COMFORTKONE KERATOCONUS ASPHERIC (HYBUFOCON A, FILOFILCON A, PEMUFOCON A) RIGID GAS PERMEABLE (RGP) DAILY WEAR CONTACT
K024164 · Metro Optics, Inc. · HQD · Feb 12, 2003 · Ophthalmic
Device Facts
Record ID
K024164
Device Name
COMFORTKONE KERATOCONUS ASPHERIC (HYBUFOCON A, FILOFILCON A, PEMUFOCON A) RIGID GAS PERMEABLE (RGP) DAILY WEAR CONTACT
Applicant
Metro Optics, Inc.
Product Code
HQD · Ophthalmic
Decision Date
Feb 12, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5916
Device Class
Class 2
Indications for Use
The ComfortKone™ Keratoconus Aspheric (hybufocon A, filofocon A, pemufocon A) Rigid Gas Permeable (RGP) Daily Wear Contact Lens (Clear and Tinted, Lathe-cut from Lens Blank) is indicated for daily wear for persons requiring Keratoconus management with otherwise non-diseased eyes. The lens may be prescribed for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and not-aphakic persons. The lens may be disinfected with a chemical disinfection system.
Device Story
ComfortKone™ is a lathe-cut, rigid gas permeable (RGP) contact lens fabricated from hybufocon A, filofocon A, or pemufocon A. Designed for keratoconus management, the lens features a 4.0 mm spherical optic zone for the cone peak, transitioning into an aspheric fitting curve, and finishing with two junctionless peripheral aspheric curves for corneal alignment. The lens acts as a refracting medium to focus light on the retina. It is intended for daily wear by patients with keratoconus and refractive ametropia. The device is prescribed by eye care professionals. Benefits include improved visual acuity and comfort for keratoconus patients. Maintenance involves chemical disinfection.
Clinical Evidence
Bench testing only. Substantial equivalence is based on comparison of physical and chemical properties (water content, Dk value, specific gravity, refractive index, light transmittance) to the predicate device.
Technological Characteristics
Lathe-cut RGP contact lens. Materials: hybufocon A, filofocon A, pemufocon A. Water content <1%. Oxygen permeability (Dk) ranges 23-50. Specific gravity 1.14-1.18. Refractive index 1.44-1.45. Light transmittance >93%.
Indications for Use
Indicated for daily wear in patients requiring keratoconus management with non-diseased eyes. Prescribed for correction of refractive ametropia (myopia, hyperopia, astigmatism) in aphakic and non-aphakic persons.
Regulatory Classification
Identification
A rigid gas permeable contact lens is a device intended to be worn directly against the cornea of the eye to correct vision conditions. The device is made of various materials, such as cellulose acetate butyrate, polyacrylate-silicone, or silicone elastomers, whose main polymer molecules generally do not absorb or attract water.
{0}------------------------------------------------
## SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. FEB 1 2 2003
The assigned 510(k) number is:
KD24164
Applicant information:
Date Prepared:
February 4, 2003
Name: Address Metro Optics, Inc. 15802 Vision Drive Pflugerville, TX 78660
Contact Person: Phone Number: . Fax:
Mr. Steve Webb Vice President (512) 251-2382 (512) 251-6554
| Official Correspondent: | Med-Vice Consulting, Inc. |
|-------------------------|---------------------------|
| Regulatory Consultant: | Ms. Deanna Werber |
| | 623 Glacier Drive |
| | Grand Junction, CO 81503 |
| Phone Number: | (970) 243-5490 |
| Fax Number: | (970) 243-5501 |
### Device Information:
Regulatory Classification: Class II
Product Code: HOD
Trade Name:
ComfortKone™ Keratoconus Aspheric (hybufocon A, filofocon A, pernufocon A) Rigid Gas Permeable (RGP) Daily Wear Contact Lens (Clear and Tinted, Lathe-cut from Lens Blank)
Purpose for 510(k)
Addition of three RGP materials to a previously cleared design and indication for use. (K990264)
Classification Name:
Lenses, Contact (other material), Daily Wear
{1}------------------------------------------------
### Equivalent Devices:
The Comfort Keratoconus Aspheric (hybufocon A. filofocon A. pernufocon A) Rigid Gas Permeable (RGP) Daily Wear Contact Lens is substantially equivalent to the predicate device identified below in terms of intended use and design.
