FDA Browser
by
Innolitics
Anesthesiology
Review Panel
Cardiovascular
Review Panel
Chemistry
Review Panel
Miscellaneous
Miscellaneous
OYC
Pump, Infusion, Insulin, To Be Used With Invasive Glucose Sensor
3
Product Code
P
14
0015
T:SLIM G4 INSULIN PUMP WITH DEXCOM G4 PLATINUM CGM
APPR
PMA Approval
P
13
0007
ANIMAS VIBE SYSTEM
APWD
PMA Approval
OZO
Automated Insulin Dosing , Threshold Suspend
3
Product Code
OZP
Automated Insulin Dosing Device System, Single Hormonal Control
3
Product Code
LHE
Controller, Closed-Loop Blood Glucose
3
Product Code
LKV
Enzyme Immunoassay, Fetal Fibronectin
3
Product Code
LMY
Monitor, Skin Resistance/Skin Temperature, For Insulin Reactions
3
Product Code
LTC
Hyaline Membrane Disease Assay
3
Product Code
MDS
Sensor, Glucose, Invasive
3
Product Code
MWQ
System, Test, Total Antioxidant Status
3
Product Code
NCT
Instrument, Glucose, Noninvasive Technology
3
Product Code
NQC
Test, Breath Analysis, Volatile Organic Compounds
F
Product Code
NUK
For Export Only - Antigen, Anti-Cardiac, Indirect Fluorescent Antibody Test System
N
Product Code
NUL
For Export Only - Antigen, Anti-Skeletal, Indirect Fluorescent Antibody Test System
N
Product Code
OTV
Research Use Only/Clinical Chemistry Devices
N
Product Code
OZQ
Automated Insulin Dosing Device System, Bihormonal Control
3
Product Code
PAQ
Test, Nr2 Antibody
N
Product Code
PAR
Test, Volatile Organic Compounds Breath Analysis
F
Product Code
PDA
Investigational Use Only - Clinical Chemistry
N
Product Code
PEJ
Salivary Estriol Test
3
Product Code
PHK
Test, Cell Mediated Immune Response, Liver And Small Bowel Transplant/Transplantation
N
Product Code
PQF
Sensor, Glucose, Invasive, Non-Adjunctive
3
Product Code
PZE
Sensor, Glucose, Invasive, Non-Adjunctive, Factory-Calibrated, User-Initiated
3
Product Code
QCD
Continuous Glucose Monitor, Implanted, Adjunctive Use
3
Product Code
QHJ
Sensor, Glucose, Implanted, Non-Adjunctive Use
3
Product Code
QRV
Setmelanotide Eligibility Gene Variant Detection System
2
Product Code
QWZ
Prognostic Test For Assessment Of Chronic Kidney Disease Progression
2
Product Code
QXC
Diabetes Digital Therapeutic Device
2
Product Code
SBA
Integrated Continuous Glucose Monitoring System With Sensor Containing Dexamethasone Acetate
2
Product Code
SCH
Ldt, Approved By Nys Clep
N
Product Code
SCE
Ivd Offered As Ldt, First Marketed Before May 6, 2024, Not Modified Beyond Scope Described In Preamble To Ldt Final Rule
N
Product Code
SCF
Ldt, Unmet Need Within An Integrated Healthcare System
N
Product Code
SCG
Modified Version Of Another Manufacturer’S Fda-Authorized Test Within Scope Described In Preamble To Ldt Final Rule
N
Product Code
SCI
Ivd Offered As Ldt, Not An Ldt Or Under A Targeted Enforcement Discretion Policy Described In Preamble To Ldt Final Rule
N
Product Code
SCJ
Ldt, Not Under A Targeted Enforcement Discretion Policy Described In Preamble To Ldt Final Rule
N
Product Code
SFU
Glucose Range Monitoring System
2
Product Code
SFI
Sensor, Glucose, Invasive, Component Of Automated Insulin Delivery System
3
Product Code
Submissions with No Product Code
Miscellaneous
Subpart B—Biological Stains
CFR Sub-Part
Subpart B—Clinical Chemistry Test Systems
CFR Sub-Part
Subpart B—Diagnostic Devices
CFR Sub-Part
Subpart C—Clinical Laboratory Instruments
CFR Sub-Part
Subpart C—Microbiology Devices
CFR Sub-Part
Subpart D—Clinical Toxicology Test Systems
CFR Sub-Part
Subpart D—Pathology Instrumentation and Accessories
CFR Sub-Part
Subpart D—Serological Reagents
CFR Sub-Part
Subpart F—General Hospital and Personal Use Therapeutic Devices
CFR Sub-Part
Subpart F—Immunological Test Systems
CFR Sub-Part
Dental
Review Panel
Ear, Nose, Throat
Review Panel
Gastroenterology and Urology
Review Panel
General Hospital
Review Panel
Hematology
Review Panel
Immunology
Review Panel
Microbiology
Review Panel
Neurology
Review Panel
Obstetrical and Gynecological
Review Panel
Ophthalmic
Review Panel
Orthopedic
Review Panel
Pathology
Review Panel
Physical Medicine
Review Panel
Radiology
Review Panel
General and Plastic Surgery
Review Panel
Toxicology
Review Panel
Medical Genetics
Review Panel
Unknown
Review Panel
Last synced on 3 April 2026 at 11:13 pm
CH
/
CH-misc
/
OYC
/
P140015
View Source
T:SLIM G4 INSULIN PUMP WITH DEXCOM G4 PLATINUM CGM
Page Type
PMA Approval
PDF
PDF
* The FDA doesn't provide PDFs for all submissions.
