EPICON (CARBOSILFOCON A) RIGID GAS PERMEABLE CONTACT LENS DAILY WEAR
K992787 · Specialty Ultravision, Inc. · HQD · Feb 2, 2000 · Ophthalmic
Device Facts
Record ID
K992787
Device Name
EPICON (CARBOSILFOCON A) RIGID GAS PERMEABLE CONTACT LENS DAILY WEAR
Applicant
Specialty Ultravision, Inc.
Product Code
HQD · Ophthalmic
Decision Date
Feb 2, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5916
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The EpiCon K (carbosilfocon A) Rigid Gas Permeable Contact Lens for Daily Wear is indicated for daily wear in aphakic or not-aphakic persons exhibiting the irregular corneal surface associated with keratoconus and/or requiring keratoconus management with otherwise nondiseased eyes. The lens may be prescribed for the correction of refractive ametropia (myopia, hyperopia and astigmatism). The lenses may be disinfected using a chemical disinfecting system.
Device Story
EpiCon K is a rigid gas permeable (RGP) contact lens designed for daily wear in patients with keratoconus or refractive ametropia. The lens is a hemispherical shell manufactured from carbosilfocon A (polymethyl methacrylate combined with a silicone-containing copolymer) using a gel flow molding process. It features a blue handling tint (D&C Green No. 6). The device is prescribed by eye care professionals to correct vision and manage irregular corneal surfaces. It functions by providing a stable refractive surface over the irregular cornea. Patients use the lens daily and maintain it using chemical disinfection systems. The device benefits patients by improving visual acuity and managing keratoconus-related corneal irregularities.
Clinical Evidence
Four-week prospective clinical evaluation at five sites with 38 keratoconic subjects. 18 subjects (47.4%) completed the study; 20 (52.6%) discontinued. Primary endpoints included slit lamp findings, subject symptoms, keratometry, refractive changes, visual acuity, and lens comfort. Results indicated performance within expected ranges for RGP lenses in keratoconic eyes. Two adverse reactions occurred, both resolving without sequelae. Clinical performance was determined to be substantially equivalent to the predicate.
Technological Characteristics
Material: carbosilfocon A (PMMA/silicone copolymer); Hydrophobic Lens Group 2. Dimensions: 10.5-14.0 mm chord diameter, 0.10-0.35 mm center thickness, 6.0-8.5 mm base curve. Power: Plano to -20.00 D. Dk (35°C): 52 x 10^-11. Refractive index: 1.49 at 20°C. Specific gravity: 1.105. Wetting angle: < 19°. Manufacturing: Gel Flow Molding. Sterilization: Chemical disinfection system.
Indications for Use
Indicated for daily wear in aphakic or non-aphakic persons with irregular corneal surfaces due to keratoconus or requiring keratoconus management in otherwise nondiseased eyes; also for correction of myopia, hyperopia, and astigmatism.
Regulatory Classification
Identification
A rigid gas permeable contact lens is a device intended to be worn directly against the cornea of the eye to correct vision conditions. The device is made of various materials, such as cellulose acetate butyrate, polyacrylate-silicone, or silicone elastomers, whose main polymer molecules generally do not absorb or attract water.
Predicate Devices
UltraCon S (carbosilfocon A) Rigid Gas Permeable Contact Lens for Daily Wear (K980197)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the word "ULTRAVISION" in large, bold, black letters. Below the word "ULTRAVISION" is the abbreviation "INC." in smaller, black letters. The font appears to be a serif font. The background is white.
510(k) Summary
| Company: | Specialty UltraVision, Inc.<br>307 Orchard City Drive, Suite 100<br>Campbell, CA 95008 |
|----------|----------------------------------------------------------------------------------------|
| Contact: | Garold L. Edwards, O.D., F.A.A.O.<br>Vice President. Technical Affairs |
Telephone: (408) 341-0700 Fax: (408) 341-0717
Date Prepared: December 6, 1999
#### Device Name:
| Common Name: | Rigid Gas Permeable Contact Lens |
|--------------------------|-------------------------------------------------------------------------------|
| Trade/Proprietary Names: | EpiCon K (carbosilfocon A) Rigid Gas<br>Permeable Contact Lens for Daily Wear |
Classification Name: Rigid Gas Permeable Contact Lens
Device Classification: Class II (21 CFR 886.5916)
# Predicate Device:
The predicate device is the UltraCon S (carbosilfocon A) Rigid Gas Permeable Contact Lens for Daily Wear, which was cleared under 510(k) K980197. This device was selected as the predicate device because it is manufactured from the same polymer, using similar lens designs, at the same manufacturing facility, under the same Quality System and using the same manufacturing equipment.
#### Description of Device:
.
