WRIGHT 1000 MONOPLACE HYPERBARIC SYSTEM

K952545 · Wright Industries · CBF · May 28, 1997 · Anesthesiology

Device Facts

Record IDK952545
Device NameWRIGHT 1000 MONOPLACE HYPERBARIC SYSTEM
ApplicantWright Industries
Product CodeCBF · Anesthesiology
Decision DateMay 28, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5470
Device ClassClass 2
AttributesTherapeutic

Device Story

Wright 1000 Mono Hyperbaric Chamber is a pressure vessel for human occupancy (PVHO) used for hyperbaric oxygen therapy (HBO). Device consists of a steel chamber with acrylic windows, a reclining chair, and integrated monitoring systems. Patient is placed inside and subjected to pressures between 2 and 3 atmospheres while breathing 100% oxygen. System includes 2-way communication, variable ventilation, and pressure/oxygen monitoring with programmable alarms. Operated by clinical staff to treat specific conditions by increasing alveolar oxygen pressure and tissue oxygen delivery. Device is manufactured to ASME and NFPA guidelines.

Clinical Evidence

No clinical data provided; device equivalence is based on design, manufacturing standards (ASME, PVHO, NFPA), and hydrostatic/safety testing.

Technological Characteristics

Constructed of SA516-70N steel; 14" PVO acrylic windows. Max operational pressure 30 PSIG; hydrostatic tested at 45 PSIG. Features include 2-way communication, variable ventilation, and programmable pressure/oxygen alarms. Manufactured per ASME and PVHO guidelines. Wiring complies with National Electrical Code. Electromagnetic and safety testing conducted.

Indications for Use

Indicated for air or gas embolism, carbon monoxide poisoning/smoke inhalation, clostridial myonecrosis, crush injury/compartment syndrome/acute traumatic ischemias, decompression sickness, problem wound healing, exceptional blood loss (anemia), necrotizing soft tissue infections, refractory osteomyelitis, radiation tissue damage (osteoradionecrosis), compromised skin grafts/flaps, thermal burns, and adjunctive hyperbaric oxygen in intracranial abscess.

Regulatory Classification

Identification

A hyperbaric chamber is a device that is intended to increase the environmental oxygen pressure to promote the movement of oxygen from the environment to a patient's tissue by means of pressurization that is greater than atmospheric pressure. This device does not include topical oxygen chambers for extremities (§ 878.5650).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} MAY 28 1997 K95-2545/53 Wright INDUSTRIES C. Wright Industries 13760 S. Gramercy PL. Gardena, CA. 90249 Phone: 310-523-5132 Fax: 310-523-1825 Contact: Charles Wright Date: 4-15-1997 510(k) Summery Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFC-401) 9200 Corporate Boulevard Rockville, Maryland 20850 Re: 510 (k) Notification Document control # K95245 K952545 Proprietary Name: Wright 1000 Mono Hyperbaric Chamber. Common Name: Hyperbaric Chamber. Calssification Name: (PVHO) Pressure Vessel for Human Occupancy. 73 CBF, CFR 868.5470 Predicate Device: HYOX System Limited 510(k) No. K874107 Attention: Christy Foreman Sincerely Yours Charles Wright Owner/Operator C. Wright Industries 13760 S. Gramercy Pl. Gardena, CA 90249 Phone (310) 523-5132 {1} # HYPERBARIC TREATMENT: Hyperbaric oxygen therapy (HBO) is a medical treatment where a patient is placed in a specially designed vessel and placed under pressure greater than sea level and breathes 100% oxygen intermittently. Alveolar oxygen pressure is increased causing a rise in plasma oxygen content which results in enhanced tissue oxygen delivery. The amount of pressure increases and the length of time under pressure are determined by the condition being treated. Treatment pressures are usually between 2 and 3 times atmospheric. treatment usually lasts from 1 to 2 hours at full pressure. The following are approved indications: 1: Air or gas embolism. 2: Carbon monoxide poisoning and smoke inhalation. Carbon monoxide complicated by cyanide poisoning. 3: Clostridial myonecrosis (Gas gangrene). 4: Crush injury, Compartment syndrome, and other acute traumatic ischemias. 5: Decompression sickness. 6: Enhancement of healing in selected problem wounds. 7: Exceptional blood loss (Anemia) 8: Necrotizing soft tissue infections (Subcutaneous tissue, Muscle, Fascia). 9: Osteomyelitis (Refractory). 10: Radiation tissue damage (Osteoradionecrosis). 11. Skin grafts and flaps (Compromised). 12. Thermal burns. 13. Adjunctive hyperbaric oxygen in intracranial abscess. {2} CONTRAINDICATIONS: Only previously listed indications are approved all others are considered contraindications. REFERENCES: 1: HYPERBARIC MEDICINE PROCEDURES. 1996 2: HYPERBARIC OXYGEN THERAPY: A COMMITTEE REPORT. 1996 3: MONOPLACE HYPERBARIC CHAMBER SAFTY GUIDELINES. 1991 {3} The Wright 1000 Mono Hyperbaric Camber: - *Max operational pressure 30 PSIG. - *Constructed out of SA516-70N steel - *The 4 Large 14" Windows Constructed out of PVO Acrylic - *Hydrostatic tested at 45 PSIG. - *Welds are X-rayed - *Welds magna Flux tested after hydrostatic test. - *Manufactured by a ASME approved house and certified by ASME. Certification No. 28,352 - *Manufactured under PVHO guidelines - *Manufactured under NFPA Guidelines - *Registered with National Board Of Pressure Vessel Inspectors. - *Wiring to National Electrical code. - *Electrical Testing: Electromagnetic and Safety testing Conducted. - *Spacious interior - *2 Way Communications - *Pressure and Oxygen monitoring With alarms. - *Programmable Treatment Pressure Alarms - *Variable ventilation - *Comfortable reclining chair {4} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAY 28 1997 Mr. Charles Wright Wright Industries 13760 South Gramercy Place Gardena, California 90249 Re: K952545 Wright 1000 Mono Hyperbaric Chamber Regulatory Class: II (two) Product Code: 73 CBF Dated: April 15, 1997 Received: April 21, 1997 Dear Mr. Wright: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {5} Page 2 - Mr. Charles Wright This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, ![img-0.jpeg](img-0.jpeg) Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...