K161860 · Nihon Kohden Corporation · CCK · Feb 24, 2017 · Anesthesiology
Device Facts
Record ID
K161860
Device Name
TG-970P CO2 Sensor Kit
Applicant
Nihon Kohden Corporation
Product Code
CCK · Anesthesiology
Decision Date
Feb 24, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1400
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Nihon Kohden TG-970P CO2 sensor kit is intended for medical purposes to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory status. Along with other methods indicated by the physician for medical diagnosis, this device is intended as an indicator of patient carbon dioxide concentration during expiration. The device is intended for use by qualified medical personnel within a hospital or clinical environment.
Device Story
The TG-970P CO2 Sensor Kit measures expired CO2 partial pressure via infrared absorption. The system consists of a CO2 sensor (infrared light source and detector), a CO2 adapter, and an airway adapter or mask. The sensor emits infrared light through the airway adapter; the detector measures light absorption to calculate CO2 concentration. The CO2 adapter converts these optical signals into digital data for the host instrument. Used in hospital or clinical settings by qualified medical personnel, the device provides real-time capnography data. This output assists clinicians in assessing patient ventilatory status and managing respiratory support. The device benefits patients by providing continuous, non-invasive monitoring of CO2 levels, facilitating timely clinical intervention.
Clinical Evidence
Bench testing only. Testing included CO2 waveform confirmation for YG-213T and YG-214T, dead space volume evaluation for YG-214T/TW, mechanical durability testing of the adapter housing, and biocompatibility evaluation for YG-211T/YG-213T/YG-214T. No clinical trials were performed; equivalence was established via design verification and validation.
Technological Characteristics
Infrared absorption spectroscopy; CO2 sensor (photo detector and light emitter); airway adapters (YG-211T, YG-213T, YG-214T) and masks (YG-232T, YG-242T); digital signal processing; biocompatible materials.
Indications for Use
Indicated for monitoring end-tidal CO2 in intubated neonates and infants (endotracheal tube size ≤ 4.0 mm) to aid in determining ventilatory status.
Regulatory Classification
Identification
A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized eagle or bird in flight, composed of three interconnected human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 24, 2017
Nihon Kohden Corporation Thomas Bento Sr. Vice President, Quality and Regulatory Affairs Nihon Kohden America, Inc. 15353 Barranca Parkway Irvine, California 92618
Re: K161860
Trade/Device Name: TG-970P CO2 Sensor Kit Regulation Number: 21 CFR 868.1400 Regulation Name: Carbon Dioxide Gas Analyzer Regulatory Class: Class II Product Code: CCK Dated: January 23, 2017 Received: January 24, 2017
Dear Thomas Bento:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{2}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Susan Runne, DDS, MA
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
Device Name Nihon Kohden TG-970P CO2 Sensor Kit
Indications for Use (Describe)
The Nihon Kohden TG-970P CO2 sensor kit is intended for medical purposes to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory status. Along with other methods indicated by the physician for medical diagnosis, this device is intended as an indicator of patient carbon dioxide concentration during expiration.
The device is intended for use by qualified medical personnel within a hospital or clinical environment.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## Traditional 510(k) Summary
| Company Name: | Nihon Kohden Corporation<br>1-31-4 Nishiochiai, 1-Chome, Shinjuku-Ku<br>Tokyo, Japan 161-18560 |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Phone number: | 81-2-59968020 |
| 510(k) Sponsor,<br>Contact: | Nihon Kohden America, Inc.<br>15353 Barranca Parkway<br>Irvine, CA 92618<br><br>Thomas Bento<br>Sr. VP Quality Assurance and Regulatory Affairs<br>Office: (949) 580-1555 x3324<br>Mobile: (949) 680-9048<br>Fax: (913) 273-0732 |
| Date Prepared: | February 22, 2017 |
| Device Name: | Nihon Kohden TG-970P CO2 Sensor Kit |
| Common Name: | Carbon Dioxide (CO2) Gas Analyzer |
## Classification:
The device has been classified as Class II per the Division of Anesthesiology Devices.
| Carbon Dioxide (CO2) Gas Analyzer | 868.1400 | CCK |
|-----------------------------------|----------|-----|
|-----------------------------------|----------|-----|
Predicate Device(s): Nihon Kohden TG-970P CO2 Sensor Kit (K122214)
#### 1.0 Description of Device
The TG-970P CO2 Sensor Kit measures the partial pressure of the expired CO2 of the patient by measuring the amount of absorbed CO2 infrared light, using a CO2 feature which absorbs infrared light. It is comprised of three main components; a CO2 Sensor (photo detector and light emitter) with cable and connector, a CO2 adapter and an airway adapter (YG-211T, YG-213T and YG-214T) or mask (YG-232T and YG-242T).
