TG-970P SENSOR KIT W/ NEW AIRWAY ADAPTER (YG-213T)
K122214 · Nihon Kohden Corp. · CCK · May 21, 2013 · Anesthesiology
Device Facts
Record ID
K122214
Device Name
TG-970P SENSOR KIT W/ NEW AIRWAY ADAPTER (YG-213T)
Applicant
Nihon Kohden Corp.
Product Code
CCK · Anesthesiology
Decision Date
May 21, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1400
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Nihon Kohden TG-970P and is intended for medical purposes to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory status. Along with other methods indicated by the physician for medical diagnosis, this device is intended as an indicator of patient carbon dioxide concentration during expiration. The device is intended for use by qualified medical personnel within a hospital or clinical environment and is available for use on any patient weighing 2kg or more.
Device Story
TG-970P CO2 Sensor Kit measures end-tidal CO2 concentration via infrared absorption. Components include CO2 sensor (photo detector/light emitter), CO2 adapter, and airway adapter. Airway adapter connects between patient airway and respirator; sensor measures CO2 levels during expiration. Data sent to adapter for display on patient monitor. Used in hospitals/clinics by qualified medical personnel. Provides real-time CO2 monitoring to assist clinicians in assessing patient ventilatory status. Benefits include continuous monitoring of respiratory function. New YG-213T airway adapter enables use on patients 2kg or more.
Clinical Evidence
Bench testing only. Verification and validation testing performed to confirm accuracy of CO2 measurement with the new YG-213T airway adapter. Testing verified performance across specified patient weight ranges and confirmed that the reduced dead space (1mL) does not impact measurement accuracy compared to the predicate.
Technological Characteristics
Infrared radiation absorption sensing principle. Components: CO2 sensor (photo detector/light emitter), CO2 adapter, airway adapter. YG-213T airway adapter dead space: 1mL. Transmissivity: >45% at 4.27um. Operating environment: 10-40°C, 30-85% RH. Single patient use. No sterilization. Complies with IEC 60601-1, ISO 10993-1, and ISO 14971.
Indications for Use
Indicated for patients weighing 2kg or more to measure end-tidal CO2 concentration in gas mixtures to aid in determining ventilatory status. Not for patients <2kg or respiration rates >150 bpm.
Regulatory Classification
Identification
A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.
{1}------------------------------------------------
## KIZZZIY
Nihon Kohden America, Inc. Revised May 3, 2013
510(k) Notification TG-970P Sensor Kit w/new Airway Adaptor
## 510(k) Summary
# MAY 2 1 2013
| Company Name: | Nihon Kohden Corporation 90 Icon Street Foothill Ranch, CA 92610 | | | |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|----------|-----|
| Device Name: | Nihon Kohden TG-970P CO2 Sensor Kit w/new Airway Adaptor | | | |
| 510(k) Sponsor, Contact: | Nihon Kohden America, Inc. 90 Icon Street Foothill Ranch, CA 92610 Steve Geerdes Director Quality Assurance and Regulatory Affairs Phone: (949) 580-1555 Ext. 3325 Fax: (949) 580-1550 | | | |
| Summary Date: | Revised 05/03/2013 | | | |
| Common Name: | Carbon Dioxide (CO2) Analyzer | | | |
| Classification Names: | <table><tr><td>Anesthesiology Device</td><td>868.1400</td><td>CCK</td></tr></table> | Anesthesiology Device | 868.1400 | CCK |
| Anesthesiology Device | 868.1400 | CCK | | |
| Predicate Device(s): | | | | |
**Predicate Device(s):**
TG-970P CO2 Sensor Kit
### 1.0 Description of Device
The TG-970P Series CO2 Sensor Kit is comprised of three main components, a CO2 Sensor (photo detector and light emitter) with cable and connector, a CO2 Adapter and an airway adapter. The CO2 Sensor incorporates an infrared light source, of specified wavelength, and an infrared detector. The photo detector and light emitter end of the CO2 sensor is connected to an airway adapter. The airway adapter is connected between the patient airway and the respirator. As the patient completes an expiratory breath the sensor measures the CO2 levels in the expiratory breath and sends that data to the CO2 adapter. The adapter then reads the data and converts the data so it can be displayed on the patient monitor.
