RESPIRATORY HUMIDIFIER, MODEL MR850

K020332 · Fisher & Paykel Healthcare, Ltd. · BTT · Jul 7, 2003 · Anesthesiology

Device Facts

Record IDK020332
Device NameRESPIRATORY HUMIDIFIER, MODEL MR850
ApplicantFisher & Paykel Healthcare, Ltd.
Product CodeBTT · Anesthesiology
Decision DateJul 7, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5450
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The Fisher & Paykel Healthcare MR850 Humidifier is intended to be used to warm and add humidity to gases delivered to patients requiring mechanical ventilation or positive pressure breathing assistance or general medical gases. The heated-wire breathing circuits are intended as conduits of breathing gas for ventilation of patients, and to maintain the temperature of humidified inspired gas, to reduce condensation. They are accessories for the Fisher & Paykel Healthcare MR850 Respiratory Gas Humidifier. The RT130 is used for flow rates between 0.3 and 4 L/min, and the RT131 is for flow rates greater than 4 L/min, for neonatal patients.

Device Story

MR850 Respiratory Humidifier is a heated pass-over humidifier; adds moisture and heat to dry medical gases. Inputs: temperature and flow parameters via sensors at chamber and patient airway ends. Operation: microprocessor-controlled PID algorithm manages heater plate and heated-wire breathing circuits to maintain gas temperature/humidity. Outputs: humidified gas at 37°C (Intubated Mode) or 31°C (Mask Mode); visual/audible alarms for high temperature, inadequate humidity, or equipment setup errors. Used in clinical settings by healthcare providers. Benefits: prevents respiratory tract damage/irritation and secretion desiccation in patients with bypassed upper airways. Features thermoplastic enclosure, serial data interface, and mounting bracket.

Clinical Evidence

Clinical verification studies confirmed safety, effectiveness, and performance. The device successfully provided required temperature and humidification output across various respiratory gas therapies. Results indicated low user intervention requirements and reduced susceptibility to user error. No new hazards were introduced by modified components.

Technological Characteristics

Thermoplastic enclosure; heated pass-over humidification; PID algorithm microprocessor control; dual-sensor temperature/flow probes; heated-wire breathing circuits. Complies with IEC 60601-1, IEC 60601-1-2, UL 2601-1, ISO 8185, and ASTM F1690. Connectivity includes serial data interface port.

Indications for Use

Indicated for patients requiring mechanical ventilation, positive pressure breathing assistance, or general medical gases, including neonatal patients. Contraindicated for patients where humidification of medical gases is not required.

