NIHON KOHDEN OGS-2001A POCKETCARE AND ACCESSORIES
K992455 · Nihon Kohden America, Inc. · DQA · Jan 7, 2000 · Cardiovascular
Device Facts
| Record ID | K992455 |
| Device Name | NIHON KOHDEN OGS-2001A POCKETCARE AND ACCESSORIES |
| Applicant | Nihon Kohden America, Inc. |
| Product Code | DQA · Cardiovascular |
| Decision Date | Jan 7, 2000 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 870.2700 |
| Device Class | Class 2 |
| Attributes | Pediatric |
Intended Use
The OGS-2001A PocketCare is intended for medical purposes to transmit radiation at a known wavelength through blood and measure the blood oxygen saturation based on the amount of reflected or scattered radiation and is intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory status.
Device Story
OGS-2001A PocketCare is a portable SpO2/CO2 monitor; inputs include optical signals from SpO2 probes and gas mixture samples for CO2 concentration; device processes signals to calculate blood oxygen saturation and CO2 levels; output displayed for medical personnel to aid in ventilatory status assessment; used in clinical or remote/transport settings; not intended as sole diagnostic basis; benefits include real-time monitoring of patient respiratory status.
Clinical Evidence
Bench testing only. Electromagnetic, environmental, safety, and performance testing conducted per design control requirements. Software validation confirmed acquisition, processing, display, and recording functions performed within specifications. No clinical data provided.
Technological Characteristics
SpO2/CO2 monitor. Materials: silicone (reusable probes), biocompatible per ISO 10993. Airway adapter for single-patient use. Non-sterile. IEC 601-1 compliant. Software-based signal acquisition and processing.
Indications for Use
Indicated for medical personnel in medical facilities or remote environments to monitor SpO2 and CO2 to aid in assessing ventilatory status. CO2 monitoring indicated for patients ≥3 years old; not recommended for patients <22 lbs (low tidal volume) or respiratory rate >60 breaths/min.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
- OLG-1100A PocketCap CO2 Monitor (K964305)
- BSM-1101 Life Scope EC Patient Monitor (K973918)
Related Devices
- K964305 — NIHON KOHDEN OLG-100A POCKETCAP POCKET CO2 MONITOR WITH ACCESSORIES · Nihon Kohden America, Inc. · Feb 19, 1997
- K063752 — LIFESENSE, MODEL LS1-9R · Nonin Medical, Inc. · May 4, 2007
- K991086 — CAPNOCHECK II, MODEL 8400 · Bci Intl., Inc. · Jun 9, 1999
- K123797 — NT1D VITAL SIGNS MONITOR · New Tech, Inc. · Aug 27, 2013
- K982969 — NONIN PULSE OXIMETER AND CARBON DIOXIDE DETECTOR, MODELS 9843, 9845 AND 9847 · Nonin Medical, Inc. · Aug 12, 1999
Submission Summary (Full Text)
{0}------------------------------------------------
NIHON KOHDEN AMERICA, INC. July 22, 1999
K992455
510(k) NOTIFICATION
OGS-2001A PocketCare
JAN - 7 2000
# SECTION 2 - 510(K) SUMMARY
Name and Address of Applicant Nihon Kohden America. Inc. Attn: Regulatory Affairs 2601 Campus Drive Irvine, California 92612-1601
Phone: (949) 250-3959 Fax: (949) 250-3210
The device has been classified as Class II by the Cardiovascular Device Classification Panel under 21 CFR Part 870.2700 "Oximeter" per 74 DQA and by the Anesthesiology Classification Panel under 21 CFR Part 868.1400 Analyzer, Gas, Carbon dioxide, Gaseous-phase" per 73 CCK.
The common name for the OGS-2001A device is SpO2/CO2 monitor.
The predicate devices are the OLG-1100A PocketCap CO2 Monitor per 510(K) #K964305 commercial distribution certification dated 02/19/1997and the BSM-1101 Life Scope EC Patient Monitor per 510(k) #K973918, commercial distribution certification dated 01/13/1998.
Nihon Kohden's model number OGS-2001A is intended for medical use to measure the blood oxygen saturation (SpO-) and to measure the concentration of carbon dioxide in a gas mixture (CO-) to aid in determining the patient's ventilatory status. This product will be available for use by medical personnel within a medical facility or in a remote environment. This device is not intended to be the sole basis of diagnosis.