The CO2 Sensor incorporates an infrared light source, of specified wavelength, and an infrared detector. It detects the infrared light from the light emitter which passes through the expired air in the airway adapter and calculates the partial pressure of the expired CO2. The photo detector and light emitter end of the CO2 Sensor is connected to the airway adapter or mask. The airway adapter is then connected between the patient airway and the respirator. The CO2 adapter processes the electrical signal of the detected infrared light into digital data which is sent to the instrument.
{5}------------------------------------------------
#### 2.0 Indications For Use
The Nihon Kohden TG-970P CO2 sensor kit is intended for medical purposes to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory status. Alonq with other methods indicated by the physician for medical diagnosis, this device is intended as an indicator of patient carbon dioxide concentration during expiration.
The device is intended for use by qualified medical personnel within a hospital or clinical environment.
#### 3.0 Technical Characteristics
The Nihon Kohden CO2 Sensor Kit, model number TG-970P, is intended for medical purposes to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory (end tidal CO2) status. The device measuring technique is through absorption of infrared radiation.
#### 4.0 Substantial Equivalence
The technology, performance and materials of the Nihon Kohden TG-970P CO2 Sensor Kit, with airway adapters YG-213T and YG-214T, is substantially equivalent to the predicate device, the Nihon Kohden TG-970P CO2 Sensor Kit with airway adapter YG-213T (K122214). Differences between the devices are minor and are described below.
- 4.1 The new airway adapter YG-214T/TW is being added and is considered substantially equivalent to YG-213T/TW. The difference between the YG-213T/TW and YG-214T/TW is that the YG-214T/TW does not have an inner connection tube, which allows it to be connected to other types of respirators that have an existing inner connection tube. There is no change to the materials. It was assessed through design verification/validation testing that this modification does not impact the intended use of the device.
- 4.2 The dead space calculation for airway adapter YG-213T/TW has been updated to better simulate the clinical setting. The connection between the adapter and the endotracheal tube overlap in the respiration circuit and the actual dead space volume is decreased by the connecting parts. Bench testing was conducted to confirm the calculation. This change does not impact the intended use of the device.
- 4.3 The device durability of the airway adapter was improved by changing the outer shape where the user holds the device for connecting and disconnecting. The thickness of this area has increased. Bench testing was conducted to evaluate this change and this modification does not impact the intended use of the device.
- 4.4 The intended use of the TG-970P has not changed as a result of the above described modifications. The neonate/infant airway adapters, YG-213T/TW and YG-214T/TW, may be used for intubated neonates and infants, using an endotracheal tube size of ≤ 4.0 mm. It has been shown that the minimal dead space in the YG-213T/TW and YG-214T/TW airway adapters allows monitoring at low tidal volumes
{6}------------------------------------------------
which helps identify the intended patient population, as described in the literature review letter and bench testing. The proposed labeling for YG-213T/TW and YG-214T/TW would remove the current patient weight requirement of 2-7kg and recommend the intended population based on the endotracheal tube size and dead space volume of the airway adapters. The physician should determine the total dead space of the capnography setup with all other connected equipment to determine the appropriate airway adaptor based on the individual patients tidal volume and weight. Warnings and Caution statements are provided in the Operators Manual on p. 3.
#### 5.0 Performance Testing
The following tests were identified and performed based on the modifications made to the device.
- Clinical evaluation for prototype of a new neonatal airway adapter with no inner ● tube, applied to patient mouth
- Confirmation of CO2 waveform with YG-213T on the bench test .
- . Design change evaluation for YG-213T shape
- . YG-214T bench test for CO2 waveform
- Airway adapter for neonate and infant YG-214T, 214-TW dead space volume ●
- Evaluation of reducing mechanical burden on the sensor part due to YG-214T ● shape change
- YG-214T/214-TW device detailed design specification conformity evaluation
- . YG-211T/YG-213T/YG-214T Biocompatibility evaluation
#### 6.0 Conclusions
The TG-970P CO2 Sensor Kit was previously cleared in K122214 and the airway adapter (YG-213T/TW) cleared with the TG-970P is capable of accurately measuring CO2 in the patient's airway during expiration. The technology of the photo detector and light emitter of the TG-970P has not changed. An additional accessory, airway adapter YG-214T/TW, has been shown to be equivalent to the predicate, YG-213T/TW.
The intended population for the YG-213T and YG-214T airway adapters is for neonates and infants. Removal of the patient weight from the labeling is supported by literature and similar products. Furthermore, a recommended endotracheal tube size has been added to the labeling. Additionally, the dead space has been updated for YG-213T, which also further identifies the intended patient population. The device durability was also improved by changing the outer shape where the user holds the device for connecting and disconnecting. Based on a literature review and verification/validation testing in compliance with the Design Control requirements, these airway adapters are shown to be equivalent in safety and effectiveness to the predicate device.
The technology, performance and materials of the Nihon Kohden TG-970P CO2 Sensor Kit, with airway adapters YG-213T and YG-214T, is substantially equivalent to the predicate device, the Nihon Kohden TG-970P CO2 Sensor Kit with airway adapter YG-213T (K122214).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.