection. Midstream format for OTC self-testing; strip and cassette formats for professional use.
Regulatory Classification
Identification
A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.
Submission Summary (Full Text)
{1}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k103574
B. Purpose for Submission:
New device
C. Measurand:
Human chorionic gonadotropin (hCG) in human urine
D. Type of Test:
Qualitative immunochromatographic assay
E. Applicant:
Zhejiang Orient Gene Biotech, Co., Ltd.
F. Proprietary and Established Names:
Pregnancy One Step Rapid Test (Strip)
Pregnancy One Step Rapid Test (Cassette)
Pregnancy One Step Rapid Test (Midstream)
G. Regulatory Information:
1. Regulation section:
21 CFR § 862.1155, Human chorionic gonadotropin (HCG) test system
2. Classification:
II
3. Product code:
LCX
{2}
4. Panel:
Chemistry (75)
H. Intended Use:
1. Intended use(s):
See below.
2. Indication(s) for use:
The Pregnancy One Step Rapid Test Strip is an in vitro diagnostic visual qualitative immunochromatographic assay designed for the rapid determination of human chorionic gonadotropin (hCG) in urine to aid in the early detection of pregnancy. For professional use only.
The Pregnancy One Step Rapid Test Cassette is an in vitro diagnostic visual qualitative immunochromatographic assay designed for the rapid determination of human chorionic gonadotropin (hCG) in urine to aid in the early detection of pregnancy. For professional use only.
The Pregnancy One Step Rapid Midstream Test is an in vitro diagnostic visual qualitative immunochromatographic assay designed for the rapid determination of human chorionic gonadotropin (hCG) in urine to aid in the early detection of pregnancy. For OTC self-testing use.
3. Special conditions for use statement(s):
Test strip and cassette formats are for prescription use.
Midstream format is for over-the-counter (OTC) use.
4. Special instrument requirements:
None
I. Device Description:
The Pregnancy One Step Rapid Test is a rapid sandwich immunoassay device designed for the qualitative determination of human chorionic gonadotropin (hCG) concentration in human urine samples, as an aid in the early detection of pregnancy. The test is available in three formats: strip, cassette, and midstream. The midstream format is intended for OTC use. The cassette and strip formats are intended for prescription use. All three configurations have the same membrane format, reagents, and flow characteristics. Devices are packaged one device per pouch with 2 devices
2
{3}
per kit.
## J. Substantial Equivalence Information:
1 and 2. Predicate device name(s):
| Name | Format | 510k |
| --- | --- | --- |
| Acon Combo Pregnancy | Cassette | k993065 |
| Acon hCG One Step Pregnancy Test Strip | Strip, Cassette | k993317 |
| Acon Spectrum Midstream | Midstream cassette | k042151 |
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate device |
| Intended Use | The Pregnancy One Step Rapid Test is an in vitro diagnostic visual qualitative immunochromatographic assay designed for the rapid determination of human chorionic gonadotropin (hCG) to aid in the detection of pregnancy. | Same |
| Sensitivity | 25 IU/mL | Same |
| Storage | 2 - 30°C | Same |
| Principle | Sandwich Immunochromatographic Assay | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate device |
| Users | Prescription and OTC | Prescription |
| Specimen | Urine | Urine, serum |
| Read Time | 5 Minutes | 3 minutes |
| Test Line | Particle Membrane Immunoassay | Colloidal Gold |
| Formats | Strip, cassette, midstream | Strip and midstream |
## K. Standard/Guidance Document Referenced (if applicable):
Guidance for Industry and FDA Reviewers/Staff: FDA Guidance for Over-the-Counter (OTC) Human Chorionic Gonadotropin (hCG) 510(k)s
{4}
CLSI EP7-A2; Interference Testing in Clinical Chemistry
# L. Test Principle:
The Pregnancy One Step Rapid Test is a qualitative, solid phase, two-site sandwich immunoassay for the detection of human chorionic gonadotropin (hCG) in urine. The membrane is pre-coated with monoclonal anti-hCG antibodies on the test band region and anti-mouse antibodies on the control band region. During testing, the urine sample reacts with the dye conjugate (mouse anti-hCG antibody-colloidal gold conjugate) which has been pre-coated on the test strip. The mixture then migrates upward on the membrane chromatographically by capillary action to react with anti-hCG antibodies on the membrane and generate a red band. Presence of the red band indicates a positive result, while its absence indicates a negative result. Regardless of the presence of hCG, as the mixture continues to migrate across the membrane to the immobilized goat anti-mouse region, a red band at the control band region will always appear. The presence of this red band serves as verification for sufficient sample volume and proper flow and as a control for the reagents.
