K230346 · Ge Medical Systems Ultrasound & Primary Care Diagnostic, LLC · IYN · Jun 20, 2023 · Radiology
Device Facts
Record ID
K230346
Device Name
Voluson SWIFT; Voluson SWIFT+
Applicant
Ge Medical Systems Ultrasound & Primary Care Diagnostic, LLC
Product Code
IYN · Radiology
Decision Date
Jun 20, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
AI/ML, Pediatric
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Follicle diameter measurement
—
Success rate >= 70%
70% or higher
223 volumes used for training; distribution by systems: Voluson E8 (20), Voluson E10 (105), Voluson P8 (29), Voluson S10 (48), Voluson S8 (21); distribution by probes: RIC5-9A-RS (98), RIC5-9-D (123), RIC6-12-D (2); distribution by countries: India (62), Germany (29), United Kingdom (29), Spain (103)
67 volumes (134 follicles) collected from Germany, India, Spain, United Kingdom, and USA
>1 (clinical experts)
Indications for Use
The Voluson SWIFT, Voluson SWIFT+ is a general-purpose diagnostic ultrasound system intended for use by qualified and trained healthcare professionals that are legally authorized or licensed by law in the country, state or other local municipality in which he or she practices for ultrasound imaging, measurement, display and analysis of the human body and fluid. The users may or may not be working under supervision or authority of a physician. Voluson SWIFT, Voluson SWIFT+ clinical applications include: Fetal Obstetrics; Abdominal (including renal and GYN/Pelvic); Pediatio; Small Organ (Breast, Testes, Thyroid, etc.); Neonatal Cephalic; Cardiac (Adult and Pediatric); Peripheral Vascular (PV); Musculo-skeletal Conventional and Superficial; Transrectal(including UrologyProstate) (TR); Transvaginal (TV). Mode of operation include: B, M, AMM (Anatonical M-Mode), PW Doppler, CW Doppler, Color Doppler, Color M Doppler, Power Doppler, HD-Flow (High Definition-Flow), Harmonic Imaging, Coded Pulse, 3D/4D Inaging mode, Elastography, B-Flow and Combined modes: BM, B/ Color, B/PWD, B/Power/PWD. The Voluson SWIFT / Voluson SWIFT+ are intended to be used in a hospital or medical clinic.
Device Story
Mobile diagnostic ultrasound system; provides real-time 2D/3D/4D imaging, measurement, and analysis. Inputs: acoustic signals via mechanical/electronic scanning transducers. Processing: digital acquisition and display; includes AI-based Auto-Caliper feature for follicle measurement. Outputs: ultrasound images and quantitative measurements displayed on console monitor. Used in hospitals/clinics by physicians, sonographers, or trained healthcare professionals. Assists clinicians in diagnostic evaluation of fetal, abdominal, cardiac, and vascular structures. Benefits: enables non-invasive visualization and quantification of anatomy/pathology to support clinical decision-making.
Clinical Evidence
No clinical studies required. Bench testing only. AI feature 'Auto-Caliper' verified using 67 3D volumes (134 follicles) collected from 6 countries. Success rate ≥70%. Accuracy evaluated by comparing AI-predicted diameters to manual ground truth; 60.2% of measurements showed <1.0mm deviation. Training dataset included 223 volumes from diverse systems (Voluson E8, E10, P8, S10, S8) and probes. Biocompatibility, cleaning, disinfection, and electrical/thermal safety verified per ISO 10993-1, IEC 60601-1, and IEC 60601-2-37.
Technological Characteristics
Mobile console with control panel and touch monitor. Supports linear, curved linear, and matrix phased array transducers. Imaging modes: B, M, AMM, PW/CW/Color/Power Doppler, HD-Flow, 3D/4D, Elastography, B-Flow. Connectivity: DICOM, networking. Safety standards: AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993-1, ISO 14971, IEC 62359. Biocompatible patient-contact materials.
