STORZ MILLENNIUM MICROSURGICL SYSTEM HIGH SPEED VITRECTOMY SYSTEM
Applicant
Storz Instrument Co.
Product Code
HQE · Ophthalmic
Decision Date
Mar 23, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.4150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Storz Millennium™ High Speed Vitrectomy System is a new device intended for removal of vitreous matter from the posterior vitreous cavity during ophthalmic vitrectomy surgery. The Storz high speed vitrectomy system consists of the Storz CX5800 module to be used with the Storz Millennium™ Microsurgical System, Storz CX5810 handpiece, and the Storz CX4804 sterile single use accessory pack.
Device Story
System performs aspiration of vitreous matter during ophthalmic vitrectomy surgery. Components include CX5800 module, CX5810 handpiece, and CX4804 disposable pack. Operates via pneumatic drive using guillotine cutting action; controlled by surgeon via foot pedal and touch screen interface. Used in clinical ophthalmic surgical settings. Output consists of physical removal of vitreous tissue; provides surgeon with precise control over aspiration and cutting parameters. Benefits include efficient vitreous removal during posterior segment procedures. System is modular and programmable for multiple surgeons.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Modular system; pneumatic drive; guillotine cutting mechanism; Venturi aspiration. Handpiece constructed of plastic; cutter pack uses AISI 304 stainless steel with anti-reflective coating. User interface via touch screen and foot pedal. Sterile, single-use disposable packs. System is networked/integrated with the Storz Millennium Microsurgical System.
Indications for Use
Indicated for removal of vitreous matter from the posterior vitreous cavity during ophthalmic vitrectomy surgery in patients requiring such procedures.
Regulatory Classification
Identification
A vitreous aspiration and cutting instrument is an electrically powered device, which may use ultrasound, intended to remove vitreous matter from the vitreous cavity or remove a crystalline lens.
Special Controls
*Classification.* Class II (special controls). The device, when it is phacofragmentation unit replacement tubing, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{1}------------------------------------------------
K980488
MAR 2 3 1998
#### Storz Millennium™ High Speed Vitrectomy System Premarket Notification
#### 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
The following information is submitted in accordance with the requirements of 21 CFR 807.92:
Contact Person:Gary Rauvola, Regulatory Affairs Group Manager Bausch & Lomb Surgical, Storz Products 3365 Tree Court Industrial Blvd. St. Louis. MO 63122-6694 Phone: (314) 225- 5051, ext. 5340.
Date Prepared: February 4, 1998
Proprietary Name: Storz Millennium™ High Speed Vitrectomy System
Common/Usual Name: Automated Vitrectomy Device
Classification Name: Vitreous Aspiration & Cutting Instrument, 86 HQE; 21CFR \$886.4150.
Device Description/Intended Use: The Storz Millennium™ High Speed Vitrectomy System is intended for the aspiration of vitreous matter from the eye during ophthalmic vitrectomy surgery. The system consists of the Storz CX5800 Millennium™ System High Speed Vitrectomy module used with the Storz Millennium™ Microsurgical System, Storz CX5810 High Speed Vitrectomy handpiece, and the Storz CX4804 High Speed Vitrectomy single use disposable pack.
Predicate Devices: The Storz CX5800 Millennium™ System High Speed Vitrectomy module is substantially equivalent in design and function to the Storz Millennium™ Microsurgical System (K961310), Promex Vitrectomy System (K961078), Scieran Technologies VIT Commander System (K961738), and the Alcon Accurus™ Vitreoretinal Surgical System. The Storz CX5810 Millennium™ High Speed Vitrectomy handpiece is substantially equivalent to the Storz Micro Vit Vitrectomy Probe (K954816 and K961310), Promex Vitrectomy handpiece (K961078), Alcon Accurus InnoVit Probe, and the Scieran Vit Commander System handpiece (K961738). The Storz CX4804 High Speed Vitrectomy pack is similar in design, composition, and function to the Storz DP4801 Vitrectomy Probe pack (K954816 and K961310), the Promex vitrectomy cutter pack (K961078), and the Alcon Accurus Total Plus Pak.
