The device is a general purpose ultrasound system intended for use by qualified and trained healthcare professionals. Specific clinical applications remain the same as previously cleared: Fetal/OB; Abdominal (including GYN, pelvic and infertility monitoring/follicle development); Pediatric; Small Organ (breast, testes, thyroid etc.); Neonatal and Adult Cephalic; Cardiac (adult and pediatric); Musculo-skeletal Conventional and Superficial; Vascular; Transvaginal (including GYN); Transrectal Modes of operation include: B, M, PW Doppler, CW Doppler, Color M Doppler, Color M Doppler, Power Doppier, Harmonic Imaging, Coded Pulse, 3D/4D Imaging mode, Elastography, Shear Wave Elastography and Combined modes: B/M, B/Color, B/PWD, B/Color/PWD, B/Power/ PWD, B/Elastography. The Voluson™ Expert 20, Voluson™ Expert 22 is intended to be used in a hospital or medical clinic.
Device Story
Mobile ultrasound console with keyboard, touch panel, and display; utilizes linear, curved, and matrix phased array transducers for real-time 3D/4D imaging. Operates in hospital/clinic settings by trained professionals. Inputs: acoustic signals from transducers; Outputs: diagnostic images, measurements, and reports. Features AI-based workflow tools (SonoPelvicFloor, SonoLyst, SonoFHR) to assist in image sorting, grading, and anatomical measurement. Healthcare providers use output for clinical decision-making in OB/GYN and general radiology. Benefits include improved workflow efficiency and diagnostic accuracy through automated detection and measurement.
Clinical Evidence
No clinical studies required. Bench testing only. Performance of AI features (SonoPelvicFloor, SonoLyst, FetalHS) validated using retrospective and prospective datasets (e.g., 40,000+ images for SonoLyst, 110 volumes for SonoPelvicFloor). Accuracy metrics reported: SonoPelvicFloor success rate ≥70% (good IQ) or ≥60% (challenging IQ); SonoLyst sorting/grading accuracy >80%; FetalHS view suggestion success rate ≥70%, heart angle measurement success rate ≥80%.
Technological Characteristics
Mobile ultrasound console; linear, curved, and matrix phased array transducers. Modes: B, M, PW/CW/Color/Power Doppler, 3D/4D, Elastography. Connectivity: DICOM. Standards: AAMI/ANSI ES60601-1, IEC60601-1-2, IEC60601-2-37, ISO10993-1, ISO14971. Software includes AI-based automated analysis features.
Indications for Use
Indicated for qualified healthcare professionals performing diagnostic ultrasound imaging and fluid flow analysis in Fetal/OB, Abdominal, GYN, pelvic, infertility monitoring, Pediatric, Small Organ (breast, testes, thyroid), Neonatal, Adult Cephalic, Cardiac, Musculo-skeletal, Vascular, Transvaginal, and Transrectal applications.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Voluson E10 Diagnostic Ultrasound System (K201768)
{1}------------------------------------------------
June 6, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
GE Medical Systems Ultrasound and Primary Care Diagnostics, LLC % Bryan Behn Regulatory Affairs Director 9900 Innovation Dr. WAUWATOSA WI 53226
# Re: K220358
Trade/Device Name: Voluson Expert 22, Voluson Expert 20, Voluson Expert 18 Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: April 28, 2022 Received: April 29, 2022
Dear Bryan Behn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{2}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Michael D. O'Hara, Ph.D. Deputy Director DHT 8C: Division of Radiological Imaging and Radiation Therapy Devices OHT 8: Office of Radiological Health Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known)
#### K220358
#### Device Name
Voluson Expert 22, Voluson Expert 20, Voluson Expert 18
#### Indications for Use (Describe)
The device is a general purpose ultrasound system intended for use by qualified and trained healthcare professionals. Specific clinical applications remain the same as previously cleared: Fetal/OB; Abdominal (including GYN, pelvic and infertility monitoring/follicle development); Pediatric; Small Organ (breast, testes, thyroid etc.); Neonatal and Adult Cephalic; Cardiac (adult and pediatric); Musculo-skeletal Conventional and Superficial; Vascular; Transvaginal (including GYN); Transrectal
