The Fisher & Paykel MR810 System is intended to provide therapeutic levels of heat and humidity to a patient's inspired respiratory gases, when using a continuous or intermittent non-invasive ventilator system or a continuous gas flow system. The MR810 System is intended for non-invasive therapies only. Addition of heat and humidity to the supply of cold and dry respiratory gases provided through non-invasive ventilation is beneficial to prevent drying of the patient airways. The MR810 System is designed for use in hospitals, long term care facilities and homes under the prescription of a qualified medical professional.
Device Story
MR810 System conditions respiratory gases by increasing water vapor content and temperature. Input: cold/dry respiratory gases from external source (e.g., ventilator). Process: electrically powered heat controller with microprocessor-based software regulates heating element in humidification chamber and heated inspiratory limb of breathing circuit. Output: heated, humidified gas delivered to patient. Used in hospitals, long-term care, or home settings; operated by healthcare professionals or patients under prescription. Heaterwire adaptor includes ambient and chamber temperature sensors to manage power delivery. LED indicator confirms correct circuit connection. Benefits patient by preventing airway drying during non-invasive ventilation.
Clinical Evidence
No clinical data. Bench testing only. Performance verified against ISO 8185:1997 and ISO 8185:2007 standards, confirming humidity output >10 mg/L and enthalpy <194 kJ/kg dry gas.
Technological Characteristics
Electrically powered respiratory humidifier. Components: heat controller, humidification chamber, heated inspiratory limb (heaterwire). Sensors: ambient and chamber temperature. Connectivity: heaterwire adaptor with LED status indicator. Software: microprocessor-based control of heaterplate and inspiratory limb power. Standards: ISO 8185, IEC 60601-1, IEC 60601-1-2, ISO 5367.
Indications for Use
Indicated for patients requiring non-invasive respiratory therapy via continuous or intermittent ventilator systems or continuous gas flow systems. Intended for use in hospitals, long-term care facilities, and homes under prescription. Contraindicated for invasive therapy (patients with bypassed airways).
Regulatory Classification
Identification
A respiratory gas humidifier is a device that is intended to add moisture to, and sometimes to warm, the breathing gases for administration to a patient. Cascade, gas, heated, and prefilled humidifiers are included in this generic type of device.
{0}------------------------------------------------
## MR810 System
### 510(k) Summary 5
**Fisher&Paykel**
HEALTHCARE
15 Maurice Paykel Place, East Tamaki P O Box 14 348, Panmure Auckland, New Zealand Telephone: +64 9 574 0100 Facsimile: +64 9 574 0158 Website: www.fphcare.com
DEC 0 4 2013
Contact person/submitter
Elizabeth Goldstein
8 November 2013
Date prepared
Contact details
Address: 15 Maurice Paykel Place
East Tamaki Auckland 2013, New Zealand
Telephone: +64 9 574 0100
Fax: +64 9 574 0148
Trade name
Common name
Classification name
Predicate devices
MR810 System
Respiratory gas humidifier
Respiratory gas humidifier II (21 CFR §868.5450), product code BTT
K953392 HC500 Servo-Controlled Heated Respiratory Humidifier K100554 880 Respiratory Humidification System
{1}------------------------------------------------
#### Device Description 5.1
The Fisher & Paykel Healthcare MR810 System is designed to condition respiratory gases for patients by raising the delivered water vapor content (humidity) and temperature of the gases.
The MR810 System consists of the following components:
- . MR810 Respiratory Humidifier
- Accessories: .
- O Breathing circuit (i.e. 900MR810 Adult Single Limb Circuit, 900MR810E Adult Dual Limb Circuit)
Note: The 900MR810 and 900MR810E breathing circuits have identical heated inspiratory limbs, however the 900MR810E also includes an expiratory limb (unheated) and a Y-piece (which connects the inspiratory and expiratory limbs).
- O Humidification chamber (as cleared in K934140 and K913368).
The device consists of an electrically powered heat controller, utilizing a microprocessor with embedded software, to control a heating element which transfer heat to the water in a humidification chamber and control power to the heated inspiratory limb of the breathing circuit.
A dryline tube (part of the breathing circuit) transports respiratory gases from a gas source (e.g. ventilator) to the humidification chamber where the gases are heated and humidified.
