K170234 · Zhanjiang Star Enterprise Co., Ltd. · BTR · Jan 17, 2018 · Anesthesiology
Device Facts
Record ID
K170234
Device Name
Endotracheal tube; Reinforced Endotracheal tube
Applicant
Zhanjiang Star Enterprise Co., Ltd.
Product Code
BTR · Anesthesiology
Decision Date
Jan 17, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5730
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Endotracheal tube is intended for oral or nasal intubation and for airway management. The Reinforced Endotracheal tube is intended for airway management by oral or nasal intubation of the trachea during anesthesia. It is a device inserted into the trachea through the mouth or nose to facilitate breathing and it may be used where the patient's neck is likely to be moved of flexed or the patient is in the prone position so that a non-reinforced tracheal tube might become kinked.
Device Story
Disposable endotracheal tubes (ETTs) and reinforced ETTs facilitate airway management and artificial ventilation. Standard ETTs consist of a PVC tube, connector, and optional cuff/inflation system; reinforced ETTs include a 304 stainless steel wire spiral for kink resistance. Used in ORs and ICUs by clinicians for oral or nasal intubation. Devices are inserted into the trachea to establish a patent airway. Cuffed variants provide a tracheal seal to route gases. Reinforced variants are used when patient positioning (e.g., prone, head/neck surgery) risks tube kinking. Devices are supplied sterile, single-use, and intended for short-term use (up to 30 days).
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including biocompatibility (ISO 10993-1), sterilization validation (ISO 11135-1), and performance testing (ISO 5361:2012) covering dimensions, cuff integrity, kink resistance, and mechanical properties.
Technological Characteristics
Materials: Polyvinyl chloride (PVC), 304 stainless steel (reinforced models). Sensing/Actuation: Passive airway management. Dimensions: 2.0mm-10.0mm ID (standard), 3.0mm-9.0mm ID (reinforced). Connectivity: None. Sterilization: Ethylene Oxide. Standards: ISO 5361:2012, ISO 10993-1, ISO 11135-1.
Indications for Use
Indicated for patients requiring airway management via oral or nasal intubation during anesthesia, artificial ventilation, or assisted breathing. Reinforced version indicated for patients where neck flexion or prone positioning risks tube kinking.
Regulatory Classification
Identification
A tracheal tube is a device inserted into a patient's trachea via the nose or mouth and used to maintain an open airway.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 17, 2018
Zhanjiang Star Enterprise Co., Ltd % Jessie You Official Correspondent Shenzhen Joyantech Consulting Co., Ltd Room 1122, No.55 Shizhou Middle Road Nanshan District Shenzhen, GD755 TW
Re: K170234
Trade/Device Name: Endotracheal Tube; Reinforced Endotracheal Tube Regulation Number: 21 CFR 868.5730 Regulation Name: Tracheal Tube Regulatory Class: Class II Product Code: BTR Dated: November 15, 2017 Received: November 30, 2017
Dear Jessie You:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{2}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Tara A. Ryan -S 2018.01.17 06:29:26 -05'00' for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K170234
Device Name
Endotracheal tube / Reinforced Endotracheal tube
Indications for Use (Describe)
The Endotracheal tube is intended for oral or nasal intubation and for airway management.
