CADD -SOLIS MEDICATION SAFETY SOFTWARE: ADMINISTRATOR AND POINT-OF-CARE

K082783 · Smiths Medical MD, Inc. · MRZ · Dec 17, 2008 · General Hospital

Device Facts

Record IDK082783
Device NameCADD -SOLIS MEDICATION SAFETY SOFTWARE: ADMINISTRATOR AND POINT-OF-CARE
ApplicantSmiths Medical MD, Inc.
Product CodeMRZ · General Hospital
Decision DateDec 17, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5725
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

The CADD®-Solis Medication Safety Software - Administrator allows use of a computer to create therapy based protocol libraries to be used with the CADD®-Solis Ambulatory Infusion Pump or CADD-Prizm® PCS II Ambulatory Infusion Pump (software revision H or higher). The CADD®-Solis Medication Safety Software -- Point of Care allows use of a computer to send therapy-based protocols developed by the CADD®-Solis Medication Safety Software - Administrator to the CADD®-Solis Ambulatory Infusion Pump and CADD-Prizm® PCS II Ambulatory Infusion Pump (software revision H or higher).

Device Story

Software system for PC/tablet; manages infusion pump programming via therapy-based protocol libraries. Comprises Administrator and Point-of-Care (POC) modules. Administrator module: creates/edits protocols, defines user access, prints barcodes, generates reports. POC module: downloads protocols to CADD-Solis/CADD-Prizm PCS II pumps, verifies settings against protocols, views history logs. Uses serial communication for data transfer; supports barcode scanners for input. Prevents duplicate entries for drugs, protocols, or users. Used in clinical settings by healthcare staff to standardize pump programming, reduce manual entry errors, and improve medication safety through protocol-based infusion management.

Clinical Evidence

No clinical data. Bench testing only; included software validation and verification of software-controlled programming functions and pump operation.

Technological Characteristics

PC-based software application; operates on standard hardware (PCs/tablets). Employs serial communication for data transfer to infusion pumps. Supports barcode scanner integration. Features include protocol library management, user access control, and report generation. Software-based logic for protocol verification and pump setting management.

Indications for Use

Indicated for use by clinicians to create, manage, and download therapy-based infusion protocols to CADD-Solis and CADD-Prizm PCS II ambulatory infusion pumps (software revision H or higher).

