CONVERTIBLE I-SERIES POSITIONING AND TRANSFER CHAIR
K071793 · Barton Medical Corp. · INO · Aug 3, 2007 · Physical Medicine
Device Facts
Record ID
K071793
Device Name
CONVERTIBLE I-SERIES POSITIONING AND TRANSFER CHAIR
Applicant
Barton Medical Corp.
Product Code
INO · Physical Medicine
Decision Date
Aug 3, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3110
Device Class
Class 2
Attributes
Therapeutic, Pediatric, 3rd-Party Reviewed
Indications for Use
The Convertible® I-1000 Positioning & Transfer Chair is an electric positioning chair with motorized positioning control that can be adjusted to various positions. The device is used to alter postural positions and to provide a means of transferring a patient horizontally from a height adjustable bed or trolley, to the chair and vice versa. The chair can then be transformed from a supine position into the sitting position without ever lifting the patient. This device may be used on a wide range of patients, including adults and children up to 1000 pounds.
Device Story
Electric positioning and transfer chair; steel tubular construction with foam-filled upholstery. Operates via hand-held push-button controller; two DC linear actuators enable transition between horizontal supine and sitting positions, plus independent tilt-in-space inclination. Powered by 24-volt rechargeable battery system. Used in clinical or home settings to facilitate horizontal patient transfer from height-adjustable beds or trolleys without patient suspension or manual lifting. Benefits include reduced physical strain on caregivers and improved patient safety during transfers. Healthcare providers operate the device to position patients for mobility or care.
Clinical Evidence
No clinical data. Bench testing only. Device underwent load and repeatability testing to verify functional performance. Biocompatibility testing for skin irritation/sensitization performed per ISO 10993-10; results categorized as negligible. Electrical safety and EMC verified per IEC 60601-1 and IEC 60601-1-2.
Technological Characteristics
Steel tubular frame; foam-filled upholstery. Dual DC linear actuators; 24V rechargeable battery power. Dimensions/form factor: positioning chair. Connectivity: none. Sterilization: N/A. Standards: IEC 60601-1, IEC 60601-1-2, ISO 10993-10, JIS T9201, EN 12531-12, JIS Z0237.
Indications for Use
Indicated for adults and children up to 1000 lbs requiring postural adjustment or horizontal transfer between a height-adjustable bed/trolley and the chair without manual lifting.
Regulatory Classification
Identification
An electric positioning chair is a device with a motorized positioning control that is intended for medical purposes and that can be adjusted to various positions. The device is used to provide stability for patients with athetosis (involuntary spasms) and to alter postural positions.
Special Controls
(1) Appropriate analysis and non-clinical testing must demonstrate that the safety controls are adequate to ensure safe use of the device and prevent user falls from the device in the event of a device failure;
(2) Appropriate analysis and non-clinical testing must demonstrate the ability of the device to withstand the rated user weight load with an appropriate factor of safety;
(3) Appropriate analysis and non-clinical testing must demonstrate the longevity of the device to withstand external forces applied to the device and provide the user with an expected service life of the device;
(4) Appropriate analysis and non-clinical testing must demonstrate proper environments of use and storage of the device to maximize the longevity of the device;
(5) Appropriate analysis and non-clinical testing (such as that outlined in the currently FDA-recognized editions of ANSI/AAMI/ES60601-1, “Medical Electrical Equipment—Part 1: General Requirements for Basic Safety and Essential Performance,” and ANSI/AAMI/IEC 60601-1-2, “Medical Electrical Equipment—Part 1-2: General Requirements for Basic Safety and Essential Performance—Collateral Standard: Electromagnetic Disturbances—Requirements and Tests”) must validate electromagnetic compatibility and electrical safety;
(6) Appropriate analysis and non-clinical testing (such as that outlined in the currently FDA-recognized editions of ANSI/AAMI/ISO 10993-1, “Biological Evaluation of Medical Devices—Part 1: Evaluation and Testing Within a Risk Management Process,” ANSI/AAMI/ISO 10993-5, “Biological Evaluation of Medical Devices—Part 5: Tests for In Vitro Cytotoxicity,” and ANSI/AAMI/ISO 10993-10, “Biological Evaluation of Medical Devices—Part 10: Tests for Irritation and Skin Sensitization”) must validate that the skin-contacting components of the device are biocompatible;
(7) Appropriate analysis and non-clinical testing (such as that outlined in the currently FDA-recognized editions of IEC 62304, “Medical Device Software—Software Life Cycle Processes”) must validate the software life cycle and that all processes, activities, and tasks are implemented and documented;
(8) Appropriate analysis and non-clinical testing must validate that the device components are found to be non-flammable;
(9) Appropriate analysis and non-clinical testing must validate that the battery in the device (if applicable) performs as intended over the anticipated service life of the device; and
(10) Adequate patient labeling is provided to the user to document proper use and maintenance of the device to ensure safe use of the device by the patient in the intended use environment.