Indications for Use
Indicated for diagnostic ultrasound imaging, measurement, and analysis of human body and fluid. Clinical applications: Fetal/Obstetrics, Abdominal (renal, GYN/Pelvic), Pediatric, Small Organ (Breast, Testes, Thyroid), Neonatal Cephalic, Adult Cephalic, Cardiac (Adult/Pediatric), Peripheral Vascular, Musculoskeletal (Conventional/Superficial), Transrectal (Urology/Prostate), and Transvaginal. Intended for use by qualified healthcare professionals in hospitals or clinics.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Voluson SWIFT, Voluson SWIFT+ Diagnostic Ultrasound System (K201828)
Reference Devices
Versana Premier Diagnostic Ultrasound System (K210438)
Voluson E Series Diagnostic Ultrasound System (K220358)
Voluson S Series Diagnostic Ultrasound System (K213642)
Submission Summary (Full Text)
{1}------------------------------------------------
June 20, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".
GE Medical Systems Ultrasound & Primary Care Diagnostic, LLC % Bryan Behn Regulatory Affairs Director 9900 Innovation Dr. WAUWATOSA WI 53226
Re: K230346
Trade/Device Name: Voluson SWIFT: Voluson SWIFT+ Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: May 16, 2023 Received: May 16, 2023
Dear Bryan Behn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Yanna S. Kang -S
Yanna Kang, Ph.D. Assistant Director Mammography and Ultrasound Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a blue circle. The circle is surrounded by swirling, wave-like shapes, also in blue, which give the impression of motion or energy. The logo is simple, recognizable, and associated with a well-known global corporation.
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known)
K230346
Device Name Voluson SWIFT, Voluson SWIFT+
#### Indications for Use (Describe)
The Voluson SWIFT, Voluson SWIFT+ is a general-purpose diagnostic ultrasound system intended for use by qualified and trained healthcare professionals that are legally authorized or licensed by law in the country, state or other local municipality in which he or she practices for ultrasound imaging, measurement, display and analysis of the human body and fluid. The users may or may not be working under supervision or authority of a physician. Voluson SWIFT, Voluson SWIFT+ clinical applications include: Fetal Obstetrics; Abdominal (including renal and GYN/Pelvic); Pediatio; Small Organ (Breast, Testes, Thyroid, etc.); Neonatal Cephalic; Cardiac (Adult and Pediatric); Peripheral Vascular (PV); Musculo-skeletal Conventional and Superficial; Transrectal(including UrologyProstate) (TR); Transvaginal (TV).
Mode of operation include: B, M, AMM (Anatonical M-Mode), PW Doppler, CW Doppler, Color Doppler, Color M Doppler, Power Doppler, HD-Flow (High Definition-Flow), Harmonic Imaging, Coded Pulse, 3D/4D Inaging mode, Elastography, B-Flow and Combined modes: BM, B/ Color, B/PWD, B/Power/PWD. The Voluson SWIFT / Voluson SWIFT+ are intended to be used in a hospital or medical clinic.
| Type of Use (Select one or both, as applicable) |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
PSC Publishing Sarvices (301) 443-6740 5
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized script, enclosed within a blue circle. The blue color is consistent throughout the logo, providing a clean and recognizable brand identity.