Modes of operation include: B, M, PW Doppler, CW Doppler, Color M Doppler, Color M Doppler, Power Doppier, Harmonic Imaging, Coded Pulse, 3D/4D Imaging mode, Elastography, Shear Wave Elastography and Combined modes: B/M, B/Color, B/PWD, B/Color/PWD, B/Power/ PWD, B/Elastography. The Voluson™ Expert 20, Voluson™ Expert 22 is intended to be used in a hospital or medical clinic.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary (K220358)
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | April 28, 2022 |
|-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE Healthcare [GE Healthcare Austria GmbH & Co OG]<br>Tiefenbach 15<br>Zipf, Austria 4871 |
| Primary Contact<br>Person: | Bryan Behn<br>Regulatory Affairs Director<br>GE Healthcare<br>T:(262)247-5502<br>F:(414)918-8275 |
| Secondary Contact<br>Person: | Roland Kuntscher<br>Regulatory Affairs Specialist<br>GE Healthcare Austria GmbH & Co OG<br>T:(+43)7682-3800-660<br>F:(+43)7682 3800-47 |
| Device:<br>Trade | Voluson Expert Series |
| Name: | Models: Voluson Expert 22, Voluson Expert 20, Voluson Expert 18 |
| Common/Usual<br>Name: | Ultrasound system |
| Classification Names:<br>Product Code: | Class II<br>Ultrasonic Pulsed Doppler Imaging System. 21CFR 892.1550, 90-IYN<br>Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO<br>Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX |
| Primary Predicate<br>Device(s): | K201768 Voluson E10 Diagnostic Ultrasound System |
| Classification Names:<br>Product Code: | Class II<br>Ultrasonic Pulsed Doppler Imaging System. 21CFR 892.1550, 90-IYN<br>Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO<br>Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX |
| Reference Predicate<br>Device(s): | K192159 Voluson E10/E8/E6<br>K173555 LOGIQ E10 |
| Classification Names:<br>Product Codes: | Class II<br>Ultrasonic Pulsed Doppler Imaging System. 21CFR 892.1550, 90-IYN<br>Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO<br>Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX |
{5}------------------------------------------------
- The systems are full-featured Track 3 ultrasound systems, primarily for Device Description: general radiology use and specialized for OB/GYN with particular features for real-time 3D/4D acquisition. They consist of a mobile console with keyboard control panel; color LCD/TFT touch panel, color video display and optional image storage and printing devices. They provide high performance ultrasound imaging and analysis and have comprehensive networking and DICOM capability. They utilize a variety of linear, curved linear, matrix phased array transducers including mechanical and electronic scanning transducers, which provide highly accurate real-time three-dimensional imaging supporting all standard acquisition modes. The following probes are the same as the predicate: RIC5-9-D, IC5-9-D. RIC6-12-D, 9L-D, 11L-D, ML6-15-D, RAB6-D, C1-6-D, C2-9-D, M5Sc-D, RM7C, eM6CG3. The following have migrated from LOGIQ E10 6S-D and L18-18i-D (K173555). The RSP6-16-D was
previously cleared on the Voluson E10 (K192159) and has been added back. The RIC10-D is a new probe and is substantially equivalent to the RIC5-9-D, it is an incremental improvement in technology.
- The device is a general purpose ultrasound system intended for use by Intended Use: qualified and trained healthcare professionals. Specific clinical applications remain the same as previously cleared:
Fetal/OB; Abdominal (including GYN, pelvic and infertility monitoring/follicle development); Pediatric; Small Organ (breast, testes, thyroid etc.); Neonatal and Adult Cephalic; Cardiac (adult and pediatric); Musculo-skeletal Conventional and Superficial; Vascular; Transvaginal (including GYN): Transrectal
Modes of operation include: B, M, PW Doppler, CW Doppler, Color Doppler, Color M Doppler, Power Doppler, Harmonic Imaging, Coded Pulse, 3D/4D Imaging mode, Elastography, Shear Wave Elastography and Combined modes: B/M, B/Color, B/PWD, B/Color/PWD, B/Power/ PWD, B/Elastography. The Voluson™ Expert 18, Voluson™ Expert 20, Voluson™ Expert 22 is intended to be used in a hospital or medical clinic.
- The Voluson Expert Series (Voluson E22/20/18) employs the same Technology: fundamental scientific technology as its predicate devices.
Determination of Comparison to Predicates Substantial Equivalence:
The proposed Voluson Expert 22/20/18 is substantially equivalent to the predicate device with regards to intended use, imaging capabilities, technological characteristics and safety and effectiveness.
