The BIOTEQUE Hemodialysis Blood Tubing Set(Models BT-102A, BT-102B and BT-190) is intended to serve as a conduit through which blood is taken from the patient, delivered to the dialyzer, and returned to the patient during hemodialysis therapy. The device is intended to be used for patients suffering from renal disease only. USERS TO INSTALL THE DEVICE: Trained nurses or the doctors. ENVIRONMENT FOR THE DEVICE TO BE USED: The hemodialysis center. SPECIAL NOTES: The hemodialysis blood tubing sets must be installed by trained nurses and doctors. The patients can not influence the use of the device.
Device Story
Bioteque Hemodialysis Blood Tubing Set (Models BT-102A, BT-102B, BT-190) functions as a conduit for extracorporeal blood circulation; transports blood from patient to dialyzer and returns it to patient during hemodialysis. Used exclusively in hemodialysis centers; installed by trained nurses or physicians. Device operation is entirely controlled by clinical staff; patient has no influence on device function.
Clinical Evidence
Bench testing only. Device conforms to ISO 10993 series (biocompatibility), ISO 11607-1 (packaging), ISO 11135 (sterilization), and USP pyrogenic standards.
Technological Characteristics
Hemodialysis blood tubing set consisting of arterial and venous lines. Conforms to ISO 10993, ISO 11607-1, ISO 11135, and USP standards. Sterilized device.
Indications for Use
Indicated for patients with renal disease requiring hemodialysis therapy.
Regulatory Classification
Identification
A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
{0}------------------------------------------------
### 510(K) SUMMARY 9 2000 FEB
K992930
.his summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA and 21 CFR §807.92
1. Submitter's Name: BIOTEQUE CORPORATION
| Address: | Suite 402, 4th Fl, No. 136, Sec.3,Jen-Ai Road, Taipei, R.O.C. |
|----------|---------------------------------------------------------------|
| Phone: | 886-2-2708-6716 |
| Fax: | 886-2-2707-6610 |
| Contact: | Mr. William Lee (General Manager) |
2. Device Name
| Trade Name: | BIOTEQUE HEMODIALYSIS BLOOD TUBING SET |
|----------------------|---------------------------------------------------------------|
| Common Name: | HEMODIALYSIS BLOOD TUBING SET |
| Classification name: | SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE |
3. Classification:
| Classification: | Class II |
|-----------------|----------|
|-----------------|----------|
4. Predicate Device:
| Predicate Device: | FRESENIUS DISPOSABLE BLOOD LINES (K853851) |
|-------------------|--------------------------------------------|
|-------------------|--------------------------------------------|
5. Device Description:
| Device Description: | Bioteque Hemodialysis Blood Tubing Set, consists of Arterial Line & Venous Line, is used as the conduit through which blood is taken from the patient, delivered to the dialyzer, and returned to the patient during dialysis therapy. |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
6. Intended Use:
| Intended Use: | INTENDED USE:<br>The BIOTEQUE Hemodialysis Blood Tubing Set(Models BT-102A, BT-102B and BT-190) is intended to serve as a conduit through which blood is taken from the patient, delivered to the dialyzer, and returned to the patient during hemodialysis therapy. The device is intended to be used for patients suffering from renal disease only.<br>USERS TO INSTALL THE DEVICE:<br>Trained nurses or the doctors.<br>ENVIRONMENT FOR THE DEVICE TO BE USED:<br>The hemodialysis center.<br>SPECIAL NOTES:<br>The hemodialysis blood tubing sets <i>must</i> be installed by trained nurses and doctors.<br>The patients can not influence the use of the device. |
|---------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|---------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
FDA 510(K) SUMMARY 【A2】 Page Revision
Page 1 of 2
{1}------------------------------------------------
7. Performance Summary: In terms of Physical specification, Chemical specification, Biological specification & Sterilization Specification, the device conforms to applicable standards included ISO 10993 series, ISO 11607-1, ISO 11135, USP Pyrogenic standards & related standards ---- etc.
#### 8. Conclusions:
The BIOTEQUE HEMODIALYSIS BLOOD TUBING SET have the same intended use and similar technological characteristics as the FRESENIUS DISPOSABLE BLOOD LINES ( K853851 ). Moreover, bench testing contained in this submission demonstrate that any differences in their technological characteristics do not raise any new questions of safety or effectiveness. Thus, the BIOTEQUE HEMODIALYSIS BLOOD TUBING SET is substantially equivalent to the predicate devices.
FDA 510(K) SUMMARY
Page Revision 【A2】
{2}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
9 2000 FEB
Bioteque Corporation c/o Mr. Allen Reich Harvest Consulting, Inc. 900 N. Switzer Canyon Dr., #142 Flaggstaff, AZ 86001
Re: K992930
Bioted® Hemodialysis Blood Tubing Sets Models BT-102A; BT-102B; and, BT-190 Dated: December 11, 1999 Received: December 14, 1999 Regulatory Class: II 21 CFR §876.5820/Procode: 78 FJK
Dear Mr. Reich:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Reculations, Title 21. Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat. and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
| | 1<br>2 | 1 |
|-------|--------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ----- | | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>- |
#### NUMBER (IF KNOWN): 510 (K)
## BIOTEQ Hemodialysis Blood Tubing Set DEVICE NAME: BIOTEQUE CORPORATION
# INDICATIONS FOR USE:
The BIOTEQ Hemodialysis Blood Tubing Set (Models BT-102A, BT-102B, and BT-190) is intended to serve as a conduit through which blood is taken from the patient, delivered to the dialyzer, and returned to the patient during hemodialysis therapy. The device is intended to be used for patients suffering from renal disease only.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation |
|--------------------------------------------------|
|--------------------------------------------------|
| Prescription Use (Per 21 CFR 801.109) | OR Over-The-Counter (Optional Format) |
|---------------------------------------|---------------------------------------|
|---------------------------------------|---------------------------------------|
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
| 510(k) Number | K992930 |
|---------------|---------|
|---------------|---------|
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.