The Ganshorn SpiroScout is intended for prescription use only to conduct simple diagnostic spirometry testing of adults and pediatric patients over 5 years old, in general practice and specialty physician, industrial and hospital settings.
Device Story
SpiroScout is a PC-based diagnostic spirometer measuring lung function parameters including vital capacity, flow/volume curves, and maximum voluntary ventilation (MVV). Input consists of patient respiratory airflow through a single-use disposable 'Spirette' mouthpiece. Device utilizes two diagonally opposing ultrasound transducers to measure flow, temperature, and humidity via ultrasound transit-time technology. System includes a handheld ScoutSensor and a base station. Operated by clinicians in general practice, specialty, industrial, or hospital settings. Output is processed by Ganshorn LFX software on a connected PC, providing clinicians with quantitative lung function data to support respiratory diagnosis and clinical decision-making.
Clinical Evidence
No clinical data or animal testing performed. Bench testing only, including compliance with AAMI ANSI ES 60601-1, IEC 60601-1-11, IEC 60601-1-2, and ATS (2005) standards.
Technological Characteristics
Ultrasound transit-time measurement; PC-based system with handheld ScoutSensor and base station; disposable Spirette mouthpiece. Flow range ± 18 l/s; volume range ± 10 liters. Power 100-240 V, 50/60 Hz. Patient contact parts (Spirette) are externally communicating, limited duration (<24 hours).
Indications for Use
Indicated for diagnostic spirometry testing in adults and pediatric patients aged 5 years and older in general practice, specialty physician, industrial, and hospital settings.
Regulatory Classification
Identification
A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a stylized symbol that resembles three human profiles facing to the right, with flowing lines beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 7, 2016
Ganshorn Medizin Electronic GmbH c/o Paul Dryden Consultant Industriestr 6-8 Niederlaauer, 97618 DE
Re: K160116
Trade/Device Name: SpiroScout Regulation Number: 21 CFR 868.1840 Regulation Name: Diagnostic Spirometer Regulatory Class: Class II Product Code: BZG Dated: September 7, 2016 Received: September 8, 2016
Dear Paul Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D. Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K160116
Device Name SpiroScout
Indications for Use (Describe)
The Ganshorn SpiroScout is intended for prescription use only to conduct simple diagnostic spirometry testing of adults and pediatric patients over 5 years old, in general practice and specialty physician, industrial and hospital settings.
Type of Use (Select one or both, as applicable)
XX Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
Page 1 of 1
PSC Publishing Services (301) 443-6740
{3}------------------------------------------------
# 510(k) Summary
October 6, 2016 Page 1 of 5
| Official Contact: | Stefan Ponto<br>Ganshorn Medizin Electronic GmbH<br>Industriestr 6-8<br>97618 Niederlauer Germany<br>Tel: +49 (9771) 6222-0<br>Fax: +49 (9771) 6222-55 |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| Proprietary or Trade Name: | SpiroScout |
| Common/Usual Name: | Diagnostic Spirometer |
| Classification Name: | Diagnostic Spirometer<br>BZG, Class II, 21 CFR 868.1840 |
| Predicate Device: | ndd Medical Technology - EasyOne 510(k) K99392 |
## Device Description:
The SpiroScout is an electronic measurement device intended to be used for measurement of lung function (determination of the respiratory flows and volume) parameters. It enables the determination of the following parameters.
#### Spirometry
Inspiratory and expiratory vital capacity (VC) Inspiratory and expiratory subdivisions
#### Flow/Volume
The measurement of the flow/volume curve enables the determination of the dynamic lung volumes. FEV1 value (forced expiratory volume within the first second of exhalation).
# Maximum Voluntary Ventilation MVV
The respiratory limit value indicates the respired volume per minute during maximally forced respiration. The respiratory limit value is also referred to as "Maximum Voluntary Ventilation".
The SpiroScout sensor operates using two diagonally opposing ultrasound transducers that alternately send and receive ultrasonic waves. The ultrasound technology simultaneously measures the flow, temperature and humidity.
The SpiroScout is a PC-based system running the Ganshorn LFX software program. The system comprises:
- the hand-held ScoutSensor
- Base station where the sensor is stored. ●
- Spirette, a single use, disposable mouthpiece and flow tube ●
#### Indications for Use:
The Ganshorn SpiroScout is intended for prescription use only to conduct simple diagnostic spirometry testing of adults and pediatric patients over 5 years old, in general practice and specialty physician, industrial and hospital settings.