Regulatory Classification

Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K082783 ### 510(k) Premarket Notification 510(k) Summary DEC 1 7 2008 ## GENERAL INFORMATION | Applicant's Name and Address: | Smiths Medical MD, Inc.<br>1265 Grey Fox Road<br>St. Paul, MN 55112 | |--------------------------------|------------------------------------------------------------------------------------------------| | Contact Person: | Phil Neururer<br>Sr. Regulatory Affairs Specialist | | Common/Usual Name: | Pump Communications System | | Proprietary Name: | CADD®-Solis Medication Safety Software | | Classification Name: | 21 CFR 880.5725, Accessories, Pump, Infusion | | Product Code | MRZ | | Equivalence Device Comparison: | CADD®-Solis Medication Safety Software, and<br>CADD-Sentry ProTM Medication Safety<br>Software | #### DEVICE DESCRIPTION II. #### CADD®-Solis Medication Safety Software The Smiths Medical MD, Inc. CADD® -Solis Medication Safety Software, a software program that operates on commercially available personal computers or similar hardware platforms such as tablets, is designed for pump programming of the CADD -Solis Planorms such as lables, is designed for pains programming so managers infusion Pump (software revision H or higher) through a therapy-based protocol database defined by the user. The CADD®-Solis Medication Safety Software consists of an Administrator and a Point-of-Care (POC) software module that employs serial communications to send and receive pump information. Both modules are compatible with barcode scanners (or similar input devices) through various PC connections. Barcode format is determined by the user; but is limited to 20 alphanumerical characters. The CADD®-Solis Medication Safety Software does not allow duplicative Drug, Protocol or User identification entries. The CADD®-Solis Medication Safety Software – Administrator module allows the user to create, cdit, and save therapy-based protocols and pump settings within user-defined protocol libraries. The Administrator user determines POC user access and library editing capabilities. Other Administrator module features include barcode printing, reports, and sending and receiving pump information. {1}------------------------------------------------ ## 510(k) Premarket Notification 510(k) Summary The CADD®-Solis Medication Safety Software – Point of Care module allows the user to download therapy based protocols to the CADD® Solis Ambulatory Infusion Pump and CADD-Prizm® PCS II Ambulatory Infusion Pump (software revision H or higher) and send and receive pump settings via serial communication. Additional features include storing and printing pump program settings and reports, verifying pump settings to established protocols and viewing history logs in the easier view of a PC monitor. #### INTENDED USE OF THE DEVICE III. ### CADD®-Solis Medication Safety Software - Administrator The CADD®-Solis Medication Safety Software - Administrator allows use of a computer to create therapy based protocol libraries to be used with the CADD®-Solis Ambulatory Infusion Pump or CADD-Prizm® PCS II Ambulatory Infusion Pump (software revision H or higher). ### CADD®-Solis Medication Safety Software - Point of Care The CADD®-Solis Medication Safety Software -- Point of Care allows use of a computer to send therapy-based protocols developed by the CADD®-Solis Medication Safety Software - Administrator to the CADD®-Solis Ambulatory Infusion Pump and CADD-Prizm® PCS II Ambulatory Infusion Pump (software revision H or higher). #### IV. DEVICE COMPARISON ### CADD®-Solis Medication Safety Software The CADD® -Solis Medication Safety Software was compared to and found to be substantially equivalent to the following commercially available predicate devices: CADD® Solis Medication Safety Software, and CADD-Sentry Pro™ Medication Safety Softwarc. #### SUMMARY OF STUDIES V. ### A. Functional Testing Test plans associated with software validation, verification of software controlled programming functions and pump operation were performed. ### B. Clinical Studies Human clinical studies were deemed not necessary to evaluate the safety or effectiveness of the CADD®-Solis Medication Safety Software. ### C. Conclusions Drawn from the Studies Based upon the information provided, the CADD®-Solis Medication Safety Software is safe, effective and performs to established specifications. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" is arranged in a circular pattern around the eagle. The eagle is black, and the text is also black. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Phil Neururer Senior Regulatory Affairs Specialist Smiths Medical MD, Incorporated 1265 Grey Fox Road St. Paul, Minnesota 55112 DEC 1 7 2008 Re: K082783 Trade/Device Name: CADD® - Solis Medication Safety Software - Administrator Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: MRZ Dated: December 11, 2008 Received: December 12, 2008 Dear Mr. Neururer: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting vour device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Mr. Neururer Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # SMITHS MEDICAL MD, INC. 510(k) Premarket Notification CADD® -Solis Medication Safety Software #### Indications for Use 510(k) Number: #### Device Name: CADD® -Solis Medication Safety Software - Administrator Indications for Use: "The CADD® -Solis Medication Safety Software - Administrator allows use of a computer to create therapy based protocol libraries to be used with the CADD®-Solis Ambulatory Infusion Pump or CADD-Prizm® PCS II Ambulatory Infusion Pump (software revision H or higher)." OR Prescription Use _____________________________________________________________________________________________________________________________________________________________ Over-The Counter Use Per 21 CFR 801.109 ### Device Name: CADD® -Solis Medication Safety Software - Point of Care Indications for Use: "The CADD"-Solis Medication Safety Software -- Point of Care allows use of a computer to send therapy-based protocols developed by the CADD® -Solis Medication Safety Software -Administrator to the CADD®-Solis Ambulatory Infusion Pump and CADD-Prizm® PCS II Ambulatory Infusion Pump (software revision H or higher)." Prescription Use OR Over-The Counter Use Per 21 CFR 801.109 (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Anton Bm (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control. Dental Devices 510(k) Number: TAB 5 - Page 1 of 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...