K230346
GE Healthcare 510(k) Premarket Notification Submission
# 510(k) Summary
| Date: | February, 8th, 2023 |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE Medical Systems Ultrasound and Primary care Diagnostics, LLC<br>9900 Innovation Dr<br>Wauwatosa, WI 53226 |
| Manufacturer: | GE Ultrasound Korea, Ltd.<br>9, Sunhwan-ro 214beon-gil, Jungwon-gu,<br>Seongnam-si, Gyeonggi-do, Republic of Korea |
| Primary Contact Person: | Bryan Behn<br>Regulatory Affairs Director<br>GE Healthcare<br>T:+1 262-247-5502 |
| Alternate Contact Person: | Karim Amrouche<br>Senior Regulatory Affairs Manager<br>GE Healthcare<br>T: +33620783855 |
| Device Trade Name: | Voluson SWIFT, Voluson SWIFT+ |
| Common/Usual Name: | Diagnostic Ultrasound System |
| Classification Names: | Class II |
| Product Code: | IYN (primary), IYO, ITX (secondary)<br>Ultrasonic Pulsed Doppler Imaging System. 21CFR 892.1550, 90-IYN;<br>Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO;<br>Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX |
| Primary Predicate Device: | K201828 Voluson SWIFT, Voluson SWIFT+ Diagnostic Ultrasound System |
| Reference Device(s): | K210438 Versana Premier Diagnostic Ultrasound System<br>K220358 Voluson E Series Diagnostic Ultrasound System<br>K213642 Voluson S Series Diagnostic Ultrasound System |
| Device Description: | The subject device is a Track 3 device, primarily intended for general-<br>purpose radiology evaluation and specialized for OB/GYN with particular<br>features for real-time 3D/4D acquisition. The Voluson SWIFT, Voluson<br>SWIFT+ provide digital acquisition, processing and display capability.<br>Voluson SWIFT, Voluson SWIFT+ consist of a mobile console with<br>control panel, full touch monitor, optional image storage and printing<br>devices. It provides high-performance ultrasound imaging and analysis<br>and have comprehensive networking and DICOM capability. It utilizes a<br>variety of linear, curved linear, matrix phased array transducer including |
| Technology: | The Voluson™ SWIFT, Voluson™ SWIFT+ employs the same<br>fundamental scientific technology as its predicate device(s). |
| Determination of | Comparison to Predicates |
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized script, enclosed within a blue circle. There are decorative swirls around the circle, giving it a dynamic and recognizable appearance. The logo is simple, yet distinctive, and is widely recognized as the symbol of the General Electric company.
mechanical and electronic scanning transducers which provide accurate real-time three-dimensional imaging supporting all standard acquisition modes.
The Voluson SWIFT, Voluson SWIFT+ is a general-purpose diagnostic Intended Use: ultrasound system intended for use by qualified and trained healthcare professionals that are legally authorized or licensed by law in the country, state or other local municipality in which he or she practices for ultrasound imaging, measurement, display and analysis of the human body and fluid. The users may or may not be working under supervision or authority of a physician. Voluson SWIFT, Voluson SWIFT+ clinical applications include: Fetal/Obstetrics; Abdominal (including renal and GYN/Pelvic); Pediatric; Small Organ (Breast, Testes, Thyroid, etc.); Neonatal Cephalic; Adult Cephalic; Cardiac (Adult and Pediatric); Peripheral Vascular (PV); Musculo-skeletal Conventional and Superficial; Transrectal(including Urology/Prostate) (TR); Transvaginal (TV). Mode of operation include: B, M, AMM (Anatomical M-Mode), PW Doppler, CW Doppler, Color Doppler, Color M Doppler, Power Doppler, HD-Flow (High Definition-Flow), Harmonic Imaging, Coded Pulse, 3D/4D Imaging mode, Elastography, B-Flow and Combined modes: B/M, B/Color, B/PWD, B/Power/PWD. The Voluson SWIFT / Voluson™ SWIFT+ are intended to be used in a hospital or medical clinic.
Substantial Equivalence: The proposed Voluson SWIFT, Voluson SWIFT+ is a new platform substantially equivalent to the predicate devices. The following is an overview of the differences between the Voluson SWIFT, Voluson SWIFT+ and the predicate Voluson SWIFT, Voluson SWIFT+ (K201828).
- The systems are all intended for diagnostic ultrasound imaging and fluid flow analysis.
- The Voluson SWIFT, Voluson SWIFT+ and predicate Voluson SWIFT, Voluson SWIFT+ have the same clinical intended use.
- . The Voluson SWIFT, Voluson SWIFT+ and predicate Voluson SWIFT, Voluson SWIFT+ have the similar imaging modes with the exception that the proposed Voluson SWIFT, Voluson
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized script, enclosed within a circular border. The color of the logo is a light blue.
SWIFT+ does not include the attenuation imaging mode.
