The Fisher & Paykel Healthcare MR850 Humidifier is a Respiratory Gas Humidifier as per 73 BTT, 21 CFR §868.5450. It is intended to add moisture to and warm breathing gases for administration to a patient. The MR850 is intended to be used to warm and add humidity to gases delivered to patients requiring mechanical ventilation, positive pressure breathing assistance or general medical gases. Gases available for medical use do not contain sufficient moisture and may damage or irritate the respiratory tract, or desiccate secretions of patients whose supraglottic airways have been bypassed. Heat is used to increase the water output of the humidifier. Heated breathing tubes are also utilized in order to increase operating efficiency and reduce excessive water and heat loss.
Device Story
MR850 is a heated passover respiratory gas humidifier; adds moisture and heat to medical gases to prevent respiratory tract damage or secretion desiccation. Device uses a heaterplate to warm water in a humidification chamber; heated breathing circuits minimize condensation and heat loss. Inputs include temperature and flow data from sensors at the chamber and patient airway ends. Microprocessor-controlled PID algorithm manages heater output to maintain target temperatures (37°C for intubated patients; 31°C for mask mode). Used in clinical settings by healthcare providers; provides visual feedback via front panel display for setup and alarm conditions (e.g., high temperature, inadequate humidity, equipment errors). Benefits include stable, controlled gas conditioning for patients on mechanical ventilation or positive pressure support.
Clinical Evidence
Clinical verification studies demonstrated the device's safety, effectiveness, and performance. The humidifier successfully provided required temperature and humidity levels across various respiratory gas therapies. Results indicated low user intervention requirements and reduced susceptibility to user error. No new hazards were introduced by modified components.
Technological Characteristics
Heated passover humidifier; thermoplastic enclosure (140x135x173mm, 2.8kg). Features heaterplate, temperature/flow probes, and heated-wire breathing circuits. Microprocessor-based PID control algorithm. Complies with IEC 60601-1, IEC 60601-1-2, UL 2601-1, ISO 8185, and ASTM F1690.
Indications for Use
Indicated for patients requiring mechanical ventilation, positive pressure breathing assistance, or general medical gases, including those with bypassed supraglottic airways (e.g., via endotracheal tube) or receiving gases via face mask.
Regulatory Classification
Identification
A respiratory gas humidifier is a device that is intended to add moisture to, and sometimes to warm, the breathing gases for administration to a patient. Cascade, gas, heated, and prefilled humidifiers are included in this generic type of device.
{0}------------------------------------------------
NOV 10 1998
Image /page/0/Picture/1 description: The image shows the logo for Fisher & Paykel Healthcare. The words "Fisher & Paykel" are on the top line, with a line underneath. The word "Healthcare" is on the second line in all caps.
K983112
Fisher & Paykel Electronics Limited
25 Carbine Road, Panmure, Auckland, New Zealand
P.O. Box 14-348, Panmure, Auckland, New Zealand
Tel: +64-9-574-0100 Fax: +64-9-574-0158
31 August, 1998
# 510(k) Summary of Safety and Effectiveness Information
#### Model No. / Name: MR850 Respiratory Humidifier
Classification Name: Humidifier, Respiratory Gas (Direct Patient Interface) - 73 BTT Anesthesiology Devices, 21 CFR §868.5450 (Class II) Fisher & Paykel, MR730 Respiratory Humidifier, K913368 Predicate Devices: Fisher & Paykel, HC500 Respiratory Humidifier, K953392 Fisher & Paykel, MR290 Humidification Chamber, K934140 Puritan-Bennett, 7200 Microprocessor Ventilator, K902506
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR §807.92:
## (a)(1) - (a)(3) Refer to information above and concluding this summary.
# (a)(4) Description of the Device
The MR850 Humidifier is a Respiratory Gas Humidifier (heated passover type) according to 21 CFR 8868,5450. Heat is used to provide an evaporated water content to dry breathing gases. Heated breathing tubes are also utilized in order to increase operating efficiency and reduce excessive water and heat loss.
The MR850 has a thermoplastic enclosure with dimensions of 140mm high × 135mm wide × 173mm deep, and weighs 2.8kg. A heaterplate is positioned in the top of the unit, where the enclosure rim and finger guard allow a humidification chamber to be added. Temperature probe and heater wire connection sockets are on the right side of the unit. A serial data interface port is located in the underside of the unit, with a mounting bracket at the back of the device.
The unit controls and displays are located on the front panel. Controls consist of power, operating mode and alarm mute buttons. A setup / alarm display indicates if a part of the equipment is incorrectly installed, or type of alarm condition occurring.
Accessories for the MR850 Humidification chamidification chambers, breathing circuits, electrical adaptors and temperature / flow probes.
The chamber slides on to the heaterplate and contains the water supply for adding humidity to breathing gases. The breathing circuit transports gases to the patient, and includes sections for connection from ventilator to humidifier, inspiratory limb to the patient, and expiratory limb for return to the ventilator. Heated wires in the inspiratory and expiratory limbs prevent condensation. The electrical adaptor supplies power to the heated wires. The temperature / flow probe has sensors at the chamber and patient airway ends of the inspiratory section for heater control.
