K972664 · Storz Instrument Co. · MRH · Oct 10, 1997 · General Hospital
Device Facts
Record ID
K972664
Device Name
STORZ MILLENNIUM VISCOUS FLUID SYSTEM
Applicant
Storz Instrument Co.
Product Code
MRH · General Hospital
Decision Date
Oct 10, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Storz Millennium™ Viscous Fluid System is a new device intended for the injection and aspiration of viscous fluids into and out of the eye during ophthalmic posterior vitreoretinal surgery. The Storz viscous fluid system consists of the Storz CX5700 module to be used with the Storz Millennium™ Microsurgical System and the Storz CX5710 sterile single use accessory pack.
Device Story
System comprises Storz CX5700 module and CX5710 disposable injection pack; used during posterior vitreoretinal surgery. Module integrates with Storz Millennium™ Microsurgical System to inject/aspirate viscous fluids (e.g., silicone oil) for retinal tamponade. Operated by surgeon via foot pedal and touch screen interface. System provides dual linear foot control for pressure/vacuum regulation (0-70 PSI injection; 0-600 mmHg aspiration). Disposable pack includes tubing, syringe, and needle assembly. Output allows surgeon to manipulate intraocular fluid dynamics to fixate retinal tears/detachments.
Clinical Evidence
No clinical data provided; substantial equivalence based on bench testing and design comparison to legally marketed predicate devices.
Indicated for injection and aspiration of viscous fluids (e.g., silicone oil) during posterior ophthalmic vitreoretinal surgery to act as a tamponade for retinal tears and detachments.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
{0}------------------------------------------------
K972664
Image /page/0/Picture/1 description: The image shows the word "storz" in white letters on a black background. The letters are stylized and bold. There is a registered trademark symbol in the upper right corner of the image. The background is a rounded rectangle.
Storz Millennium™ Viscous Fluid System Premarket Notification
0CT 1 0 1997
#### 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
The following information is submitted in accordance with the requirements of 21 CFR 807.92:
Contact Person: Patrick G. Balsmann, Regulatory Affairs Associate Storz, 3365 Tree Court Industrial Bldv., St. Louis, MO 63122-6694 Phone: (314) 225- 5051, ext. 5538.
Date Prepared: July 15, 1997
Proprietary Name: Storz Millennium™ Viscous Fluid System
Common/Usual Name: Viscous Fluid Injection/Aspiration System(s).
Classification Name: Infusion Pump, 86 MRH; 21CFR §880.5725.
Device Description/Intended Use: The Storz Millennium™ Viscous Fluid System is intended for the injection and aspiration of viscous fluids into and out of the eye during posterior ophthalmic vitreoretinal surgery. The system is designed to inject and aspirate viscous fluids, such as silicone oil which act as a tamponade in fixating large retinal tears and detachments. The Storz viscous fluid system consists of the Storz CX5700 Millennium™ Viscous Fluid System Module and the Storz CX5710 Viscous Fluid Injection Pack. The Storz CX5700 module is a self-contained pump system used as an individual module with the Storz Millennium™ Microsurgical System for the injection and aspiration of viscous fluids. The Storz CX5710 is the disposable accessory pack intended for single use to be used with the Storz viscous fluid module.
Predicate Device: The Storz CX5700 Millennium™ Viscous Fluid System Module is similar in design and function to the Storz Millennium™ Microsurgical System (K961310), Escalon® Viscous Fluid System (K963434), the Richard James Viscous Fluid Transfer System (K902835), and the Alcon Accurus™ Vitreoretinal Surgical System. The Storz CX5710 Viscous Fluid Injection Pack is similar in design, composition, and function to the Escalon® Fluid Delivery Pack (K963434.)
Predicate Comparison: A chart comparing characteristics of the Storz CX5700 Millennium™ Viscous Fluid System Module and the Storz CX5710 Viscous Fluid Injection Pack to those of the predicate devices is attached.