# M. Performance Characteristics (if/when applicable):
# 1. Analytical performance:
# a. Precision/Reproducibility:
Reproducibility of each device type was evaluated at three sites by three users per site. Each device format was evaluated—test strip, cassette, and midstream. Testing was performed by spiking negative urine samples with purified hCG, traceable to WHO International $4^{\text{th}}$ Standard, to $15\mathrm{mIU/mL}$ , 20 mIU/mL, 25 mIU/mL, 30 mIU/mL, 35 IU/mL of hCG over 3 days. hCG concentrations were confirmed by a FDA cleared commercially available chemiluminescent immunoassay. Healthcare professionals (HCP) performed testing on the cassette and test strip. Lay users performed the midstream testing. All samples were blinded to both the HCPs and lay users. Inter-site and day-to-day precision by device format results are shown below:
Test Strip
| | Site | | | | | | Total | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | 1 | | 2 | | 3 | | | |
| | | + | - | + | - | + | - | + | - |
| HCG concentration (mIU/mL) | 15 | 0 | 27 | 0 | 27 | 0 | 27 | 0 | 81 |
| | 20 | 1 | 26 | 0 | 27 | 1 | 26 | 2 | 79 |
| | 25 | 13 | 14 | 12 | 15 | 14 | 13 | 39 | 42 |
| | 30 | 26 | 1 | 27 | 0 | 27 | 0 | 80 | 1 |
| | 35 | 27 | 0 | 27 | 0 | 27 | 0 | 81 | 0 |
{5}
Cassette
| | Site | | | | | | Total | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | 1 | | 2 | | 3 | | | |
| | | + | - | + | - | + | - | + | - |
| HCG concentration (mIU/mL) | 15 | 0 | 27 | 0 | 27 | 0 | 27 | 0 | 81 |
| | 20 | 0 | 27 | 1 | 26 | 2 | 25 | 3 | 78 |
| | 25 | 11 | 16 | 13 | 14 | 13 | 14 | 37 | 44 |
| | 30 | 26 | 1 | 27 | 0 | 27 | 0 | 80 | 1 |
| | 35 | 27 | 0 | 27 | 0 | 27 | 0 | 81 | 0 |
Midstream
| | Site | | | | | | Total | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | 1 | | 2 | | 3 | | | |
| | | + | - | + | - | + | - | + | - |
| HCG concentration (mIU/mL) | 15 | 0 | 27 | 0 | 27 | 0 | 27 | 0 | 81 |
| | 20 | 0 | 27 | 2 | 25 | 0 | 27 | 2 | 79 |
| | 25 | 12 | 15 | 15 | 12 | 11 | 16 | 38 | 43 |
| | 30 | 27 | 0 | 26 | 1 | 26 | 1 | 79 | 2 |
| | 35 | 27 | 0 | 27 | 0 | 27 | 0 | 81 | 0 |
Precision data was also compared among the three formats e.g., strip, cassette, and midstream, to demonstrate that the performance of each test format, was similar across all test sites. Results are summarized below:
| | Configuration | | | | | | Total | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | Test Strip | | Cassette | | Midstream | | | |
| | | + | - | + | - | + | - | + | - |
| HCG concentration (mIU/mL) | 15 | 0 | 81 | 0 | 81 | 0 | 81 | 0 | 243 |
| | 20 | 2 | 79 | 3 | 78 | 2 | 79 | 7 | 236 |
| | 25 | 39 | 42 | 37 | 44 | 38 | 43 | 114 | 129 |
| | 30 | 80 | 1 | 80 | 1 | 79 | 2 | 239 | 4 |
| | 35 | 81 | 0 | 81 | 0 | 81 | 0 | 234 | 0 |
b. Linearity/assay reportable range:
Not applicable. This is a qualitative device.
