STORZ DP4800 DELUXE , STORZ DP4801 BASIC W/REFLUX, DP4801 BASIC POSTERIOR VITRECTOMY PACK WITH VARIABLE PORT CUTTER & RX

K954816 · Storz Instrument Co. · HQE · Mar 4, 1996 · Ophthalmic

Device Facts

Record IDK954816
Device NameSTORZ DP4800 DELUXE , STORZ DP4801 BASIC W/REFLUX, DP4801 BASIC POSTERIOR VITRECTOMY PACK WITH VARIABLE PORT CUTTER & RX
ApplicantStorz Instrument Co.
Product CodeHQE · Ophthalmic
Decision DateMar 4, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.4150
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The STORZ DP4801, DP4800, and DP4801VP surgical accessory packs contain a vitrectomy probe and attached tubing set to be used in conjunction with the STORZ PREMIERE® or DAISY® Microsurgical Systems in the performance of posterior vitrectomy.

Device Story

Surgical accessory packs containing vitrectomy probe, tubing set, and collection cassette; used with STORZ PREMIERE® or DAISY® Microsurgical Systems. Probe utilizes pneumatic guillotine cutting mechanism; 20-gauge needle size. Used in ophthalmic surgery by surgeons. Safety feature: probe cutter opens when turned off. Single-patient use; provided sterile in PETG tray with TYVEK lid. Facilitates vitreous cutting and aspiration during posterior vitrectomy.

Clinical Evidence

No clinical data; bench testing only.

Technological Characteristics

Type 304 stainless steel probe; pneumatic drive mechanism; guillotine cutting action; 20-gauge needle. Sterile, single-patient use. Packaging: PETG tray with TYVEK lid.

Indications for Use

Indicated for patients undergoing posterior vitrectomy procedures.

Regulatory Classification

Identification

A vitreous aspiration and cutting instrument is an electrically powered device, which may use ultrasound, intended to remove vitreous matter from the vitreous cavity or remove a crystalline lens.

Special Controls

*Classification.* Class II (special controls). The device, when it is phacofragmentation unit replacement tubing, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.

Predicate Devices

Submission Summary (Full Text)

{1} LOTZ MAR - 4 1996 K954816 Premarket Notification STORZ Instrument Company # 510(k) SUMMARY OF SAFETY AND EFFICACY The following information is submitted in accordance with the requirements of 21 CFR 807.92: Contact Person: Manager, Regulatory Affairs Date Prepared: October 16, 1995 Proprietary Names: STORZ DP4801 Basic Posterior Vitrectomy Pack; STORZ DP4800 Deluxe Posterior Vitrectomy Pack; STORZ Basic Posterior Vitrectomy Pack with Variable Port Cutter. Common/Usual Name: Ophthalmic surgical system accessory packs for posterior Vitrectomy Classification Name: Instrument, vitreous cutting and aspirating, AC-powered (86HQE) Device Description/Intended Use: The STORZ DP4801, DP4800, and DP4801VP surgical accessory packs contain a vitrectomy probe and attached tubing set to be used in conjunction with the STORZ PREMIERE® or DAISY® Microsurgical Systems in the performance of posterior vitrectomy. Predicate Device: The STORZ DP4801, DP4800, and DP4801VP are substantially equivalent to the STORZ MVS-3201, DP4400 and DP4400VP anterior vitrectomy packs (K953575). Predicate Comparison: A chart comparing the STORZ DP4801, DP4800, and DP4801VP posterior vitrectomy packs to the predicate device, demonstrating substantial equivalence, is attached. Submitted by: Michael H. Southworth Storz Instrument Company 3365 Tree Court Industrial Blvd. St. Louis, MO 63122 Ph. 314/225-5051 Court Industrial Blvd. MO 63122-6094 5051 491 STORZ STI 225-2365 {2} # DEVICE COMPARISON CHART | DEVICE CHARACTERISTIC | Storz DP4801 PACK | Storz DP4800 PACK | Storz DP4801VP PACK | PREDICATE DEVICES: Storz MVS3201 PACK Storz DP4400 PACK Storz DP4400VP PACK | | --- | --- | --- | --- | --- | | Pack Contents | Vitrectomy Probe with Tube Set and Collection Cassette | Vitrectomy Probe with Tube Set, Collection Cassette plus Deluxe Accessories | Vitrectomy Probe with Tube Set and Collection Cassette | Vitrectomy Probe with Tube Set | | Indication For Use | Posterior Vitrectomy | Posterior Vitrectomy | Posterior Vitrectomy | Anterior Vitrectomy | | Opthalmic Surgical System to be Used | STORZ PREMIERE® and DAISY® Systems | STORZ PREMIERE® and DAISY® Systems | STORZ PREMIERE® and DAISY® Systems | STORZ PREMIERE®, Protegé, and DAISY® Systems | | Probe Cutting Mechanism Type | Guillotine | Guillotine | Guillotine | Guillotine | | Probe Drive Mechanism Type | Pneumatic | Pneumatic | Pneumatic | Pneumatic | | Does Probe Cutter Open When Turned Off As A Safety Measure? | YES | YES | YES | YES | | Probe Needle Size | 20 Gauge | 20 Gauge | 20 Gauge | 20 Gauge | | Patient Contact Materials | Type 304 Stainless Steel | Type 304 Stainless Steel | Type 304 Stainless Steel | Type 304 Stainless Steel | | Instructions For Use Included In Pack? | YES | YES | YES | YES | {3} # DEVICE COMPARISON CHART | DEVICE CHARACTERISTIC | Storz DP4801 PACK | Storz DP4800 PACK | Storz DP4801VP PACK | PREDICATE DEVICES: Storz MVS3201 PACK Storz DP4400 PACK Storz DP4400VP PACK | | --- | --- | --- | --- | --- | | Labeled For Single Patient Use Only? | YES | YES | YES | YES | | Labeled Per 21CFR §801.109(b)? | YES | YES | YES | YES | | Provided Sterile? | YES | YES | YES | YES | | Packaging | PETG Tray with TYVEK Lid | PETG Tray with TYVEK Lid | PETG Tray with TYVEK Lid | PETG Tray with TYVEK Lid |
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...