K253855 · Intervention Tech, Inc. ; dba Corvention · OZT · May 6, 2026 · Cardiovascular
Device Facts
Record ID
K253855
Device Name
KardiaPSI Balloon Catheter
Applicant
Intervention Tech, Inc. ; dba Corvention
Product Code
OZT · Cardiovascular
Decision Date
May 6, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1255
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The KardiaPSI™ Balloon Catheter is indicated for Balloon Aortic Valvuloplasty.
Device Story
Over-the-wire coaxial catheter with non-compliant balloon at distal tip; 120cm working length; dual-lumen design (one for balloon inflation/deflation, one for guidewire). Used in clinical settings by physicians for aortic valve procedures. Operates via connection to third-party syringe inflation device to deliver radiopaque contrast media/saline mixture. Two radiopaque marker bands facilitate fluoroscopic positioning across aortic valve. Nominal inflation pressure 4ATM. Higher rated burst pressure compared to predicate. Benefits include mechanical dilation of aortic valve; device performance verified via bench testing.
Clinical Evidence
Bench testing only. Verification included dimensional, kink, tensile, corrosion, fluoroscopy, simulated use, and user validation per ISO 10555-1:2023 and ISO 10555-4:2023. Balloon testing included dimensional, inflation, deflation, fatigue, and rated burst. Sterility validated per ISO 11135:2014; packaging per ISO 11607-1:2019. Biocompatibility testing performed per ISO 10993 series (cytotoxicity, sensitization, reactivity, systemic toxicity, hemolysis, complement activation) and ASTM standards (F2382, F2888) plus in vivo thromboresistance.
Technological Characteristics
Over-the-wire coaxial catheter; non-compliant balloon; 120cm length. Materials and manufacturing similar to predicate. Nominal inflation pressure 4ATM. Radiopaque marker bands. Sterilization via ISO 11135:2014. Standards: ISO 10555-1:2023, ISO 10555-4:2023, ISO 10993 series, ASTM F2382, ASTM F2888, USP <151>, ISO 11607-1:2019.
Indications for Use
Indicated for Balloon Aortic Valvuloplasty. Patient population not specified.
Regulatory Classification
Identification
A balloon aortic valvuloplasty catheter is a catheter with a balloon at the distal end of the shaft, which is intended to treat stenosis in the aortic valve when the balloon is expanded.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) The device must be demonstrated to be biocompatible.
(2) Sterility and shelf life testing must demonstrate the sterility of patient-contacting components and the shelf life of these components.
(3) Non-clinical performance evaluation must demonstrate that the device performs as intended under anticipated conditions of use, including device delivery, inflation, deflation, and removal.
(4) In vivo evaluation of the device must demonstrate device performance, including the ability of the device to treat aortic stenosis.
(5) Labeling must include a detailed summary of the device-related and procedure-related complications pertinent to the use of the device.
In combination with the general controls of the FD&C Act, the NuMED NuCLEUS and NuCLEUS-X BAV Catheters are subject to the following special controls:
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 6, 2026
Intervention Tech, Inc. ; dba Corvention
% Natasha Bond
Regulatory Consultant to Corvention
Ikigai Medical Development
2198 Zinnia Way
Golden, Colorado 80401
Re: K253855
Trade/Device Name: KardiaPSI™
Regulation Number: 21 CFR 870.1255
Regulation Name: Balloon Aortic Valvuloplasty Catheter
Regulatory Class: Class II
Product Code: OZT
Dated: April 6, 2026
Received: April 6, 2026
Dear Natasha Bond:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K253855 - Natasha Bond
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{3}
K253855 - Natasha Bond
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JULIE B. MACKEL -S
Digitally signed by
JULIE B. MACKEL -S
Date: 2026.05.06
15:34:56 -04'00"
For
Jennifer Kevit, PhD
Acting Assistant Director
DHT2B: Division of Circulatory Support, Structural, and Vascular Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K253855 | |
| Device Name KardiaPSI™ Balloon Catheter | |
| Indications for Use (Describe) The KardiaPSI™ Balloon Catheter is indicated for Balloon Aortic Valvuloplasty. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{5}
K253855
# KardiaPSI™ Balloon Catheter 510(k) Summary
Applicant Contact Details (refer 21 CFR 807.92(a)(1)):
Applicant Name: Intervention Tech, Inc; dba Corvention
Applicant Address: 2201 N. Gemini Dr, Ste 138
Applicant Contact Phone # 928-707-4031
Applicant Contact: Mr. Ryan Smith, Vice President of Engineering