{1}------------------------------------------------
510(k) Summary continued - Fisher & Paykel MR850 Respiratory Humidifier
The MR850 Humidifier has two operating modes. Intubated Mode is used for patients who have bypassed upper airways, and delivers humidified gas to the patient at 37°C (body temperature). Mask Mode is used for patients receiving breathing gases via a face mask, and delivers humidified gas to the patient at 31°C. The MR850 Humidifier monitors temperature and flow parameters, and equipment integrity, in order to maintain stable performance conditions and notify the user of high delivered temperature, inadequate delivered humidity, or incorrect equipment set-up conditions.
### (a)(5) Statement of the Intended Use
The MR850 Humidifier is intended to add moisture to, and to warm, the breathing gases for administration to a patient. Gases available for medical use do not contain sufficient moisture and may damage or irritate the respiratory tract, or desiccate secretions of patients whose supraglottic airways have been bypassed.
This may be indicated for patients requiring mechanical ventilation, positive pressure breathing assistance, or general medical gases. These gases may be delivered by face mask or through bypassing the upper airways, for example use of an endotracheal tube.
### (a)(6) Technological Characteristics Summary
The technological characteristics of the MR850 Humidifier are equivalent to the predicate devices listed above.
The MR850 is equivalent to the MR730 and HC500 Humidifiers in terms of: type (heated passover humidification), configuration (chamber, heated wire breathing circuits, dual-sensor temperature probe), power usage (same heater system ratings), performance (same temperature and humidity output), control method (electronic and PID algorithm microprocessor), and uses equivalent materials and some common components.
The MR850 is equivalent to the 7200 Ventilator in terms of the use of heated-wire flow-sensing technology for respiratory gas flow-sensing purposes.
### (b)(1) Discussion of the Non-Clinical Tests
Non-clinical testing of the MR850 Humidifier has been carried out covering mechanical, electrical and thermal safety, environmental conditions and electromagnetic compatibility, functional verification, and performance capacity and accuracy.
The MR850 meets the requirements of the IEC 60601-1 and IEC 60601-1-2 electromedical and EMC standards, and the relevant USA deviations to these in UL 2601-1. It meets the mechanical, electrical and environmental testing requirements of the 1993 FDA Reviewer Guidance for Premarket Notification Submissions. It complies with performance and safety requirements of the ISO 8185 and ASTM F1690 (USA) particular standards for Humidification Systems.
{2}------------------------------------------------
510(k) Summary continued - Fisher & Paykel MR850 Respiratory Humidifier
# (b)(2) Discussion of the Clinical Tests
Clinical verification studies on the MR850 Humidifier demonstrated the safety, effectiveness and performance of the device. The humidifier was able to provide required temperature and humidification output across a variety of respiratory gas therapies. It required a low level of user intervention and had reduced susceptibility to user error factors. Modified technological components fulfilled their purpose of safety and effectiveness improvements, and did not introduce further hazards to user or patient.
# (b)(3) Conclusions Demonstrating Safety, Effectiveness and Performance
The testing carried out for the MR850 Humidifier indicates that it meets design and performance functional requirements. Clinical verification studies demonstrate the successful use of the humidifier and its ability to provide effective humidity levels. The proposed device meets the requirements of international and USA medical electrical equipment standards for safety, and key performance and safety requirements from the particular standard for humidification systems.
This information indicates that the MR850 Humidifier is equivalent to or better than the predicate devices in terms of safety, effectiveness and performance.
signed:
Chris Mander Fisher & Paykel Healthcare Ltd
te: 31 August 1998
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle-like symbol composed of three human profiles facing right, with flowing lines suggesting movement or wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 1 0 1998
Mr. Chris Mander Fisher & Paykel Healthcare 25 Carbine Road P.O. Box 14-348 Panmure, Auckland NEW ZEALAND
Re: K983112 Respiratory Humidifier - Model MR850 Requlatory Class: II (two) Product Code: 73 BTT Dated: August 31, 1998 Received: September 4, 1998
Dear Mr. Mander:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to leqally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
#### Page 2 - Mr. Chris Mander
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diaqnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Thomas J. Callahan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Fisher & Paykel Healthcare. The logo consists of the words "Fisher & Paykel" in a bold, sans-serif font, with an ampersand (&) between the two names. Below the names is the word "HEALTHCARE" in a slightly smaller, sans-serif font. A horizontal line is placed above the word "HEALTHCARE" and below the company name.
[510(k)] Number:
K983/2
Fisher & Paykel Electronics Limited
25 Carbine Road, Panmure, Auckland, New Zealand
P.O. Box 14-348, Panmure, Auckland, New Zealand
Tel: +64-9-574-0100 Fax: +64-9-574-0158
31 August, 1998
# Fisher & Paykel MR850 Respiratory Humidifier
# PREMARKET NOTIFICATION 510(k) INDICATIONS FOR USE STATEMENT
The Fisher & Paykel Healthcare MR850 Humidifier is a Respiratory Gas Humidifier as per 73 BTT, 21 CFR §868.5450. It is intended to add moisture to and warm breathing gases for administration to a patient.
The MR850 is intended to be used to warm and add humidity to gases delivered to patients requiring mechanical ventilation, positive pressure breathing assistance or general medical gases.
Gases available for medical use do not contain sufficient moisture and may damage or irritate the respiratory tract, or desiccate secretions of patients whose supraglottic airways have been bypassed. Heat is used to increase the water output of the humidifier. Heated breathing tubes are also utilized in order to increase operating efficiency and reduce excessive water and heat loss.
Concurrence of CDRH, Office of Device Evaluation (ODE)
gtame
: Douglas
11-9-98
Prescription Use (Per 21 CFR §801.109)
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.