The inspiratory limb of the breathing circuit transports the heated and humidified gases from the humidification chamber to the patient. The inspiratory tube of the 900MR810 and 900MR810E breathing circuits is electrically heated by means of a heaterwire placed internally to the tube, which is controlled by the MR810 respiratory humidifier.
The MR810 respiratory humidifier also includes a built-in heaterwire adaptor. The heaterwire adaptor makes an electrical connection with the heated inspiratory limb of the breathing circuit for the purpose of powering the heated limb. The heater wire adaptor incorporates an ambient (room) temperature sensor and a chamber temperature sensor. The heaterwire adaptor also includes an embedded LED to visually indicate that the heated breathing circuit has been connected correctly.
The expiratory limb of the breathing circuit transports expired gas from the patient when a return flow of expired gases from the patient to the gas source is required (the 900MR810E Adult Dual Limb Breathing Circuit should be used as it includes an expiratory limb).
{2}------------------------------------------------
#### 5.2 Intended Use
The Fisher & Paykel MR810 System is intended to provide therapeutic levels of heat and humidity to a patient's inspiratory gases, when using a continuous or 'intermittent non-invasive ventilator system or a continuous gas flow system.
The MR810 System is intended for non-invasive therapies only. Addition of heat and humidity to the supply of cold and dry respiratory gases provided through non-invasive ventilation is beneficial to prevent drying of the patient airways.
The MR810 System is designed for use in hospitals, long term care facilities and homes under the prescription of a qualified medical professional.
# 5.2.1 Indications for Use Comparison
The indications for Use Statements of the subject device, MR810 System, and the predicate device, HC500 Servo-Controlled Heated Respiratory Humidifier (K953392) are identical with the exception of the following:
- · The subject device identifies compatible ventilators by function/intended purpose rather than examples of device types;
- · The subject device is intended for use for non-invasive therapy only, whereas the predicate device is intended for use for both non-invasive of invasive therapies.
Refer to Table 1 for a side-by-side comparison of the Indications for Use Statements of the subject and predicate devices.
| Device feature | MR810 | HC500 (predicate) | Comment |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use | The Fisher & Paykel MR810<br>System is intended to<br>provide therapeutic levels of<br>heat and humidity to a<br>patient's inspired respiratory<br>gases, when using a<br>continuous or intermittent<br>non-invasive ventilator<br>system or a continuous gas<br>flow system. | The intended use of the<br>HC500 is to provide<br>therapeutic levels of heat<br>and humidity to a patient's<br>inspired respiratory gases,<br>when using an artificial<br>ventilation system. This<br>includes use with systems<br>such as portable volume<br>ventilation systems,<br>pressure support ventilation<br>and continuous positive<br>airway pressure (CPAP)<br>devices. | Identical intended use -<br>identification of all<br>artificial ventilator<br>systems that the MR810<br>System is to be used with<br>using the systems'<br>function/intended<br>purpose rather than<br>examples of device<br>types. |
| | | These systems may bypass<br>the patient airway (using an<br>endotracheal tube) | The subject device<br>(MR810'System) is not<br>recommended for<br>invasive therapy (i.e. for<br>patients with bypassed<br>airways). |
Table I Comparison of MR810 and HC500 Indications for Use
{3}------------------------------------------------
Fisher & Paykel Healthcare
| The MR810 System is<br>intended for non-invasive<br>therapy only. | or use mask ventilation. | Identical therapy -<br>addition of the term<br>'non-invasive therapy' to<br>describe 'mask<br>ventilation'. |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Providing heat and<br>humidification to these<br>gases counteracts the<br>effects of bypassing the<br>nose, pharynx and trachea,<br>where this function would<br>not normally be carried out<br>by the body. | The subject device<br>(MR810 System) is not<br>recommended for<br>invasive therapy (i.e. for<br>patients with bypassed<br>airways). | |
| Addition of heat and<br>humidity to the supply of<br>cold and dry respiratory<br>gases provided through non-<br>invasive ventilation is<br>beneficial to prevent drying<br>of the patient airways. | Addition of heat and<br>humidity to the supply of<br>cold and dry respiratory<br>gases provided through<br>mask ventilation is<br>similarly beneficial to<br>prevent drying of the<br>patient airways. | Identical operating<br>principle - use of the<br>term 'non-invasive<br>therapy' to describe<br>'mask ventilation' to be<br>consistent with the<br>therapy statement above<br>('...intended for non-<br>invasive therapy only'). |
| The MR810 System is<br>designed for use in hospitals,<br>long term care facilities and<br>homes under the prescription<br>of a qualified medical<br>professional. | The HC500 is primarily<br>intended for a home-use<br>situation, typically for<br>patients requiring long-term<br>respiratory assistance of a<br>non-critical nature (i.e. not<br>requiring hospital or<br>intensive care levels of<br>attention). | The environment for use<br>of the subject device<br>(MR810 System) has<br>been extended to include<br>hospital and long term<br>care facilities, under the<br>prescription of a<br>qualified medical<br>professional. |
#### 5.3 Technological Characteristics Comparison
# 5.3.1 Comparison to HC500 (K953392)
The subject device, MR810 System, and the predicate device, HC500 Servo-Controlled Heated Respiratory Humidifier (K953392), control the heaterplate temperature identically, however the related user controls (i.e. buttons) have been simplified in the subject device. Refer to Table 2 for a side-by-side comparison of the technological characteristics of the subject device, MR810 System, and the predicate device, HC500 Servo-Controlled Heated Respiratory Humidifier (K953392).
