The measurement extension adds physiological measurements (Mainstream / Sidestream CO2 and optionally temperature, invasive blood pressure, cardiac output) to a dedicated host device. It is intended for use by trained healthcare professionals for adult, pediatric, and neonatal patients in a hospital environment and for transport inside hospitals. The measurement extension can only function when it is connected to a dedicated host device. The measurement extension adds physiological measurements (Microstream CO2 and optionally temperature and invasive blood pressure) to a dedicated host device. It is intended for use by trained healthcare professionals for adult, pediatric, and neonatal patients in a hospital environment and for transport inside hospitals. The measurement extension can only function when it is connected to a dedicated host device. The Integrated Pulmonary Index (IPI) is intended for use with adult and pediatric (1 to 12 years) patients only. The IPI is an adjunct to and not intended to replace vital sign monitoring.
Device Story
Compact, display-less measurement modules (867040, 867041) attach to dedicated host monitors; provide physiological data including CO2 (mainstream/sidestream/microstream), temperature, invasive blood pressure, and cardiac output. Operated by trained healthcare professionals in hospitals/transport. Modules utilize host device power and settings; store trend data on host. Processed waves, numerics, and alerts transmitted to host for clinical display. Output supports real-time patient monitoring and clinical decision-making regarding respiratory and hemodynamic status. Integrated Pulmonary Index (IPI) provides adjunct pulmonary assessment for pediatric/adult patients. Benefits include continuous, reliable physiological monitoring to assist in patient management.
Clinical Evidence
Bench testing only. No clinical data. Verification and validation included electrical safety (AAMI/ANSI ES60601-1), EMC (IEC 60601-1-2), respiratory gas monitoring (ISO 80601-2-55), clinical thermometers (ISO 80601-2-56), and invasive blood pressure (IEC 60601-2-34). Environmental and mechanical testing (7M1/7M3) confirmed performance under hospital conditions. Functional and regression testing verified system-level performance against predicate specifications.
Technological Characteristics
Compact modules; PBT and PC housing materials. Sensing: CO2 (mainstream/sidestream/microstream), temperature, invasive blood pressure, cardiac output. Energy: Powered by host device. Connectivity: Proprietary interface to host monitor. Software: AAMI ANSI IEC 62304 compliant. Standards: IEC 60601-1-2, AAMI/ANSI ES60601-1, IEC 60601-2-34, ISO 80601-2-55, ISO 80601-2-56.
Indications for Use
Indicated for adult, pediatric, and neonatal patients in hospital environments and during hospital transport. IPI indicated for adult and pediatric (1-12 years) patients only. Contraindications: Not intended to replace vital sign monitoring; requires connection to a dedicated host device.
Regulatory Classification
Identification
A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 26, 2017
Philips Medizin Systeme Boeblingen GmbH Markus Stacha Sr. Regulatory Affairs Engineer Hewlett-Packard-Str. 2 Boeblingen, D-71034 GERMANY
Re: K172904
Trade/Device Name: IntelliVue Capnography Extension 867040, IntelliVue Microstream Extension 867041 Regulation Number: 21 CFR 868.1400 Regulation Name: Carbon Dioxide Gas Analyzer Regulatory Class: Class II Product Code: CCK, BZQ, DXG, DRQ, FLL Dated: October 25, 2017 Received: October 27, 2017
Dear Markus Stacha:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Tina Kiang
-s
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below
Indications for Use
510(k) Number: K172904
Device Name: IntelliVue Capnography Extension 867040 and Microstream Extension 867041 Indications for Use (Describe)
Indications for Use statement for the Capnography Extension 867040:
The measurement extension adds physiological measurements (Mainstream / Sidestream CO2 and optionally temperature, invasive blood pressure, cardiac output) to a dedicated host device. It is intended for use by trained healthcare professionals for adult, pediatric, and neonatal patients in a hospital environment and for transport inside hospitals.
The measurement extension can only function when it is connected to a dedicated host device.
Indications for Use statement for the Microstream Extension 867041:
The measurement extension adds physiological measurements (Microstream CO2 and optionally temperature and invasive blood pressure) to a dedicated host device. It is intended for use by trained healthcare professionals for adult, pediatric, and neonatal patients in a hospital environment and for transport inside hospitals.
The measurement extension can only function when it is connected to a dedicated host device.
The Integrated Pulmonary Index (IPI) is intended for use with adult and pediatric (1 to 12 years) patients only. The IPI is an adjunct to and not intended to replace vital sign monitoring.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
| <span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including
the time to review instructions, search existing data sources, gather and maintain the data needed and
complete and review the collection of information. Send comments regarding this burden estimate or any
other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, providing a strong contrast that makes the word stand out.
## 510(k) Summary
This summary of 510(k) information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 C.F.R. §807.92.
- 1. Submitter of this premarket notification
Markus Stacha Philips Medizin Systeme Boeblingen GmbH Hewlett-Packard-Str. 2 D-71034 Boeblingen, Germany Tel: +49 7031 463-2840 Fax: +49 7031 463-2442 e-mail: markus.stacha@philips.com
This summary was prepared on Dec 22, 2017.
- 2. The name and classification of the device
Trade name: IntelliVue Capnography Extension 867040 and Microstream Extension 867041 Common name: Multifunction Patient Monitor Module.
Primary regulation and classification: CFR Title 21 Part 868 - Anesthesiology Devices, Section 868.1400 'Carbon dioxide gas analyzer'. Device Class II. Product Code: CCK. Secondary Product Codes: BZQ, DXG, DRQ, DSA, FLL.