The Reinforced Endotracheal tube is intended for airway management by oral or nasal intubation of the trachea during anesthesia. It is a device inserted into the trachea through the mouth or nose to facilitate breathing and it may be used where the patient's neck is likely to be moved of flexed or the prone position so that a non-reinforced tracheal tube might become kinked.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;"> <span style="font-family: Symbol;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span style="font-size: 10pt;"> <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Version: A/0
# VOL_006:001_510(k) Summary
## 1. Submission Sponsor
| Applicant Name | Zhanjiang Star Enterprise Co., Ltd |
|----------------|------------------------------------------------------------|
| Address | NO.1, West Jinhua Rd. Zhanjiang, Guangdong, P.<br>R. CHINA |
| Post Code | 524094 |
| Phone No. | 86-759-2701619 |
| Fax No. | 86-759-2706838 |
| Contact Person | Rongsheng Tang |
| Email | Tangrs@starcompany.com.cn |
| Date Prepared | 01/19/2017 |
## 2. Submission correspondent
| Name | Shenzhen Joyantech Consulting Co., Ltd |
|----------------|----------------------------------------------------------------------------------------------------|
| Address | Room 1122, No.55 Shizhou Middle Road, Nanshan<br>District,<br>Shenzhen,<br>Guangdong,<br>P.R.China |
| Image | Image: Company Logo |
| Post Code | 518000 |
| Phone No. | 86-755-86069197 |
| Contact Person | Jessie You; Field Fu |
| Email | Jessie@cefda.com; cefda13485@163.com |
## 3. Devices Identification
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
| Trade name | Endotracheal tube |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common name | Tracheal tube |
| Classification | II |
| Classification name | Tracheal tube |
| Regulation number | 21 CFR 868.5730 |
| Product code | BTR |
| 510(k) review panel | Anesthesiology |
| Performance standards | The performance was evaluated in accordance with<br>ISO 5361: 2012.<br>Biocompatibility tests were done in conformance<br>with relevant requirements of ISO10993. |
{5}------------------------------------------------
Version: A/0
| Trade name | Reinforced Endotracheal tube |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common name | Tracheal tube |
| Classification | II |
| Classification name | Tracheal tube |
| Regulation number | 21 CFR 868.5730 |
| Product code | BTR |
| 510(k) review panel | Anesthesiology |
| Performance standards | The performance was evaluated in accordance with<br>ISO 5361: 2012.<br>Biocompatibility tests were done in conformance<br>with relevant requirements of ISO10993. |
#### 4. Legally Marketed Predicate Devices
| Trade name | Disposable Tracheal Tube |
|-------------------|----------------------------------------------------|
| Regulation number | 21 CFR 868.5730 |
| Regulation name | Tracheal tube |
| Regulatory class | II |
| 510 (k) number | K140228 |
| Product code | BTR |
| Manufacturer | Suzhou Weikang Medical Apparatus Co., Ltd |
| Trade name | BEVERTM Reinforced Endotracheal Tube |
| Regulation number | 21 CFR 868.5730 |
| Regulation name | Tracheal tube |
| Regulatory class | II |
| 510 (k) number | K111406 |
| Product code | BTR |
| Manufacturer | Hangzhou Bever Medical Devices Company,<br>Limited |
### 5. Device Description
The Endotracheal tubes are used for oral or nasal intubation and for airway management, which are mainly for patients that require repeated anesthesia, artificial ventilation and assisted breathing. The Endotracheal tube is available in cuffed and uncuffed variants. The cuff is intended to provide a seal against the trachea, ensuring that inspiratory and expiratory gasses are routed through the tube and not allowed to escape to the patient's upper airway. The structure of Cuffed Endotracheal tube contains a main tube (patient end, machine end and murphy eye), cuff, and inflating system (including inflating tube, valve and pilot balloon). The uncuffed Endotracheal tube contains main tube (patient end, shaft and machine end). All variants have a Murphy eye. The products are primarily made from polyvinyl chloride (PVC) materials. The Endotracheal tubes are available in a number of sizes: the nominal inside
{6}------------------------------------------------
diameter of uncuffed Endotracheal tubes are from 2.0mm to 10.0mm; the nominal inside diameter of cuffed Endotracheal tubes are from 2.5mm to 10.0mm. The Endotracheal tubes are disposable and are supplied sterile. The contact time should be no more than 30 days.
The reinforced Endotracheal tubes are a tracheal tube with additional metal wire spiral reinforcement to provide kink-resistance. The products are always used during operations where a high degree of flexibility is required from the tube. The reinforced Endotracheal tubes are also available in cuffed and uncuffed variants, and nominal inside diameter of both the uncuffed and cuffed reinforced Endotracheal tubes are from 3.0mm to 9.0mm. The only difference between reinforced Endotracheal tubes and Endotracheal tubes is the metal wire spiral inside the tube, which is made by 304 stainless steel and used for reinforcement that maintains the patency of the lumen.