| Device Feature | MR810 | HC500 (predicate) | Comment |
|------------------------------------|-------------------------------------------------------------------------------|-------------------------------------------------------------------------------|-----------|
| Heaterplate temperature<br>control | MR810 controls<br>heaterplate temperature<br>by adjusting delivered<br>power. | HC500 controls<br>heaterplate temperature<br>by adjusting delivered<br>power. | Identical |
| Table 2 Comparison of the MR810 and HC500 Technological Characteristics | | | | | |
|-------------------------------------------------------------------------|--|--|--|--|--|
|-------------------------------------------------------------------------|--|--|--|--|--|
{4}------------------------------------------------
| Device Feature | MR810 | HC500 (predicate) | Comment |
|----------------------------------------------------|--------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Switchable Heaterwire<br>mode | Connection of<br>compatible heated circuit<br>changes operating mode<br>to Heaterwire mode. | Front panel button<br>("Heaterwire" button)<br>switches operating mode<br>between Heaterwire<br>mode and Non-Hose<br>Heated mode | Heaterwire mode is the<br>recommended mode of<br>operation for the MR810.<br>Disconnection of<br>compatible heated circuit<br>changes mode to Non-<br>Hose Heated mode. |
| Patient-end set<br>temperature:<br>Heaterwire mode | Temperature setting:<br>LOW: 28 to 29°C<br>MED: 28 to 30°C<br>HIGH: 31 to 32°C<br>(Flow 5 to 20 L/min) | 32 to 39 °C<br>(dial allows intermediate<br>values to be set).<br>(Flow <30 L/min) | MR810 patient<br>temperature is dependent<br>on both ambient<br>temperature and<br>Temperature Setting |
| | Heaterplate maximum<br>temperature:<br>$\leq$ 70 °C<br>(Software limit) | Heaterplate maximum<br>temperature:<br>$\leq$ 100 °C<br>(Software limit) | HC500 maximum'<br>heaterplate temperature<br>is set by solder pads on<br>the PCB. Optional limits<br>of 50, 80 and 90°C can<br>be set. |
| Heaterplate set<br>temperature | Temperature setting:<br>LOW: 45°C<br>MED: 60°C<br>HIGH: 70°C | Adjustable in the range<br>45 to 80 °C | MR810 controls to a<br>fixed value for the<br>heaterplate temperature<br>dependent on the<br>Temperature Setting |
# 5.3.2 Comparison to MR880 (K100554)
The subject device, MR810 System, and the predicate device, MR880 Respiratory Humidification System (K100554), set the power delivered to the heaterwire identically (i.e. using an ambient sensor). The subject device controls the desired heaterplate setpoint to maintain the chamber output temperature whilst the predicate device controls the chamber output temperature directly.
