MICRO DIARY SPIROMETER
K100928 · Viasys Respiratory Care, Inc. · BZG · Mar 7, 2011 · Anesthesiology
Device Facts
| Record ID | K100928 |
| Device Name | MICRO DIARY SPIROMETER |
| Applicant | Viasys Respiratory Care, Inc. |
| Product Code | BZG · Anesthesiology |
| Decision Date | Mar 7, 2011 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 868.1840 |
| Device Class | Class 2 |
| Attributes | Pediatric |
Intended Use
The Micro Diary Spirometer is used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs. Specifically, the Micro Diary Spirometer measures the following lung function parameters: FEV1, FVC, FEV6 and PEF.
Device Story
Micro Diary Spirometer is a compact, battery-operated, handheld recording spirometer. Device utilizes a digital turbine volume transducer; exhaled breath rotates a vane; rotation frequency/count proportional to flow rate and volume. Measures FEV1, FVC, FEV6, and PEF. Used in hospitals, clinics, or patient homes under medical practitioner supervision. Operated by patients or clinicians. Device records test data for later review and supports transfer to compatible computers. Healthcare providers use output to monitor lung function and manage respiratory conditions like asthma or COPD. Benefits include portable, longitudinal tracking of pulmonary function.
Clinical Evidence
No clinical data. Bench testing only; device verified for accurate measurement of FEV1, FVC, FEV6, and PEF. Electrical safety and electromagnetic compatibility testing passed.
Technological Characteristics
Handheld diagnostic spirometer; turbine-based digital volume transducer; battery-powered; 5-button keypad; B&W LCD display. Data recording and transfer capability. Technological design, materials, and energy source identical to predicate.
Indications for Use
Indicated for patients 4 years and older requiring lung function measurements, including those with asthma or chronic obstructive pulmonary disorder, provided they can follow practitioner instructions.
Regulatory Classification
Identification
A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.
Predicate Devices
- Micro Diary Spirometer (K965042)
Related Devices
- K103575 — SPIRO PD · Pmd Healthcare · Nov 9, 2011
- K991412 — VITALOGRAPH MICRO · Vitalograph, Inc. · Feb 4, 2000
- K160253 — Vitalograph Model 6300 micro · Vitalograph (Ireland) , Ltd. · Aug 26, 2016
- K200550 — Vitalograph Model 2120 In2itive eDiary · Vitalograph Ireland, Ltd. · Jul 27, 2020
- K965042 — MICRODL DIARYCARD SPIROMETER · Micro Direct, Inc. · Jun 17, 1997
Submission Summary (Full Text)
{0}------------------------------------------------
## 510(k) Summary
Image /page/0/Picture/1 description: The image shows the logo for CareFusion. The logo consists of a circular icon on the left and the word "CareFusion" on the right. The icon contains a shield shape inside the circle.
K100925
CareFusion 22745 Savi Ranch Pa Yorba Linda, CA 92 714-283-2228 te 714-283-8424 fax www.carefusion.com
# 510(k) Summary
MAR - 7 2011
Device Name: Device Trade Name: Common/Classification Name: Diagnostic Spirometer Micro Diary Spirometer Diagnostic Spirometer
Contact Person and Submitter: Merritt Girgis CareFusion 22745 Savi Ranch Parkway Yorba Linda, CA 92887 Phone: (714) 919-3286 Fax: (714)-283-8420 Email: merritt.girgis@carefusion.com
#### Substantial Equivalence
The Micro Diary Spirometer is substantially equivalent to the previous version of the Micro Diary Spirometer (K965042).
#### Description of the device
The Micro Diary Spirometer is a compact, hand held, and battery operated recording spirometer. The device features a five button keypad and an LCD display. The device utilizes a digital volume transducer. As a patient's exhaled breath is passed through the transducer, the vane rotates. The number of rotations is proportional to the volume of air passed through the transducer, and the frequency of rotation is proportional to the flow rate.
#### Indications for Use Statement
The Micro Diary Spirometer is used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs. Specifically, the Micro Diary Spirometer measures the following lung function parameters: FEV1, FVC, FEV6 and PEF.
#### Target Population
The device can be used on patients who require lung function measurements. These patients are usually suffering from diseases such as asthma and chronic obstructive pulmonary disorder. It can be utilized for patients from 4 years and older, providing that they are able to follow the medical practitioner's instructions.
