The 27+ Retina Instrument Set instruments, along with the Constellation Vision System or Next Generation Laser System, are intended for the following uses: 27+ UltraVit Probe: Vitreous aspiration & cutting. Membrane cutting & aspiration. Lens removal. Dissect tissue in the eye. (The 27+ UltraVit Probe is intended to be used with the Alcon Constellation Vision System.) 27+ Flex-Tip Laser Probe: Retinal photocoagulation, panretinal photocoagulation and intravitreal photocoagulation of vascular and structural abnormalities of the retina and choroid including: Proliferative and nonproliferative retinopathy (including diabetic); Choroidal neovascularization secondary to age-related macular degeneration; Retinal tears and detachments; Macular edema; Retinopathy of prematurity; Choroidal neovascularization; Leaking microaneurysms. Iridotomy/Iridectomy for treatment of Chronic/Primary Open Angle Glaucoma (COAG,POAG) and Refractory Glaucoma. Trabeculoplasty for treatment of Chronic/Primary Open Angle Glaucoma (COAG,POAG) and refractory Glaucoma. And other laser treatments including: Internal sclerostomy; Lattice degeneration; Central and Branch Retinal Vein Occlusion; Suturelysis; Vascular and pigmented skin lesions. (The 27+ Flex-Tip Probe is intended to be used with 532 nm laser systems such as the Alcon Constellation Vision System, or the Alcon PurePoint Laser System.) 27+ Endoilluminator Probe: Endoillumination (The 27+ Endoilluminator Probe is intended to be used with the Alcon Constellation Vision System) 27+ Valved Entry System: Scleral incision; Canulae for posterior segment instrument access; Venting (of valved cannulae) 27+ Infusion Cannula: Posterior segment infusion (liquid or gas)
Device Story
27+ Retina Instrument Set comprises five surgical accessories: UltraVit probe, Flex-Tip laser probe, endoilluminator, valved entry system, and infusion cannula. Devices are modified versions of larger-gauge (25+) CONSTELLATION accessories, featuring smaller diameter tips for less invasive vitreoretinal surgery. Used in clinical settings by ophthalmologists during posterior segment procedures. Instruments interface with Alcon CONSTELLATION Vision System or PurePoint Laser System. Laser probe delivers 532 nm laser energy for photocoagulation; UltraVit probe performs mechanical vitreous cutting/aspiration. Devices facilitate access, illumination, infusion, and tissue manipulation. Benefits include reduced invasiveness due to smaller gauge size. Single-use, sterile, EO-sterilized.
Clinical Evidence
Bench testing only. Biocompatibility testing performed per ISO 10993 standards (cytotoxicity, irritation, sensitization, systemic toxicity). Sterilization validated per EN ISO 11135-1:2007. Functional testing confirmed performance requirements met.
Technological Characteristics
Vitreoretinal surgical instruments; 27-gauge diameter. Materials compliant with ISO 10993. Energy sources: pneumatic (aspiration/cutting), 532 nm laser (photocoagulation), light source (endoillumination). Single-use, EO-sterilized (SAL 10^-6). Manufactured per 21 CFR 820 and ISO 14971.
Indications for Use
Indicated for patients requiring vitreoretinal surgery, including treatment of proliferative/nonproliferative retinopathy (diabetic), age-related macular degeneration, retinal tears/detachments, macular edema, retinopathy of prematurity, glaucoma (COAG/POAG/refractory), retinal vein occlusion, and other vascular/structural retinal/choroidal abnormalities.
Regulatory Classification
Identification
An ophthalmic photocoagulator is an AC-powered device intended to use the energy from an extended noncoherent light source to occlude blood vessels of the retina, choroid, or iris.
CONSTELLATION® Vision System (and accessories) (K101285)
Submission Summary (Full Text)
{0}------------------------------------------------
OCT 1 7 2011
Alcon Research, Ltd. Traditional 510(k) Premarket Notification 27+ Retina Instrument Set
March 31, 2011
## PREMARKET NOTIFICATION 510(k) SUMMARY 5.
This 510(k) Summary is provided here and is also included in Attachment B.