Regulatory Classification

Identification

A respiratory gas humidifier is a device that is intended to add moisture to, and sometimes to warm, the breathing gases for administration to a patient. Cascade, gas, heated, and prefilled humidifiers are included in this generic type of device.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows a handwritten alphanumeric string. The string is "KO20332". The characters are written in a cursive style with varying stroke thicknesses. The image has a white background. Image /page/0/Picture/1 description: The image shows the logo for Fisher & Paykel Healthcare. The words "Fisher & Paykel" are in a bold, sans-serif font on the top line, with an ampersand between the two words. Below that, the word "HEALTHCARE" is in a similar font, but smaller and not bolded. There is a line above the word "HEALTHCARE". ile: +64 9 574 0158 Website: www.fphcare.com # JUL 0 7 2003 ## 510(K) Summary Contact person Brett Whiston Date prepared 4 April 2003 Trade name MR850 Respiratory Humidifier Common name Respiratory Humidifier Respiratory Gas Humidifier (21 CFR § 868.5450) Classification name MR850 Respiratory Humidifier, K983112 Predicate device Airlife Breathing Circuit, K000697 {1}------------------------------------------------ #### Description of device The MR850 Humidifier is a Respiratory Gas Humidifier (heated pass-over type) according to 21 CFR §868.5450. Heat is used to provide evaporated water content to dry breathing gases. Heated breathing tubes are also utilized in order to increase operating efficiency and reduce excessive water and heat loss. The MR850 has a thermoplastic enclosure with dimensions of 140 mm high × 135 mm wide × 173 mm deep, and weighs 2.8 kg. A heater plate is positioned in the top of the unit, where the enclosure rim and finger guard allow a humidification chamber to be added. Temperature probe and heater wire connection sockets are on the right side of the unit. A serial data interface port is located in the underside of the unit, with a mounting bracket at the back of the device. The unit controls and displays are located on the front panel. Controls consist of power, operating mode and alarm mute buttons. A setup / alarm display indicates if a part of the equipment is incorrectly installed, or type of alarm condition occurring. Accessories for the MR850 Humidification chambers, breathing circuits, electrical adaptors and temperature / flow probes. The chamber slides on to the heater plate and contains the water supply for adding humidity to breathing gases. The breathing circuit transports gases to the patient, and includes sections for connection from ventilator to humidifier, inspiratory limb to the patient, and expiratory limb for return to the ventilator. Heated wires in the inspiratory and expiratory limbs prevent condensation. The electrical adaptor supplies power to the heated wires. The temperature / flow probe has sensors at the chamber and patient airway ends of the inspiratory section for heater control. The MR850 Humidifier has two operating modes. Intubated Mode is used for patients who have bypassed upper airways, and delivers humidified gas to the patient at 37 ℃ (body temperature). Mask Mode is used for patients receiving breathing gases via a face mask, and delivers humidified gas to the patient at 31 °C. The MR850 Humidifier monitors temperature, flow parameters and equipment integrity, in order to maintain stable performance conditions. It will also notify the user of high delivered temperature. inadequate delivered humidity, or incorrect equipment set-up conditions. #### Intended use The MR850 Humidifier is intended to add moisture to, and to warm, the breathing gases for administration to a patient. Gases available for medical use do not contain sufficient moisture and may damage or irritate the respiratory tract, or desiccate secretions of patients whose supraglottic airways have been bypassed. This may be indicated for patients requiring mechanical ventilation, positive pressure breathing assistance, or general medical gases. These gases may be delivered by face mask or through bypassing the upper airways, for example use of an endotracheal tube. {2}------------------------------------------------ #### Technological Characteristics Summary The technological characteristics of the MR850 Humidifier are equivalent to the predicate devices. The MR850 is equivalent to the MR730 and HC500 Humidifiers in terms of: type (heated passover humidification), configuration (chamber, heated wire breathing circuits, dualsensor temperature probe), power usage (same heater system ratings), performance (same temperature and humidity output), control method (electronic and PID algorithm microprocessor), and uses equivalent materials and some common components. The MR850 is equivalent to the 7200 Ventilator in terms of the use of heated-wire flowsensing technology for respiratory gas flow-sensing purposes. #### Discussion of the Non-Clinical Tests Non-clinical testing of the MR850 Humidifier has been carried out covering mechanical, electrical and thermal safety, environmental conditions and electromagnetic compatibility, functional verification, and performance capacity and accuracy. The MR850 meets the requirements of the IEC 60601-1 and IEC 60601-1-2 electro-medical and EMC standards, and the relevant USA deviations to these in UL 2601-1. It meets the mechanical, electrical and environmental testing requirements of the 1993 FDA Reviewer Guidance for Premarket Notification Submissions. It complies with performance and safety requirements of the ISO 8185 and ASTM F1690 (USA) particular standards for Humidification Systems with the exception of clauses 51.6.2 and 51.7. #### Discussion of the Clinical Tests Clinical verification studies on the MR850 Humidifier demonstrated the safety, effectiveness and performance of the device. The humidifier was able to provide required temperature and humidification output across a variety of respiratory gas therapies. It required a low level of user intervention and had reduced susceptibility to user error factors. Modified technological components fulfilled their purpose of safety and effectiveness improvements, and did not introduce further hazards to user or patient. #### Conclusions Demonstrating Safety, Effectiveness and Performance The testing carried out for the MR850 Humidifier indicates that it meets design and performance functional requirements. Clinical verification studies demonstrate the successful use of the humidifier and its ability to provide effective humidity levels. The proposed device meets the requirements of international and USA medical electrical equipment standards for safety, and key performance and safety requirements from the particular standard for humidification systems. This information indicates that the MR850 Humidifier is equivalent to or better than the predicate devices in terms of safety, effectiveness and performance. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three curved lines representing its wings and body. The eagle is positioned within a circular border that contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a sans-serif font. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUL 0 7 2003 Mr. Brett Whiston Regulatory Affairs Fisher & Paykel Healthcare Limited 15 Maurice Paykel Place, East Tamaki P.O. Box 14348, Panmure Auckland, New Zealand Re: K020332 Trade/Device Name: MR 850 Respiratory Humidifier Regulation Number: 21 CFR 868.5450 Regulation Name: Respiratory Gas Humidifier Regulatory Class: II Product Code: BTT Dated: April 4, 2003 Received: April 8, 2003 Dear Mr. Whiston: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA); it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Mr. Brett Whiston Please be advised that FDA`s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely vours. Susan Rmse Susan Runner, DDS, MA Interim Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ### Indications for use statement K020332 510(k) Number Device Trade Name Respiratory Humidifier The Fisher & Paykel Healthcare MR850 Humidifier is intended to be used to warm and add humidity to gases delivered to patients requiring mechanical ventilation or positive pressure breathing assistance or general medical gases. The heated-wire breathing circuits are intended as conduits of breathing gas for ventilation of patients, and to maintain the temperature of humidified inspired gas, to reduce condensation. They are accessories for the Fisher & Paykel Healthcare MR850 Respiratory Gas Humidifier. The RT130 is used for flow rates between 0.3 and 4 L/min, and the RT131 is for flow rates greater than 4 L/min, for neonatal patients. Concurrence of CDRH, Office of Device Evaluation (ODE) Michael Bazaral 510(k) Number: K020332 Prescription Use (per 21 CFR 801.109)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...