{6}------------------------------------------------
## New Model Names and Model differences:
New model names Voluson Expert 18, Voluson Expert 20 and Voluson Expert 22 are similar in hardware . Voluson Expert 18 is lower version and not all probes or functions are available. Voluson Expert 20 is mid version and product with complete configuration with all the probes and functions of software with exception of 4D electronically probe eM6CG3. The high-end model Voluson Expert 22 allows to drive additionally 4D electronically probe eM6CG3.
- . The systems are all intended for diagnostic ultrasound imaging and fluid flow analysis.
- . The proposed Voluson Expert 22/20/18 and predicate Voluson E10 systems have the same clinical intended use.
- The proposed Voluson Expert 22/20/18 and predicate Voluson E10 systems have the same imaging modes.
- The proposed Voluson Expert 22/20/18 and predicate Voluson . • E10 system transducers are equivalent. Four new transducers RSP6-16-D, L8-18i-D, 6S-D and RIC10-D were added and three transducers removed to the proposed system.
- There is no change to the system indications for use.
- The systems are manufactured with materials which have . been evaluated and found to be safe for the intended use of the device.
- . The systems have acoustic power levels which are below the applicable FDA limits.
- . The proposed Voluson Expert Series 22/20/18 and predicate Voluson E10 system have similar capability in terms of performing measurements, capturing digital images, reviewing and reporting studies.
- The proposed Voluson Expert Series 22/20/18 and predicate . systems have been designed in compliance with approved electrical and physical safety standards.
- . There proposed Voluson Expert Series 22/20/18 and predicate Voluson E10 system Software Features are equivalent. Some minor improvements to the existing Software features have been implemented into the proposed system.
- The proposed Voluson Expert Series 22/20/18 adds additional . AI software features SonoPelvicFloor and SonoLyst/ Sonolyst Live (workflow improvement) to the system.
{7}------------------------------------------------
- The proposed Voluson Expert Series 22/20/18 adds additional ● software features Shadow Reduction . Adapt and Sono FHR (Performance improvement) to the system.
- The proposed Voluson Expert Series 22/20/18 migrated from (K192159) the following software features Elastography, Shear Wave Elastography , CW Doppler Mode and Acquisition mode Contrast.
## Summary of Non-Clinical Tests:
The device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic, and mechanical safety, and has been found to conform to applicable medical device safety standards. The Voluson Expert Series 22/20/18 and its applications comply with voluntary standards:
- AAMI/ANSI ES60601-1, Medical Electrical Equipment Part . 1: General Requirements for Safety, 2005/(R)2012 And A1:2012
- IEC60601-1-2 Medical Electrical Equipment Part 1-2: . General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility Requirements and Tests, 2014
- IEC60601-2-37, Medical Electrical Equipment - Part 2-37: Particular Requirements for the Safety of Ultrasonic Medical Diagnostic and Monitoring Equipment, 2015
- ISO10993-1, Biological Evaluation of Medical Devices- Part 1: ● Evaluation and Testing- Third Edition, 2009
- ISO14971, Application of risk management to medical devices: ● Third edition 2019
- NEMA PS 3.1 3.20 (2022a), Digital Imaging and ● Communications in Medicine (DICOM) Set. (Radiology)
The following quality assurance measures are applied to the development of the system:
- . Risk Analysis
- . Requirements Reviews
- Design Reviews
- . Testing on unit level (Module verification)
- . Integration testing (System verification)
- Performance testing (Verification) .
{8}------------------------------------------------
- . Safety testing (Verification)
- Final Acceptance Testing (Validation) .
Transducer materials and other patient contact materials are biocompatible.
## AI Summary of Testing
#### SonoPelvic Floor:
Summary test statistics or other test results including acceptance criteria or other information supporting the appropriateness of the characterized performance.
- . Tested on datasets marked as Good in Image Quality assessment, the success rate of each AI component of the feature (MHD plane alignment, LH contour and measurements) is expected to be 70% or higher. On datasets that are marked as challenging in image quality measure the success rate of each AI component of the feature should be 60% or higher.
- The number of individual patients images were collected from: ● 70+
- The number of samples, if different from above, and the ● relationship between the two: 110 3D/4D Volumes
Demographic distribution:
- . Gender: Female
- Age: Reproductive age, specific age not collected
- Ethnicity/Country; Europe, Asia and South Africa ●
Information about clinical subgroups and confounders present in the dataset:
- . During testing, a differentiation is made between good IO (Image Quality) and challenging IQ datasets.