- The systems are manufactured with materials which have been evaluated and found to be safe for the intended use of the device.
- The systems have acoustic power levels which are below the applicable FDA limits.
- . The proposed Voluson SWIFT, Voluson SWIFT+ and predicate Voluson SWIFT, Voluson SWIFT+ systems have similar capability in terms of performing measurements, capturing digital images, reviewing, and reporting studies.
- The proposed Voluson SWIFT, Voluson SWIFT+ and predicate ● Voluson SWIFT. Voluson SWIFT+ systems have been designed in compliance with approved electrical and physical safety standards.
- The probes supported in the proposed Voluson SWIFT, Voluson SWIFT+ and predicate Voluson SWIFT, Voluson SWIFT+ are identical except:
- The addition of a new transducer 9L-RS (already cleared on predicate Versana Premier K210438).
- The addition of a new transducer 12S-RS (already cleared on predicate Versana Premier K210438).
- The absence of the E8C-RS transducer on the proposed Voluson SWIFT, Voluson SWIFT+
- The following software features have been added compared to the ● predicate Voluson SWIFT, Voluson SWIFT+ (K201828) : Shadow Reduction Button, Autolive, RadiantFlow, Volume with color, SpineTrace, Fractional Limb Volume, Timer for BPP, Fetal Heart Support, SonoPelvicFloor, Auto-Caliper, Reminders, SW Download, AVURI
- The following software features have been migrated from Voluson ● E series (K220358): Shadow Reduction, Fetal Heart Support, SonoPelvicFloor, SW download, AVURI
- The following software features have been migrated from Voluson ● S Series (K213642) : Radiantflow, Volume with color, Fractional Limb Volume.
- The proposed Voluson SWIFT, Voluson SWIFT+ adds the ● following new features: Autolive, Spine Trace, Time for BPP (Biophysical Profile), Auto-Caliper, reminders.
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized script, enclosed within a blue circle. The circle is surrounded by wave-like shapes, also in blue, giving the logo a dynamic and fluid appearance.
Summary of Non-Clinical Tests:
The device has been evaluated for acoustic output, biocompatibility, cleaning, and disinfection effectiveness as well as thermal, electrical, electromagnetic and mechanical safety, and has been found to conform with applicable medical device safety standards. The Voluson SWIFT, Voluson SWIFT+ complies with voluntary standards:
- AAMI/ANSI ES60601-1, Medical Electrical Equipment Part 1: ● General Requirements for Safety, 2005/(R)2012 And A1:2012
- IEC 60601-1-2 Medical Electrical Equipment Part 1-2: General ● Requirements for Basic Safety and Essential Performance -Collateral Standard: Electromagnetic Disturbance - Requirements and Tests. Edition 4.0. 2014
- IEC 60601-2-37, Medical Electrical Equipment - Part 2-37: Particular Requirements for the Safety of Ultrasonic Medical Diagnostic and Monitoring Equipment, Edition 2.1, 2015
- ISO 10993-1, Biological Evaluation of Medical Devices-Part 1: Evaluation and Testing Within A Risk Management Process, Fourth edition, 2009
- ISO 14971, Application of risk management to medical devices, ● 2007
- NEMA PS 3.1 3.20, Digital Imaging and Communications in ● Medicine (DICOM) Set. (Radiology), 2016
- IEC 62359, Ultrasonics - Field characterization - Test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields, Edition 2.1, 2017
The following quality assurance measures are applied to the development of the system:
- Risk Analysis
- Requirements Reviews ●
- Design Reviews ●
- Testing on unit level (Module verification) ●
- Integration testing (System verification) ●
- Performance testing (Verification) ●
- Safety testing (Verification) ●
Transducer materials and other patient contact materials are biocompatible.
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized script, enclosed within a blue circle. The circle is surrounded by wave-like shapes, also in blue, giving the logo a dynamic and fluid appearance.
## Summary of Clinical Tests:
The subject of this premarket submission, Voluson SWIFT, Voluson SWIFT+, did not require clinical studies to support substantial equivalence.
