K955543 · Fukuda Denshi USA, Inc. · IYO · Feb 11, 1997 · Radiology
Device Facts
Record ID
K955543
Device Name
ULTRASOUND SYSTEM MODEL UF-3500
Applicant
Fukuda Denshi USA, Inc.
Product Code
IYO · Radiology
Decision Date
Feb 11, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1560
Device Class
Class 2
Attributes
Pediatric
Indications for Use
This device is intended to be used for applications in fetal, abdominal, intra-operative, pediatric, small organ, transrectal, and peripheral vessel scanning. Ultrasonic probes are available to obtain images either transrectally or intraoperatively. It is intended to be used by or on the order of a physician or similarly qualified health care professional. The UF-3500 is intended to be used in a doctors office and all hospital environments; ER ICU, CCU, OR, etc. This device is intended to be used on any patient; neonate, pediatric, or adult; where the placement and positioning of the transducer does not interfere with or complicate the treatment of the patient. This device is not intended for home use.
Device Story
Portable general purpose ultrasound scanner; projects pulsed sound beams into body tissue; measures acoustic pulse duration/depth/location of tissue interfaces. Input: ultrasound transducer signals. Processing: microprocessor-based signal processing; calculates distance, area, circumference, volume, angle, and fetal metrics (gestational age, expected date of confinement, fetal weight). Output: 16-level greyscale B-Mode image on 7-inch monitor. Used in clinics/hospitals (ER, ICU, CCU, OR) by physicians/technicians. Healthcare providers use images/measurements for diagnostic assessment of fetal/abdominal abnormalities. Benefits: non-invasive visualization of internal structures for clinical management.
Clinical Evidence
Bench testing only. Validation included environmental testing per FDA 1993 guidance, acoustic output measurement per AIUM 1992 standards, and electrical safety testing per ANSI/AAMI ES1-1993. Hazard analysis and software validation conducted to verify design specifications.
Technological Characteristics
Portable ultrasound scanner; microprocessor-controlled; linear transducers; 7-inch greyscale monitor. Connectivity: supports up to two probes (one active). Electrical safety: UL 601, CSA 22.2, AAMI standards. Acoustic output: compliant with AIUM 1992 standards. Software: embedded microprocessor-based measurement and signal processing.
Indications for Use
Indicated for fetal, abdominal, intra-operative, pediatric, small organ, transrectal, and peripheral vessel scanning in neonate, pediatric, and adult patients. Used by or on order of physician/qualified healthcare professional in clinical/hospital settings.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
FUKUDA DENSHI FF sonic model UF-4500 General Purpose Ultrasound Scanner (K922208)
Submission Summary (Full Text)
{0}
FEB 11 1997
K955543
# 510(k) Summary
## FUKUDA DENSHI FF sonic model UF-3500
General Purpose Ultrasound Scanner
## Submitter:
FUKUDA DENSHI AMERICA CORP.
11725 NE 65th St., Building C
Redmond WA 98052-4911
Tel: 206/881-7737
Fax: 206/869-2018
## Contact Person:
David J. Geraghty
Regulatory Affairs Manager
FUKUDA DENSHI AMERICA CORP.
11725 NE 65th St., Building C
Redmond WA 98052-4911
Tel: 206/881-7737
Fax: 206/869-2018
## Date Prepared:
December 4, 1995
Revised October 23, 1996
## Device Name:
### Proprietary Name:
FF sonic model UF-3500 General Purpose Ultrasound Scanner
### Common Name:
General Purpose Ultrasound Scanner
### Classification Name:
System, Imaging, Pulsed Echo, Ultrasonic
## Legally Marketed Device:
FUKUDA DENSHI FF sonic model UF-4500 General Purpose Ultrasound Scanner (K922208)
UF-3500 510(k) Additional Information
SUMMAR~1-jan16.doc
K955543
January 16, 1997
Page 1 of 5
{1}
# Description:
The model UF-3500 is a portable, General Purpose Ultrasound Scanner. This ultrasonic device is designed to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interface and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. The resulting information is displayed on the system's 7 inch diagonal video monitor as a 16 level greyscale B-Mode image.
The system provides user flexibility for different measurement and calculations that are incorporated into the unit. General measurements include distance, area, circumferential length, volume, and angle. Fetal calculations include gestational week, expected date of confinement, and fetal weight. All fetal measurements are derived from either Tokyo University's system, Osaka University's system or the American/European system; the same systems used on the UF-4500.
