The Fisher & Paykel MR810 System is intended to provide therapeutic levels of heat and humidity to a patient's inspired respiratory gases, when using a continuous or intermittent ventilator system or a continuous gas flow system. The MR810 System is intended for both non-invasive therapies. Addition of heat and humidity to the supply of cold and dry respiratory gases provided through non-invasive or invasive ventilation is beneficial to prevent drying of the patient airways. The MR810 System is for use for adult and pediatric patients requiring a flow range ≥5 L/min. The MR810 System is designed for use in hospitals, long term care facilities and homes under the prescription of a qualified medical professional.
Device Story
MR810 System is an electrically-powered heated passover humidifier; provides heat and humidity to inspired respiratory gases to prevent airway drying. Inputs: cold/dry gases from ventilator/gas source. Operation: gases pass through humidification chamber; heated/humidified; delivered via breathing circuit. Inspiratory limb includes heater wire powered by humidifier to maintain gas temperature. Used in hospitals, long-term care, and homes; operated by clinicians or patients under prescription. Output: conditioned respiratory gas. Healthcare providers use device to support ventilation therapy; benefits patient by maintaining airway moisture. System includes humidifier unit, breathing circuits, and reusable humidification chambers.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, including system-level performance per ISO 8185 and basic safety/essential performance per IEC 60601-1 and IEC 60601-1-2.
Technological Characteristics
Heated passover humidifier; mains-powered. Components: humidifier unit, breathing circuits, reusable humidification chambers. Features: three temperature settings, two operation modes (heater-wire/non-heater-wire), fault indicators, and automatic shutdown. Complies with ISO 8185, IEC 60601-1, and IEC 60601-1-2.
Indications for Use
Indicated for adult and pediatric patients requiring flow range ≥5 L/min who need heated and humidified inspired respiratory gases during non-invasive or invasive ventilation or continuous gas flow therapy.
Regulatory Classification
Identification
A respiratory gas humidifier is a device that is intended to add moisture to, and sometimes to warm, the breathing gases for administration to a patient. Cascade, gas, heated, and prefilled humidifiers are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of three human profiles facing right, with flowing lines representing connection and support.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 5, 2015
Fisher & Paykel Healthcare Ltd Elizabeth Goldstein Regulatory Affairs Specialist 15 Maurice Paykel Place, East Tamaki Auckland, 2013 New Zealand
Re: K143646 Trade/Device Name: MR810 System Regulation Number: 21 CFR 868.5450 Regulation Name: Respiratory Gas Humidifier Regulatory Class: II Product Code: BTT Dated: December 18, 2014 Received: January 7, 2015
Dear Ms. Goldstein,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin Keith Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control. and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K143646
Device Name MR810 System
### Indications for Use (Describe)
The Fisher & Paykel MR810 System is intended to provide therapeutic levels of heat and humidity to a patient's inspired respiratory gases, when using a continuous or intermittent ventilator system or a continuous gas flow system.
The MR810 System is intended for both non-invasive therapies. Addition of heat and humidity to the supply of cold and dry respiratory gases provided through non-invasive or invasive ventilation is beneficial to prevent drying of the patient airways.
The MR810 System is for use for adult and pediatric patients requiring a flow range ≥5 L/min.
The MR810 System is designed for use in hospitals, long term care facilities and homes under the prescription of a qualified medical professional.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary 5
Image /page/3/Picture/4 description: The image shows the logo for Fisher & Paykel Healthcare. Below the logo is the address "15 Maurice Paykel Place, East Tamaki". The logo is in blue and the address is in black.
P O Box 14 348, Panmure Auckland, New Zealand Telephone: +64 9 574 0100 Facsimile: +64 9 574 0158 Website: www.fphcare.com
Establishment registration number: 9611451
| Contact person/submitter | Elizabeth Goldstein |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| Date prepared | 5 March 2015 |
| Contact details | Address: 15 Maurice Paykel Place<br>East Tamaki<br>Auckland 2013, New Zealand<br>Telephone: +64 9 574 0100<br>Fax: +64 9 574 0158 |
| Trade name | MR810 System |
| Common name | Respiratory gas humidifier |
| Classification name | Respiratory gas humidifier<br>II (21 CFR §868.5450), product code BTT |
| Predicate device | K131957 MR810 System |
| Reason for submission | Expanded indications |
{4}------------------------------------------------
#### Device Description 5.1
The Fisher & Paykel MR810 System is intended to heat and humidify breathing gases and therefore meets the definition of a respiratory gas humidifier in accordance with 21 CFR §868.5450 and the FDA guidance document Heated Humidifier Review Guide.
The MR810 System consists of the following components:
- . MR810 Respiratory Humidifier;
- . Accessories:
- Breathing circuit (e.g. 900MR810 Adult Single Limb Circuit, 900MR810E o Adult Dual Limb Circuit); and
- Humidification chamber (e.g. MR370 reusable humidification chamber). O
The MR810 respiratory humidifier is an electrically-powered heat controller which utilizes a microprocessor with embedded software to control the heating elements.
Unconditioned (i.e. cold, dry) gases are delivered to the MR810 System from a gas source (e.g. ventilator) via the dryline component of the breathing circuit. The unconditioned gases are transported to the humidification chamber where they are heated and humidified.
The heated and humidified gases then travel down the inspiratory limb of the breathing circuit. The inspiratory limb of the 900MR810/900MR810E breathing circuits is electrically heated by means of a heaterwire in order to maintain the temperature of the gases. The heater wire is powered by the MR810 respiratory humidifier via a built-in heaterwire adaptor.
