MEDIKRO SPIROSTAR, M929 Medikro Spirostar USB and M921 Medikro SpiroStar DX
K120577 · Medikro OY · BZG · Jun 24, 2015 · Anesthesiology
Device Facts
Record ID
K120577
Device Name
MEDIKRO SPIROSTAR, M929 Medikro Spirostar USB and M921 Medikro SpiroStar DX
Applicant
Medikro OY
Product Code
BZG · Anesthesiology
Decision Date
Jun 24, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1840
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Medikro Spirometer is a device that measures lung air volume and airflow rate for pulmonary disease diagnosis and screening. These measurements provide information about a patient's pulmonary function which may be compared with normal values or the patient's previous values. The device is designed for adult and pediatric patients, hospital and clinic use only.
Device Story
Medikro SpiroStar (USB/DX models) measures lung air volume and airflow rate; utilizes disposable SpiroSafe flow transducer connected via pressure tube; requires nose clip. Device connects to PC via USB or serial port; powered by host PC. Medikro Spirometry Software performs measurements; calculates volume based on patient demographics (age, gender, race); compares results against reference values. Used in hospitals/clinics by healthcare professionals. Output includes flow-volume/volume-time curves, pulmonary function metrics (FVC, FEV1, FEF25-75%, PEF), and post-bronchodilator change quantification. Enables clinical assessment of pulmonary function and disease screening.
Clinical Evidence
Bench testing only. Device performance validated against American Thoracic Society (ATS) 24 standard waveforms. Testing confirmed compliance with ATS recommendations for FVC, FEV1, FEF25-75%, PEF, flow resistance, and BTPS conditions. Electrical safety and EMC testing performed per IEC 60601-1, IEC 60601-1-2, and IEC 60601-1-4 standards.
Technological Characteristics
Diagnostic spirometer; disposable flow transducer; USB or serial PC interface; powered via PC port. Software-based analysis of lung volume/airflow. Complies with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), and IEC 60601-1-4 (programmable electrical medical systems). ATS/ERS compliant.
Indications for Use
Indicated for adult and pediatric patients requiring pulmonary function measurement for disease diagnosis and screening in hospital or clinical settings.
Regulatory Classification
Identification
A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.
Welch Allyn CardioPerfect Workstation, Software Version 1.5.0 (K052158)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features the department's name in a circular arrangement around an emblem. The emblem is a stylized representation of a human figure, composed of three overlapping profiles, symbolizing health and well-being.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 24, 2015
Medikro Oy Mr. Mikko Eloranta Managing Director Kellolahdentie 27 FI-70460 Kuopio Finland
Re: K120577
Trade/Device Name: Medikro SpiroStar USB and Medikro SpiroStar DX Regulation Number: 21 CFR 868.1840 Regulation Name: Diagnostic Spirometer Regulatory Class: Class II Product Code: BZG Dated: June 15, 2015 Received: June 18, 2015
Dear Mr. Eloranta:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Eloranta
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
| 510(k) Number: | K120577 |
|----------------|------------------------------------------------------------|
| Device Name: | Medikro® SpiroStar |
| Models: | M929, Medikro SpiroStar USB and M921, Medikro SpiroStar DX |
Indications For Use:
Medikro Spirometer is a device that measures lung air volume and airflow rate for pulmonary disease diagnosis and screening. These measurements provide information about a patient's pulmonary function which may be compared with normal values or the patient's previous values.
The device is designed for
- . adult and pediatric patients,
- hospital and clinic use only.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801.109)
OR
Over-The-Counter Use _ _ _ _ _ _ _ _
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
# 510(k) Summary
### General Information
| Date: | 24.6.2015 |
|-------------------------|--------------------------------------------------------------------------------------|
| Trade Names: | M929, Medikro SpiroStar USB and<br>M921, Medikro SpiroStar DX |
| Common Name: | Diagnostic Spirometer |
| Classification Name: | Spirometer, Diagnostic |
| Classification: | Class II |
| Manufacturer: | Medikro Oy<br>Pioneerinkatu 3<br>FI-70800 Kuopio<br>Finland<br>tel: +358 17 283 3000 |
| Corresponding Official: | Mikko Eloranta, Managing Director<br>tel: +358 17 283 3000<br>fax: +358 17 283 3300 |
| Predicate Devices: | Caird Technology Spirometer, K971336 |
| | Welch Allyn CardioPerfect Workstation, Software Version<br>1.5.0, K052158 |
#### Device Description
Medikro® SpiroStar spirometer runs on a personal computer with Microsoft Windows operating systems.
Medikro® SpiroStar spirometer unit is connected to PC via USB port (USB model) or via serial port (DX model). SpiroSafe disposable flow transducer is connected to the spirometer unit via pressure tube. A nose clip is used to prevent air flow from nose during measurements. Optional calibration syringe is used for recommended daily volume calibrations.
Medikro® SpiroStar spirometers are used to measure lung air volume and airflow rate. Medikro® Spirometry Software is used to perform the measurement and calculate the measurement volume based on chosen reference value. The reference value is based on patient's gender, race and age. User can analyze the results in different presentations and create a final report based on the results and patient information.
{4}------------------------------------------------
The spirometers take all the power that it needs from the USB or serial port, so no other external or internal power supply is needed.
