STORZ DP5100 ANTERIOR COLLECTION CASSETTE PACK

K961874 · Storz Instrument Co. · HQC · Aug 6, 1996 · Ophthalmic

Device Facts

Record IDK961874
Device NameSTORZ DP5100 ANTERIOR COLLECTION CASSETTE PACK
ApplicantStorz Instrument Co.
Product CodeHQC · Ophthalmic
Decision DateAug 6, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.4670
Device ClassClass 2

Indications for Use

The Storz DP5100 surgical accessory pack contains a single collection cassette. This component is intended to be used as an accessory to the Storz DAISY® (K854508), Storz Protegé® (K921758 and K950114) and Storz PREMIERE® (K894278, K921460, K946227) Ophthalmic Microsurgical Systems as a reservoir for aspirated fluid and tissue collected during anterior segment ophthalmic surgical procedures.

Device Story

Single-use collection cassette; accessory to Storz DAISY, Protegé, and PREMIERE ophthalmic microsurgical systems. Functions as reservoir for aspirated fluid/tissue during anterior segment surgery. Used in clinical/surgical settings by ophthalmic surgeons. Device collects surgical waste; facilitates fluid management during phacofragmentation procedures.

Clinical Evidence

Bench testing only; no clinical data provided.

Technological Characteristics

Anterior collection cassette; material: Phillips K-Resin; single-patient use; sterile; sterilization method: Gamma Irradiation; packaging: Tyvek and PET/PE pouch.

Indications for Use

Indicated for use as a reservoir for aspirated fluid and tissue during anterior segment ophthalmic surgical procedures.

Regulatory Classification

Identification

A phacofragmentation system is an AC-powered device with a fragmenting needle intended for use in cataract surgery to disrupt a cataract with ultrasound and extract the cataract.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1} K961874 storz AUG 6 1996 # Premarket Notification ## Storz Instrument Company ### 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS The following information is submitted in accordance with the requirements of 21 CFR 807.92: **Contact Person:** Regulatory Affairs Associate **Date Prepared:** May 9, 1996 **Proprietary Name:** Storz DP5100 Anterior Collection Cassette Pack **Common/Usual Name:** Ophthalmic microsurgical system accessory pack for anterior segment surgical procedures. **Classification Name:** System, Phacofragmentation (86 HQC). **Device Description/Intended Use:** The Storz DP5100 surgical accessory pack contains a single collection cassette. This component is intended to be used as an accessory to the Storz DAISY® (K854508), Storz Protegé® (K921758 and K950114) and Storz PREMIERE® (K894278, K921460, K946227) Ophthalmic Microsurgical Systems as a reservoir for aspirated fluid and tissue collected during anterior segment ophthalmic surgical procedures. **Predicate Device:** The modified Storz DP5100 is substantially equivalent to the existing Storz DP5100 Ambulatory Surgery Center (ASC) Collection Cassette Pack (K894278, K921758), and to the Storz DP4101 Basic Irrigation / Aspiration Pack (K953078). **Predicate Comparison:** A chart comparing the Storz DP5100, as modified, to the predicate devices, demonstrating substantial equivalence, is attached. **Submitted by:** Audrey Swearingen Storz Instrument Company STORZ* 3365 Tree Court Industrial Blvd St. Louis, MO 63122-6694 (314) 225-5051 FAX (314) 225-7365 75 {2} 76 2100217 # DEVICE COMPARISON CHART ## Storz DP5100 Anterior Collection Cassette Pack | Device | Storz DP5100, modified | Storz DP5100, existing | Storz DP4101 | | --- | --- | --- | --- | | Intended Use | Collection of Aspirant during Anterior Segment Ophthalmic Surgery | Collection of Aspirant during Anterior Segment Ophthalmic Surgery | Delivery of Irrigant and Collection of Aspirant during Anterior Segment Ophthalmic Surgery | | Components | Anterior Collection Cassette, Instruction Sheet | Anterior Collection Cassette | I/A Tubing, Test Chamber, Anterior Collection Cassette, Instruction Sheet | | Single Patient Use? | Yes | Yes | Yes | | Cassette Material | Phillips K-Resin® | LEXAN™ | Phillips K-Resin® | | Provided Sterile? | Yes | Yes | Yes | | Sterilization Method | Gamma Irradiation | Gamma Irradiation | Gamma Irradiation | | Packaging | Tyvek® and PET/PE Pouch, Six pouches per Box | Tyvek® and PET/PE Pouch, Six Pouches per Box | High-impact Polystyrene Tray With Tyvek® Lid, Six Packs per Box |
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...