{4}------------------------------------------------
#### Patient Population:
Adult and pediatric patients 5 years of age and older
#### Environment of Use:
General practice and specialty physician, industrial and hospital settings
## Contraindications:
None
#### Device Comparison
The following table compares the proposed device to the predicate ndd Medical Technology – EasyOne - K993921
| Technical<br>feature/specification | ndd EasyOne<br>K993921 | Proposed<br>Ganshorn SpiroScout |
|--------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The ndd Medical Technologies EasyOne<br>Spirometer is intended for prescription use<br>only to conduct simple diagnostic spirometry<br>testing of adults and pediatric patients over 4<br>years old, in general practice and specialty<br>physician, industrial and hospital settings. | The Ganshorn SpiroScout is intended for<br>prescription use only to conduct simple<br>diagnostic spirometry testing of adults and<br>pediatric patients over 5 years old, in general<br>practice and specialty physician, industrial and<br>hospital settings |
| Technology | Ultrasound transit-time measurement | Ultrasound transit-time measurement |
| Environment of use | General practice and specialty physician,<br>industrial and hospital settings.as is the<br>predicate. | General practice and specialty physician,<br>industrial and hospital settings |
| Patient Population | EasyOne can be used for adults and pediatric<br>patients over 4 years old | SpiroScout can be used for adults and pediatric<br>patients over 5 years old |
| Accessory | Disposable spirette respiratory tube | Single use, disposable spirette respiratory tube |
| Energy Type | 100-240 V / 50-60Hz | 100 to 240 V, 50/60 Hz |
| Materials and type of | Spirette | Spirette |
| Patient contact | Externally communicating (Indirect), Tissue<br>and Surface Contact, Mucosa, limited exposure | Externally communicating (Indirect), Tissue and<br>Surface Contact, Mucosa, limited exposure |
| Hardware components | EasyOne device | SpiroScout |
| | Spirette | Spirette |
| | USB Cradle | Base station |
| Parameters- Flow | | |
| Flow Range | $\pm$ 16 1/s | $\pm$ 18 1/s |
| Flow Accuracy | Flow: $\pm$ 2% or 0.020 l/s, (except PEF)<br>PEF: $\pm$ 5% or 0.200 l/s<br>MVV: $\pm$ 5% or 5 l/s | Flow: < $\pm$ 2 % or 0.03 1/s<br>PEF: < $\pm$ 3 % or 0.15 1/s<br>MVV: < $\pm$ 5 % or 5 1/s |
| Flow Resistance | approx. 0.02 cm H2O/l/s | approx. 0.02 cmH2O/l/s |
| Parameters Volume | | |
| Volume Range | $\pm$ 12 liters | $\pm$ 10 liters |
| Technical<br>feature/specification | ndd EasyOne<br>K993921 | Proposed<br>Ganshorn SpiroScout |
| Volume accuracy | $\pm$ 2% or 0.05 liters | +/- 2% or 0.05 liters |
| Parameters- Calculated | | |
| Measurement programs<br>(see legend below) | FVC, FVL, SVL, MVV, post testing | Slow Spirometry, Forced Spirometry, MVV |
| Forced Spirometry | FVC - Forced Vital Capacity (expiratory) | FVC - Forced Vital Capacity (expiratory) |
| Forced Spirometry | FIVC - Forced Vital Capacity (inspiratory) | FIVC - Forced Vital Capacity (inspiratory) |
| Forced Spirometry | FEV1 - Forced Expiratory Volume<br>(1 s) | FEV1 - Forced Expiratory Volume<br>(1 s) |
| Forced Spirometry | FEV6 - Forced Expiratory Volume | FEV6 - Forced Expiratory Volume |
| | (6 s) | (6 s) |
| | FEV1/FVC - ratio of FEV1 to FVC | FEV1/FVC - ratio of FEV1 to FVC |
| | FEV1/VCmax - ratio of FEV1 to Vcmax | FEV1/VCmax - ratio of FEV1 to Vcmax |
| | FEV1/FEV6 - ratio of FEV1 to FEV6 | FEV1/FEV6 - ratio of FEV1 to FEV6 |
| | FEF 25 (MEF 75) | FEF 25 (MEF 75) |
| | FEF 50 (MEF 50) | FEF 50 (MEF 50) |
| | FEF 75 (MEF 25) | FEF 75 (MEF 25) |
| | FEF 25-75 | FEF 25-75 |
| | PEF - Peak Expiratory Flow | PEF - Peak Expiratory Flow |
| | PIF - Peak Inspiratory Flow | PIF - Peak Inspiratory Flow |
| Slow Spirometry | VT - Tidal Volume Tidal volume at rest | VT - Tidal Volume Tidal volume at rest |
| | ERV - Expiratory Reserve Volume | ERV - Expiratory Reserve Volume |
| | IRV - Inspiratory Reserve Volume | IRV - Inspiratory Reserve Volume |
