NIHON KOHDEN CO2 SENSOR KIT, MODEL TG-970P

K083456 · Nihon Kohden America, Inc. · CCK · Mar 2, 2009 · Anesthesiology

Device Facts

Record IDK083456
Device NameNIHON KOHDEN CO2 SENSOR KIT, MODEL TG-970P
ApplicantNihon Kohden America, Inc.
Product CodeCCK · Anesthesiology
Decision DateMar 2, 2009
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 868.1400
Device ClassClass 2
AttributesPediatric

Intended Use

The Nihon Kohden TG-970P Series CO2 Sensor Kit is intended for medical purposes to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory status. Along with other methods indicated by the physician for medical diagnosis, this device is intended as an indicator of patient carbon dioxide concentration during expiration.

Device Story

The TG-970P Series CO2 Sensor Kit measures carbon dioxide concentration in gas mixtures via infrared radiation absorption. The device is intended for use with intubated patients weighing 7kg or more. It functions as an indicator of end-tidal CO2 during expiration to assist clinicians in assessing patient ventilatory status. The system includes a single-patient-use disposable airway adapter. Output is provided to the clinician to aid in medical diagnosis alongside other clinical methods. The device is intended for prescription use.

Clinical Evidence

Bench testing only. No clinical data provided. Safety and efficacy were verified through electromagnetic, environmental, and performance testing procedures.

Technological Characteristics

Infrared radiation absorption sensing principle. Single-patient-use disposable airway adapter. Non-sterile. Class II device (21 CFR 868.1400).

Indications for Use

Indicated for intubated patients weighing 7kg or more to monitor end-tidal CO2 concentration during expiration. Not indicated for patients weighing less than 7kg or those with respiration rates exceeding 150 breaths per minute.

Regulatory Classification

Identification

A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## NIHON KOHDEN AMERICA, INC. # SPECIAL 510(k) NOTIFICATION TG-970P Series CO2 Sensor Kit 08345 # SECTION 2 - 510(K) SUMMARY # Name and Address of Applicant Nihon Kohden America, Inc. 90 Icon Street Foothill Ranch, CA 92610 Contact: Jack Coggan 、 Director, Regulatory Affairs (949) 580-1555 ex. 3325 Fax: (949) 580-1550 ### Trade/Device Name: Nihon Kohden TG-970P Series CO2 Sensor Kit # MAR - 2 2009 #### Common or Usual Name: Capnometer, Carbon Dioxide (CO2) Analyzer, Carbon Dioxide (CO2) Indicator #### Classification Name: The device has been classified as Class II by the Division of Anesthesiology Devices and Anesthesiology Classification Panel under 21 CFR Part 868.1400 "Analyzer. Gas. Carbon dioxide, Gaseous-phase" as per part 73 CCK. #### Legally Marketed Predicate Device: Nihon Kohden TG-920P CO2 Sensor Kit as per 510(k) K040875 commercial distribution certification dated October 15, 2004. #### Intended Use: The Nihon Kohden TG-970P Series CO2 Sensor Kit is intended for medical purposes to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory status. Along with other methods indicated by the physician for medical diagnosis, this device is intended as an indicator of patient carbon dioxide concentration during expiration. #### A summary of the technological characteristics of the device: The Nihon Kohden CO2 Sensor Kit, model number TG-970P Series, is intended for medical purposes to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory (end tidal CO2) status. The device measuring technique is through absorption of infrared radiation. The airway adapter is a Single -Patient Use disposable product. The device is intended as an indicator of patient carbon dioxide concentration during expiration for intubated patients. The device is intended for use with patients weighing 7kg or more. The device is not recommended for patients with low tidal volume such as patients weighing less than 7kg or patients with a respiration rate greater than 150 breaths per minute. {1}------------------------------------------------ ## 510(k) Summary: - The device is not sterile. ● - . The device performance and specifications are consistent with all requirements for this device type. To date, no performance standards or special controls are known or established for this type of device. The device was subject to electromagnetic, environmental, safety and performance testing procedures. These tests verified the safety and efficacy of the device under intended operation for this device. Therefore, Nihon Kohden believes that the TG-970P Series CO2 Sensor Kit device is substantially equivalent to Nihon Kohden's predicate device TG-920P CO2 Sensor Kit. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" around the perimeter. Inside the circle is an abstract image of an eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # MAR - 2 2009 Mr. Jack Coggan Director of Regulatory Affairs/Quality Assurance Nihon Kohden America, Incorporated 90 Icon Street Foothill Ranch, California 92610-1601 Re: K083456 Trade/Device Name: TG-970P Series CO2 Sensor Kit Regulation Number: 21 CFR 868.1400 Regulation Name: Carbon Dioxide Gas Analyzer Regulatory Class: II Product Code: CCK Dated: January 29, 2009 Received: February 2, 2009 Dear Mr. Coggan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ #### Page 2 - Mr. Coggan Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Suite y. Michlene Dns. Ginette Y. Michaud, M.D. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## NIHON KOHDEN AMERICA, INC. # SPECIAL 510(k) NOTIFICATION TG-970P Series CO2 Sensor Kit G. Indications for Use Statement: 510(k) Number (if known): Device Name: TG-970P Series CO2 Sensor Kit ﺷﻴﺐ Indications of Use: The Nihon Kohden TG-970P Series CO2 Sensor Kit is intended for medical purposes to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory status. Along with other methods indicated by the physician for medical diagnosis, this device is intended as an indicator of patient carbon dioxide concentration during expiration. Over The Counter Use Prescription Use X AND/OR (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) CONCURRENCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE) Souge L. Mocheil m.d. (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: K083456
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...