- 3. The new Philips IntelliVue Capnography Extension 867040 is substantially equivalent to the previously cleared Philips Capnography Extension M3014A marketed pursuant to K050762. The new Philips IntelliVue Microstream Extension 867041 is substantially equivalent to the previously cleared Philips Microstream Extension M3015B marketed pursuant to K993383 and K113441.
- 4. Description of the device
The new IntelliVue Capnography Extension 867040 and Microstream Extension 867041 are modifications of the legally marketed Philips Capnography Extension M3014A and Microstream Extension M3015B, respectively.
IntelliVue Capnography Extension 867040 and Microstream Extension 867041 are compact measuring devices without display and control elements that, in combination with dedicated host devices, add the following measurement parameters:
867040: Capnography (Mainstream / Sidestream CO2) and optionally temperature, invasive blood pressure, and cardiac output (C.O.)
867041: Microstream CO2 and optionally temperature and invasive blood pressure. IntelliVue Capnography Extension 867040 and Microstream Extension 867041 use settings and power of the connected host device. Trend data and measurement settings from the measurements in the extensions are stored in the connected host device. The measurement extensions send processed waves, numerics and basic alerts to the host device.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is the only element in the image and it is centered.
- 5. Intended Use
Indications for Use statement for the Capnography Extension 867040:
The measurement extension adds physiological measurements (Mainstream / Sidestream CO2 and optionally temperature, invasive blood pressure, cardiac output) to a dedicated host device. It is intended for use by trained healthcare professionals for adult, pediatric, and neonatal patients in a hospital environment and for transport inside hospitals. The measurement extension can only function when it is connected to a dedicated host device.
Indications for Use statement for the Microstream Extension 867041:
The measurement extension adds physiological measurements (Microstream CO2 and optionally temperature and invasive blood pressure) to a dedicated host device. It is intended for use by trained healthcare professionals for adult, pediatric, and neonatal patients in a hospital environment and for transport inside hospitals.
The measurement extension can only function when it is connected to a dedicated host device.
The Integrated Pulmonary Index (IPI) is intended for use with adult and pediatric (1 to 12 years) patients only. The IPI is an adjunct to and not intended to replace vital sign monitoring.
The Intended Use and Indications for Use for the new Capnography Extension 867040 are the same as those for the predicate Capnography Extension M3014A. The only difference is the Continuous Cardiac Output (CCO) measurement option. Whereas the predicate measurement extension M3014A supports CCO measurement together with the dedicated host device, the new measurement extension does not support this measurement.
The Intended Use and Indications for Use for the new Microstream Extension 867041 are the same as those for the predicate Microstream Extension M3015B.
## 6. Technological Characteristics
The new IntelliVue Capnography Extension 867040 and Microstream Extension 867041 have the same relevant technological characteristics as the predicates M3014A and M3015B with regard to energy sources, portability, and robustness.
Compared to the predicates M3014A and M3015B, in the new 867040 and 867041 several electronic components were replaced by state-of-the-art components and the form factor was adapted to fit the new housing. The former common PCA for invasive pressure and temperature measurements was divided into two PCAs.
The plastic materials of the housings of the predicates M3014A and M3015B have been changed in the new 867040 and 867041 from ABS and PC-FR to PBT and PC, in order to enhance mechanical and chemical resistance.
Biocompatibility aspects are not affected because the devices do not have contact with patients.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame. The background is plain white.
The fundamental scientific technology of the new IntelliVue Capnography Extension 867040 and Microstream Extension 867041 is the same as that of the predicates Capnography Extension M3014A and Microstream Extension M3015B.
The differences in technological characteristics between the new and the predicate devices do not raise different questions of safety or effectiveness.
- 7. Summary of V&V activities
The new IntelliVue Capnography Extension 867040 and Microstream Extension 867041 have been subject of a series of V&V activities:
- Compliance with relevant recognized consensus standards:
- . IEC 60601-1-2: 2007 (Ed. 3) (Electromagnetic Compatibility)
- AAMI / ANSI ES60601-1:2005/(R)2012 and A1:2012 (Ed. 3.1) (Basic safety and ● essential performance)
- IEC 60601-2-34: 2011 (Ed. 3) (Invasive blood pressure monitoring equipment)
- ISO 80601-2-55: 2011 (Ed. 1) (Respiratory gas monitors)
- ISO 80601-2-56: 2009 (Ed. 1) (Clinical thermometers) ●
- AAMI ANSI IEC 62304:2006 (Ed. 1) (Software life cycle processes) All applicable requirements have been met.
- Testing as required by the Hazard Analysis.
All specified pass/fail criteria have been met. The test results have confirmed the effectiveness of the implemented design risk mitigation measures.
- -Additional environmental testing (temperature, humidity) and mechanical testing (mechanical classes 7M1 and 7M3 and chemical resistance against disinfecting and cleaning agents)
All specified test requirements have been met. The tests have confirmed that the new devices work according to their specifications and indicated claims during tests simulating general hospital conditions, handling and transport in hospital environments, disinfection, and storage.
- -Performance bench testing of C.O. signal acquisition All specified test requirements have been met. The new devices perform according to the specified criteria that are the same as those for the predicate devices.
- Unit level tests, functional system level tests, and regression system level tests. -The new devices perform according to the specified criteria. All specified test requirements have been met. The test results demonstrate that modified and previously available device functions work correctly according to the specifications and labeling claims.
- 8. Conclusion
Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the new devices with respect to the predicates. Testing comprised electrical and mechanical tests, EMC tests, environmental tests, performance tests, functional and regression tests.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, which makes the blue letters stand out.
Pass/Fail criteria were based on the specifications cleared for the predicate devices and test results showed substantial equivalence.
The results demonstrate that the new devices meet all defined reliability requirements and performance claims.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.