The submitter of the 510(k) is:
Martin A. Kaufman Director, Regulatory Affairs Alcon Research, Ltd. 18500 Alton Parkway Irvine, CA 92618 Phone: (949) 753-6250 Fax: (949) 753-6237
Devices Subject to this 510(k):
Trade Name: 27+ Retina Instrument Set Common Name: Vitreoretinal Surgical Accessories Classification Name: See below
| Modified<br>Instrument | 21 CFR | Classification name | Class | Product<br>Code | Predicate<br>510(k) | Date<br>Cleared |
|---------------------------------|----------|-----------------------------------------------|-------|-----------------|---------------------|----------------------|
| 27+ UltraVit®<br>Probe | 886.4150 | Vitreous aspiration and<br>cutting instrument | II | HQE | K093305 | 04/02/10 |
| 27+ Flex-Tip<br>Laser Probe | 886.4390 | Ophthalmic laser | II | HQB | K062624 | 11/05/07 |
| 27+<br>Endoilluminator<br>Probe | 876.1500 | Endoscope and<br>accessories | II | MPA | K063583<br>K101285 | 05/09/08<br>11/12/10 |
| 27+ Valved Entry<br>System | 886.4350 | Manual ophthalmic<br>surgical instrument | I | NGY | n/a | |
| 27+ Infusion<br>Cannula | 886.4350 | Manual ophthalmic<br>surgical instrument | I | HMX | n/a | |
{1}------------------------------------------------
Alcon Research, Ltd. Traditional 510(k) Premarket Notification 27+ Retina Instrument Set
#### 5.1. Predicate Devices:
The legally marketed devices(s) to which we are claiming equivalence to are:
| Predicate Date<br>510(k)<br>Cleared | Predicate Device | Predicate Instrument | Modified Instrument |
|--------------------------------------------|-----------------------------------------------------------------------------------------------|---------------------------------|------------------------------|
| K093305<br>04/02/10 | Enhanced UltraVit® Probe | 25+ UltraVit®<br>Probe | 27+ UltraVit® Probe |
| K062624<br>11/05/07 | Next Generation Laser and<br>accessories (PurePoint®<br>laser) | 25+ Flex-Tip Laser<br>Probe | 27+ Flex-Tip Laser<br>Probe |
| K063583<br>05/09/08<br>K101285<br>11/12/10 | Alcon Vision System (and<br>accessories)<br>CONSTELLATION® Vision<br>System (and accessories) | 25+<br>Endoilluminator<br>Probe | 27+ Endoilluminator<br>Probe |
| n/a | n/a | 25Ga Valved<br>Entry System | 27+ Valved Entry<br>System |
| n/a | n/a | 25Ga Infusion<br>Cannula | 27+ Infusion Cannula |
#### 5.2. Device Description
The five instruments comprising the 27+ Retina Instrument Set are modified versions of approved larger gauge CONSTELLATION accessories but with smaller diameter instrument tips for less invasive surgery. All other features are exactly the same as the respective predicate devices cleared in K093305, K062624, K063583, and K101285.
{2}------------------------------------------------
## 5.3. Indications for Use:
The 27+ Retina Instrument Set instruments, along with the CONSTELLATION Vision System or Next Generation Laser System, are intended for the following uses:
| 27+ UltraVit® Probe | Vitreous aspiration & cutting. Membrane cutting & aspiration. Remove a crystalline lens. Dissect tissues in the eye. |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 27+ Flex-Tip Laser<br>Probe | Retinal photocoagulation, panretinal photocoagulation and<br>intravitreal photocoagulation of vascular and structural<br>abnormalities of the retina and choroid including: Proliferative and nonproliferative retinopathy (including diabetic); Choroidal neovascularization secondary to age-related macular degeneration; Retinal tears and detachments; Macular edema; Retinopathy of prematurity; Choroidal neovascularization; Leaking microaneurysms. Iridotomy/Iridectomy for treatment of Chronic/Primary Open Angle<br>Glaucoma (COAG, POAG) and Refractory Glaucoma. Trabeculoplasty for treatment of Chronic/Primary Open Angle<br>Glaucoma (COAG, POAG) and refractory Glaucoma. And other laser treatments including: Internal sclerostomy; Lattice degeneration; Central and Branch Retinal Vein Occlusion; Suturelysis; Vascular and pigmented skin lesions. |
| 27+ Endoilluminator<br>Probe | Endoillumination |
| 27+ Valved Entry<br>System | Scleral incision |
| 27+ Infusion Cannula | Posterior segment infusion |
{3}------------------------------------------------
## 5.4. Brief Summary of Nonclinical test and Results:
Biocompatibility evaluations of materials coming in contact with the patient or patient fluid path have been performed to the following standards:
| Standard # | Title |
|--------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| 10993-1: 2009<br>AAMI/ANSI/ISO | Biological Evaluation of Medical Devices -- Part 1: Evaluation and<br>testing within the Risk Management Process |
| 10993-5: 2009<br>AAMI/ANSI/ISO | Biological Evaluation of Medical Devices -- Part 5: Tests for In Vitro<br>Cytotoxicity |
| 10993-7: 2008<br>AAMI/ANSI/ISO | Biological Evaluation of Medical Devices - Part 7: Ethylene Oxide<br>Sterilization Residuals |
| 10993-10:<br>2002/R:2008<br>AAMI/ANSI/BE78 | Biological Evaluation of Medical Devices -- Part 10: Tests for Irritation<br>and Delayed-type Hypersensitivity |
| 10993-10:2009<br>EN ISO | Biological Evaluation of Medical Devices - Part 10: Tests for Irritation<br>and Delayed -type Hypersensitivity |
| 10993-11:2009<br>AAMI/ANSI/ISO | Biological Evaluation of Medical Devices -- Part 11: Tests for systemic<br>toxicity |
| 10993-12:2009<br>AAMI/ANSI/ISO | Biological Evaluation of Medical Devices --Part 12: Sample Preparation<br>and Reference.Materials |
These accessories are provided sterile and are intended for single use only. These products are EO sterilized and the process has been validated to a SAL of 106 per FDA Recognized Consensus Standard EN ISO 11135-1:2007: Sterilization of health care products - Ethylene oxide - Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices.