**Submitted by:**
Robert H. Johnson
Patrick G. Balsmann Regulatory Affairs
1 × 4
{1}------------------------------------------------
#### Storz CX5700 Millennium™ Viscous Fluid System Module Device Comparison Chart
(
(
| Device<br>Description | Storz CX5700<br>Viscous Fluid<br>System | Storz<br>Millennium™<br>Microsurgical<br>System | Alcon Accurus™<br>Vitreoretinal<br>Surgical System | Escalon® Viscous<br>Fluid System | Richard James<br>Viscous Fluid<br>Transfer System |
|----------------------------------------------------|----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| 510(k) | current | K961310 | unknown | K963434 | K902835 |
| Intended Use | Posterior segment<br>ophthalmic<br>surgery - injection<br>& aspiration of<br>viscous fluids. | Anterior & posterior<br>segment ophthalmic<br>surgery. | Posterior segment<br>ophthalmic surgery. | Posterior ophthalmic<br>surgery - injection<br>& aspiration of<br>viscous fluids. | Posterior ophthalmic<br>surgery - injection &<br>aspiration of viscous<br>fluids. |
| Modular<br>Design | Yes | Yes. | No. | No. | No. |
| Ophthalmic<br>Features | Viscous fluid<br>injection/<br>aspiration. | Irrigation/aspiration,<br>phacoemulsification/<br>fragmentation,<br>vitrectomy, scissors,<br>bipolar, illumination,<br>& IOP control. | Viscous fluid<br>injection/aspiration,<br>Irrigation/aspiration,<br>fragmentation,<br>vitrectomy, scissors,<br>bipolar, illumination,<br>& IOP control. | Viscous fluid<br>injection/aspiration. | Viscous fluid<br>injection/aspiration. |
| Programmable<br>For Multiple<br>Surgeons | Yes. | Yes. | Yes. | No. | No. |
| Device<br>Description | Storz CX5700<br>Viscous Fluid<br>System | Storz<br>MillenniumTM<br>Microsurgical<br>System | Alcon AccurusTM<br>Vitreoretinal<br>Surgical System | Escalon® Viscous<br>Fluid System | Richard James<br>Viscous Fluid<br>Transfer System |
| 510(k) | current | K961310 | unknown | K963434 | K902835 |
| Disposable<br>Accessories | Yes. | Yes. | Yes. | Yes. | Yes. |
| Viscous Fluid<br>User Interface | Foot pedal &<br>touch screen. | Foot pedal & touch<br>screen. | Foot pedal & touch<br>screen. | Foot pedal &<br>control knob. | Foot pedal & control<br>knob. |
| Foot<br>Controller -<br>Dual Linear | Yes | Yes. | No. | No. | No. |
| Viscous Fluid<br>Injection,<br>Pressure<br>Range | 0 - 70 PSI | Not Applicable. | 0 - 80 PSI. | 0 - 70 PSI. | 0 - 70 PSI. |
| Viscous Fluid<br>Injection,<br>Aspiration<br>Range | 0 - 600 mmHg. | Not Applicable. | 0 - 600 mmHg. | 0 - 600 mmHg. | 0 - 650 mmHg. |
ﻟ
{2}------------------------------------------------
# Storz CX5700 Millennium™ Viscous Fluid System Module
Comparison
(
ﺴﺮ ﻣ
{3}------------------------------------------------
#### Storz CX5710 Viscous Fluid Injection Pack Device Comparison Chart
ﻟﻤﺴﺘﻌﻤﺎ
-
| Device Description | Storz CX5710 Viscous Fluid Injection Pack<br>Escalon® Fluid Delivery Pack |
|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) | current<br>K963434 |
| Pack Components | Modified hose barb, tubing, male luer with locking nut, filter, 10cc syringe adaptor, 10cc modified piston, needle/cover assembly, 10cc standard piston, & 10cc syringe assembly.<br>Hose barb, tubing, male luer with locking nut, filter, 10cc syringe adaptor, 10cc modified piston, needle/cover assembly, 10cc standard piston, & 10cc syringe assembly. |
| Applicable Ophthalmic Surgical Unit | Storz Millennium™ Microsurgical System<br>Escalon® Viscous Fluid System |
| Provided Sterile | Yes.<br>Yes. |
| Labeled For Single Patient Use | Yes.<br>Yes. |
| Patient Contact Material | AISI 303 stainless steel.<br>AISI 303 stainless steel. |
| Packaging | Sealed Tyvek® pouches.<br>10 packs/box.<br>Sealed Tyvek® pouches.<br>10 packs/box. |
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of three human profiles facing to the right. The profiles are stacked on top of each other, with the top profile being the largest and the bottom profile being the smallest. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular pattern.
OCT 10 1997
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Patrick G. Balsmann Domestic Regulatory Affairs Associate Storz Instrument Company 3365 Tree Ct. Industrial Blvd. St. Louis, MO 63122-6694
Re: K972664
> Trade Name: Storz Millennium™ Viscous Fluid System Regulatory Class: II Product Code: 86 MRH Dated: July 15, 1997 Received: July 16, 1997
Dear Mr. Balsmann:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}------------------------------------------------
Page 2 - Mr. Patrick G. Balsmann
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
A. Roepke Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a logo for Storz. The logo is in black and white and features the word "storz" in a stylized font. The word is set against a black background with rounded corners. There is a registered trademark symbol to the right of the word.
## Storz Millennium™ Viscous Fluid System Premarket Notification
### INDICATIONS FOR USE
Device Name:
Storz Millennium™ Viscous Fluid System
Indications for Use:
The Storz Millennium™ Viscous Fluid System is a new device intended for the injection and aspiration of viscous fluids into and out of the eye during ophthalmic posterior vitreoretinal surgery. The Storz viscous fluid system consists of the Storz CX5700 module to be used with the Storz Millennium™ Microsurgical System and the Storz CX5710 sterile single use accessory pack.
Enette TK Beers
ivision of Ophthalmic 510(k) Number
**Prescription Use**
**(Per 21 CFR 801.109)**
(114
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.