The Sotair™ device is intended for use with a manual resuscitator with air-tight connections, in non-breathing patients that require flow controlled ventilation using ambient air or a supplemental oxygen source. The Sotair device is a single use, disposable device added to a manual resuscitator and can be used for in-hospital, emergency, and transport care. The adult Sotair device comprises a flow-limiting valve that limits the inspiratory flow enabling providers to ventilate approximately 55 liters-per-minute (LPM). The flow-limiting valve is intended to minimize gastric inflation during manual ventilation. The Sotair device can be disabled by removing the manual resuscitator to its conventional operation. The Sotair device is intended for adult use only.
Device Story
Sotair is a single-use, disposable flow-limiting valve accessory for manual resuscitators; connects via standard ISO 5356-1 15/22 mm interfaces. Device operates mechanically; when provider squeezes resuscitator bag with excessive force, internal flapper valve with blade design activates to limit inspiratory flow to approximately 55 LPM. Used in-hospital, emergency, and transport care by clinicians. Minimizes gastric inflation during manual ventilation; can be disabled by removal. Benefits patient by reducing risk of gastric insufflation during resuscitation. Performance validated via mechanical testing and human factors/usability studies.
Clinical Evidence
No clinical data. Evidence consists of bench testing and human factors/usability validation. Mechanical testing included auditory feedback, leakage, resistance, drop testing, water immersion, and valve functionality. Biocompatibility testing performed per ISO 10993 and ISO 18562. Summative usability study confirmed device meets criteria for intended users and environments.
Technological Characteristics
Flow-limiting valve; mechanical flapper valve with blade design. Materials: medical-grade polycarbonate. Dimensions: compatible with ISO 5356-1:2015 15/22 mm connectors. Energy source: manual (operator-powered). Connectivity: none (standalone accessory). Sterilization: not specified. Standards: ISO 10993, ISO 18562, ISO 14971, ISO 5356-1, ISO 10651-4.
Indications for Use
Indicated for non-breathing adult patients requiring flow-controlled ventilation via manual resuscitator in hospital, emergency, or transport settings.
Regulatory Classification
Identification
A manual emergency ventilator is a device, usually incorporating a bag and valve, intended to provide emergency respiratory support by means of a face mask or a tube inserted into a patient's airway.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Safe BVM Corporation % Sara Toyloy President Fabrica Consulting, LLC 73 Lincoln Drive Sausalito, California 94965
Re: K212905
Trade/Device Name: Sotair™ Device Regulation Number: 21 CFR 868.5915 Regulation Name: Manual Emergency Ventilator Regulatory Class: Class II Product Code: BTM Dated: July 14, 2022 Received: July 18, 2022
Dear Sara Toyloy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Brandon Blakely, Ph.D. Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory, and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212905
Device Name Sotair™ Device
#### Indications for Use (Describe)
The Sotair™ device is intended for use with a manual resuscitator with air-tight connections, in non-breathing patients that require flow controlled ventilation using ambient air or a supplemental oxygen source. The Sotair device is a single use, disposable device added to a manual resuscitator and can be used for in-hospital, emergency, and transport care. The adult Sotair device comprises a flow-limiting valve that limits the inspiratory flow enabling providers to ventilate approximately 55 liters-per-minute (LPM). The flow-limiting valve is intended to minimize gastric inflation during manual ventilation. The Sotair device can be disabled by removing the manual resuscitator to its conventional operation. The Sotair device is intended for adult use only.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 5. 510(k) Summary
| 1. Submitter: | SafeBVM, Corporation<br>Chesterfield, MO 63017 USA |
|------------------|---------------------------------------------------------------------------------------------------------|
| Phone Number: | +1 (314) 973-6606 |