The UF-3500 can be used with any of eight (8) different linear transducers, only one of which may be active at any one time. An option is provided that will allow two probes to be simultaneously connected to the UF-3500; again, only one of which may be active at any time.
# Intended Use:
This device is intended to be used for applications in fetal, abdominal, intra-operative, pediatric, small organ, transrectal, and peripheral vessel scanning. Ultrasonic probes are available to obtain images either transrectally or intraoperatively. It is intended to be used by or on the order of a physician or similarly qualified health care professional. The UF-3500 is intended to be used in a doctors office and all hospital environments; ER ICU, CCU, OR, etc. This device is intended to be used on any patient; neonate, pediatric, or adult; where the placement and positioning of the transducer does not interfere with or complicate the treatment of the patient. This device is not intended for home use.
UF-3500 510(k) Additional Information
SUMMAR~1-jan16.doc
K955543
January 16, 1997
Page 2 of 5
{2}
The UF-3500 applies ultrasound energy though the abdominal wall to obtain an image of the fetus and abdominal organs that can be used to determine the gestational age of the fetus or to detect abnormalities of the fetus or abdominal organs.
The UF-3500, when used with the transrectal probe, applies ultrasound energy though the rectal wall to obtain an image of the abdominal organs that can be used to detect abnormalities of these organs.
## Technological Characteristics
The UF-3500 incorporates a microprocessor in the same manner as the predicate device. Transducers are limited to linear probes, again the same probes available for the predicate device. (Transrectal and intraoperative probes are added to the UF-4500 under a separate 510(k)filing.)
These technological differences do not effect the safety or efficacy of the device. Any safety issues that may be raised by a software controlled medical device are the same issues already addressed by the predicate device and are addressed in the systems hazard analysis and in the system validation.
## Testing:
Laboratory testing was conducted to validate and verify the FUKUDA DENSHI FF sonic model UF-3500 General Purpose Ultrasound Scanner met all design specifications and was substantially equivalent to the FUKUDA DENSHI model UF-4500. This testing consisted of all environmental testing identified in the FDA's DCRND November 1993 "Reviewer Guidance Document for Premarket notification Submissions" Draft Guidance Document. Additional testing was performed to demonstrate compliance with the voluntary AIUM "Acoustic Output Measurement and Labeling Standard for Diagnostic Ultrasound Equipment", 1992 and ANSI/AAMI ES1-1993, "Safe current limits for electromedical apparatus". Finally, a hazard analysis of the system and its software was performed and testing was conducted to validate the systems overall operation.
Although the device is neither life supporting nor life sustaining, diagnostic information derived from the use of
UF-3500 510(k) Additional Information
SUMMAR~1-jan16.doc
K955543
January 16, 1997
Page 3 of 5
{3}
the device may be critical to the proper management of the patient.
The other area of concern is the safety of the patient in that the device intentionally inputs energy, projects a pulsed sound beam, into body tissue. The UF-3500's acoustic out power levels were tested and found to be below the pre-amendment acoustic power intensity levels listed in the "Revised 510(k) Diagnostic Ultrasound Guidance for 1993" document.
So, the areas of risk for this device are the same as other devices in this class, and are the following:
- Electrical shock
Excessive electrical chassis leakage current can disturb the normal electrophysiology of the heart, and possibly leading to the onset of cardiac arrhythmias.
- Excessive Ultrasound Energy
Excessive ultrasound energy can cause localized heating and possible tissue damage.
- Misdiagnosis
- Inadequate design of the signal processing and measurement circuitry or program can lead to generation of inaccurate diagnostic data. If inaccurate diagnostic data are used in managing the patient, the physician may prescribe a course of treatment that places the patient at risk unnecessarily.
- Inadequate design of the device's software, used to make various measurements, can lead to generation of inaccurate diagnostic data. If inaccurate diagnostic data are used in managing the patient, the physician may prescribe a course of treatment that places the patient at risk unnecessarily.
The design of the UF-3500 has taken into account all the above. The device is designed to meet UL 601, CSA 22.2 and AAMI standards for electrical safety for medical equipment to prevent the possibility of excessive electrical leakage current to the patient.
UF-3500 510(k) Additional Information
SUMMAR~1-jan16.doc
K955543
January 16, 1997
Page 4 of 5
{4}
# Conclusion:
The conclusions drawn from the laboratory testing of the FUKUDA DENSHI FF sonic model UF-3500 General Purpose Ultrasound Scanner demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed predicate device, the FUKUDA DENSHI model UF-4500.
UF-3500 510(k) Additional Information
SUMMAR~1-jan16.doc
K955543
January 16, 1997
Page 5 of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.