CTS-310(B)
K984161 · Shantou Institute of Ultrasonic Instuments · IYO · May 5, 1999 · Radiology
Device Facts
| Record ID | K984161 |
| Device Name | CTS-310(B) |
| Applicant | Shantou Institute of Ultrasonic Instuments |
| Product Code | IYO · Radiology |
| Decision Date | May 5, 1999 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 892.1560 |
| Device Class | Class 2 |
| Attributes | Pediatric |
Intended Use
Ultrasonic pulsed echo imaging and measurement for fetal imaging and other abdominal as well as pediatric, small organ, neonatal cephalic, adult cephalic, and heart.
Device Story
CTS-310B is a portable electronic linear/convex scanning ultrasonotomograph; utilizes ultrasonic pulsed echo technology to generate tomographic images of fetal, abdominal, pediatric, small organ, and cardiac structures. System inputs include signals from EZU-PL12 (linear) or EZU-PC3A (convex) transducers; digital scan converter (DSC) processes signals for display on a 9-inch video monitor. Clinicians operate the system via keyboard and trackball to perform measurements (distance, area, circumference, time interval, velocity, depth, heart rate) and calculations (volume, pregnant weeks, heart function). Used in clinical settings for diagnostic imaging; output assists physicians in anatomical assessment and clinical decision-making. Benefits include non-invasive visualization of internal structures.
Clinical Evidence
Bench testing only. No clinical data provided. Substantial equivalence is based on comparative technical specifications, performance parameters (e.g., frame rate, detecting depth, grey scale), and adherence to safety standards (IEC 601-1).
Technological Characteristics
Portable electronic scanning ultrasonotomograph; 80-element linear/convex probes; 4-step dynamic focusing with variable aperture; 16 grey scale levels; 512x512x4 bit digital scan converter; PAL/NTSC video output; 100-234V power supply; IEC 601-1 Class I Type B safety compliance; trackball-based cursor control.
Indications for Use
Indicated for fetal, abdominal, pediatric, small organ (thyroid, breast, testes), neonatal cephalic, adult cephalic, and cardiac imaging using B-mode, M-mode, and combined modes. Prescription use only.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
- Hitachi Medical Corporation EUB-310 (K862867)
Related Devices
- K013327 — CTS-415 · Shantou Institute of Ultrasonic Instuments · Dec 14, 2001
- K041455 — SSI-600/SSI-800 DIAGNOSTIC ULTRASOUND IMAGING SYSTEM · Sonoscape Company Limited · Jun 10, 2004
- K963065 — CTS-285 WITH EZU-PL21 AND EZU-PC3B TRANSDUCERS · Shantou Institute of Ultrasonic Instuments · May 29, 1997
- K963068 — CTS-200 WITH EZU-PL21 TRANSDUCER · Shantou Institute of Ultrasonic Instuments · May 29, 1997
- K012772 — CTS-485 ULTRASOUND IMAGING SYSTEM · Shantou Institute of Ultrasonic Instuments · Oct 1, 2001
Submission Summary (Full Text)
{0}------------------------------------------------
KG84161
韶声仪器研究
5 1999 MAY
### Tab 11
### Premarket Notification [510(k)] Summary
September 7,1998
CTS-310B with EZU-PL12 and EZU-PC3A Transducers Trade Name:
Diagnostic Ultrasound System Common Name:
Ultrasonic Pulsed Echo Imaging System, 90 IYO Classification Name:
(per 21 CFR section 892.1560)
Shantou Institute of Ultrasonic Instruments
Manufacturer's Name:
Address:
Shantou Sez, 515041, China
#2, Jinsha Road, M.,
Corresponding Official: Mr. Jinzhong Yao
Title: President
Fax: (86) 754-8251499 Telephone: (86) 754-8250150
Predicate: Hitachi Medical Corporation EUB-310, K862867
Device Description: Model CTS-310B is a linear/convex electronic scanning ultrasonotomograph with a built-in digital scan converter (DSC). The unit allows heart, abdominal organic and fetal tomographic images to be observable on a video monitor. The main unit is portable and is separable from other equipment to be carried for its use at another place as well as being usable in combination with a 9-inch video monitor and a special photographic unit.
Intended Use: Ultrasonic pulsed echo imaging and measurement for fetal imaging and other abdominal as well as pediatric, small organ, neonatal cephalic, adult cephalic, and heart.
Technological Characteristics: See the attached "Comparison List" of the SIUI CTS-310B and the Hitachi EUB-310.