| Sponsor Contact: | Haris Shekhani, MD<br>President, SafeBVM<br>Email: haris@safebvm.com<br>Phone: +1 (314) 973-6606 |
| Correspondent: | Sara Toyloy<br>Fabrica Consulting, LLC<br>Email: sara@fabricaconsulting.com<br>Phone: +1 (415) 297-8826 |
| Date Prepared: | August 17, 2022 |
- 2. Device:
| Proprietary Name: | Sotair™ Device |
|-------------------|-----------------------------|
| Common Name: | Flow-Limiting Valve |
| Regulation: | 21 CFR § 868.5915 |
| Classification: | Manual emergency ventilator |
| Device Class: | Class 2 |
| Device Code: | BTM |
| Review Panel: | Anesthesiology |
### 3. Predicate Device:
Smart Bag MO (K021328), O-Two Systems International, Inc. No reference devices were used in this submission
## 4. Device Description:
The Sotair™ device is a universal safety accessory device that is compatible with all manual resuscitators and airway interfaces that have standard ISO 5356-1:2015 15/22 mm connections. The Sotair device's flow-limiting function has been tested with the following bag resuscitators:
| Model Number | Description/Brand Name | Manufacturer |
|---------------------------|---------------------------------------------------|--------------|
| 520 211 00X, 520 211 00XX | Ambu Spur II, adult | Ambu |
| 2442-BVMPAD | Curaplex Resuscitation Bags, adult premium<br>BVM | Curaplex |
{4}------------------------------------------------
| Model Number | Description/Brand Name | Manufacturer |
|-----------------------|-----------------------------------------------------------------------|----------------------------|
| 301-558XEA | Curaplex® VentiSure2™ BVM Manual Resuscitator, Adult | Curaplex |
| 04-2K80XX | AirLife Adult Manual Resuscitator | Vyaire Medical/Carefusion |
| AF5140MBX, AF5140MBXX | Sunmed/ventlab AirFlow Standard, Size 5 | Sunmed/Ventlab |
| 10564XX, 10585XX | Mercury Medical CPR 2 - Small Adult Bag | Mercury Medical |
| 1056XXX, 10582XX | Adult CPR-2 Bag | Mercury Medical |
| 850XX | 1st Response Adult Manual Resuscitator with Oxygen Reservoir Bag | Smith's Medical |
| 8451X1 | The BAG II Resuscitator Adult w/ Mask #5 | Laerdal |
| 562013 | Adult BagEasy Resuscitator w/ Mask | Westmed |
| L670-0X0XX | Disposable Bag Mask Resuscitator | Allied Healthcare Products |
| 53710X | RUSCH, Manual Pulmonary Resuscitator with oxygen reservoir bag, Adult | Teleflex |
The Sotair device is not compatible with manual resuscitators that have a built-in flow limiting device.
## 5. Indications for Use:
The Sotair device is intended for use with a manual resuscitator with air-tight connections, in nonbreathing patients that require flow-controlled ventilation using ambient air or a supplemental oxygen source. The Sotair device is a single use, disposable device added to a manual resuscitator and can be used for in-hospital, emergency, and transport care. The adult Sotair device comprises a flow-limiting valve that limits the inspiratory flow enabling user to ventilate approximately 55 liters-per-minute (LPM). The flow-limiting valve is intended to minimize gastric inflation during manual yentilation. The Sotair device can be disabled by removing the device, thereby returning the manual resuscitator to its conventional operation. The Sotair device is for adult use only.
### 6. Comparison of Indications for Use and Technological Characteristics with the Predicate Device:
### Comparison of Indications for Use:
The Sotair device is a flow-limiting valve with the primary function of minimizing gastric inflation during manual ventilation. This intended use statement is the same as the predicate device with regard to the Smart Bag MO's flow-limiting-valve which provides the same primary function of minimizing gastric inflation during manual ventilation. Small differences exist between the two intended uses wherein the Sotair device has a statement indicating that "the device is intended for
{5}------------------------------------------------
use in conjunction with a manual resuscitator which does not have a flow-limiting valve" whereas the predicate Smart Bag MO indicates that "the Smart Bag MO includes a flow-limiting valve that limits the inspiratory flow". Additional differences exist between the two devices including the maximum ventilation rate of ~55 LPM and the targeted patient population (adult use only) for Sotair, however, these differences do not impact the substantial equivalence of the Sotair device as compared to the predicate device.
