IVAC MEDSYSTEM III, INFUSION PUMP W/DRUG EDITING SOFTWARE KIT

K961486 · Ivac Medical Systems · MRZ · Apr 17, 1998 · General Hospital

Device Facts

Record IDK961486
Device NameIVAC MEDSYSTEM III, INFUSION PUMP W/DRUG EDITING SOFTWARE KIT
ApplicantIvac Medical Systems
Product CodeMRZ · General Hospital
Decision DateApr 17, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5725
Device ClassClass 2

Indications for Use

The intended use for the DESK is to allow users the ability to add, modify, or delete drug names and dosing parameters to the existing drug list for the Dose Rate Calculator feature in the MedSystem III Infusion Pump. The intended use for the MedSystem III Infusion Pump is to deliver a wide variety of fluids over a broad range of infusion rates, at high levels of accuracy.

Device Story

DESK is PC-based software (Windows) for customizing drug lists in MedSystem III Infusion Pump's Dose Rate Calculator (DRC). User accesses pump's factory default drug list via Field Maintenance Software (FMS); adds, modifies, or deletes drug names and dosing parameters; downloads updated list to pump. MedSystem III delivers fluids/drugs at various rates. DESK enables permanent customization of drug lists compared to existing one-by-one manual entry. Used by clinical staff/technicians in healthcare settings to manage infusion parameters. Benefits include standardized, customized drug libraries for improved infusion accuracy and workflow efficiency.

Clinical Evidence

No clinical data; bench testing only.

Technological Characteristics

PC-based software (Windows) accessory. Operates via Field Maintenance Software (FMS) interface to communicate with MedSystem III Infusion Pump. No changes to pump hardware. Connectivity via FMS download to pump.

Indications for Use

Indicated for delivery of drugs and fluids (e.g., bloods, lipids, saline, dextrose) via enteral, intravenous, and epidural routes. No specific patient population or contraindications specified.

Regulatory Classification

Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

Predicate Devices

Submission Summary (Full Text)

{0} APR 17 1996 K961486 # 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS 1. SUBMITTER'S NAME: IVAC Medical Systems, Inc. 10221 Wateridge Circle San Diego, California 92121 (619) 458-7563 CONTACT PERSON: Renée L. Fluet Regulatory Affairs Specialist DATE PREPARED: April 12, 1996 2. DEVICE NAME: Proprietary Name: IVAC MedSystem III™ Infusion Pump with Drug Editing Software Kit (DESK) Common Name: Infusion Pump Classification Name: Infusion Pump 3. PREDICATE DEVICE: IVAC MedSystem III™ Infusion Pump 4. DEVICE DESCRIPTION: The DESK is a PC-based software tool operated under Microsoft Windows. This software tool will be used to customize the drug list for the Dose Rate Calculator (DRC) feature that is already available in the legally marketed MedSystem III Infusion Pump. The DESK allows the user to access the factory default drug list in the infusion pump via the Field Maintenance Software (FMS). After access the user can add, modify, or delete drug names and dosing parameters. The modified list can then be downloaded into the MedSystem III Infusion Pump via the FMS. The FMS is already available as a legally marketed accessory to the MedSystem III Infusion Pump. 5. INTENDED USE: The intended use for the DESK is to allow users the ability to add, modify, or delete drug names and dosing parameters to the existing drug list for the Dose Rate Calculator feature in the MedSystem III Infusion Pump. The intended use for the MedSystem III Infusion Pump is to deliver a wide variety of fluids over a broad range of infusion rates, at high levels of accuracy. 2 {1} 3 6. TECHNOLOGICAL CHARACTERISTICS: The MedSystem III Infusion Pump has not changed. The DESK is a software accessory with the same technological characteristics as the FMS software accessory. The primary function of the MedSystem III Infusion Pump with or without the DESK is the same: to deliver a wide variety of fluids over a broad range of infusion rates. A similar feature to the DESK is already available in the legally marketed MedSystem III Infusion Pump. This feature allows the user to customize the drug list on a single basis. DESK allows the user to customize the drug list on a permanent basis. The technological characteristics of the devices are the same and raise no new questions of safety and effectiveness. IVAC Medical Systems, Inc. concludes that the MedSystem III Infusion Pump with or without the DESK are substantially equivalent. {2} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 APR 17 1998 Ms. Renee L. Fluet *Regulatory Affairs Specialist ALARIS Medical™ Systems, Incorporated Corporate Office 10221 Wateridge Circle San Diego, California 92121-2733 Re: K961486 Trade Name: IVAC MedSystem III™, (MSIII) Infusion Pump with Drug Editing Software Kit, (DESK) Regulatory Class: II Product Code: MRZ Dated: January 14, 1998 Received: January 20, 1998 Dear Ms. Fluet: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of {3} Page 2 - Ms. Fluet the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Timothy A. Ulatowski Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4} IVAC MEDICAL SYSTEMS INDICATIONS FOR USE 510(k) Number (if known): _________________________________ Device Name: MedSystem III Infusion Pump with Drug Editing Software Kit (DESK) Indications for Use: The intended use of the MedSystem III Infusion Pump is to deliver a wide variety of drugs and fluids (e.g., bloods, lipids, saline, dextrose) over a broad range of infusion rates, at high levels of accuracy. IVAC’s policy is to not indicate specific drugs and/or fluids. Specific intended uses include enteral, intravenous, and epidural deliveries. The Drug Editing Software Kit (DESK) allows the user to customize the resident drug list for the Dose Rate Calculator (DRC) feature that is available in the legally marketed MedSystem III Infusion Pump. The current system allows custom changes on a one by one basis. The DESK system allows users to make permanent custom changes to the resident drug list. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use ☐ OR Over-The Counter Use ☐ (Per 21 CFR 801.109) (Division Sign-Off) (Division of Dental Infection Control, and General Hospital Devices) 510(k) Number 11961986 (4) Number _________________________ Corporate 10221 Wateridge Circle San Diego, California 92121-1579 P.O. Box 85335 San Diego, California 92186-5335 (619) 458-7000 Web Site: http://www.ivac.com/~ivac/ 5 ISO 9001 Quality System O
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...