K220881 · Toray Industries, Inc. · OZT · Dec 18, 2022 · Cardiovascular
Device Facts
Record ID
K220881
Device Name
INOUE BALLOON A
Applicant
Toray Industries, Inc.
Product Code
OZT · Cardiovascular
Decision Date
Dec 18, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1255
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
INOUE BALLOON A is intended for balloon aortic valvuloplasty (BAV) in patients with aortic stenosis.
Device Story
Inoue Balloon A is a balloon aortic valvuloplasty catheter used for treating aortic stenosis. The device is inserted percutaneously over a guidewire. It features a coaxial shaft and a distally mounted balloon composed of a latex and fiber mesh complex. The operator inflates the balloon using a diluted contrast medium via a syringe and extension tube; the balloon undergoes staged inflation from an hourglass to a barrel shape. A radiopaque marker assists in fluoroscopic positioning across the aortic valve. The balloon is semi-compliant, with diameter controlled by infusion volume. The device is intended for use by clinicians in a hospital setting to dilate the aortic valve, potentially improving valve function and patient outcomes.
Clinical Evidence
Bench testing only. Performance data included radio-detectability, surface properties, corrosion resistance, tensile force, leakage, hub integrity, size designation, guidewire compatibility, sheath compatibility, burst strength, fatigue, inflation/deflation times, torquing, bending, over-inflation, and catheter removal. Biocompatibility testing was performed per ISO 10993-1 for short-term blood contact (<24 hours).
Technological Characteristics
Coaxial shaft construction; distally mounted semi-compliant balloon made of latex and fiber mesh complex. Radiopaque marker for fluoroscopy. Sterilized by ethylene oxide (SAL 10^-6). Biocompatibility per ISO 10993-1.
Indications for Use
Indicated for balloon aortic valvuloplasty (BAV) in patients with aortic stenosis.
Regulatory Classification
Identification
A balloon aortic valvuloplasty catheter is a catheter with a balloon at the distal end of the shaft, which is intended to treat stenosis in the aortic valve when the balloon is expanded.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) The device must be demonstrated to be biocompatible.
(2) Sterility and shelf life testing must demonstrate the sterility of patient-contacting components and the shelf life of these components.
(3) Non-clinical performance evaluation must demonstrate that the device performs as intended under anticipated conditions of use, including device delivery, inflation, deflation, and removal.
(4) In vivo evaluation of the device must demonstrate device performance, including the ability of the device to treat aortic stenosis.
(5) Labeling must include a detailed summary of the device-related and procedure-related complications pertinent to the use of the device.
In combination with the general controls of the FD&C Act, the NuMED NuCLEUS and NuCLEUS-X BAV Catheters are subject to the following special controls:
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 18, 2022
Toray Industries, Inc. % Yuya Shizume Manager Toray Industries (America), Inc. 461 Fifth Avenue, 9th Floor New York, New York 10017
Re: K220881
Trade/Device Name: Inoue Balloon A Regulation Number: 21 CFR 870.1255 Regulation Name: Balloon aortic valvuloplasty catheter Regulatory Class: Class II Product Code: OZT Dated: November 10, 2022 Received: November 14, 2022
Dear Mr. Shizume:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jaime Raben -S
Jaime Raben, Ph.D. Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220881
Device Name INOUE BALLOON A
Indications for Use (Describe)
INOUE BALLOON A is intended for balloon aortic valvuloplasty (BAV) in patients with aortic stenosis.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
#### I. SUBMITTER
Submission sponsor Toray Industries, Inc. 1-1, Nihonbashi-muromachi 2-chome, Chuo-ku, Tokyo 103-8666, JAPAN Phone: +81-3-3245-8606 Fax: +81-3-3245-8612 Email: ra-global-md.toray.ug(@mail.toray Contact: Satoshi Kakuyama, General Manager, Global Development and Regulatory Affairs Dept.
Submission correspondent Toray Industries (America), Inc. 461 Fifth Avenue, 9th Floor New York, NY 10017 Phone: 212-922-3711 Fax: 212-972-4279 Email: medical.tam.mb(@mail.toray Contact: Shunsuke Kobayashi, Director, Life Science
Date Prepared: March 11, 2022
# II. DEVICE
Name of Device: INOUE BALLOON™ A Common or Usual Name: Balloon Aortic Valvuloplasty Catheter Classification Name: Catheter, Balloon Aortic Valvuloplasty (21 CFR 870.1255) Regulatory Class: II Product Code: OZT
#### III. PREDICATE DEVICE
NuMED NuCLEUS and NuCLEUS-X BAV Catheters, K082776/DEN080015
{5}------------------------------------------------
## IV. DEVICE DESCRIPTION
INOUE BALLOON™ A is intended for the treatment of patients with aortic valvuloplasty (BAV). The device is transported over a guidewire as it is inserted through a percutaneous entry site and expanded with a predetermined amount of the diluted contrast medium by use of specified syringe with extension tube connected to the balloon inflation luer-lock hub. This results in the staged inflation of the balloon from hour-glass to barrel shape. The balloon is stretched and made thinner by pushing the inner tube in during passing thorough the introducer sheath. A radiopaque marker is provided for fluoroscopic positioning of the device across the valve.
## V. INDICATIONS FOR USE
INOUE BALLOON™ A is intended for balloon aortic valvuloplasty (BAV) in patients with aortic stenosis.
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The INOUE BALLOON™ A is substantially equivalent in both function and use to the predicate devices, NuMED NuCLEUS and NuCLEUS-X BAV Catheters (K082776/DEN080015).
The subject device has the same intended use as and has equivalent technology to the predicate device. The predicate device is a Class II legally marketed device.
The subject device has the same construction as the predicate device; that has a coaxial shaft construction with a distally mounted balloon of the subject device consists of a complex of latex and the fiber mesh has semi-compliant characteristics which differs from non-compliant characteristics of the predicate device. Due to the semi-compliant balloon characteristics, the diameter of inflated balloon is controlled by the infusion volume. Since the balloon burst strength is almost the same as that of the predicate device, it can be said that the function of the subject device as a balloon catheter for aortic valve dilation is substantially equivalent to that of the predicate device.
{6}------------------------------------------------
# VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
- Radio-detectability
- Surface ●
- Corrosion resistance
- Peak tensile force
- Freedom from leakage
- Hubs
- Designation of nominal size
- Diameter of the largest guidewire
- Sheath compatibility
- Balloon minimum burst strength test
- Balloon fatigue test for freedom from leakage and damage on inflation
- Balloon inflation and deflation times test
- Torquing
- Balloon preparation test
- Bending test
- Balloon over inflation test
- Catheter removal test
- Usability evaluation
In addition, the device was tested for biocompatibility per ISO 10993-1 for short duration contact with blood (<24 hours). The device is sterilized by ethylene oxide to an SAL 10tº level. These performances are similar to that described by the predicate device.
The preclinical testing showed that the device meets specifications before and after aging indicating that the device is as safe and effective as the predicate device.
# VIII. CONCLUSIONS
The INOUE BALLOONTM A is substantially equivalent to the predicate devices (NuMed NuCLEUS and NuCLEUS-X BAV Catheters).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.