Information about equipment and protocols used to collect images:
- . Mix of data from across six different probe models and five different Console variants. The data collection protocol was standardized across all data collection sites.
Information about how the reference standard was derived from the dataset (i.e. the "truthing" process)
- . For the testing process, the results are generated by the AI software and the same are verified as Pass or Fail by a certified sonographer/Clinician. The results are then aggregated to yield an accuracy metric for the AI algorithm.
Description of how independence of test data from training data was ensured.
{9}------------------------------------------------
- The volumes used for test/validation purpose is completely ● distinct from the ones used during training process and there is no overlap between the two.
## SonoLyst:
Summary test statistics or other test results including acceptance criteria or other information supporting the appropriateness of the characterized performance
- . For SonoLystIR the sorting accuracy is higher than 80% on a test data set containing 40000+ images. For SonoLystX the grading accuracy is higher than 80% on a test data set containing 9500+ images. For SonoLystLive the accuracy is higher than 80% on a test data set containing 5500+ images.
- The number of individual patients images were collected from: . 5000+ exams
- The number of samples, if different from above, and the ● relationship between the two: SonoLyst was tested on 40000+ images derived from the collected exams. The exams contain multiple standard views of the fetal anatomy and cine loops.
Demographic distribution
- Gender: Female
- Age: Gestational Age of fetuses: 18-24 weeks
- Ethnicity/Country: Exams from United Kingdom, Austria, . India, USA
Information about clinical subgroups and confounders present in the dataset:
- . During testing, a differentiation was made between retrospective and prospective data collection from GE and non-GE scanners, evaluating the generalization performance.
Information about equipment and protocols used to collect images.
- Mix of data from across five different console variants, 4 ● Voluson GE, 1 non-GE. Mix of data from retrospective data collection in clinical practice and prospective data collection.
Information about how the reference standard was derived from the dataset (i.e. the "truthing" process):
- 1. The images were curated (sorted and graded) by a single Sonographer
- 2. The images were sorted and graded by SonoLyst. This process resulted in some images being reclassified during sorting.
- Where they differed from the ground truth, the sorted 3.
{10}------------------------------------------------
images from step 2 were reviewed by a 5-sonographer review panel, in order to determine the sorting accuracy of the system. The sorting process resulted in some images being reclassified based upon the majority view of the panel.
- 4. Where they differed from the ground truth, the graded images from step 1 were reviewed by a 5-sonographer review panel, in order to determine the grading accuracy of the system.
Description of how independence of test data from training data was ensured:
- . The exams used for test/training validation purpose are separated from the ones used during training process and there is no overlap between the two.
# FetalHS:
Summary test statistics or other test results including acceptance criteria or other information supporting the appropriateness of the characterized performance.
- The success rate of the 4CH view and 3VT view suggestion within cines is expected to be 70% or higher. The success rate of the suggested heart angle measurement is expected to be 80% or higher.
- The number of individual patients images were collected from: ● 250+ patients
- The number of samples, if different from above, and the ● relationship between the two: 400+ cines
Demographic distribution
- . Gender: Female
- Age: Gestational age 18-25 weeks ●
- Ethnicity/Country : USA, India, Japan, Germany, Austria ●
Information about clinical subgroups and confounders present in the dataset.
- . Cines were grouped into the sub-groups 4CH view present / not present, 3VT view present / not present
Information about equipment and protocols used to collect images.
- . Mix of data from across six different probe models and three different console variants. The data collection protocol was standardized across all data collection sites.
Information about how the reference standard was derived from the dataset (i.e. the "truthing" process).
- For the testing process, the results are generated by the AI
{11}------------------------------------------------
software and the same are verified as Pass or Fail by certified sonographers/clinicians. The results are then aggregated to yield an accuracy metric for the AI algorithm.
Description of how independence of test data from training data was ensured.
- . The data sets used for test/validation purpose are completely distinct from the ones used during training process and there is no overlap between the two.
# Summary of Clinical Tests:
The subject of this premarket submission, Voluson Expert Series 22/20/18 did not require clinical studies to support substantial equivalence.
- Conclusion: GE Healthcare considers the Voluson Expert Series 22/20/18 to be as safe, as effective, and performance is substantially equivalent to the predicate device(s).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.