Predicate device manufacturer: Metro Optics of Austin, Inc. 1.) 15802 Vision Drive Pflugerville, TX 78660 510(k) 990264
Device name: ComfortKone™, Keratoconus, Daily Wear
510(k) #: K990264
Device Description:
The ComfortKone" Keratoconus Aspheric Contact Lens is fabricated from the hydrophobic contact lens materials (hybufocon A, filotocon A, pemufocon A). When placed on the human cornea, the Comfort Keratoconus Aspheric rigid gas permeable contact lenses act as a refracting medium to focus light rays upon the retina.
The Comfort Kone" is a aspheric contact lens. It is designed to provide optimum comfort and visual acuity to the keratoconus patient. The ComfortKone™ lens design begins with a spherical 4.0 mm ontic zone that fits the peak of the cone and provides for good visual acuity. The lens then thattens into the aspheric curve, which is considered the fitting curve of the lens. The aspheric curve will vary in rate of change depending on how far the keratoconus has advanced and creates optimal corneal alignment. design finishes with two junctioniess peripheral aspheric curves to maintain alignment.
Metro Optics of Austin, Inc., has been granted referencing rights from the button manufacturers. The physical properties of the lenses can be referenced in the corresponding 510(k)'s.
> Contamac HyBrid FS (hybufocon A) reference 510(k) K021977 Innovision Hydro2 (filofocon A) reference 510(k) K000485 Innovision Accucon (pemufocon A) reference 510(k) K944223
{2}------------------------------------------------
### Intended Use:
The Comfort Kone™ Keratoconus Aspheric (hybufocon A, filofocon A, persufocon A) Rigid Gas The Connortone - Romact Lens (Clear and Tinted, Lathe-cut from Lens Blank) is indicated for daily wear for persons requiring Keratocomus management with otherwise non-diseased eyes. The lens may be prescribed for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in appakic and not-arbakic persons. The lens may be disinfected with a chemical disinfection system.
#### Substantial Equivalence:
The new device will be manufactured according to specified process controls and a Quality Management System certified to OSR guidelines. The new device will undergo manufacturing, packaging and other bysich berther to QGP devices currently marketed and distributed by Metro Optics, Inc. in the USA. The established safety profile (pre-clinical toxicology and manufacturing/chemistry data) of the device is equivalent to the Metro Optics, Inc. ComfortKone™, Keratoconus, Daily Wear SION #K990264. Being similar with respect to indications for use, materials, physical construction and safety & effectiveness to the predicate device, this meets the requirements per section 510(k) of the act regarding substantial equivalence and does not raise different questions of safety and effectiveness than the predicate device identified above.
The following matrix illustrates that the production method, lens function and material of the Comfort Keratoconus Aspheric (hybufocon A, filofocon A, pemufocon A) Rigid Gas Permeable (RGP) Daily Wear Contact Lens are substantially equivalent to the predicate device. In addition, the indication, intended use, production method, water content, polymer, Dk value, specific gravity, are as well substantially equivalent to the predicate device.
{3}------------------------------------------------
# SUMMARY OF SAFETY AND EFFECTIVENESS
Substantial Equivalence Matrix
.
: 上一
.
.
1. 1. 1.