Hook effect studies were performed on three lots of test strips at hCG concentrations of 50, 100, 200, 300, 500 IU/mL of hCG. No hook effect was observed at $< 300$ IU/mL.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The devices are traceable to the WHO 4th International Standard for hCG.
{6}
Closed pouch real time stability studies were performed separately on three lots of each format—test strips, cassette and midstream. The devices were stored at 2-8°C, and 25-30°C for 18 months. Samples were tested were tested every 2 months with the WHO 4th International Standard for hCG at 4 concentrations around the cutoff. Based on these studies, closed pouch stability is 18 months when stored at 2-30°C.
Open pouch studies were performed over 5 days at 25-30°C and ≥ 85% relative humidity for various concentrations of hCG. Based on these studies, the device was stable for one day once opened, however, labeling recommends that testing take place immediately after opening the pouch.
d. Detection limit:
See cutoff studies M.1.f.
e. Analytical specificity:
An interference study was carried out with two lots of the test strips on urine samples containing 5 mIU/mL and 50 mIU/mL of hCG. These pools were further spiked with 2 concentrations of potential interfering substances at therapeutic and toxic concentrations following CLSI-EP7A. 10 replicates of each interferent at each hCG concentration were analyzed for each lot (n=20). Interference was defined as ≥ 95% agreement with the control samples. Complete agreement (100%) was observed for the control samples at 5 mIU/mL and 50 mIU/mL of hCG. The highest interferent concentration demonstrating tested, showing no interference is summarized below:
| Acetaminophen | 21.1 mg/dL |
| --- | --- |
| Acetylsalicylic acid | 64.8 mg/dL |
| Ampicillin | 5.9 mg/dL |
| Ascorbic acid | 6340 mcg/dL |
| Atropine | 926.1 mcg/dL |
| Caffeine | 6214.4 mcg/dL |
| Tetracycline | 1777.6 mcg/dL |
| Bilirubin | 18.71 mg/dL |
| Hemoglobin | 2 g/L |
| Glucose | 1000 mg/dL |
| Total Protein | 120 g/L |
| Albumin | 60 g/L |
A cross-reactivity study was carried out by adding known amounts of potential cross reactants of LH, FSH, and TSH to a total of 90 negative urine samples to evaluate the test result lines of 3 lots of test strips. The cross-reactivity results are shown below:
{7}
7
| | Lot I | Lot II | Lot III |
| --- | --- | --- | --- |
| LH 300 mIU/mL | None | None | None |
| FSH 1000 mIU/mL | None | None | None |
| TSH 1000 mIU/mL | None | None | None |
No cross-reactivity was observed for urine samples up to the following concentrations: LH = 300 mIU/mL, FSH = 1000 mIU/mL, and TSH = 1000 mIU/mL.
Cross-reactivity studies to β-core hCG were also performed using 10 strips from 3 lots of test strips (n=30). 0.5, 10 and 200 mcg/mL of β-core hCG were added to samples containing 5 mIU/mL and 50 mIU/mL of hCG. Results are summarized below and expressed as the number of positive samples per lot number for each hCG concentration tested.
β-core hCG Concentrations
| | 0.5 mic/mL | | | 10 mic/mL | | | 200 mic/mL | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| hCG | Lot 1 | Lot 2 | Lot 3 | Lot 1 | Lot 2 | Lot 3 | Lot 1 | Lot 2 | Lot 3 |
| 5 mIU/L | 0/10 | 0/10 | 0/10 | 0/10 | 0/10 | 0/10 | 0/10 | 0/10 | 0/10 |
| 50 mIU/L | 10/10 | 10/10 | 10/10 | 10/10 | 10/10 | 10/10 | 10/10 | 10/10 | 10/10 |
There was no cross-reactivity with β-core hCG up to 200 mcg/mL
The effects of pH and specific gravity were evaluated by testing 10 replicates of urine samples spiked with hCG at pH 2-9 and specific gravity of 1.003-1.030. No interference was observed for pH or specific gravity. However, labeling states that low concentrations of hCG may not be detected in very dilute urine and, if pregnancy is suspected, repeat testing should be performed after 48 hours.