{1}------------------------------------------------
# COMPARIS ON LIST OF SIUI PRODUCT AND HITACHI PRODUCT
| | Performance | CTS-310 (SIUI) | EUB-310A (HITACHI) |
|---------------------------|--------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| main<br>unit | scanning<br>mode | electronic linear scanning<br>(compatible 80 elements linear probe) electronic convex sector<br>(compatible convex probe of 80 elements) | electronic linear scanning<br>(compatible 80 elements linear probe) electronic convex sector<br>(compatible convex probe of 80 elements) |
| | display mode | B mode B/B ode, M mode, B/M mode<br>simultaneously | B mode, B/B mode, M mode, B/M<br>mode simultaneously |
| | measure-<br>ment | in B mode display: distance, area<br>and circumference<br>in M mode display: time interval,<br>velocity, depth and heart rate | in B mode display: distance, area<br>and circumference<br>in M mode display: time interval,<br>velocity, depth and heart rate |
| | calculation | area, circumference, volume, heart<br>rate, pregnant weeks and heart<br>function | area, circumference, volume, heart<br>rate, pregnant weeks and heart<br>function |
| | focusing<br>mode | 4-steps dynamic focusing with<br>variable aperture and lens focusing | 4-steps dynamic focusing with<br>variable aperture and lens focusing |
| | scanning<br>width | linear scanning: 3.5MHz probe<br>104mm, 5MHz probe 61mm<br>convex sector scanning probe:<br>sector angle 60° | linear scanning: 3.5MHz probe<br>104mm, 5MHz probe 61mm<br>convex sector scanning probe:<br>sector angle 60° |
| | transmitting<br>voltage | pulse height 130V | pulse height 130V |
| | transmitting<br>pulse width | 3.5MHz pulse width 140 μs<br>5MHz pulse width 100 μs | 3.5MHz pulse width 140 μs<br>5MHz pulse width 100 μs |
| | detecting<br>depth | 3.5MHz probe maximum depth:<br>210mm<br>5MHz probe maximum depth:<br>140mm | 3.5MHz probe maximum depth:<br>210mm<br>5MHz probe maximum depth:<br>140mm |
| | zoom | 3.5MHz probe: x1.0, x1.2, x1.5, x2.0<br>selectable as well as depth shift<br>5MHz probe: x1.0, x1.5, x2.0<br>selectable as well as depth shift | 3.5MHz probe: x1.0, x1.2, x1.5, x2.0<br>selectable as well as depth shift<br>5MHz probe: x1.0, x1.5, x2.0<br>selectable as well as depth shift |
| | frame rate | the maximum is 40 frame / second | the maximum is 40 frame / second |
| | grey scale | 16 | 16 |
| | memory | 512 x 512 x 4 bit | 512 x 512 x 4 bit |
| | coordinate<br>transforma-<br>tion function | monitor can display electronic linear<br>scanning image or convex sector<br>scanning image | monitor can display electronic linear<br>scanning image or convex sector<br>scanning image |
| Performance | | CTS-310 (SIUI) | EUB-310A (HITACHI) |
| main<br>unit | video output | PAL or NTSC system TV signal<br>(confirmed in order) | PAL or NTSC system TV signal<br>(confirmed in order) |
| | power supply | 100V, 110V, 117V, 200V, 220V or<br>234V, ±10%, 50/60Hz, about<br>250W | 100V, 110V, 117V, 200V, 220V or<br>234V, ±10%, 50/60Hz, about<br>250W |
| | monitor | 9" black and white monitor | 5.5" and 9" black and white monitor |
| | volume<br>and<br>weight | 395 (W) x 1180 (H) x 728 (L)mm<br>approx. 65 kg | 405(W) X 710(1) X 1310(h)mm<br>approx. 40kg |
| | cursor shift | by trackball on keyboard | by joystick on keyboard |
| | electric<br>apparatus<br>safty<br>standard | conform to requirement of I class B<br>type apparatus of IEC 601-1 isolate<br>resistor testing: testing voltage<br>1000V L-L, L-G ≥ 10MΩ | conform to requirement of I class B<br>type apparatus of IEC 601-1 isolate<br>resistor testing: testing voltage<br>1000V L-L, L-G≥10MΩ |
| | | leakage current: U*-G≤500μA<br>P-G≤100μA | leakage current: U*-G≤500μA<br>P-G≤100µA |
| | | voltage resistance testing: L-G, P-G<br>1500V 2mA, no sparking or arcing in<br>1 minute<br>work normally when voltage changes<br>± 10% | voltage resistance testing: L-G, P-G<br>1500V 2mA, no sparking or arcing in<br>1 minute<br>work normally when voltage changes<br>± 10% |
| | probe ultrasound | linear probe: | linear probe: |
| transmitting<br>frequency | | EZU-PL12 80 elements,<br>5MHz<br>scanning width 61mm<br>EZU-PC3A 80 elements,<br>40R, 3.5MHz | EZU-PL11 80 elements, 3.5MHz<br>scanning width 104mm<br>EZU-PL12 80 elements, 5MHz |
| | | 3.5MHz scanning angle 60° | scanning width 61mm<br>EUU-L11S 80 elements, 3.5MHz<br>scanning width 84mm<br>convex sector probe: EZU-PC3A 80<br>elements, 40R, 3.5MHz, scanning<br>angle 60°<br>EZU-PC2A 80 elements, 40R,<br>5MHz, scanning angle 60°<br>EUP-V12A 40 elements, 40R, 5MHz, |
| | | | scanning angle 40° (transvaginal<br>probe) |
| operation | | temperature 5-40°C, relative | temperature 5-40 °C, relative |
| environment | | humidity 30-85% (no water drop) | humidity 30-85% (no water drop) |
| storage environment | | temperature -10-60°C, relative<br>humidity 30-95% (no water drop) air<br>pressure 700-1060mB | temperature -10-60°C, relative<br>humidity 30-95% (no water drip) air<br>pressure 700-1060mB |
| note | | | |
{2}------------------------------------------------
# COMPARIS ON LIST OF SIUI PRODUCT AND HITACHI PRODUCT
*U means main unit.