| Slow Spirometry or<br>Forced Spirometry | VCmax - Maximum Vital Capacity<br>of an SVC, FVC, FVL | VCmax - Maximum Vital Capacity<br>of an Slow or Forced Spirometry |
| Slow Spirometry | VC ex - Expiratory Vital Capacity<br>of a slow spirometric test | VC ex - Expiratory Vital Capacity<br>of a slow spirometric test |
| | VC in - Inspiratory Vital Capacity<br>of a slow spirometric test | VC in - Inspiratory Vital Capacity<br>of a slow spirometric test |
| | VC - Vital Capacity<br>of a slow spirometric test | VC - Vital Capacity<br>of a slow spirometric test |
| | IC - Inspiratory Capacity | IC - Inspiratory Capacity |
| MVV | MVV - Maximum Voluntary Ventilation | MVV - Maximum Voluntary Ventilation |
{5}------------------------------------------------
#### 510(k) Summary October 6, 2016 Page 3 of 5
Legend: The predicate ndd EasyOne and Ganshorn SpiroScout use different terms for the same programs. The equivalence of these terms is described in the table below.
| | ndd Term | Ganshorn Term |
|-------------------------------|--------------------------------------------------------|----------------------------------|
| Measurement Program<br>Legend | FVC (Forced Vital Capacity), FVL (Flow<br>Volume Loop) | Forced Spirometry |
| | SVL (Slow Volume Loop) | Slow Spirometry |
| | MVV (Maximal Volume Ventilation) | MVV (Maximal Volume Ventilation) |
# Summary of Performance Testing
#### Non-clinical
- Compliance with AAMI ANS ES 60601-1
- Compliance with IEC 60601-1-11
- Compliance with IEC 60601-1-2 .
- . ATS (2005)
{6}------------------------------------------------
#### 510(k) Summary October 6, 2016 Page 4 of 5
## Biocompatibility:
No Biocompatibility testing was done. The only patient contacting part is the Spirette that was purchased in the final finished form from the manufacturer of the Spirette of the predicate device ndd EasyOne 510(k) K993921.
Per G95-1 and ISO 10993-1:2009, this component is considered as:
- External Communicating (Indirect gas pathway) ●
- . Tissue / Bone / Dentin communicating
- Duration of Use limited (<24 hours) .
And
- Surface Contact
- Mucosal membrane
- Duration of Use limited (< 24 hours) ●
#### Animal
No animal testing was performed
Clinical: No clinical testing was performed
# Comparison to Predicate Devices:
#### Indications for Use
As in Comparison of Indications For Use above, we conclude that the indications for use for the SpiroScout and the predicate are substantially equivalent.
Prescriptive - The SpiroScout and predicate are prescription devices.
Design and Technology – As above design and technology are substantially equivalent to the predicate.
Performance and Specifications -As above the performance and specifications are substantially equivalent to the predicate.
Compliance with Standards - The SpiroScout complies with the currently recognized safety and EMC standards (AAMI ANSI ES 60601-1, IEC 60601-1-2). The predicate complied with IEC 601-1.
Patient Population - The SpiroScout and predicate are indicated for pediatric and adult patients.
Environment of Use - The SpiroScout is for use in general practice and specialty physician, industrial and hospital settings.as is the predicate.
Differences: There are some differences:
- Patient population is 5 and up vs. 4 yo and up
- This has been changed based upon FDA request O
- Difference in flow ranges where the proposed device has a high flow range which is consistent with the ● recommendations of ATS spirometer designs
The differences between the proposed device and the predicate device do not raise any new safety and efficacy concerns.
{7}------------------------------------------------
#### 510(k) Summary October 6, 2016 Page 5 of 5
# Substantial Equivalence Conclusion
Based upon the performance testing and comparison to the legally marketed predicate devices for use, technology, and performance we believe we have demonstrated that the SpiroScout is substantially equivalent in safety and effectiveness to the predicate devices.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.