NIHON KOHDEN OLG-100A POCKETCAP POCKET CO2 MONITOR WITH ACCESSORIES
Device Facts
| Record ID | K964305 |
|---|---|
| Device Name | NIHON KOHDEN OLG-100A POCKETCAP POCKET CO2 MONITOR WITH ACCESSORIES |
| Applicant | Nihon Kohden America, Inc. |
| Product Code | CCK · Anesthesiology |
| Decision Date | Feb 19, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 868.1400 |
| Device Class | Class 2 |
| Attributes | Pediatric |
Intended Use
The Nihon Kohden PocketCap Pocket CO₂ Monitor, model number OLG-1100A, is intended for medical purposes to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory status. The device is intended as an indicator of patient carbon dioxide concentration during expiration and is not intended as the sole basis for medical diagnosis. The device will be available for use only by medical personnel within a hospital or other medical facility, at a remote site, or during transport from a remote site to a hospital or other medical facility.
Device Story
Pocket-sized CO2 monitor; measures expired gas CO2 concentration; aids assessment of patient ventilatory status. Used by medical personnel in hospitals, remote sites, or during transport. Input: gas mixture samples; Output: CO2 concentration display. Not intended as sole diagnostic basis. Provides real-time monitoring to assist clinical decision-making regarding ventilation. Benefits: portable, immediate feedback on patient respiratory status.
Clinical Evidence
No clinical data. Bench testing only: electromagnetic, environmental, safety, and performance testing conducted. Software validation performed for acquisition, processing, and display functions. Results confirmed device performance within specifications.
Technological Characteristics
Gas analyzer (CO2); portable form factor. Software-based acquisition, processing, and display. Non-sterile. Electromagnetic and environmental testing performed to verify operation.
Indications for Use
Indicated for monitoring CO2 concentration during expiration to assess ventilatory status in clinical settings. Contraindicated for patients <3 years old, <22 lbs, or with respiratory rates ≥60 breaths per minute.
Regulatory Classification
Identification
A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.
Predicate Devices
- Nellcor N-60 Stat-Cap Airway CO₂ Indicator (K915494)
Related Devices
- K992455 — NIHON KOHDEN OGS-2001A POCKETCARE AND ACCESSORIES · Nihon Kohden America, Inc. · Jan 7, 2000
- K981114 — MICROCAP · Oridion Medical , Ltd. · Jun 25, 1998
- K062115 — CARBON DIOXIDE MONITOR, MODEL OLG-2800A · Nihon Kohden America, Inc. · Dec 27, 2006
- K023820 — CO2 CLIP · Bregas AB · May 7, 2003
- K983114 — MAGNATEK80 · Magna Advanced Technologies, Inc. · Sep 22, 1998