Refer to Table 3 for a side-by-side comparison of the technological characteristics of the subject device, MR810 System, and the predicate device, MR880 Respiratory Humidification System (K100554).
| Device feature | MR810 | MR880 (predicate) | Comment |
|--------------------------------------------|--------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| Ambient (room)<br>temperature sensor | Over moulded feature on<br>heaterwire adapter | Over moulded feature on<br>heaterwire adapter | Identical. |
| Heaterwire control using<br>ambient sensor | MR810 uses ambient<br>temperature to set the<br>power delivered to the<br>heaterwire | MR880 uses ambient<br>temperature to set the<br>power delivered to the<br>heaterwire. | Identical.<br>Sensor used to set<br>delivered power to the<br>heaterwire contained in<br>inspiratory limb |
| Table 3 Comparison of the MR810 and MR880 Technological Characteristics | |
|-------------------------------------------------------------------------|--|
|-------------------------------------------------------------------------|--|
{5}------------------------------------------------
| Device feature | MR810 | MR880 (predicate) | Comment |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Chamber output control | MR810 controls<br>heaterplate temperature<br>to maintain desired<br>chamber output<br>temperature. | MR880 controls chamber<br>output temperature. | MR880 controls chamber<br>output temperature<br>directly. |
| Breathing circuit - Adult | • Single-heated<br>breathing circuits (i.e.<br>900MR810,<br>900MR810E)<br>Length: 1.5m (60")<br>Diameter: 22mm (0.9") | • Single-heated<br>breathing circuits (i.e.<br>RT241)<br>Length: 1.8m (70.9")<br>Diameter: 10.6mm (0.4") | The MR810 and MR880<br>Systems are both offered<br>with single-heated<br>breathing circuit options<br>only.<br>The breathing circuits<br>used with the MR880<br>System are longer and<br>narrower than those used<br>with the MR810,<br>however both circuits are<br>single-heated and are<br>fitted with standard<br>22mm connectors as per<br>ISO 5356-1. |
#### 5.4 Non-Clinical Tests
The MR810 System is compliant with the same product standards as the previously cleared HC500 Servo-Controlled Heated Respiratory Humidifier (K953392) as follows:
Table 3 Comparison of MR810 and HC500 Standards
| Standard | MR810 | HC500 (Predicate<br>device) | Comment |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|---------------------------------------------------------|-----------|
| ISO 8185: 1997<br>Humidifiers for medical<br>use - General<br>requirements for<br>humidification systems | Compliant with ISO<br>8185:1997 | Compliant with ISO<br>8185:1997 | Identical |
| IEC 60601-1: 1988<br>Medical Electrical<br>Equipment - Part 1:<br>General Requirements<br>for Safety, Amendment<br>1, 1991, Amendment 2,<br>1995 | Compliant with IEC<br>60601-1:1998 +A1:1991<br>+A2:1995 | Compliant with IEC<br>60601-1:1998 +A1:1991<br>+A2:1995 | Identical |
| IEC 60601-1-2: 2001<br>Medical electrical<br>equipment. Part 1:<br>General requirements<br>for safety. 2. Collateral<br>standard:<br>Electromagnetic<br>compatibility -<br>requirements and tests,<br>Amendment 1 | Compliant with IEC<br>60601-1-2:2001 | Compliant with IEC<br>60601-1-2:2001 | Identical |
{6}------------------------------------------------
The MR810 System is compliant with the same product standards as the previously cleared MR880 Respiratory Humidification System (K100554) as follows:
Table 4 Comparison of MR810 and MR880 Standards
| Standard | MR810 | MR880 (predicate) | Comment |
|----------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|-----------|
| ISO 5367:2000<br>Breathing tubes intended<br>for use with anaesthetic<br>apparatus and ventilators | Single-heated breathing<br>circuit accessories (i.e.<br>900MR810,<br>900MR810E) compliant<br>with ISO 5367:2000 | Single-heated breathing<br>circuit accessories (i.e.<br>RT241) compliant with<br>ISO 5367:2000 | Identical |
Compliance of the subject device, MR810 System, and the:
- · Predicate device HC500 Servo-Controlled Heated Respiratory Humidifier (K953392); and
- · Single-heated breathing circuit accessories provided with the MR880 Respiratory Humidification System (K100554)
to the same device standards supports substantial equivalence of these products.
In addition, testing to ISO 8185:2007, the particular standard for humidification systems, supports performance of the subject device in accordance with the intended use and substantial equivalence to the predicate device HC500 Servo-Controlled Heated Respiratory Humidifier (K953392).
As summarized in Table 5 below, the humidity output of both the subject device, MR810 System, and the predicate device, HC500 Servo-Controlled Heated Respiratory Humidifier (K953392), is >10 mg/L and the enthalpy is of both the subject and predicate devices is <194 kJ/kg dry gas and therefore the performance of the devices is substantially equivalent.