When a return flow of expired gases from the patient to the gas source is required, the expired gases are transported via the expiratory limb of the breathing circuit. In other cases, the expired gases from the patient are exhaled into the environment, as per normal, unassisted breathing.
#### Intended Use and Indications for Use 5.2
#### Intended Use 5.2.1
The Fisher & Paykel Healthcare MR810 System is intended to provide therapeutic levels of heat and humidity to a patient's inspired respiratory gases to prevent drying of the patient's airways.
The intended use of the subject device is identical to the intended use of the predicate device.
# 5.2.2 Indications for Use
The Fisher & Paykel MR810 System is intended to provide therapeutic levels of heat
{5}------------------------------------------------
and humidity to a patient's inspired respiratory gases, when using a continuous or intermittent ventilator system or a continuous gas flow system.
The MR810 System is intended for both non-invasive therapies. Addition of heat and humidity to the supply of cold and dry respiratory gases provided through non- invasive or invasive ventilation is beneficial to prevent drying of the patient airways.
The MR810 System is for use for adult and pediatric patients requiring a flow range ≥ 5 L/min.
The MR810 System is designed for use in hospitals, long term care facilities and homes, under the prescription of a qualified medical professional.
The indications for use of the subject device and the predicate device are identical with the exception of the applications for use, where:
- . The indications for use statement for the subject device. MR810 System, has been revised to specify the intended patient population and operating flow range.
- . The predicate device, MR810 System (K131957), is intended for use for non- invasive applications only.
- . The subject device, MR810 System, is intended for use for both non-invasive and invasive applications.
Performance bench testing has been carried out on the MR810 System which verifies that the system is suitable for use for invasive applications.
#### Technological Characteristics Comparison 5.3
The subject device, MR810 System, is a modification of the predicate device, MR810 System (K131957).
The subject device is identical to the predicate device with respect to:
- Principle device design: The subject device, MR810 System, and the predicate . device, MR810 System (K131957), have identical principle device designs. Both the subject and predicate devices are heated passover humidifiers.
- . Patient-contacting components (i.e. material biocompatibility): The subject device, MR810 System, and the predicate device, MR810 System (K131957), have identical patient-contacting components (i.e. breathing circuits, e.g. 900MR810 and 900MR810E, and humidification chambers, e.g. MR370).
- . Energy source: The subject device, MR810 System, and the predicate device, MR810 System (K131957), use an identical energy source. Both the subject and predicate respiratory humidifier devices are powered via mains electricity and the heated breathing circuits are powered by the respiratory humidifier devices.
- . Operation: The operation of the subject device, MR810 System, and the predicate device, MR810 System (K131957), is identical. Both the subject device and the predicate device have three temperature settings which can be selected by the operator, and two modes of operation which are specific to the type of breathing
{6}------------------------------------------------
circuit used (i.e. heater-wire mode, non heater-wire mode).
- Safety features: The subject device, MR810 System, and the predicate device, . MR810 System (K131957), have identical safety features. Both the subject device and predicate device have a "see manual" fault indicator, heater wire indicator and shut down mechanisms which render the system inoperable during fault conditions.
- Sterility: The subject device, MR810 System, and the predicate device, MR810 . System (K131957), are identical in terms of sterility. No components of the subject device or the predicate device are sold sterile, nor are they intended to be sterilized prior to use.
- Reusable accessories: The subject device, MR810 System, and the predicate device, . MR810 System (K131957), can be used with identical reusable accessories. Both the subject device and predicate device can be used with cleared Fisher & Paykel Healthcare reusable breathing circuits (e.g. 900MR810, 900MR810E) and cleared Fisher & Paykel Healthcare reusable humidification chambers (e.g. MR370).
Minor revisions have been made to sub-components of the MR810 System assembly since submission of K131957. Performance bench testing has been carried out on the modified MR810 System which verifies that the revisions made to the assembly subcomponents does not raise any new questions of safety or effectiveness when compared with the predicate device.
#### Non-Clinical Performance Data 5.4
Bench testing has been carried out on the MR810 System to demonstrate that the differences in indications for use and assembly sub-components does not raise any new questions or safety of effectiveness when compared with the predicate device. MR810 System (K131957).
The following bench testing demonstrates subject and predicate devices are substantially equivalent:
- System-level performance and safety testing in accordance with ISO 8185 Respiratory tract humidifiers for medical use - Particular requirements for respiratory humidification systems.
- . Basic safety and essential performance testing in accordance with IEC 60601-1 Medical electrical equipment – Part 1: General requirements for basic safety and essential performance.
- . Basic safety and essential performance in accordance with IEC 60601-1-2 Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.
{7}------------------------------------------------
#### 5.5 Clinical Performance Data
The subject device, MR810 System, is a modification to the legally marketed predicate device, MR810 System (K131957).
Clinical performance data is not required as bench testing was sufficient to demonstrate substantial equivalence to the predicate device.
#### Conclusions 5.6
The subject device, MR810 System, is a modification to the legally marketed predicate device, MR810 System.
The subject device and the predicate device are identical in terms of the intended use, environment for use, principle device design, patient-contacting components and materials, biocompatibility, energy source, operation, safety features, sterility and reusable accessories.
Since submission of K131957, the indications for use of the MR810 System have been expanded to include invasive applications. In addition, minor revisions have been made to MR810 System assembly sub-components. The differences between the subject device and the predicate device do not constitute significant changes in the materials, design, energy source or other features, as demonstrated through bench testing, and therefore the MR810 System is substantially equivalent to the predicate device, MR810 System (K131957).
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.