K083456
Depending on which airway adapter is used, the TG-970 can be used with patients weighing 2kg or more. The device is not recommended for patients weighing less than 2kg or patients with a respiration rate greater than 150 breaths per minute.
{2}------------------------------------------------
## Nihon Kohden America, Inc. Revised May 3, 2013
## 510(k) Notification TG-970P Sensor Kit w/new Airway Adaptor
Image /page/1/Figure/2 description: This image shows the parts of a CO2 adapter. The parts include a photo detector and light emitter, a CO2 adapter connector, a JG-970P CO2 adapter, and a TG-221T CO2 sensor. The image also shows an airway adapter, a transparent film, and the connection points to the mask or tracheal tube and to the resuscitation bag or respirator.
#### 2.0 -Intended Use of Device
The Nihon Kohden TG-970P and is intended for medical purposes to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory status. Along with other methods indicated by the physician for medical diagnosis, this device is intended as an indicator of patient carbon dioxide concentration during expiration.
The device is intended for use by qualified medical personnel within a hospital or clinical environment and is available for use on any patient weighing 2kg or more.
#### 3.0 Technical Characteristics
The Nihon Kohden CO2 Sensor Kit, model number TG-970P Series, is intended for medical purposes to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory (end tidal CO2) status. The device measuring technique is through absorption of infrared radiation.
| | Modified Device | Predicate Device | Comments |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product | TG-970P with YG-213T<br>airway adapter | TG-970P with YG-211T<br>airway adapter (K#083456) | |
| Intended Use | To measure the | Same | |
| | Modified Device | Predicate Device | Comments |
| Product | TG-970P with YG-213T<br>airway adapter | TG-970P with YG-211T<br>airway adapter (K#083456) | |
| | concentration of carbon<br>dioxide in a gas mixture to<br>aid in determining the<br>patient's ventilatory status | | |
| Use frequency | Single Patient Use | Same | |
| Shelf life | 36 months | Same | |
| Population | YG-213T: For patients 2kg to<br>7kg. | For patients more than<br>approx. 7kg. | Technology & materials<br>are substantially equivalent<br>to the predicate device.<br>Validation/verification<br>testing confirms that the<br>accuracy of the measured<br>gas is substantially<br>equivalent to the predicate<br>device. No new questions<br>of safety or effectiveness<br>were raised |
| Dead space | 1mL | 4mL | The change in dead space<br>is due to smaller airway<br>adapter.<br>Validation/verification<br>testing confirms that the<br>smaller dead space does<br>not affect the accuracy of<br>the measured gas and is<br>substantially equivalent to<br>the predicate device. No<br>new questions of safety or<br>effectiveness were raised |
| Function | Airway adapter with windows<br>which infrared light of the CO2<br>sensor kit can pass through is<br>connected to respiratory<br>circuit. | Same | |
| Specifications | | | |
| Transmissivity | More than approx 45%<br>(at 4.27um±0.02um) | Same | |
| | Modified Device | Predicate Device | Comments |
| Product | TG-970P with YG-213T<br>airway adapter | TG-970P with YG-211T<br>airway adapter (K#083456) | |
| Optical length | 6mm±0.2mm | Same | |
| Ratio of<br>antifogging<br>performance :<br>fog | Ratio of fog: less than 5%(at<br>15°Cto35°C 20% to 80RH 10<br>minutes later) | Same | |
| Ratio of<br>antifogging<br>performance<br>: Wet | Ratio of Wet: less than ±8%<br>(at 15°C to 35°C, 20% to 80RH<br>10 minutes later) | Same | |
## Technical Comparison
Page 2 of 2
{3}------------------------------------------------
## Nihon Kohden America, Inc. Revised May 3, 2013
# 510(k) Notification
TG-970P Sensor Kit w/new Airway Adaptor
Page 3 of 3
·
{4}------------------------------------------------
Nihon Kohden America, Inc. Revised May 3, 2013
510(k) Notification
TG-970P Sensor Kit w/new Airway Adaptor
| <b>Environment</b> | | | |
|-----------------------|---------------------------------------------------|----------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Facility | Hospitals, Clinics, Physician office | Same | |
| | | Same | |
| | | Same | |
| Operating environment | 10 to 40 degree C<br>30 to 85%RH, non condensing | Same | |
| Storage environment | -20 to 65 degree C<br>10 to 95%RH | Same | |
| Chemical resistance | This product cannot be<br>cleaned or disinfected. | Same | |
| Sterilization | No | No | |
| Continuation use time | 72 hours | 24 hours | Validation/verification testing confirms that the accuracy of the measured gas is not changed with longer continuous use time.<br>No new questions of safety or effectiveness were raised |
### 4.0 Data Summary
·
.
