OZT · Balloon Aortic Valvuloplasty

Cardiovascular · 21 CFR 870.1255 · Class 2

Overview

Product CodeOZT
Device NameBalloon Aortic Valvuloplasty
Regulation21 CFR 870.1255
Device ClassClass 2
Review PanelCardiovascular

Identification

A balloon aortic valvuloplasty catheter is a catheter with a balloon at the distal end of the shaft, which is intended to treat stenosis in the aortic valve when the balloon is expanded.

Classification Rationale

Class II (special controls). The special controls for this device are:

Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) The device must be demonstrated to be biocompatible. (2) Sterility and shelf life testing must demonstrate the sterility of patient-contacting components and the shelf life of these components. (3) Non-clinical performance evaluation must demonstrate that the device performs as intended under anticipated conditions of use, including device delivery, inflation, deflation, and removal. (4) In vivo evaluation of the device must demonstrate device performance, including the ability of the device to treat aortic stenosis. (5) Labeling must include a detailed summary of the device-related and procedure-related complications pertinent to the use of the device.

In combination with the general controls of the FD&C Act, the NuMED NuCLEUS and NuCLEUS-X BAV Catheters are subject to the following special controls:

Recent Cleared Devices (20 of 22)

Showing 20 most recent of 22 cleared devices.

RecordDevice NameApplicantDecision DateDecision
K253855KardiaPSI Balloon CatheterIntervention Tech, Inc. ; dba CorventionMay 6, 2026SESE
K260437Valvuloplasty Balloon Catheter (ValvuloPro)Dongguan TT Medical, Inc.Mar 20, 2026SESE
K241562OutFlo Aortic Valve Dilatation Balloon CatheterDisa MedinotecMar 11, 2025SESE
K240967ValvuloPro Valvuloplasty Balloon CatheterDongguan TT Medical, Inc.Jul 31, 2024SESE
K231814ValvuloPro Valvuloplasty Balloon CatheterDongguan TT Medical, Inc.Oct 12, 2023SESE
K220881INOUE BALLOON AToray Industries, Inc.Dec 18, 2022SESE
K172932True Flow Valvuloplasty Perfusion CatheterC.R. Bard, Inc.Dec 13, 2017SESE
K152613True Flow Valvuloplasty Perfusion CatheterC.R. Bard, Inc.Jan 19, 2016SESE
K152150V8 Balloon Aortic Valvuloplasty CatheterIntervalve, Inc.Dec 18, 2015SESE
K150667True Dilatation Balloon Valvuloplasty CatheterC.R. Bard, Inc.Jun 17, 2015SESE
K142083TRUE FLOW VALVULOPLASTY PERFUSION CATHETERC.R. Bard, Inc.Apr 17, 2015SESE
K150343V8 Transluminal BAV CatheterIntervalve, Inc.Apr 3, 2015SESE
K142199Ascendra Balloon Aortic Valvaloplasty CatheterEdwards LifesciencesOct 29, 2014SESE
K141985TRUE BA V BALLOON VALVULOPLASTY CATHETERC.R. Bard, Inc.Sep 18, 2014SESE
K140241EDWARDS TRANSFEMORAL BALLOON CATHETEREdwards Lifesciences, LLCJul 3, 2014SESE
K133607V8 TRANSLUMINAL BAV CATHETERIntervalve, Inc.Feb 11, 2014SESE
K133569TRUE DILATATION BALLOON VALVULOPLASTY CATHETER, 20MM X 4.5 CM, 22MM X 4.5CM, 24 MM X 4.5CM, 26MM X 4.5CMBard Peripheral Vascular, Inc.Dec 20, 2013SESE
K132728V8 TRANSLUMINAL BAV CATHETERIntervalve, Inc.Nov 22, 2013SESE
K123111V8 TRANSLUMINAL BAV CATHETERIntervalve, Inc.Jan 31, 2013SESE
K121083TRUE DILATATION PERCUTANEOUS TRANSLUMINAL VALVULOPLASTY CATHETER, 20MM X 4.5CM, TRUE DILATATION PERCUTANEOUS TRANSLUMINALoma Vista MedicalOct 11, 2012SESE

Top Applicants

Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...