Applicant Contact email rsmith@corvention.com
Subject Device (refer 21 CFR 807.92(a)(2)):
Trade Name: KardiaPSI™ Balloon Catheter
Common Name: Valvuloplasty Balloon Catheter
Classification Name: Balloon Aortic Valvuloplasty Catheter
Submission #: K253855
Submission Type: Traditional 510(k)
Product Regulation #: 21-CFR-870.1255
Product Code: OZT
Legally Marketed Predicate Device (refer 21 CFR 807.92(a)(3)):
Predicate 510(k)#: K122012
Predicate Trade name: Z-Med & Z-Med-II Balloon catheters
Predicate Sponsor: B. Braun
Predicate Product Code: OZT
Corvention Inc
Page 1 of 3
{6}
K253855
Device Description (refer 21 CFR 807.92(a)(4)):
The KardiaPSI™ Balloon Catheter is an over-the-wire coaxial catheter with a non-compliant balloon fixed at the distal tip. The catheter working length is 120cm long and has two lumens: one lumen is used to inflate and deflate the balloon (closed ended) and the other permits the use of a guidewire to position the catheter (open ended). The balloon inflation luer-lock hub (angled) connects to a currently marketed third-party syringe inflation device to deliver radiopaque contrast media (diluted with saline) for inflation. The guidewire luer-lock hub (straight) connects to the guidewire lumen. The balloon is non-compliant and is designed to reach a known diameter and length when inflated to the specified nominal inflation pressure of 4ATM (common across all sizes of balloons in the product matrix). Two radiopaque marker bands are provided for fluoroscopic positioning of the device across the aortic valve. These bands are positioned at the proximal and distal balloon shoulders.
Intended Use / Indication for Use (refer 21 CFR 807.92(a)(5)):
The KardiaPSI™ Balloon Catheter is indicated for Balloon Aortic Valvuloplasty.
Indication for Use Comparison (refer 21 CFR 807.92(a)(5)):
A comparison of the characteristics of the proposed device and the predicate device shows there are no differences between the subject device and the predicate device with respect to clinical aspects, indications and intended use. Both devices have the same principle and mechanism of operation. The KardiaPSI™ Balloon Catheter is determined to be substantially equivalent to the predicate device.
Technological Comparison (refer 21 CFR 807.92(a)(6)):
Technological aspects were assessed for substantial equivalence. The subject device is available in a range of sizes that are completely within the dimensional range of the sizes offered by the predicate device for both catheter and balloon dimensions. The shape, materials of construction and manufacturing processes are similar to the predicate device. The catheter configuration, user interface and nominal inflation pressures are all comparable to predicate. The balloon rated burst pressure for the subject device is greater than for the predicate. This technological difference has been tested against the recognized methods per ISO10555-4:2023 and does not raise new questions of safety or effectiveness.
Corvention Inc
{7}
K253855
Non-Clinical and / or Clinical Tests Summary & Conclusions (refer 21 CFR 807.92(b)):
Bench Verification for the catheter including dimensional verification, kink, tensile, corrosion, fluoroscopy, simulated use and user validation, per ISO 10555-1: 2023 and ISO 10555-4: 2023.
Bench verification for the balloon included dimensional, inflation, deflation, fatigue and rated burst. per ISO 10555-1: 2023 and ISO 10555-4: 2023.
Sterility Validation per ISO 11135: 2014. Sterile barrier packaging effectiveness was confirmed per ISO 11607-1:2019.
Biocompatibility testing was completed:
- Cytotoxicity per ISO-10993-5
- Sensitization per ISO-10993-10
- Intracutaneous Reactivity per ISO-10993-23
- Systemic Toxicity per ISO-10993-11
- Pyrogenicity per USP <151>
- Hemolysis per ISO-10993-4
- Complement Activation per ISO-10993-4
- Partial Thromboplastin Time per ASTM F2382
- Platelet & Leukocyte Count Assay per ASTM F2888
- In Vivo Thromboresistance Study per ISO-10993-4
Conclusion:
Corvention considers the KardiaPSI™ Balloon Catheter to be substantially equivalent to the predicate device. This conclusion is based upon the fact that the devices have identical Intended Use, and there are no clinical, technical or biological differences that, when tested against recognized consensus standards, raise new questions of safety or effectiveness. Nonclinical testing demonstrated that the KardiaPSI™ Balloon Catheter performs substantially similar to the predicate device, except in providing a higher rated burst pressure and does not introduce new questions of safety or effectiveness.
Corvention Inc
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.