## 6. Indications for Use Statement
The Endotracheal tube is intended for oral or nasal intubation and for airway management.
The Reinforced Endotracheal tube is intended for airway management by oral or nasal intubation of the trachea during anesthesia. It is a device inserted into the trachea through the mouth or nose to facilitate breathing and it may be used where the patient's neck is likely to be moved of flexed or the patient is in the prone position so that a non-reinforced tracheal tube might become kinked.
## 7. Substantial Equivalence Discussion
| Table 1: Substantial equivalence comparison between Endotracheal tube and |
|---------------------------------------------------------------------------|
| Disposable Tracheal Tube. |
| Items | The proposed device-<br>Endotracheal tube | The predicate device-<br>Disposable Tracheal Tube<br>K140228 |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification | II<br>21 CFR 868.5730<br>BTR | II<br>21 CFR 868.5730<br>BTR |
| Indications for use | The Endotracheal tube is<br>intended for oral or nasal<br>intubation and for airway<br>management. | Disposable endotracheal tube is<br>intended for oral or nasal<br>intubation and for airway<br>management. |
| Structure and<br>composition | The uncuffed tube:<br>1. Main tube<br>2. Connector<br>The cuffed tube:<br>1. Main tube<br>2. Cuff | The tracheal tube without a cuff:<br>1. Tubular body<br>2. Standard connector<br>The cuffed tracheal tube:<br>1. Tubular body<br>2. Cuff |
| | 3. Inflating system (contains<br>inflating tube, valve and<br>pilot balloon) | 3. One-way valve<br>4. Pilot balloon<br>5. Inflating tube<br>6. Standard connector |
| Specifications | 1. Cuffed Endotracheal tubes: 2.5mm-10.0mm;<br>2. Uncuffed Endotracheal<br>tubes: 2.0mm-10.0mm | 1. Cuffed Endotracheal tubes: 2.0mm-10.0mm;<br>2. Uncuffed Endotracheal<br>tubes: 2.0mm-10.0mm |
| Materials | The products are primarily<br>made from polyvinyl chloride<br>(PVC) materials. | The disposable tracheal tubes<br>are made from polyvinyl chloride<br>material (PVC). |
| Mechanism of<br>action | The Endotracheal tube<br>intubation is the first aid basic<br>operation of severe surgical<br>and ICU departments. The<br>basic principle is inserting the<br>Endotracheal tube through the<br>mouth, throat into the patient's<br>trachea and to build artificial<br>respiration channel to ensure<br>patient breathes smoothly. The<br>single use endotracheal tube<br>has a smooth surface, and<br>there is little mucosal injury for<br>the trachea. | As the device functions in airway<br>management/gas transport for<br>anesthesia or resuscitation, it is<br>required to be flexible and<br>resistant to kinking. The airway<br>management of patients is<br>achieved by inserting the<br>endotracheal tube into trachea<br>through oral or nasal intubation. |
| Biocompatibility | The proposed device is<br>considered as external<br>communicating device in<br>contact with tissue less than 30<br>days, and evaluation was<br>conducted in accordance with<br>#G95-1 and ISO 10993-1,<br>which contains:<br>1. In Vitro Cytotoxicity<br>2. Skin Sensitization<br>3. Oral mucosa Irritation<br>4. Oral mucosa Prolonged<br>Irritation<br>5. Genotoxicity Test<br>6. Implantation Test<br>7. Acute Systemic Toxicity<br>Test | The device is considered as<br>external communicating device<br>in contact with tissue with<br>cumulative contact up to 24<br>hours, and evaluation was<br>conducted in accordance with<br>#G95-1 and ISO 10993-1. The<br>battery of testing included the<br>following tests:<br>1. Cytotoxicity<br>2. Sensitization<br>3. Irritation |
| | 8. Subchronic Systemic<br>Toxicity Test | |
| | 9. Endotoxin Test | |
| | 10. Pyrogen | |
| | 11. EO and ECH Residual Test | |
| | 12. DEHP Residual Test | |
| Sterilization | Ethylene Oxide | Ethylene Oxide |