:
{3}------------------------------------------------
Public Health Service
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines representing hair or clothing.
MAY 5 1999 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Shantou Institute of Ultrasonic Instruments C/O Robert J. Morton, M.S. President & Certified Radiological Physicist Quality & Regulatory Services 1106 Chiltern Drive Walnut Creek, California 94596
Re: K984161 SIUI CTS-310B Ultrasound Imaging System Dated: April 1, 1999 Received: April 5, 1999 Regulatory Class: II 21 CFR 892.1560/Procode: 90 IYO
Dear Mr. Morton:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the SIUI CTS-310B Ultrasound Imaging System, as described in your premarket notification:
Transducer Model Number
#### EZU-PL12 EZU-PC3A
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (OS) for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic QS inspections, the FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification does not affect any obligation you may have under sections 531 and 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page -2 - Mr. Morton
This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:
> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
If you have any questions regarding the content of this letter, please contact Rodrigo Perez at (301) 594-1212.
Sincerely yours,
Orind be. Seymon
for
CAPT Daniel B. Sch ltz, M.D. Acting Director Division of Reproductive,
Abdominal, Ear, Nose and Throat, And Radiological Devices Office of Device Evaluation Center for Devices and Radiological Devices
Enclosure(s)
{5}------------------------------------------------
Page 2 Re: K984161 SIUI CTS-310B Ultrasound Imaging System
# Diagnostic Ultrasound Indications for Use Form Device Name: SIUI CTS-310B
| Clinical Application | A | B | M | PWD | CWD | Color Doppler | Amplitude Doppler | Color Velocity Imaging | Combined (specify) |
|----------------------------------|---|---|---|-----|-----|---------------|-------------------|------------------------|--------------------|
| Ophthalmic | | | | | | | | | |
| Fetal | | N | N | | | | | | N |
| Abdominal | | N | | | | | | | |
| Intraoperative (specify) | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | |
| Pediatric | | N | N | | | | | | N |
| Small Organ (specify) | | N | | | | | | | |
| Neonatal Cephalic | | | | | | | | | |
| Adult Cephalic | | | | | | | | | |
| Cardiac | | N | N | | | | | | N |
| Transesophageal | | | | | | | | | |
| Transrectal | | | | | | | | | |
| Transvaginal | | | | | | | | | |
| Transurethrat | | | | | | | | | |
| Intravascular | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | |
| Laparoscopic | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | |
| Musculo-skeletal<br>superficial | | | | | | | | | |
| Other (specify) | | | | | | | | | |
N=new indication
Additional Comments: thyroid gland, breast, testes____________________________________________________________________________________________________________________________
PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED.
Concurrence of CDRH, office of Device Evaluation (ODE)
Yhrist A. Lezon
---
(Division Sign-Off)
Prescription Use (Per 21 CFR 801.109)
Division of Reproductive, Abdominal, ENT, and Radiological I
510(k) Number K984110
{6}------------------------------------------------
# Scanhead Indications for Use Form Device Name: EZU-PL12
| Clinical Application | | Mode of Operation | | | | | | | | | |
|----------------------------------|---|-------------------|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|--|
| | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) | |
| Ophthalmic | | | | | | | | | | | |
| Fetal | | N | N | | | | | | | | |
| Abdominal | | N | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | | |
| Pediatric | | N | | | | | | | | | |
| Small Organ (specify) | | N | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethrat | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Perpheral Vascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | |
| Musculo-skeletal<br>superficial | | | | | | | | | | | |
| Other (specify) | | | | | | | | | | | |
N = new indication
Additional Comments: thyroid gland, breast, testes____________________________________________________________________________________________________________________________
..............................................................................................................................................................................
Concurrence of CDRH, office of Device Evaluation (ODE)
am
Prescription Use (Per 21 CFR 801.109)
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
510(k) Number K984161
{7}------------------------------------------------
## Scanhead Indications for Use Form Device Name: EZU-PC3A
| Clinical Application | | Mode of Operation | | | | | | | | | |
|--------------------------------------------|---|-------------------|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|--|
| | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) | |
| Ophthalmic | | | | | | | | | | | |
| Fetal | | N | N | | | | | | N | | |
| Abdominal | | N | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | | |
| Pediatric | | N | N | | | | | | N | | |
| Small Organ (specify) | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | N | N | | | | | | N | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethrat | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | |
| Musculo-skeletal<br>superficial | | | | | | | | | | | |
| Other (specify) | | | | | | | | | | | |
| N = new indication<br>Additional Comments: | | | | | | | | | | | |
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Concurrence of CDRH, office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
510(k) Number K984161
Devices