#### Environment of Use
The environment of use is the hospital or a doctor's or medical practitioner's office or clinic or the patient's home.
{1}------------------------------------------------
### Technological Characteristics
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
The updated version of the Micro Diary Spirometer has the same technological characteristics (i.e., design, material, chemical composition, energy source) as the predicate device, which is previous version of the Micro Diary Spirometer (K965042).
### Comparison of Technological Characteristics
As a result of tooling and component obsolescence, it was necessary to produce a direct replacement for the previous version of the Micro Diary Spirometer using current tooling and available copnoponents. Both devices perform standard spirometric measurements, and both are capable of recording the test results for review at a later time. The MicroDiary (new) and the older MicroDiary perform exactly the same functions and tests.
| | Micro Diary (new) | Micro Diary (previous) |
|-------------------------------|--------------------------------------------------------------------|------------------------------------------------------------------|
| Parent Company | CareFusion | CareFusion(Formerly<br>Cardinal Health, Viasys,<br>Micro Medical |
| Device Classification<br>name | Diagnostic Spirometer | Diagnostic Spirometer |
| 510(k) Number | New | K965042 |
| Indications for Use | Measures lung function<br>as follows: FEV1, FVC,<br>FEV6, and PEF. | Measures lung function<br>as follows: FEV1, FVC,<br>FEV6, PEF |
| Display Screen | LCD Display, B&W | LCD Display, B&W |
| User interface | 5 Keys – on, up, down,<br>enter, and delete | 5 Keys – on, up, down,<br>enter, and delete |
| Measurement Transducer | Turbine | Turbine |
| Power supply | Battery | Battery |
### Nonclinical Testing
The Micro Diary Spirometer was tested to assure that it accurately measures lung function across the following standard spirometric measures: FEV1, FVC, FEV6, and PEF. All tests were passed successfully. Electrical Safety and Electromagnetic Compatibility testing was also performed and passed.
#### Clinical Testing
No clinical testing was submitted for device.
#### Conclusion
The testing has demonstrated that the device is as safe, as effective, and performs as well as or better than the legally marketed Micro Diary Spirometer (K965042).
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Merritt Girgis Manager, Regulatory Affairs CareFusion 22745 Savi Ranch Parkway Yorba Linda, California 92887
MAR - 7 2011
- Re: K100928
Trade/Device Name: Micro Diary Spirometer Regulation Number: 21 CFR 868.1840 Regulation Name: Diagnostic Spirometer Regulatory Class: II Product Code: BZG Dated: March 2, 2011 Received: March 3, 2011
Dear Mr. Girgis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registerion and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Anthony D. Avena
Anthony Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## EXHIBIT A Statement of Indications for Use
510(k) Number (if known): Not known.
Device Name: Micro Diary Spirometer
Indications for Use: The Micro Diary Spirometer measures lung function using the following standard spirometric measures: FEV1, FVC, FEV6, and PEF. The device also records the test data for later review and has the ability to transfer these test records to a compatible computer.
Target Population: Patients requiring lung function measurements.
Environment of Use: Under the supervision of a medical practitioner.
AND/OR Over-The-Counter Use Prescription Use × (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
L. Schultheis
(Division Sign-Off) (Division Sign-On)
Division of Anesthesiology, General Hospital
Division of Anest Franks Devices Division Controi, Dental Devices
510(k) Number: K/00928
{5}------------------------------------------------
### Statement of Indications for Use
510(k) Number: K100928.
Device Name: Micro Diary Spirometer
Indications for Use: The Micro Diary Spirometer is used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs. Specifically, the Micro Diary Spirometer measures the following lung function parameters: FEV1, FVC, FEV6 and PEF.
Target Population: The device can be used on patients who require lung function measurements. These patients are usually suffering from diseases such as asthma and chronic obstructive pulmonary disorder. It can be utilized for patients from 4 years and older, providing that they are able to follow the medical practitioner's instructions.
Environment of Use: The environment of use is the hospital or a doctor's or medical practitioner's office or clinic or the patient's home.
X Prescription Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
L. Schutter
(Division Sign-Off)
Division of Anesthesiology, General Hospital
Infection Control and Dental Devices
§10(k) Number:
K 100 928