Predicate Comparisons: Charts comparing characteristics of the Storz CX5800 Millennium™ High Speed Vitrectomy module, Storz CX5810 Millennium™ High Speed Vitrectomy handpiece, and the Storz CX4804 pack to those of the predicate devices are attached.
Submitted by:
Gary Rauvole
Gary Rauvola Regulatory Affairs Group Manager
{2}------------------------------------------------
Image /page/1/Picture/0 description: The image is a black and white halftone pattern. The pattern is arranged in a semi-circular shape, with the dots becoming smaller and more sparse towards the edges of the semi-circle. The dots are densely packed in the center of the semi-circle, creating a darker area. The overall effect is a gradient from dark to light, with the darkest area in the center and the lightest areas at the edges.
## Storz CX5800 Millennium™ High Speed Vitrectomy Module Device Comparison Chart
| Device<br>Description | Storz CX5800<br>High Speed<br>Vitrectomy<br>Module | Storz<br>MillenniumTM<br>Microsurgical<br>System | Alcon AccurusTM<br>Vitreoretinal<br>Surgical System | Promex<br>Vitrectomy System | Scieran Technologies Vit<br>Commander System | Device<br>Description | | Storz<br>MicroVit®<br>Vitrectomy<br>Probe | Alcon Accurus™<br>InnoVit™<br>Probe | Promex Vitrectomy<br>System | Scieran Technologies<br>Vit Commander<br>System |
|------------------------------------------|----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|------------------------------------------------|---------------------------------|--|-------------------------------------------|-------------------------------------|--------------------------------------------------|--------------------------------------------------|
| 510(k) | current | K961310 | unknown | K961078 | K961738 | 510(k) | | K954816<br>K961310 | unknown | K961078 | K961738 |
| Intended Use | Posterior segment<br>ophthalmic<br>surgery. | Anterior & posterior<br>segment ophthalmic<br>surgery. | Posterior segment<br>ophthalmic surgery. | Posterior segment<br>ophthalmic surgery. | Posterior segment ophthalmic<br>surgery. | Intended Use | | Posterior Vitreous<br>Removal | Posterior Vitreous<br>Removal | Posterior Vitreous<br>Removal | Posterior Vitreous<br>Removal |
| Modular<br>Design | Yes. | Yes. | No. | No. | Yes. | Cuts Per Minute | | 30 - 750 cpm | 100-1200 cpm | up to 2000 cpm | 400-2500 cpm |
| Ophthalmic<br>Features | Vitrectomy | Viscous fluid<br>injection/aspiration,<br>Irrigation/aspiration,<br>phacoemulsification<br>/ fragmentation,<br>vitrectomy, scissors,<br>bipolar,<br>illumination, & IOP<br>control. | Viscous fluid<br>injection/aspiration,<br>Irrigation/aspiration,<br>fragmentation.<br>vitrectomy, scissors,<br>bipolar,<br>illumination, & IOP<br>control. | Vitrectomy,<br>aspiration. | Irrigation/aspiration,<br>vitrectomy, bipolar. | Sterilization | | Single use<br>Disposable | Single use<br>Disposable | Autoclavable<br>handpiece with<br>disposable tip | Autoclavable<br>handpiece with<br>disposable tip |
| Programmable<br>For Multiple<br>Surgeons | Yes. | Yes. | Yes. | No. | No. | Handpiece Body | | Plastic | Plastic | Stainless steel and<br>aluminum | Titanium |
| Disposable<br>Accessories | Yes. | Yes. | Yes. | Yes. | Yes. | Type of<br>Handpiece Drive | | Pneumatic | Pneumatic | Electric | Electric |
| User Interface | Foot pedal &<br>touch screen. | Foot pedal & touch<br>screen. | Foot pedal & touch<br>screen. | Foot pedal & control<br>knobs. | Foot pedal & control knobs. | Associated<br>Aspiration system | | Venturi | Venturi | Venturi | Diaphragm |
| Foot<br>Controller -<br>Dual Linear | Yes. | Yes. | No. | No. | No. | Cutting Action | | Guillotine | Guillotine | Guillotine | Guillotine |
{3}------------------------------------------------
## Storz CX5810 High Speed Vitrectomy Handpiece Device Comparison Chart
1. 1. 1. 1. 1. 1.