| Performance | The performance of<br>the endotracheal tubes was<br>evaluated in accordance with<br>ISO 5361: 2012. The testing<br>items contain: | Essential performance of<br>the shaft of the tube are<br>evaluated according to the<br>following standard respectively:<br>-ISO 5361: 2012<br>The testing items contain: |
| | 1. Size designation; | 1. Size designation; |
| | 2. Dimensions (including basic<br>dimensions, inside diameter,<br>outside diameter, tracheal<br>tubes connectors, machine<br>end, inside diameter of the<br>machine end, patient end,<br>opening at the patient end); | 2. Dimensions (including basic<br>dimensions, inside diameter,<br>outside diameter, tracheal tubes<br>connectors, machine end, inside<br>diameter of the machine end,<br>patient end, opening at the<br>patient end); |
| | 3. Tracheal tube bevel<br>(including angle of bevel, free<br>from sharp edges); | 3. Tracheal tube bevel (including<br>angle of bevel, free from sharp<br>edges); |
| | 4. Tracheal tubes cuffs<br>(including tube collapse, free of<br>sharp edges); | 4. Tracheal tubes cuffs<br>(including tube collapse, free of<br>sharp edges); |
| | 6. Inflating system for cuffs<br>(including outside diameter,<br>angle between the inflating<br>tube and at the tracheal tube at<br>the point of separation, pilot<br>balloon); | 6. Inflating system for cuffs<br>(including outside diameter,<br>angle between the inflating tube<br>and at the tracheal tube at the<br>point of separation, pilot<br>balloon); |
| | 7. Curvature of the tube<br>(including radius of curvature,<br>maintain intended shape); | 7. Curvature of the tube<br>(including radius of curvature,<br>maintain intended shape); |
| | 8. Kink resistance; | 8. Kink resistance; |
| | 9. Additional requirement for<br>Murphy eye; | 9. Additional requirement for<br>Murphy eye; |
| | 10. Sterility assurance; | 10. Sterility assurance; |
| | 11. Marking.<br>The results showed that the<br>endotracheal tubes meet the | 11. Marking.<br>-ISO 11990-1: 2011 |
| requirement of ISO 5361. | The testing results demonstrate that the proposed devices meet the requirements of the standards listed above. | |
{7}------------------------------------------------
{8}------------------------------------------------
Version: A/0
{9}------------------------------------------------
| Table 2: Substantial equivalence comparison between Reinforced Endotracheal |
|-----------------------------------------------------------------------------|
| tube and BEVERTM Reinforced Endotracheal Tube. |
| Items | The proposed device-<br>Reinforced Endotracheal<br>tube | The predicate device-<br>BEVER™ Reinforced<br>Endotracheal Tube<br>K111406 | |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------|
| Classification | II<br>21 CFR 868.5730<br>BTR | II<br>21 CFR 868.5730<br>BTR | |
| Indications for<br>use | The Reinforced Endotracheal<br>tube is intended for airway<br>management by oral or nasal<br>intubation of the trachea during<br>anesthesia. It is a device<br>inserted into the trachea<br>through the mouth or nose to<br>facilitate breathing and it may<br>be used where the patient's<br>neck is likely to be moved of<br>flexed or the patient is in the<br>prone position so that a<br>non-reinforced tracheal tube<br>might become kinked. | The BEVER™ Reinforced<br>Endotracheal Tube is indicated<br>for airway management by oral<br>or nasal intubation of the trachea<br>during anesthesia. The product<br>may be used where the patient's<br>neck is likely to be moved or<br>flexed or the patient is in the<br>prone position so that a<br>non-reinforced tracheal tube<br>might become kinked. | |
| Structure and<br>composition | The cuffed tube:<br>1. Main tube<br>2. Cuff<br>3. Inflating system (contains<br>inflating tube, valve and<br>pilot balloon)<br>4. Reinforced stainless steel<br>The uncuffed tube:<br>1. Main tube<br>2. Reinforced stainless steel | The cuffed tube:<br>1. Main tube<br>2. Cuff<br>3. Inflating system (including<br>inflating tube, valve and pilot<br>balloon)<br>4. Steel reinforcement<br>5. Standard connector<br>The uncuffed tube:<br>1. Main tube<br>2. Steel reinforcement<br>3. Standard connector | |
