FLOWMATE LTE

K954759 · Spirometrics Medical Equipment Co. · BZG · Apr 12, 1996 · Anesthesiology

Device Facts

Record IDK954759
Device NameFLOWMATE LTE
ApplicantSpirometrics Medical Equipment Co.
Product CodeBZG · Anesthesiology
Decision DateApr 12, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.1840
Device ClassClass 2

Intended Use

Intended use for the Flowmate LTE is for Pulmonary Function Testing (measuring a person's ability to move air into and out of their lungs). The volume and flow rates measured for the entire effort and then reported as numerical results with a printout of the effort if selected.

Device Story

Flowmate LTE is a portable diagnostic spirometer used in clinical settings to assess pulmonary function. It utilizes a flow-sensing pneumotach to measure differential pressure as air passes through the device. The system performs automatic calculations of test results, including FVC, FEV1, and MVV, and selects the best test based on ATS criteria. It is powered by rechargeable Ni-Cad batteries and features a compact, high-impact ABS plastic housing. Healthcare providers use the device to obtain numerical results and printed reports of patient respiratory efforts. Cross-contamination is managed via an external filter system. The device aids clinicians in evaluating lung health and respiratory capacity.

Clinical Evidence

Bench testing only. Performance evaluated against ATS recommendations using 24 standard waveforms for FVC, FEV1, FEF 25-75, and PEF. MVV testing conducted using a 2-liter volume at flow rates from 40 to 245 L/M. Results demonstrate accuracy and precision within ATS criteria, with volume accuracy of +/- 3% and max flow rate of +/- 16 L/s.

Technological Characteristics

Flow-sensing pneumotach; measures differential pressure. Materials: high-impact UL-rated ABS plastic. Power: rechargeable Ni-Cad batteries. Capacity: 10 Liters. Accuracy: +/- 3% volume, +/- 16 L/s flow. Back pressure: < 1.5 cm H2O. Connectivity: printer output. Calibration: field-calibrated via 1-9 Liter syringe. Cross-contamination control: external filter system.

Indications for Use

Indicated for patients requiring pulmonary function testing to measure air movement into and out of the lungs. Applicable to individuals undergoing diagnostic or screening spirometry.

Regulatory Classification

Identification

A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} SPIROMETRICS MEDICAL EQUIPMENT COMPANY K954759 "Dedicated to Improving the Quality of Life" # 510(k) SUMMARY Submitted by: New VED Inc. d.b.a. Spirometrics Medical Equipment Co. 415 Rodman Road Auburn, ME 04210 207-784-0906, Fax 207-784-1481 Contact Person: Donald Henton, Official Correspondent Date Prepared: October 10, 1995 - Classification Name: Diagnostic Spirometer - Common or Usual Name: Diagnostic or Screening Spirometer - Proprietary Name: Flowmate LTE - Establishment Registration Number: 1720605 - Classification: Diagnostic Spirometers have been classified as Class II devices. - Performance Standard: None applicable Spirometrics has been marketing a flow sensing pneumotach with its models Flowmate and PC Flow +. The proposed product is substantially equivalent to the Flowmate 2500 premarket notification number K863953/A, and PC Flow + 3350 premarket notification number K900673 both manufactured by Spirometrics, the Renaissance from Puritan Bennett premarket notification number K911143, and the Flash Portable Spirometry System from Tamarac Systems Corporation premarket notification number unknown. For a description of the device see Exhibit 1. Description of the device: - Flow sensing pneumotach and base that measures differential pressure as air passes through it (the same principle as the Flowmate, PC Flow +, Renaissance, and Flash devices) - Compact design molded from high impact UL rated ABS plastic, rechargeable Ni-Cad batteries for portable operation, - Automatic calculations of test results, and best test selection based on ATS criteria 510(k) SUMMARY 415 Rodman Road • Auburn, ME 04210 • (207) 784-0906 Page 1 of 2 {1} - Testing capabilities include; FVC, FEV1, FEV3, FEV1/FVC%, FEF.2-1.2, FEF25-75%, FEF75-85, PEF, FEF50, FIVC, FEF50/FIF50, PIF, MVV, Methacholine Challenge, and Lung Age - Testing Capabilities: Capacity 10 Liters - Volume Accuracy +/- 3% - Max. Flow Rate +/- 16 L/s - Back Pressure Less than 1.5 cm H2O - Population Predicted Normals included: Composite - Knudson - Crapo (ITS) - ECCS - HSU - Customization allows user to change report format, input temperature and barometric readings, and select printer driver for printer being used - Reports printed on standard 8 1/2 " x 11" paper - Field calibration with a 1 to 9 Liter calibration syringe - Cross-contamination control via external filter system Intended use for the Flowmate LTE is for Pulmonary Function Testing (measuring a person's ability to move air into and out of their lungs). The volume and flow rates measured for the entire effort and then reported as numerical results with a printout of the effort if selected. Waveform performance testing showed that the Flowmate LTE exceeded the ATS recommendations for FVC, FEV1, FEF 25-75 and PEF accuracy at all of the 24 waveforms except #17. Waveform #17 takes over 20 seconds to complete. The Flowmate LTE accumulates volume for up to 15 seconds the same as the Flowmate 2500. Therefore, some of Waveform #17's volume was produced after the Flowmate LTE stopped recording data from the effort, which accounts for the variation between the actual delivered volume and the measured volume (See Exhibit 2). Taking the volume at the 15 second mark generated by waveform #17 (5.662 Liters) the accuracy is within the ATS recommendation at - 1.8%. MVV testing used a 2 Liter volume at flow/rates from 40 to 245 L/M also showed acceptable results compared with the ATS criteria for accuracy and precision to the delivered values (See Exhibit 3). Our conclusion is that the Flowmate LTE is substantially equivalent to the predicate device the Flowmate. It is also substantially equivalent to the PC Flow +, Renaissance, and Flash Portable Spirometry System. 510(k) SUMMARY Page 2 of 2
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...