Note: ISO 8185 requires that the predicate device also meet additional performance requirements for humidification systems intended for use with bypassed airways. These additional requirements are not applicable to the subject device as it is intended for noninvasive use only.
| Device feature | Requirement | MR810 | HC500 (predicate) | Comment |
|----------------------------------------------------------------------------|-------------------------------------------------------------------------------------|-------------------------------------|-------------------------------------|-----------|
| Humidity ·<br>performance,<br>Heaterwire mode -<br>for non-invasive<br>use | ≥ 10 mg/L<br>(as required by ISO<br>8185:1997 for all<br>humidification<br>systems) | > 10 mg/L<br>Flow: 5 to 60<br>L/min | > 10 mg/L<br>Flow: 5 to 60<br>L/min | Identical |
Table 5 Comparison of MR810 and HC500 humidification system performance as per ISQ 8185:1997
{7}------------------------------------------------
| Device feature | Requirement | MR810 | HC500 (predicate) | Comment |
|--------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Humidity<br>performance,<br>Heaterwire mode-<br>for invasive use<br>(i.e. patients with<br>bypassed airways) | ≥ 33 mg/L<br>(as required by ISO<br>8185:1997) | Not applicable -<br>the MR810 System<br>is intended for non-<br>invasive use only | ≥ 33 mg/L | The MR810 not<br>intended for use<br>with bypassed<br>airways and<br>therefore the<br>humidity<br>performance<br>requirements (as<br>per ISO 8185) are<br>different |
| Enthalpy | < 194 kJ/kg dry gas<br>(as required by ISO<br>8185:1997 for all<br>humidification<br>systems) | < 194 kJ/kg dry gas | < 194 kJ/kg dry gas | Identical |
#### 5.5 Clinical Tests
Not applicable - no clinical testing was performed with respect to the MR810 System.
#### 5.6 Conclusion
The comparison of the intended use and temperature control system demonstrate that the MR810 System is substantially equivalent to the predicate device, Fisher & Paykel Healthcare HC500 Servo-Controlled Heated Respiratory Humidifier (K953392), and the comparison of the temperature sensor and hardware control demonstrate that the MR810 System is substantially equivalent to the predicate device, Fisher & Paykel Healthcare 880 Respiratory Humidification System (K100554). In addition, bench testing demonstrates that the MR810 System conforms to the particular standard for humidification systems, ISO 8185, supporting performance in accordance with the intended use.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo features a stylized eagle with three swooping lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 4, 2013
Fisher & Paykel Healthcare Ms. Elizabeth Goldstein Regulatory Affairs Specialist 15 Maurice Paykel Place, East Tamaki P.O. Box 14 348, Panmure Auckland, New Zealand
Re: K131957
Trade/Device Name: MR810 System Regulation Number: 21 CFR 868.5450 Regulation Name: Respiratory Gas Humidifier Regulatory Class: II Product Code: BTT Dated: October 31, 2013 Received: November 4, 2013
# Dear Ms. Goldstein:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{9}------------------------------------------------
# Page 2 - Ms. Golstein
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Image /page/9/Picture/8 description: The image shows the name "Erin DKeith" in a stylized font. The letters "Erin" are in a bold, sans-serif font, while the letters "DKeith" are in a more decorative font with outlines and patterns. The "D" and "K" are particularly stylized, with the "D" having a circular pattern inside and the "K" having a geometric design.
Erin Keith M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{10}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
510(k) Number (if known) K131957
Device Name MR810 System
### Indications for Use (Describe)
The Fisher & Paykel MR810 System is intended to provide therapeutic levels of heat and humidity to a patient's inspired respired respired respired respired respired respired gases, when using a continuous or internittent non-invasive ventilator system or a continuous gas flow system.
The MR810 System is intended for non-invasive therapies only. Addition of heat and humidity to the supply of cold and dry respiratory gases provided through non-invasive ventilation is beneficial to prevent drying of the patient airways.
The MR810 System is designed for use in hospitals, long term care facilities and homes under the prescription of a qualified medical professional.
Type of Use (Select one or both, as applicable)
[x] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
:: :: FOR FDA USE ONLY . . . . . :・・・・・・ : : : . . . . Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) Digitally signed by James J. Lee DN c=US, o=U.S. Government, ou=HHS, ou=FDA, ou=People, cn=James J. Lee, 0:9.2342:19200300.100.1.1=2000954859
Date: 2013.11.26 23:26:32 -05'00'
FORM FDA 3881 (9/13)
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.