ಕ್ಷಗಳ
ੂ ్లు
:
ई
{4}------------------------------------------------
# Storz CX4804 High Speed Vit Cutter Pack Device Comparison Chart
100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000
| Device<br>Description | | Storz CX4804<br>High Speed<br>Vit Cutter<br>Pack | Storz<br>DP4801<br>Vitrectomy<br>Probe Pack | Alcon Accurus™ Total Plus™ Pak | Promex Pack |
|-------------------------------------------|--|------------------------------------------------------------|--------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|
| 510(k) | | current | K954816<br>K961310 | unknown | K961078 |
| Pack Components | | High speed vitrectomy<br>cutter with aspiration<br>tubing. | Vit cutter, aspiration<br>tubing, collection<br>cassette. | Cassette with drain bag, vitrectomy probe,<br>fiber optic illuminator probe, aspiration<br>tubing, infusion spike, infusion cannula. | High speed vitrectomy<br>cutter with aspiration<br>tubing. |
| Applicable<br>Ophthalmic<br>Surgical Unit | | Storz Millennium<br>Microsurgical System. | Storz Millennium<br>Microsurgical System.<br>Storz Premiere<br>Microsurgical System. | Alcon Accurus Vitreoretinal Surgical<br>System. | Promex Vitrectomy System. |
| Provided Sterile | | Yes. | Yes. | Yes. | Yes. |
| Labeled For<br>Single Patient Use | | Yes. | Yes. | Yes. | Yes. |
| Patient Contact<br>Material | | AISI 304 stainless steel<br>with anti-reflective steel | AISI 304 stainless<br>steel. | Stainless steel | AISI 304 stainless steel. |
| Packaging | | Sealed Tyvek trays.<br>Six trays/box. | Sealed Tyvek trays.<br>Six trays/box. | Unknown | Unknown |
ど .
{5}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 2 3 1998
Mr. Gary Rauvola Regulatory Affairs Group Manager Bausch & Lomb Surgical, Storz Products 3365 Tree Court Industrial Blvd. St. Louis, MO 63122-6694
Re: K980488
> Trade Name: Storz Millennium Microsurgical System High Speed Vitrectomy Regulatory Class: II Product Code: 86 HQE Dated: February 4, 1998 Received: February 9, 1998
Dear Mr. Rauvola:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
Image /page/4/Picture/10 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread.
{6}------------------------------------------------
Page 2 - Mr. Gary M. Rauvola
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
A Ralph Rosenthal
A. Ralph Rosenthal, M.I Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## Storz Millennium™High Speed Vitrectomy System Premarket Notification
## INDICATIONS FOR USE
Device Name:
Storz Millennium™ High Speed Vitrectomy System
Indications for Use:
The Storz Millennium™ High Speed Vitrectomy System is a new device intended for removal of vitreous matter from the posterior vitreous cavity during ophthalmic vitrectomy surgery. The Storz high speed vitrectomy system consists of the Storz CX5800 module to be used with the Storz Millennium™ Microsurgical System, Storz CX5810 handpiece, and the Storz CX4804 sterile single use accessory pack.
Prescription USE ✓
Dennis L. McCarthy
vision Sign-Off)
L. vision of Ophthalmic Devices 510(k) Number_K 960 488 stice Cou
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.