{0}------------------------------------------------
K071793
AUG - 3 2007
Image /page/0/Picture/1 description: The image shows the logo for BARTON MEDICAL CORPORATION. The logo is in black and white and features the company name in a bold, sans-serif font. The word "BARTON" is on the top line, and "MEDICAL CORPORATION" is on the bottom line in a smaller font. The logo has a slightly distressed or textured appearance.
www.bartonmedical.com
5725 Hwy 290 West, Suite 103 Austin, TX 78735 Tel: (512) 476-7199 Fax: (512) 476-7190
## 510(k) Summary
Submitter Information:
Barton Medical Corporation 5725 Hwy 290 West Suite #103 Austin, TX 78735 512-176-7199 Phone 512-476-7190 Fax
Contact Person:
Lawrence W. Swol Chief Quality Officer 512-476-7199 Phone 512-476-7190 Fax Larry. Swol@BartonMedical.com July 19, 2007
Manufacture:
Takano Co., Ltd. 137 Miyada-Mura, Kamuna-Gun Nagano 399-4301 Japan
Device Information:
Trade/Proprietary Name: I-1000 Convertible® I-Series Positioning and Common/Usual Name: Transfer Chair Chair, Electric, Positioning Classification Name:
Legally Marketed Predicate Device:
Altimate Medical EasyStand Evolv; K062402, Sept. 21, 2006 Invacare Model LC Series Lift-Out Chair, K002171, Aug. 10, 2000
Device Description:
The Convertible® Positioning and Transfer Chair construction is made of steel tubular components, on to which is fitted foam filled upholstery. The mechanism allows the Convertible® Positioning and Transfor Chair to transform from the horizontal supine position into a chair, which can then be tilted. This is accomplished through the
510 (k) Summary
{1}------------------------------------------------
hand-held controller, which in turn operates two DC linear actuators. The first DC actuator allows the Chair and patient profile to change from a sitting position to a horizontal or supine position suitable for transferring. The second DC linear actuator allows the I-1000 to be independently inclined to the rear (Tilt-in-Space). A hand-held push button controller is used to engage actuator motion and vary the I-1000's position. The electric system is composed of a 24-volt rechargeable battery, an external battery charger, and a control unit. Recharging of the battery is accomplished via the external battery charger.
The Convertible® Positioning & Transfer Chair is ideal for moving those patients who are unable to move themselves. The Convertible® Positioning and Transfer Chair have been specifically designed to provide a means of transferring a patient horizontally from a height adjustable bed or trolley, to the chair and vice versa. The patient will always be supported either on the bed and/or the chair and is never suspended. To achieve this, The Barton® Positioning and Transfer System (PTS®) must be used.
#### Intended Use:
The Convertible® I-1000 Positioning & Transfer Chair is an electric positioning chair with motorized positioning control that can be adjusted to various positions. The device is used to alter postural positions and to provide a means of transferring a patient horizontally from a height adjustable bed or trolley, to the chair and vice versa. The chair can then be transformed from a supine position into the sitting position without ever lifting the patient. This device may be used on a wide range of patients, including adults and children up to 1000 pounds.
#### Substantial Equivalence:
The Convertible® I-1000 Positioning and Transfer Chair and the predicate devices have similar, and in many cases, the same:
- Intended use,
- Basic construction,
- Principals of operation,
- Electrical and mechanical characteristics, and
- General safety and EMC compliance.
#### 510 (k) Summary
5-3
{2}------------------------------------------------
## Performance Standards
Although no performance standards or special controls have been developed under Section 514 of the FDC Act for electric positioning chairs, Barton Medical Corporation has chosen to test the Convertible® I-1000 Positioning and Transfer Chair against self imposed load and repeatability test requirements. Representative sample for the device underwent load and repeatability testing to verify functional and performance characteristics.
#### Biocompatibility
Materials that may contact patients used on the Convertible® I-1000 Positioning and Transfer Chair are biocompatible. The material was evaluated for primary skin irritation in accordance with in the guidelines of the requirements of ISO 10993: Biological Evaluation of Medical Devices, Part 10: Tests for Irritation and Sensitization. Under the conditions of the study, no erythema and no edema were observed. The response of the test article was categorized as negligible.
### Electromagnetic Compatibility and Electrical Safety
The Convertible® I-1000 I-Series Positioning and Transfer Chair meet the applicable requirements of IEC 60601-1 General Safety and IEC 60601-1-2 EMC.
#### Cited Standards
The Convertible® I-1000 Positioning and Transfer Chair conform to relevant portions of applicable guidance and standards that include the following:
- IEC 60601-1:1998 General Requirements for Safety
- IEC 60601-1-2:2001 Medical Electrical Equipment, -Electromagnetic Compatibility
- JIS T9201:1998 Manually Propelled Wheelchairs -
- ISO 10993-10:2003 Biological Evaluation of Medical Devices -– Part 10: Tests for Irritation and Sensitization
- EN 12531-12:1999 Castors and wheels. Hospital bed castors. ー Test condition 12: Test of castors for hospital beds
JIS Z0237:2000 Adhesive Strength Measurement Conditions 】 510(k) Summary చ్-4
{3}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Barton Medical Corporation % Intertek Testing Services NA, Inc. 2307 East Aurora Road Unit B7 Twinsburg, Ohio 44087 ATTN: Daniel W. Lehtonen
AUG - 3 2007
Re: K071793
Trade/Device Name: Convertible® I-Series (I-1000) Positioning and Transfer Chair Regulation Number: 21 CFR 890.3110 Regulation Name: Electric Positioning Chair Regulatory Class: Class II Product Code: INO. FMR Dated: July 24, 2007 Received: July 25, 2007
Dear Mr. Lehtonen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance, of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 -- Mr. Daniel W. Lehtonen
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
For
ark N. Melkerson
Mark N. Mel Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
## 510(k) Number (if known): K071793
Device Name:
Convertible® I-Series Positioning and Transfer Chair, Model I-1000
Indications for Use:
The Convertible® I-1000 I-Series Positioning and Transfer Chair is an electric positioning chair with motorized positioning control that can be adjusted to various positions. The device is used to alter postural positions and to provide a means of transferring a patient horizontally from a height adjustable bed or trolley, to the chair and vice versa. The chair can then be transformed from a supine position into the sitting position without ever lifting the patient.
This device may be used on a wide range of patients, including adults and children up to 1000 pounds.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use × (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|---------|
| | |
| (Division Sign-Off) | |
| Division of General, Restorative | |
| and Neurological Devices | Page |
| 510(k) Number | K071793 |
4-2
1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.