| Specifications | Reinforced cuffed/uncuffed<br>Endotracheal tubes: 3.0mm- | 1. Cuffed BEVER™ Reinforced<br>Endotracheal tubes: | |
| | 9.0mm | 3.0mm-10.0mm; | |
| | | 2. Uncuffed<br>BEVERTM Reinforced | |
| | | Endotracheal tubes: | |
| | | 2.0mm-7.0mm | |
| Materials | The products are primarily<br>made from polyvinyl chloride<br>(PVC) materials, and the<br>reinforced stainless steel is<br>made by stainless steel. | Tube is made of PVC material<br>and stainless steel wire. | |
| Mechanism of<br>action | The Reinforced Endotracheal<br>tube intubation is the first aid<br>basic operation of severe<br>surgical and ICU departments.<br>The basic principle is inserting<br>the tube through the mouth,<br>throat into the patient's trachea<br>and to build artificial respiration<br>channel to ensure patient<br>breathes smoothly. The tube<br>has a smooth surface, and<br>there is little mucosal injury for<br>the trachea. The additional<br>metal wire spiral reinforcement<br>to provide kink-resistance. | The BEVERTM Reinforced<br>Endotracheal Tube is an<br>Endotracheal tube with<br>additional metal wire spiral<br>reinforcement to provide<br>kink-resistance. The product is<br>typically used during operations<br>where a high degree of flexibility<br>is required from the tube, for<br>instance prone position, head<br>and neck surgery, and oral<br>surgery, which is indicated for<br>airway management by oral or<br>nasal intubation of the trachea<br>during anesthesia. | |
| Biocompatibilit<br>y | The proposed device is<br>considered as external<br>communicating device in<br>contact with tissue less than 30<br>days, and evaluation was<br>conducted in accordance with<br>#G95-1 and ISO 10993-1,<br>which contains:<br>1. In Vitro Cytotoxicity<br>2. Skin Sensitization<br>3. Oral mucosa Irritation<br>4. Oral mucosa Prolonged<br>Irritation<br>5. Genotoxicity Test<br>6. Implantation Test<br>7. Acute Systemic Toxicity | Biocompatibility testing was<br>performed based on ISO 10993<br>standards. The subject device<br>passed the following<br>biocompatibility testing:<br>1. Cytotoxicity<br>2. Sensitization<br>3. Irritation<br>4. Genotoxicity<br>5. Implantation | |
| | 8. Subchronic Systemic | | |
| | Toxicity Test | | |
| | 9. Endotoxin Test | | |
| | 10. Pyrogen | | |
| | 11. EO and ECH Residual Test | | |
| | 12. DEHP Residual Test | | |
| Sterilization | Ethylene Oxide | | Ethylene Oxide |
| Performance | The performance of | | The dimension, design, material, |
| | the endotracheal tubes was | | sterility, packaging and labeling |
| | evaluated in accordance with | | of BEVERTM Reinforced |
| | ISO 5361: 2012. The testing | | Endotracheal tubes are |
| | items contain: | | conformed with ISO 5361: 1999. |
| | 1. Size designation; | | The testing items contain: |
| | 2. Dimensions (including basic | | 1. Size designation; |
| | dimensions, inside diameter, | | 2. Dimensions (including basic |
| | outside diameter, tracheal | | dimensions, inside diameter, |
| | tubes connectors, machine | | outside diameter, tracheal tubes |
| | end, inside diameter of the | | connectors, machine end, inside |
| | machine end, patient end, | | diameter of the machine end, |
| | opening at the patient end); | | patient end, opening at the |
| | 3. Tracheal tube bevel | | patient end); |
| | (including angle of bevel, free | | 3. Tracheal tube bevel (including |
| | from sharp edges); | | angle of bevel, free from sharp |
| | 4. Tracheal tubes cuffs | | edges); |
| | (including tube collapse, free of | | 4. Tracheal tubes cuffs |
| | sharp edges); | | (including tube collapse, free of |
| | 6. Inflating system for cuffs |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.