{5}------------------------------------------------
Testing of the Nihon Kohden TG-970P with new YG-213T Neonatal/Infant Airway Adapter was performed in compliance with Nihon Kohden Corporation design control process. Testing included:
The device is in compliance with the following voluntary industrial standards:
Medical Electrical Equipment
| IEC 60601-1 | Part 1: General requirements for safety 1998-12 |
|--------------------------|------------------------------------------------------------------------------------|
| IEC 60601-1, Amendment 1 | Part 1: General Requirements for safety, Amendment 1 1991-11 |
| IEC 60601-1, Amendment 2 | Part 1: General Requirements for safety, Amendment 2 1995-03 |
| ISO 10993-1 | Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing, 2003-08 |
Medical Devices - Application of Risk Management to Medical Devices, 2007
#### 5.0 Conclusions
ISO 14971
TG-970P CO2 Sensor Kit was previously cleared in K083456 and the airway adaptor cleared with the TG-970P (YG-211T) is capable of accurately measuring CO2 in the patient's airway during expiration on patients 7kg or larger. A new airway adapter (YG-213T) is being added to the TG-970P. This will allow the TG-970P to accurately measure CO2 in the patient's airway during expiration on patients 2kg or larger.
The technology of the photo detector and light emitter of the TG-970P has not changed. The only change is the new YG-213T airway adapter which has under gone verification and validation testing.
Based on the comparison in the technical comparison chart above and confirmed by verification/validation testing in compliance with the Design Control requirements. The intended use, technology and materials of the YG-213T Airway Adapter was shown to be equivalent in safety and effectiveness to the predicate devices YG-211T and YG-211TW which are accessories of the TG-970P CO2 Sensor Kit K083456 (TG-970P). No new questions of safety or effectiveness are raised.
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized symbol that resembles a human figure embracing a bird, representing the department's mission to protect and promote the health and well-being of Americans.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
## May 21, 2013
Mr. Steve Geerdes Director of Regulatory Affairs and Quality Assurance Nihon Kohden Corporation 90 Icon Street FOOTHILL RANCH CA 92610
Re: K122214
Trade/Device Name: Nihon Kohden TG-970P Regulation Number: 21 CFR 868.1400 Regulation Name: Carbon Dioxide Gas Analyzer Regulatory Class: II Product Code: CCK Dated: May 8, 2013 Received: May 9, 2013
### Dear Mr. Geerdes:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kwame O. Ulmer -S for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital,
Respiratory, Infection Control and Dental Devices
Office of Device Evaluation Center for Devices and Radiological Health
### Enclosure
{8}------------------------------------------------
## Indications for Use Form
510(k) Number (if known): K122214
Nihon Kohden TG-970P Device Name:
Indications for Use:
The Nihon Kohden TG-970P and is intended for medical purposes to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory status. Along with other methods indicated by the physician for medical diagnosis, this device is intended as an indicator of patient carbon dioxide concentration during expiration.
The device is intended for use by qualified medical personnel within a hospital or clinical environment and is available for use on any patient weighing 2kg or more.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division Sign-Off Office of In Device Evaluation 510(k)
Lester W. Schultheis Jr
2013.05.21[13:40:16-04'00']
510(k) Number:
Page 1 of
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.