Technological characteristics affecting clinical performance are similar to that of predicate devices previously listed. The 27+ Retina Instrument Set has been developed and will be manufactured in compliance with 21 CFR 820 and ISO 14971:2007. Non-clinical testing noted above has demonstrated that the functional requirements have been met and that the modified devices are equivalent to the predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus, which is a symbol of medicine, with three lines representing the snakes and a base representing the staff. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Alcon Research, Ltd. c/o Mr. Martin Kaufman, RAC Director, Regulatory Affairs 15800 Alton Parkway Irvine, CA 92816
0CT 1 7 2011
Re: K110951
Trade/Device Name: 27+ Retina Instrument Set Regulation Number: 21 CFR 886.4150 Regulation Name: Vitreous aspiration and cutting instrument Regulatory Class: II Product Codes: HQE, HMX, HQB, MPA, NGY Dated: October 6, 2011 Received: October 7, 2011
Dear Mr. Kaufman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Dr to and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contractly und warranties. We remind you, however, that device labeling must be truthful and not midlers]
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other madon a color incant or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{5}------------------------------------------------
Page 2 - Mr. Martin Kaufman, RAC
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kesia Alexander
for
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological,
and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
510(k) Number (if known): K110951
Device Name: 27+ Retina Instrument Set
Indications for Use:
The 27+ Retina Instrument Set instruments, along with the Constellation Vision System or Next Generation Laser System, are intended for the following uses:
| 27+ UltraVit Probe | Vitreous aspiration & cutting. Membrane cutting & aspiration. Lens removal. Dissect tissue in the eye. (The 27+ UltraVit Probe is intended to be used with the Alcon<br>Constellation Vision System.) |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 27+ Flex-Tip Laser<br>Probe | Retinal photocoagulation, panretinal photocoagulation and<br>intravitreal photocoagulation of vascular and structural<br>abnormalities of the retina and choroid including: Proliferative and nonproliferative retinopathy (including<br>diabetic); Choroidal neovascularization secondary to age-related<br>macular degeneration; Retinal tears and detachments; Macular edema; Retinopathy of prematurity; Choroidal neovascularization; Leaking microaneurysms. Iridotomy/Iridectomy for treatment of Chronic/Primary Open<br>Angle Glaucoma (COAG,POAG) and Refractory Glaucoma. Trabeculoplasty for treatment of Chronic/Primary Open Angle<br>Glaucoma (COAG,POAG) and refractory Glaucoma. And other laser treatments including: Internal sclerostomy; Lattice degeneration; Central and Branch Retinal Vein Occlusion; Suturelysis; Vascular and pigmented skin lesions. (The 27+ Flex-Tip Probe is intended to be used with 532 nm laser<br>systems such as the Alcon Constellation Vision System, or the<br>Alcon PurePoint Laser System.) |
Bob
(Division Sign-Off) (Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K110951
{7}------------------------------------------------
Page 2/2
| 27+ Endoilluminator<br>Probe | • Endoillumination<br>(The 27+ Endoilluminator Probe is intended to be used with the<br>Alcon Constellation Vision System) |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------|
| 27+ Valved Entry<br>System | • Scleral incision<br>• Canulae for posterior segment instrument access<br>• Venting (of valved cannulae) |
| 27+ Infusion Cannula | • Posterior segment infusion (liquid or gas) |
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Rk
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K110951
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.