### Indications for Use
Medikro Spirometer is a device that measures lung air volume and airflow rate for pulmonary disease diagnosis and screening. These measurements provide information about a patient's pulmonary function which may be compared with normal values or the patient's previous values.
The device is designed for
- . adult and pediatric patients,
- . hospital and clinic use only.
## Nonclinical tests
Following tests were made to the devices:
- IEC 60601-1, Medical Electrical Equipment - Part 1: General Requirements for Safety, 1988; Amendment 1, 1991-11, Amendment 2, 1995. (General)
- IEC 60601-1-2, Medical Electrical Equipment Part 1-2: General Requirements for ● Safety - Collateral standard: Electromagnetic Compatibility - Requirements and Tests (Edition 2:2001 with Amendment 1:2004; Edition 2.1 (Edition 2:2001 consolidated with Amendment 1:2004)). (General)
- IEC 60601-1-4:2000 Consol. Ed. 1.1. Medical electrical equipment - Part 1-4: General requirements for safety -- Collateral standard: Programmable electrical medical systems, edition 1.1. (General)
- American Thoracic Society and European Respiratory Society (Eur Respir J, 2005, ● Vol 26, pp. 948-968. No. 5 in SERIES "ATS/ERS TASK FORCE: STANDARDISATION OF LUNG FUNCTION TESTING'')
#### Conclusion for Nonclinical tests
Complete IEC 60601-1 testing was carried out for the equipment. Full test reports were made for both M929 Medikro® SpiroStar USB and M921 Medikro® SpiroStar DX (Serial). The equipment was found to comply with standard requirements.
IEC 60601-1-2 EMC testing was made for both sensor units (USB and DX). EMC testing was made according to IEC/EN 60601-1-2 requirements and includes both emissions and immunity testing. Testing was performed at Savonia Polytechnic EMC laboratory by the laboratory test engineer. Full test reports were made for both M929, Medikro® SpiroStar USB and M921, Medikro® SpiroStar DX (Serial).
IEC 60601-1-4 testing was made and the equipment was found to comply with standard requirements.
{5}------------------------------------------------
American Thoracic Society (ATS) 24 standard waveforms test was made for both units (USB and DX) and both units meet the recommendations for:
- . Measuring FVC
- Measuring FEV1
- Measuring FEF25-75% ●
- . Measuring PEF
- Resistance to flow
- Testing under BTPS conditions .
Testing was done for the Medikro Oy measurement devices against recommendations published by the American Thoracic Society. The essential performance criteria's for spirometers is well defined by the ATS. Spirometers meeting ATS criteria for essential performance are substantially equal. The predicate devices are also designed to meet these ATS criteria.
| Feature | Predicate Device<br>510(K) Number | Predicate Device<br>510(K) Number | New Device |
|---------------------------------------------|------------------------------------------------------|------------------------------------------------------|---------------------------------------------------|
| | K971336 | K052158 | Medikro® SpiroStar |
| Device<br>classification | Spirometer,<br>diagnostics | Spirometer,<br>diagnostics | Spirometer,<br>diagnostics |
| Target population | Adult and<br>pediatric patients | Adult and pediatric<br>patients | Adult and pediatric<br>patients |
| Use environment | Hospital/Clinical | Hospital/Clinical | Hospital/Clinical |
| Indices measured | FEV1, FVC,<br>FEV1/FVC,<br>PEF, FEF25-75%<br>+others | FEV1, FVC,<br>FEV1/FVC,<br>PEF, FEF25-75%<br>+others | FEV1, FVC,<br>FEV1/FVC, PEF,<br>FEF25-75% +others |
| Real-time display of<br>each blow | Yes | Yes | Yes |
| Feedback about test<br>quality | Yes | Yes | Yes |
| Flow-volume<br>displayed & printed | Yes | Yes | Yes |
| Volume-time<br>printed | Yes | Yes | Yes |
| Quantifies post<br>bronchodilator<br>change | Yes | Yes | Yes |
| Prints predicted<br>flow-volume curve | Yes | Yes | Yes |
# Key Technological Characteristics
{6}------------------------------------------------
| Feature | Predicate Device | Predicate Device | New Device |
|------------------------------------------|-------------------------------------------------|-------------------------------------------------|-------------------------------------------------|
| | 510(K) Number<br>K971336 | 510(K) Number<br>K052158 | Medikro® SpiroStar |
| Includes<br>interpretation<br>software | Yes | Yes | Yes |
| Results download<br>to clinical software | Yes | Yes | Yes |
| Patient result<br>storage capacity | Depending the size of the data<br>storage media | Depending the size of the data storage<br>media | Depending the size of the data storage<br>media |
| Portable/not<br>portable | Portable with<br>laptop PC | Portable with laptop<br>PC | Portable with mobile<br>PC |
| Power source and<br>connection to PC | PC Serial<br>connection | PC USB or Serial<br>connection | PC USB or Serial<br>connection |
| ATS compatible | Yes | Yes | Yes |
| Flow Transducer | Disposable | Disposable | Disposable |
The New Device has the same intended use and safety characteristics as the predicate devices.
### Conclusion
Based on completed performance testing our device demonstrates that it is substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.