### Comparison of Technological Characteristics:
The Sotair device (subject device) differs from the predicate device with regard to the actual mechanism that is used to limit the flow rate. Specifically, the Sotair device uses a flapper valve with a blade design whereas the Smart Bag MO (predicate device) uses a t-shaped piston with spring design. While the actual morphology differs between the two devices, the mechanism of action is the same for both devices, i.e., the flow-limiting valve becomes activated when the provider squeezes the bag forcefully (i.e., too hard). The Sotair device and the Smart Bag MO mechanisms are both actuated with excessive force and enable the provider to limit inspiratory flows delivered to the patient. When activated, the flow-limiting valve mechanisms of both devices move within their respective housing to limit flow. Both valves are brightly colored to aid in visualization, and their plastic components consist of similar medical-grade polycarbonate material.
## 7. Performance Data:
To demonstrate that the differences in technological characteristics specific to the flow-limiting valve mechanism do not impact the substantial equivalence of the Sotair device (subject device) to the Smart Bag MO device (predicate device), performance testing was completed to support the determination of substantial equivalence as follows:
### Biocompatibility Testing:
Biological safety testing on the final, finished Sotair device was conducted in accordance with ISO 10993 and ISO 18562. The Sotair device has contact with the humidified breathing gas pathway for a limited duration (<24 hours) according to ISO 10993-1 and ISO 18562-1. While the actual clinical use is typically brief, under extreme conditions, the device could be used for up to 6 hours. The Sotair device was evaluated for the following endpoints through testing: Cytotoxicity, Sensitization, Intracutaneous Reactivity, Acute Systemic Toxicity, Material-mediated Pyrogenicity, Volatile Organic Compound (VOC) emissions, Particulates, and Leachable Substances in condensate via ISO 10993-18. Toxicological Risk Assessments were conducted according to ISO 10993-17 and ISO 18562-1. Overall, the testing met the requirements of ISO 10993 and ISO 18562.
### Mechanical Testing:
The following mechanical tests were completed at Time 2-years (accelerated) to support the substantial equivalence determination:
- Auditory Feedback, Bounce Back and Valve Penetration Testing ●
- Forward and Backward Air Leakage and Expiratory and Inspiratory Resistance Testing ●
{6}------------------------------------------------
- . Drop Testing
- . Water Immersion Testing
- Valve Functionality Testing
- Bag Valve Connectors* ●
- . Maximum Threshold Testing
- Dead Space Testing* ●
- Shelf Life Testing (comprised of the applicable the mechanical testing described above) ●
- Resuscitator Compatibility Testing ●
- Accessory Compatibility Testing .
*Time Zero testing only.
The Sotair device performance testing considered the applicable sections of the following two standards where appropriate:
- ISO 5356-1:2015 Anaesthetic and respiratory equipment conical connectors . - Part 1: Cones and sockets.
- ISO 10651-4:2002 Lung ventilators Part 4: Particular requirements for operator-● powered resuscitators.
#### Device Risk Assessment:
The Sotair device risk assessment was conducted in accordance with ISO 14971 Medical Devices - Application of Risk management to Medical Devices.
#### Human Factors / Usability Testing:
SafeBVM conducted a simulated-use Human Factors Engineering / Usability Engineering (HFE/UE) Summative / Validation Study (summative study) on the Sotair device. This study was conducted in accordance with the Guidance for Industry and Food and Drug Administration Staff - Applying Human Factors and Usability Engineering to Medical Devices; February 3rd, 2016. The study concluded that the Sotair device met the criteria for use with the intended users, uses, and use environments. The overall result of the Usability Studies confirm that the Sotair device is designed, functions and meets the expectations and need for use.
#### 8. Conclusions:
The performance data demonstrated that the Sotair device met the product specifications and also performed equivalent to the predicate device in the Valve Functionality Testing which included direct comparison to the Smart Bag MO. The simulated-use Human Factors Engineering / Usability Engineering (HFE/UE) Summative / Validation Study (summative study) demonstrated that the critical tasks and severity level were considered acceptable with the analysis of users, uses and use environments. Overall, this data demonstrates that the Sotair device (subject device) is substantially equivalent to the Smart Bag MO (predicate device).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.