The EpiCon K (carbosilfocon A) Rigid Gas Permeable Contact Lens for Daily Wear is available as a hybrid lens design with spherical and aspherical curves on its surfaces. The lens material (carbosilfocon A) is composed of polymethyl methacrylate (PMMA) combined with a proprietary copolymer that contains silicone. D&C Green No. 6 is incorporated into the polymer, resulting in a blue handling tint. The EpiCon K Rigid Gas Permeable Contact Lens for Daily Wear is a hemispherical shell of the following dimensions:
- . Chord Diameter: 10.5 mm to 14.0 mm
- Center Thickness: 0.10 mm to 0.35 mm
- Base Curve:
6.0 to 8.5 mm
- Plano to-20.00 D
- Powers: . . Peripheral Curves: A. B. C. D. E. F and G
1323 44™ AVENUE N.E. CALGARY, ALBERTA CANADA T2E 6L5 TEL: 800-668-9251 403-291-1919 FAX: 403-291-6477
{2}------------------------------------------------
# Comparison to Predicate Device
| PARAMETER | EpiCon K Rigid Gas<br>Permeable Contact Lens for<br>Daily Wear | UltraCon S Rigid Gas<br>Permeable Contact Lens for<br>Daily Wear |
|----------------------------|----------------------------------------------------------------|------------------------------------------------------------------|
| material | carbosilfocon A | carbosilfocon A |
| material<br>classification | Hydrophobic Lens Group 2 | Hydrophobic Lens Group 2 |
| indication for use | myopia, hyperopia and<br>astigmatism | myopia, hyperopia and<br>astigmatism |
| water content | 0.5% | 0.5% |
| light transmittance | 96.5% | 96.5% |
| Dk (35° C) | 52 x 10-11 | 52 x 10-11 |
| powers | Plano to -20.00 Diopters | +20.00 to -20.00 Diopters |
| color | blue visibility | blue visibility |
| specific gravity | 1.105 | 1.105 |
| refractive index | 1.49 at 20° | 1.49 at 20° |
| wetting angle | < 19° | < 19° |
| Method of<br>manufacture | Gel Flow Molding | Gel Flow Molding |
### Indications for Use:
The EpiCon K (carbosilfocon A) Rigid Gas Permeable Contact Lens for Daily Wear is indicated for daily wear in aphakic persons exhibiting the irregular corneal surface associated with keratoconus and/or requiring keratoconus management with otherwise nondiseased eyes. The lens may be prescribed for the correction of refractive ametropia (myopia, hyperopia and astigmatism). The lenses may be disinfected using a chemical disinfecting system.
# Description of Safety and Substantial Equivalence:
Biocompatibility, toxicology, physicochemical and leachability studies were previously performed on the carbosilfocon A polymer and cleared under K980197. Results of that testing indicate that the lens material is non-toxic and biocompatible.
# Clinical Testing:
A four week clinical evaluation of the EpiCon K (carbosilfocon A) Rigid Gas Permeable Contact Lens for Daily Wear was performed at five investigational sites in accordance with current Good Clinical Practices and established regulations. The study assessed the clinical performance of the EpiCon K lens in subjects with bilateral keratoconus. Subjects were seen at an Initial Visit. which was a diagnostic fitting visit. Subjects were seen at a Dispensing Visit, and after 1, 2 and 4 weeks of lens wear. The EpiCon K lens was dispensed to 38 keratoconic subjects, and 18 subjects (47.4%) completed the 4 weeks of the study period while 20 subjects (52.6%) discontinued. Two adverse reactions were reported during the study period, both of which resolved without sequelae upon discontinuation of lens wear.
Based on the clinical data analysis, it was determined that the incidence of positive slit lamp findings, subject symptoms, problems and complaints, keratometry and refractive changes, visual acuity with the lens, lens wearing time and lens comfort were within the expected range for keratoconic eyes wearing RGP lenses.
Clinical performance of the EpiCon K lens was substantially equivalent to the clinical performance of the predicate lens.
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white circular logo for the U.S. Department of Health & Human Services. The logo features the department's symbol, which is a stylized human figure with three lines representing the head, body, and legs. The symbol is positioned to the right of the text, which reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement around the symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
#### 2 2000 FEB
Gerald L. Edwards, O.D., F.A.A.O. Vice President, Technical Affairs Specialty UltraVision, Inc. 307 Orchard City Drive, Suite 100 Campbell, California 95008
Re: K992787
Trade Name: EpiCon K (Carbosilifocon A) Rigid Gas Permeable Contact Lens for Daily Wear (Keratoconus Design) Regulatory Class: Class II Product Code: 86HQD Dated: December 6, 1999 Received: December 7, 1999
Dear Dr. Edwards:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verity such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2 - Gerald L. Edwards, O.D., F.A.A.O.
· This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
・
ARalph Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# INDICATIONS STATEMENT
### Device Name:
EpiCon K (carbosilfocon A) Rigid Gas Permeable Contact Lens for Daily Wear
### Indications for Use:
The EpiCon K (carbosilfocon A) Rigid Gas Permeable Contact Lens for Daily Wear is indicated for daily wear in aphakic or not-aphakic persons exhibiting the irregular corneal surface associated with keratoconus and/or requiring keratoconus management with otherwise nondiseased eyes. The lens may be prescribed for the correction of refractive ametropia (myopia, hyperopia and astigmatism). The lenses may be disinfected using a chemical disinfecting system.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
Over-the-Counter Use OR
Linda Dee Cohn, MD
510(k) Nur
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.