The CO2 function of the device is intended for patients three (3) years and above and is not recommended for patients with low tidal volume such as patients weighing less than 22 pounds or patients with a respiratory rate greater than 60 breaths per minute. The CO2 function will be available for all other patient populations.
The device complies with IEC 601-1 subclause 56.3(c) implemented by 21 CFR Part 868 Performance Standard for Electrode Lead Wires and Patient Cables. To date, no other special controls or performance standards are known or established for this device.
The OGS-2001A device is not sterile. The airway adapter is for single patient use only. Reusable or single patient use SpO2 probes are available.
The device material components have been determined to be non-contacting per the quidance provided by BS EN 30993: 1994 / ISO 10993: 1993 which replaces the Tripartite Guidance. The reusable SpO2 probe surface is silicone, known to be biocompatible by over 30 years of various medical device uses per the open literature. The disposable, SpO2 probes are the same probes as used with the predicate device BSM-1101 (K973918). Therefore, good laboratory practice studies were not required per 21 CFR 58.
The OGS-2001A PocketCare was subjected to electromagnetic, environmental, safety and performance testing procedures per Nihon Kohden design control requirements. These tests verified the operation of the device. Software validation tested the operation of the software functions of acquiring. processing, displaying and recording all functions of the results confirmed that the device performed within specifications.
Therefore based on the above, Nihon Kohden believes that the OGS-2001A PocketCare is substantially equivalent to the combination of the predicate devices.
{1}------------------------------------------------
# SECTION 3 - PROPOSED LABELING
#### A. Intended Use
The OGS-2001A PocketCare is intended for medical purposes to transmit radiation at a known wavelength through blood and measure the blood oxygen saturation based on the amount of reflected or scattered radiation and is intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory status.
### B. Device/Package Labels
Drawings of the device and the proposed product labels are provided in attachment 1.
### C. Proposed Packaging
Packaging for the OGS-2001A PocketCare is depicted in Attachment 2.
## D. Instructions for Use
The proposed instructions for use are provided with each packaged device and are presented in Attachment 8.
### ய் Advertisement/Promotional Literature
To date no advertisement or promotional literature has been created for the OGS-2001A PocketCare device for distribution in the United States.
# F. Contraindications, Precautions & Warnings
Warnings and cautions are listed in the Operator's Manual as shown in Attachment 3.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles three overlapping human profiles or faces, with flowing lines that suggest movement or progress.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN - 7 2000
Mr. Bonnie Bishop Nihon Kohden America, Inc. 2601 Campus Drive Irvine, CA 92612
Re: K992455 Nihon Kohden OGS-2001A PocketCare Regulatory Class: II (two) 74 DQA Product Code: Dated: November 5, 1999 November 8, 1999 Received:
Dear Mr. Bishop:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual reqistration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2 - Mr. Bonnie Bishop
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
jocconi H. Westendorp fes,
Celia M. Witten, Ph.D., M.D. Acting Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
NIHON KOHDEN AMERICA, INC. July 22, 1999
G. Indications for Use Statement
510(k) Number (if known): __
Device Name: OGS-2001A PocketCare
Indications for Use:
Nihon Kohden's model number OGS-2001A is intended for medical use to measure the blood oxygen saturation (SpO2) and to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory status (CO2).
This product will be available for use by medical personnel within a medical facility or in a remote environment including transport. This device is not intended to be the sole basis of diagnosis.
The CO2 function of the device is intended for patients three (3) years and above and is not recommended for patients with low tidal volume such as patients weighing less than 22 pounds or patients with a respiratory rate greater than 60 breaths per minute. The CO2 function will be available for all other patient populations.
Joan Awher
PRESCRIPTION
Image /page/4/Picture/12 description: The image shows a hand-drawn symbol that resembles a stylized 'X' above a horizontal line. The 'X' is formed by two diagonal lines that intersect in the middle. The horizontal line is positioned directly below the 'X', creating a base or foundation for the symbol. The drawing appears to be simple and lacks any shading or color.
ાર્વદ
OVER - THE - COUNTER USE -