K063752 · Nonin Medical, Inc. · DQA · May 4, 2007 · Cardiovascular
Device Facts
Record ID
K063752
Device Name
LIFESENSE, MODEL LS1-9R
Applicant
Nonin Medical, Inc.
Product Code
DQA · Cardiovascular
Decision Date
May 4, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The LifeSense® Model LS1-9R Capnography/Pulse Oximeter monitor indicated for use in simultaneously measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2), end tidal carbon dioxide (ETC02), respiration and pulse rate of well or poorly perfused adult, pediatric and infant patients. It is intended for use in environments where patients require continuous, non-invasive monitoring of these parameters by a healthcare professional, e.g. hospitals, medical facilities, post-operative care, patient transport, home-use, or any emergency medical services and environments.
Device Story
Portable monitor for continuous, non-invasive measurement of SpO2, ETCO2, respiration, and pulse rate. Inputs include SpO2 sensors and ETCO2 sampling lines (airway adapter, nasal cannula, or nasal cannula with oxygen input). Device processes physiological signals to display real-time numerical values and waveforms on an LCD touch panel. Used by healthcare professionals in hospitals, post-operative care, transport, and home environments. Provides clinicians with continuous monitoring data to assist in patient assessment and clinical decision-making. Benefits include early detection of respiratory or oxygenation changes in intubated or spontaneously breathing patients.
Clinical Evidence
Bench and clinical testing performed to demonstrate functional equivalence to predicate devices. No specific performance metrics (e.g., sensitivity, specificity) provided in the summary document.
Technological Characteristics
Portable LCD monitor with touch panel interface. Measures SpO2 via pulse oximetry and ETCO2 via sampling lines (airway adapter/cannula). Connectivity includes display of graphs and numerical parameters. Intended for continuous monitoring of adult, pediatric, and infant patients.
Indications for Use
Indicated for continuous, non-invasive monitoring of SpO2, ETCO2, respiration, and pulse rate in well or poorly perfused adult, pediatric, and infant patients in clinical, transport, or home-use settings.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
K063752
# SECTION 2. SUMMARY AND CERTIFICATION
#### 510(k) Summary A.
MAY - 4 2007
| Submitter: | Nonin Medical, Inc. |
|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Lori M. Roth<br>Clinical/Regulatory Specialist<br>Nonin Medical, Inc.<br>13700 1st Ave. North<br>Plymouth, MN 55441-5443 |
| Date Prepared: | December 15, 2006 |
| Trade Name: | LifeSense® Model LS1-9R Capnography/Pulse Oximeter<br>Monitor |
| Classification Name:<br>and Number: | Class II, 21 CFR 868.1400 |
| Product Code: | DQA, CCK |
| Predicate Device(s): | Nonin's LifeSense® Model LS1-9R is substantially equivalent to the MicroCap Plus/NPB-75 Capnograph/Pulse Oximeter manufactured by Oridion Medical, LTD cleared by the FDA under K024300 on 4/03/03 and the Capnocheck® Plus manufactured by BCI International, Inc. cleared under K970209 on 4/18/97. |
| Device Description: | The LifeSense® Model LS1-9R is a lightweight, portable Capnography/Pulse Oximeter monitor indicated for use in simultaneously measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2), end tidal carbon dioxide (ETC02), respiration and pulse rate of adult, pediatric, and infant patients. It is intended for use in any environment where patients require continuous, non-invasive monitoring of intubated or spontaneous breathing patients. |
| | When measuring ETCO2 the patient is attached to the monitor by a simple sample line that can be either an airway adapter for an endotracheal tube, a nasal cannula, or a nasal cannula with oxygen input. |
| | The liquid crystal display (LCD) monitor of the Model LS1-9R displays factory default or operator defined |
{1}------------------------------------------------
parameters and alarm settings, ETCO2 and respiration graphs, battery status and fault messages. Operator defined settings can be made on the touch panel display.
LifeSense Model LS1-9R monitor is intended for prescription use with adult, pediatric, and infant patients.
Intended Use: The LifeSense® Model LS1-9R Capnography/Pulse Oximeter monitor indicated for use in simultaneously measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2), end tidal carbon dioxide (ETC02), respiration and pulse rate of well or poorly perfused adult, pediatric and infant patients. It is intended for use in environments where patients require continuous, non-invasive monitoring of these parameters by a healthcare professional, e.g. hospitals, medical facilities, post-operative care, patient transport, home-use, or any emergency medical services and environments.
Safety Testing: Nonin's Model LS1-9R Capnography/Pulse Oximeter monitor has undergone both bench and clinical testing in order to demonstrate that it has appropriate functional features and is substantially equivalent to the predicate devices.
Conclusion: Nonin's LifeSense Model LS1-9R is substantially equivalent to the MicroCap Plus/NPB-75 Capnograph/Pulse Oximeter manufactured by Oridion Medical, LTD cleared by the FDA under K024300 and the Capnocheck® Plus manufactured by BCI International, Inc. under K970209.
Functional and
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes forming its wings. The eagle is positioned to the right of a circular border containing the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES USA".
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Lori M. Roth Clinical/Regulatory Specialist Nonin Medical, Incorporated 13700 1st Avenue North Plymouth, Minnesota 55441-5443
MAY - 4 2007
Re: K063752
Trade/Device Name: Nonin LifeSense® Model LS1-9R Capnography/ Pulse Oximeter Monitor Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DOA, CCK Dated: April 25, 2007 Received: April 27, 2007
Dear Ms. Roth:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
### Page 2 – Ms. Roth
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Susan Runne
Chiu Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{4}------------------------------------------------
## Indications for Use Statement
### 510(K) Number: K063752
### Device Name:
### Nonin LifeSense® Model LS1-9R Capnography/Pulse Oximeter Monitor _____________________________________________________________________________________________________________
### Indications for Use:
The LifeSense® Model LS1-9R Capnography/Pulse Oximeter monitor indicated for use in simultaneously measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2), end tidal carbon dioxide (ETCO2), respiration and pulse rate of well or poorly perfused adult, pediatric and infant patients. It is intended for use in environments where patients require continuous, non-invasive monitoring of these parameters by a healthcare professional, e.g. hospitals, medical facilities, post-operative care, patient transport, home-use, or any emergency medical services and environments.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|---------------------------------------------|
| | |
| | <div style="text-align:left;">Cheffto</div> |
| (Division Sign-Off) | |
| Division of Anesthesiology, General Hospital, | |
| Infection Control, Dental Devices | |
| | |
| 510(k) Number: | K063752 |
Nonin Medical, Inc.
LifeSense® Model LS1-9R Traditional 510(K): Premarket Notification Confidential
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.