f. Assay cut-off:
To confirm the cut-off with the intended use population, studies were performed separately from the precision tests for the test strip, cassette, cartridge and midstream formats. The sponsor followed CLSI EP12-A and identified the cut-off as 25 mIU/mL. Negative human urine samples were spiked with purified hCG traceable to the WHO International 4th Standard to 15 mIU/mL (-40% cutoff), 20 mIU/mL (-20% cutoff), 25 mIU/mL, 30 mIU/mL (+20% cutoff), 35 mIU/mL (+40% cutoff). hCG concentrations were confirmed by a commercially available chemiluminescent method. Studies were performed at 3 point-of-care (POC) sites by 3 healthcare professionals (HCP) and 3 lay users at each site. All samples were masked prior to analysis. HCPs performed the testing on the test strip and cassette
{8}
formats. Lay users performed testing midstream format and collected the remaining sample in a cup for additional testing. Three lots of test strips, cassettes and midstream devices were used across all sites. Results are expressed as the number of positive samples to the total number of samples analyzed. See below:
| | Strip | Cassette | Midstream |
| --- | --- | --- | --- |
| 15 mIU/mL (-40% cutoff) | 0/81 | 0/81 | 0/81 |
| 20 mIU/mL (-20% cutoff) | 1/81 | 2/81 | 1/81 |
| 25 mIU/mL (cutoff) | 39/81 | 37/81 | 38/81 |
| 30 mIU/mL (+20% cutoff) | 80/81 | 80/81 | 79/81 |
| 35 mIU/mL (+40% cutoff) | 81/81 | 81/81 | 81/81 |
# 2. Comparison studies:
# a. Method comparison with predicate device:
Comparison studies between the test strip, cassette and predicate device were conducted by one healthcare professional each at 3 point of care sites on 3 lots of each device format. Each site collected 50-60 samples $(N = 167)$ . The samples were collected from women who fit the following categories: childbearing age, peri-menopausal, suspected of being pregnant, e.g. within days of missing the expected menses, within the first 30 days of pregnancy, and the first trimester of pregnancy. All samples were masked prior to analysis. Users performed the testing according to the labeling.
Results are summarized below:
| | Predicate device | | | |
| --- | --- | --- | --- | --- |
| | | Positive | Negative | Total Agreement |
| Test Strip n=167 | Positive | 72 | 2 | 97.0 % |
| | Negative | 3 | 90 | |
| | Total | 75 | 92 | |
| | Predicate device | | | |
| --- | --- | --- | --- | --- |
| | | Positive | Negative | Total Agreement |
| Cassette n=167 | Positive | 72 | 1 | 97.6 % |
| | Negative | 3 | 91 | |
| | Total | 75 | 92 | |
A lay user study was performed using the midstream format of the Orient
{9}
Gene Pregnancy HCG Test. Lay users were randomly recruited from the intended use population and did not include healthcare professionals, or laboratorians. Evaluation took place at 3 sites with three lots of devices. 15-20 individuals per site were enrolled in the study (N=56). All were suspected, but not confirmed to be pregnant. Each lay user performed the testing unassisted according to the labeling. In addition to midstream testing, the participants collected their sample in a cup for subsequent blinded testing by healthcare professionals with the Orient Gene Pregnancy HCG test strip format and the predicate device.
| | HCP | Test strip | | |
| --- | --- | --- | --- | --- |
| | | Positive | Negative | Total Agreement |
| Midstream Lay user n=56 | Positive | 18 | 2 | 94.6% |
| | Negative | 1 | 35 | |
| | Total | 19 | 37 | |
| | Predicate | device | | |
| --- | --- | --- | --- | --- |
| | | Positive | Negative | Total Agreement |
| Midstream Lay user n=56 | Positive | 20 | 0 | 98.8% |
| | Negative | 1 | 35 | |
| | Total | 21 | 35 | |
b. Matrix comparison:
Not applicable. The device is for urine only.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
Each lay user was given a questionnaire to assess the readability of the labeling. The results of the questionnaire reflected that the consumers found the test easy to use and that they did not have trouble understanding the labeling and interpreting the results.
{10}
4. Clinical cut-off:
Not applicable.
5. Expected values/Reference range:
Negative results are expected in healthy non-pregnant women. Healthy pregnant women have hCG present in their urine. The amount of hCG will vary greatly with gestational age and between individuals. The test can be used as early as the first day of the missed period.
A toll free telephone number with hours of operation has been provided in the labeling for technical assistance.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.