:
: 上
11. 1
: 1
1 : 1 : 1 :
:
:
.
and the comments of
:
| | Characteristic | New Device<br>ComfortKone<br>Keratoconus Aspheric<br>(hybufocon A, filofocon A,<br>pemufocon A) | Predicate Device<br>ComfortKone<br>Keratoconus Aspheric<br>(paflufocon C) |
|-----|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1.) | INDICATION | Daily wear, Rigid Gas Permeable<br>(RGP) contact lens | Daily wear, Rigid Gas Permeable<br>(RGP) contact lens |
| 2.) | INTENDED USE | Daily wear for persons requiring<br>Keratoconus management. The lens<br>may be prescribed for the correction of<br>refractive ametropia (myopia, hyperopia,<br>and astigmatism) in aphakic and not<br>aphakic persons with non-diseased<br>eyes. | Daily wear for persons requiring<br>Keratoconus management. The lens<br>may be prescribed for the correction of<br>refractive ametropia (myopia, hyperopia,<br>and astigmatism) in aphakic and not<br>aphakic persons with non-diseased<br>eyes. |
| 3.) | PRODUCTION<br>METHOD | Lathe-Cut | Lathe-Cut |
| 4.) | RGP MATERIAL<br>common name | HyBrid FS<br>Hydro2<br>Accucon | Fluroperm 30 |
| a. | Water Content | (hybufocon A) <1%<br>(filofocon A) < 1%<br>(pemufocon A) < 1% | < 1% |
| b. | Polymer/USAN | HyBrid FS (hybufocon A)<br>Hydro2 (filofocon A)<br>Accucon (pemufocon A) | Fluroperm 30 (paflufocon C) |
| c. | Specific Gravity | (hybufocon A) 1.18<br>(filofocon A) 1.14<br>(pemufocon A) 1.16 | (paflufocon C) 1.14 |
| d. | Oxygen Permeability | (hybufocon A) = 23<br>(filofocon A) = 50<br>(pemufocon A) = 25 | 30 |
| e. | Wetting Angle | (hybufocon A) = immeasurable *<br>(filofocon A) = <5 **<br>(pemufocon A) = <25 ** | 12.8 |
| f. | Light Transmittance | (hybufocon A) = >93%<br>(filofocon A) = >93%<br>(pemufocon A) = >93% | >93% |
| g. | Refractive Index | (hybufocon A) = 1.44<br>(filofocon A) = 1.45<br>(pemufocon A) = 1.45 | 1.47 |
** receding angle mathod in the course igle medioxi
11:11:11:1
: |
. . . . . . . . .
.
. .
.
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of three human profiles facing right, stacked on top of each other. The profiles are rendered in black and have a flowing, abstract design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the image.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## FEB 1 2 2003
Metro Optics, Inc. c/o Ms. Deanna Werber MedVice Consulting, Inc. 623 Glacier Drive Grand Junction, CO 81503
Re: K024164
ComfortKone™ Keratoconus Aspheric (hybufocon A, filofocon A, pemufocon A) Rigid Gas Permeable (RGP) Daily Wear Contact Lens (Clear and Tinted, Lathe-cut from Lens Blank) Regulation Number: 21 CFR 886.5916 Regulation Name: Rigid Gas Permeable Contact Lens Regulatory Class: Class II Product Code: HQD Dated: December 9, 2002 Received: December 17, 2002
Dear Ms. Werber:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
A. halyi Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
**Device Name:**
Comfort Kone™ Keratoconus Aspheric (hybufocon A, filofocon A, pemufocon A) Rigid Gas Permeable (RGP) Daily Wear Contact Lens (Clear and Timed, Lathe-cut from Lens Blank)
### INDICATIONS FOR USE:
The ComfortKone™ Keratoconus Aspheric (hybufocon A, filofocon A, pemufocon A) Rigid Gas Permeable (RGP) Daily Wear Contact Lens (Clear and Tinted, Lathe-cut from Lens Blank) is indicated for daily wear for persons requiring Keratoconus management with otherwise non-diseased eyes. The lens may be prescribed for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and not-aphakic persons. The lens may be disinfected with a chemical disinfection system.
# (PLEASE DO OT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Prescription Use
(Per 21 CFR 801.109)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Of
Over-The-Counter Use
(Optional Format 1-2-96)
Myra Smith
(Division Sign-Off) Division of Ophthalmic Devices
510(k) Number KO24169
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.