K992016 · Amuchina Intl., Inc. · LIF · Jan 10, 2000 · Gastroenterology, Urology
Device Facts
Record ID
K992016
Device Name
AMUKIN D
Applicant
Amuchina Intl., Inc.
Product Code
LIF · Gastroenterology, Urology
Decision Date
Jan 10, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5820
Device Class
Class 2
Intended Use
Amukin-D is a chlorine-based, high-level disinfectant for reprocessing hemodialyzers.
Device Story
Amukin-D is a chlorine-based chemical disinfectant used for the reprocessing of hemodialyzers. It functions as a high-level disinfectant to ensure the safety of hemodialysis equipment. The device is intended for use in clinical settings by healthcare professionals involved in dialysis procedures. By effectively disinfecting hemodialyzers, the product helps maintain equipment hygiene and patient safety during subsequent dialysis treatments.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Chlorine-based chemical disinfectant solution. Formulated for high-level disinfection of hemodialyzers.
Indications for Use
Indicated for high-level disinfection of hemodialyzers during reprocessing.
Regulatory Classification
Identification
A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
Related Devices
K043244 — FRESENIUS OPTIFLUX F20NR, FF18NR, F16NR · Fresenius Medical Care North America · Dec 20, 2004
K033505 — RENACLEAN SH DIALYZER CLEANING SYSTEM · Minntech Corp. · Apr 27, 2004
K182367 — DIASAFEplusUS · Fresenius Medical Care Renal Therapies Group, LLC · Jan 10, 2019
K211035 — U9000 Plus Ultrafilter · Baxter Healthcare Corporation · Oct 19, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 1 0 2000
Mr. Ludovico Giavotto President Amuchina International. Inc. 8-8 Metropolitan Court Gaithersburg, MD 20878
Re: K992016
Revised Labeling for Amukin-D Disinfectant for Hemodialyzers Dated: October 12, 1999 Received: October 12, 1999 Unclassified Procode: 78 LIF
Dear Mr. Giavotto:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{1}------------------------------------------------
Page 1 of 1
510(k) Number (if known): _ K992016
Device Name: Amukin-D
Indications For Use:
Amukin-D is a chlorine-based, high-level disinfectant for reprocessing hemodialyzers.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Signature
(Optional Format 3-10-98)
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Dev 510(k) Number.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.