INO · Chair, Positioning, Electric

Physical Medicine · 21 CFR 890.3110 · Class 2

Overview

Product CodeINO
Device NameChair, Positioning, Electric
Regulation21 CFR 890.3110
Device ClassClass 2
Review PanelPhysical Medicine

Identification

An electric positioning chair is a device with a motorized positioning control that is intended for medical purposes and that can be adjusted to various positions. The device is used to provide stability for patients with athetosis (involuntary spasms) and to alter postural positions.

Classification Rationale

Class II. The electric positioning chair is exempt from premarket notification procedures in subpart E of part 807 of this chapter, subject to § 890.9 and the following conditions for exemption:

Special Controls

(1) Appropriate analysis and non-clinical testing must demonstrate that the safety controls are adequate to ensure safe use of the device and prevent user falls from the device in the event of a device failure; (2) Appropriate analysis and non-clinical testing must demonstrate the ability of the device to withstand the rated user weight load with an appropriate factor of safety; (3) Appropriate analysis and non-clinical testing must demonstrate the longevity of the device to withstand external forces applied to the device and provide the user with an expected service life of the device; (4) Appropriate analysis and non-clinical testing must demonstrate proper environments of use and storage of the device to maximize the longevity of the device; (5) Appropriate analysis and non-clinical testing (such as that outlined in the currently FDA-recognized editions of ANSI/AAMI/ES60601-1, “Medical Electrical Equipment—Part 1: General Requirements for Basic Safety and Essential Performance,” and ANSI/AAMI/IEC 60601-1-2, “Medical Electrical Equipment—Part 1-2: General Requirements for Basic Safety and Essential Performance—Collateral Standard: Electromagnetic Disturbances—Requirements and Tests”) must validate electromagnetic compatibility and electrical safety; (6) Appropriate analysis and non-clinical testing (such as that outlined in the currently FDA-recognized editions of ANSI/AAMI/ISO 10993-1, “Biological Evaluation of Medical Devices—Part 1: Evaluation and Testing Within a Risk Management Process,” ANSI/AAMI/ISO 10993-5, “Biological Evaluation of Medical Devices—Part 5: Tests for In Vitro Cytotoxicity,” and ANSI/AAMI/ISO 10993-10, “Biological Evaluation of Medical Devices—Part 10: Tests for Irritation and Skin Sensitization”) must validate that the skin-contacting components of the device are biocompatible; (7) Appropriate analysis and non-clinical testing (such as that outlined in the currently FDA-recognized editions of IEC 62304, “Medical Device Software—Software Life Cycle Processes”) must validate the software life cycle and that all processes, activities, and tasks are implemented and documented; (8) Appropriate analysis and non-clinical testing must validate that the device components are found to be non-flammable; (9) Appropriate analysis and non-clinical testing must validate that the battery in the device (if applicable) performs as intended over the anticipated service life of the device; and (10) Adequate patient labeling is provided to the user to document proper use and maintenance of the device to ensure safe use of the device by the patient in the intended use environment.

Recent Cleared Devices (20 of 53)

Showing 20 most recent of 53 cleared devices.

RecordDevice NameApplicantDecision DateDecision
K131909DALTON LIFT CHAIRDalton Instrument Corp.Feb 24, 2014SESE
K123834EASYSTAND BANTAM MEDIUMAltimate Medical, Inc.Aug 13, 2013SESE
K112944FRESENIUS MEDICAL TREATMENT CHAIR T688 SERIESFresenius Medical Care South Asia Pacific Pty , Ltd.Dec 30, 2011SESE
K101723COMPACT WHEELCHAIR PLATFORMDesign Specific FletcherDec 23, 2010SESE
K091578LIFT CHAIR, MODEL T3Pride Mobility Products Corp.Dec 2, 2009SESE
K091727MEDICAL TREATMENT CHAIR, T400, T500,T600Fresenius Medical Care South Asia Pacific Pty , Ltd.Jul 6, 2009SESE
K091242BANTAMAltimate Medical, Inc.Jul 6, 2009SESE
K073565STANDARD WHEELCHAIR PLATFORMDesign Specific , Ltd.Feb 6, 2009SESE
K081053PIVOT ASSISTPivot Assist, LLCOct 9, 2008SESE
K080533WINCO MODEL POWER SERIES LIFT CHAIRWinco, Inc.Aug 11, 2008SESE
K071793CONVERTIBLE I-SERIES POSITIONING AND TRANSFER CHAIRBarton Medical Corp.Aug 3, 2007SESE
K070950LIFT CHAIR, MODEL C5Pride Mobility Products Corp.Apr 16, 2007SESE
K062402EASYSTAND EVOLVInvacare Corp.Sep 21, 2006SESE
K060212MERITS MODEL E SERIES LIFT CHAIRSMerits Health Products Co., Ltd.Mar 7, 2006SESE
K031045AEROLET TOILET LIFT, MODEL AER-01Economic Holland B.V.Jul 10, 2003SESE
K023521EASY LIFEBerkline Corp.Nov 19, 2002SESE
K022407MULTI FUCTIONAL REHABILITATION CHAIRRaye'S, Inc.Aug 16, 2002SESE
K002318UNITED #1018-FREEDOM SERIES, UNITED #2018-COMFORT SERIES, UNITED #3018-ELEGANCE SERIES, UNITED #5018-MASTER SERIESUnited Chair Mfg., LLCOct 11, 2000SESE
K001937MEDLINE ELECTRONIC POSITIONING CHAIRMedline Industries, Inc.Sep 22, 2000SESE
K002171INVACARE LIFT CHAIRInvacare Corp.Aug 10, 2000SESE

Top Applicants

Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...