Applicant
Tandem Diabetes Care, Inc.
Trade Name
T:SLIM G4 INSULIN PUMP WITH DEXCOM G4 PLATINUM CGM
Generic Name
Pump, infusion, insulin, to be used with invasive glucose sensor
Decision Date
9/8/2015
Days to Decision
418 days
Decision Code
APPR
FDA Decision
Approval
FDA Browser
by
Innolitics
Anesthesiology
Cardiovascular
Chemistry
Miscellaneous
OYC
Pump, Infusion, Insulin, To Be Used With Invasive Glucose Sensor
P
14
0015
T:SLIM G4 INSULIN PUMP WITH DEXCOM G4 PLATINUM CGM
P
13
0007
ANIMAS VIBE SYSTEM
OZO
Automated Insulin Dosing , Threshold Suspend
OZP
Automated Insulin Dosing Device System, Single Hormonal Control
LHE
Controller, Closed-Loop Blood Glucose
LKV
Enzyme Immunoassay, Fetal Fibronectin
LMY
Monitor, Skin Resistance/Skin Temperature, For Insulin Reactions
LTC
Hyaline Membrane Disease Assay
MDS
Sensor, Glucose, Invasive
MWQ
System, Test, Total Antioxidant Status
NCT
Instrument, Glucose, Noninvasive Technology
NQC
Test, Breath Analysis, Volatile Organic Compounds
NUK
For Export Only - Antigen, Anti-Cardiac, Indirect Fluorescent Antibody Test System
NUL
For Export Only - Antigen, Anti-Skeletal, Indirect Fluorescent Antibody Test System
OTV
Research Use Only/Clinical Chemistry Devices
OZQ
Automated Insulin Dosing Device System, Bihormonal Control
PAQ
Test, Nr2 Antibody
PAR
Test, Volatile Organic Compounds Breath Analysis
PDA
Investigational Use Only - Clinical Chemistry
PEJ
Salivary Estriol Test
PHK
Test, Cell Mediated Immune Response, Liver And Small Bowel Transplant/Transplantation
PQF
Sensor, Glucose, Invasive, Non-Adjunctive
PZE
Sensor, Glucose, Invasive, Non-Adjunctive, Factory-Calibrated, User-Initiated
QCD
Continuous Glucose Monitor, Implanted, Adjunctive Use
QHJ
Sensor, Glucose, Implanted, Non-Adjunctive Use
QRV
Setmelanotide Eligibility Gene Variant Detection System
QWZ
Prognostic Test For Assessment Of Chronic Kidney Disease Progression
QXC
Diabetes Digital Therapeutic Device
SBA
Integrated Continuous Glucose Monitoring System With Sensor Containing Dexamethasone Acetate
SCH
Ldt, Approved By Nys Clep
SCE
Ivd Offered As Ldt, First Marketed Before May 6, 2024, Not Modified Beyond Scope Described In Preamble To Ldt Final Rule
SCF
Ldt, Unmet Need Within An Integrated Healthcare System
SCG
Modified Version Of Another Manufacturer’S Fda-Authorized Test Within Scope Described In Preamble To Ldt Final Rule
SCI
Ivd Offered As Ldt, Not An Ldt Or Under A Targeted Enforcement Discretion Policy Described In Preamble To Ldt Final Rule
SCJ
Ldt, Not Under A Targeted Enforcement Discretion Policy Described In Preamble To Ldt Final Rule
SFU
Glucose Range Monitoring System
SFI
Sensor, Glucose, Invasive, Component Of Automated Insulin Delivery System
Submissions with No Product Code
Subpart B—Biological Stains
Subpart B—Clinical Chemistry Test Systems
Subpart B—Diagnostic Devices
Subpart C—Clinical Laboratory Instruments
Subpart C—Microbiology Devices
Subpart D—Clinical Toxicology Test Systems
Subpart D—Pathology Instrumentation and Accessories
Subpart D—Serological Reagents
Subpart F—General Hospital and Personal Use Therapeutic Devices
Subpart F—Immunological Test Systems
Dental
Ear, Nose, Throat
Gastroenterology and Urology
General Hospital
Hematology
Immunology
Microbiology
Neurology
Obstetrical and Gynecological
Ophthalmic
Orthopedic
Pathology
Physical Medicine
Radiology
General and Plastic Surgery
Toxicology
Medical Genetics
Unknown
CH
/
CH-misc
/
OYC
/
P140015
View Source
T:SLIM G4 INSULIN PUMP WITH DEXCOM G4 PLATINUM CGM
Page Type
PMA Approval
PDF
PDF
* The FDA doesn't provide PDFs for all submissions.
Applicant
Tandem Diabetes Care, Inc.
Trade Name
T:SLIM G4 INSULIN PUMP WITH DEXCOM G4 PLATINUM CGM
Generic Name
Pump, infusion, insulin, to be used with invasive glucose sensor
Decision Date
9/8/2015
Days to Decision
418 days
Decision Code
APPR
FDA Decision
Approval