## AI Verification Summary
#### Auto-Caliper:
Summary test statistics or other test results including acceptance criteria or other information supporting the appropriateness of the characterized performance.
- . Data used for both training and validation has been collected across multiple geographical sites using different systems and probe configuration to represent the variations in target population.
- The verification for the Auto-Caliper AI feature for 2D caliper ● placement is performed by clinical experts following a specific protocol :
- Visualization of the follicles on slices parallel to the A plane of ● acquisition from the datasets
- . Click on the follicle on a slice in which the given follicle is expected to be the largest.
- The steps are repeated for all follicles in each volume. ●
- The output of the caliper placement on the follicles on the plane in which the clinician clicked.
- The outputs are evaluated by the clinical expert and the assessment is documented as Pass/No result/Fail
- The success rate of the AI feature is 70% or higher
Data Collection and Distribution:
- Total Datasets: 67
- Total follicles count across all volumes : 410
- Country: Germany, India, Spain, United Kingdom, USA
Quantitative evaluation:
- Values were obtained from the test pool consisting of 67 volumes, 2 follicles evaluated per volume for a total of 134 follicles.
- Mixture of small and large diameter follicles were selected for the ● evaluation
{9}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized script, enclosed within a blue circle. Three curved shapes surround the circle, adding a sense of motion or energy to the design. The logo is simple, recognizable, and represents the brand identity of General Electric.
- . The deviation of the measurements predicted by the Auto-Caliper tool from the manual measurements were calculated to obtain the accuracy of the feature.
Demographic distribution, gender, age and ethnicity :
- . Data acquired from women examined in regular clinical practice In compliance with privacy health information.
- Data acquired from 6 countries and five systems to ensure sufficient diversity.
Information about clinical subgroups and confounders :
- . Data collected from multiple system configurations and different probes from multiples geographical sites.
- . The acquired datasets have a variation in shapes and size of the follicles having diameters ranging from 2mm to 25 mm.
- Data acquired from 6 countries and five systems to ensure ● sufficient diversity.
Equipments and protocols :
- Data is provided by external clinical partners who de-identified the data.
- . Original data is collected in the form of 3D volumes to preserve the flexibility to re-process data retrospectively during scan conversion.
- A total of 223 volumes were used for training having the ● following distributions
- Distribution by Systems: Voluson E8 (20), Voluson E10 -(105), Voluson P8 (29), Voluson S10 (48), Voluson S8 (21)
- Distribution by Probes: RIC5-9A-RS (98), RIC5-9-D -(123), RIC6-12-D (2)
- Distribution by Countries: India (62), Germany (29), -United Kingdom (29), Spain (103).
Information about the reference standard and dataset ("truthing process"):
- Datasets contained in the collections for training and verification are mutually exclusive and stem from a mutually exclusive list of patients.
- Data has been collected from patients from regular clinical ● practice from multiple sites thereby capturing the actual representative population. There is no explicit bias/preference enforced during the collection process.
{10}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized script, enclosed within a blue circle. The circle is surrounded by swirling, decorative elements, also in blue, giving the logo a classic and recognizable appearance.
- To ensure correct and reliable "truthing" process of the training, a curation protocol has been developed by clinical experts to be followed by curators, in addition, during and after the data curation process, a senior sonographer reviewed a random subset of the curated dataset for clinical accuracy.
| Absolute difference between Al predicted<br>diameter and Ground truth diameter (in mm) | Percentage of data Pool |
|----------------------------------------------------------------------------------------|-------------------------|
| <0.1 | 5.3 |
| 0.1-0.5 | 32.8 |
| 0.5-1.0 | 22.1 |
| 1.0-2.0 | 25.2 |
| 2.0-3.0 | 5.3 |
| 3.0-5.0 | 3.1 |
| >5.0 | 6.1 |
- Conclusion: GE Healthcare considers the Voluson SWIFT, Voluson SWIFT+ to be as safe, as effective, and performance is substantially equivalent to the predicate device(s).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.