K140559 · Sechrist Industries, Inc. · CBF · May 30, 2014 · Anesthesiology
Device Facts
Record ID
K140559
Device Name
SECHRIST MONOPLACE HYPERBARIC OXYGEN CHAMBER
Applicant
Sechrist Industries, Inc.
Product Code
CBF · Anesthesiology
Decision Date
May 30, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 868.5470
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use of the Sechrist Model 3300H/HR and Model 3600H/HR Hyperbaric Chambers are to administer 100% oxygen at pressure greater than ambient, up to 3 atmospheres absolute (30 psi) of pressure.
Device Story
Monoplace hyperbaric oxygen chambers administer 100% oxygen at pressures up to 3 ATA. System comprises pressure vessel, control panel, and pneumatic system. Operator selects pressurization/depressurization rates and pressure set points via rotary valves. Device includes door safety interlocks, over-pressure relief valves, and emergency vent/shutoff controls. Two-way intercom allows communication between patient and attendant. Used in clinical settings by trained personnel. Provides hyperbaric oxygen therapy for 13 specific indications. Benefits include controlled oxygen delivery for wound healing, ischemia, and other conditions. Output is patient exposure to high-pressure oxygen; healthcare providers monitor treatment via pneumatic pressure gauges and cycle counters.
Clinical Evidence
Bench testing only. Validation included system design validation, ASME Boiler & Pressure Vessel Code Section VIII Division 1 compliance, and PVHO-1 safety standard testing (1.1 x MAWP). Performance verified through calibration procedures and functional testing of relief valves, door interlocks, and gas flow rates.
Technological Characteristics
Materials: Acrylic viewport, aluminum cover, stainless steel tie rods/hinge, alloy steel pressure-bearing components. Energy: Pneumatic (50-70 psig supply). Dimensions: 32.5-35.5 inch internal diameter. Connectivity: Standalone pneumatic system. Safety: ASME PVHO-1 compliant, door interlocks, dual over-pressure relief valves. Controls: Manual rotary pneumatic valves for pressure, rate, and purge flow.
Indications for Use
Indicated for patients requiring hyperbaric oxygen therapy for: air or gas embolism, carbon monoxide poisoning (with or without cyanide poisoning), clostridial myonecrosis (gas gangrene), crush injury/compartment syndrome/acute traumatic ischemias, decompression sickness, arterial insufficiencies (central retinal artery occlusion, problem wound healing), severe anemia, intracranial abscess, necrotizing soft tissue infections, refractory osteomyelitis, delayed radiation injury (soft tissue/bony necrosis), compromised grafts/flaps, and thermal burns.
Regulatory Classification
Identification
A hyperbaric chamber is a device that is intended to increase the environmental oxygen pressure to promote the movement of oxygen from the environment to a patient's tissue by means of pressurization that is greater than atmospheric pressure. This device does not include topical oxygen chambers for extremities (§ 878.5650).
Predicate Devices
Sechrist Model 4100H/HR Hyperbaric Chamber (K100268)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for Sechrist Industries, Inc. The word "SECHRIST" is written in large, bold, black letters above a globe. Below the globe, the words "INDUSTRIES, INC." are written in smaller, black letters.
MAY 3 0 2014
# 510(k) Summary
April 25, 2014 Date Prepared
#### 1. Submitter Device Information
| 510(k) Owners Name | Sechrist Industries, Inc.<br>4225 E. La Palma Avenue<br>Anaheim, Ca 92807<br>Phone: 714-579-8400 |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Information | Victor C. Arellano<br>Director Regulatory Affairs, Product Safety & Certification<br>Phone: 714-579-8344<br>Fax: 714-579-8393<br>email: VArellano@SechristUSA.com |
#### 2. Device Information
| Trade/Proprietary Name | Model 3300H/HR Monoplace Hyperbaric Oxygen<br>Chamber<br>Model 3600H/HR Monoplace Hyperbaric Oxygen<br>Chamber |
|------------------------|----------------------------------------------------------------------------------------------------------------|
| Device Common Name | Monoplace Hyperbaric Chamber |
| Classification Name | Hyperbaric Chamber (21 CFR 868.5470, Product Code<br>CBF) |
#### 3. Predicate Device Information
Legally marketed device to which we are claiming equivalence is the Sechrist Model 4100H/HR Hyperbaric Chamber originally cleared under 510(k) K100268
#### 4. General Description
A hyperbaric oxygen chamber is a pressure vessel and control system that is designed to provide patient exposure to a very high oxygen concentration at higher than normal atmospheric pressure. Titration of the oxygen exposure is controlled by selecting the pressure achieved within the pressure vessel. Pressurization and de-pressurization rates are selected to minimize patient discomfort while increasing and decreasing the chamber pressure. Typical monoplace chambers are capable of pressurizing to 3 ATA (29.4 psig above atmospheric pressure). The typical pressurization and de-pressurization rates are in the range of 0.4 to 5.0 psig/minute.
4225 E. La Palma Avenue · Anaheim, California 92807 Telephone (714) 579-8400 •Fax (714) 579-0814 • Website www.sechristusa.com
{1}------------------------------------------------
The Sechrist Models 3300H/HR and 3600H/HR Hyperbaric Chambers are monoplace pressure chambers designed to administer 100% oxygen at pressure greater than ambient, up to 3 atmospheres absolute (30 psig) of pressure. These hyperbaric chambers consist of a mechanical and pneumatic system capable of controlling the operator defined pressurization profiles.
These Hyperbaric Chambers are constructed with a horizontal seamless acrylic viewport which varying internal diameter sizes.
| Models | Internal Diameter |
|------------------------------|----------------------|
| 3300H/HR Hyperbaric Chambers | 32.50 inches (83 cm) |
| 3600H/HR Hyperbaric Chambers | 35.50 inches (90 cm) |
Each model is equipped with a clear anodized aluminum cover on one end and the other end with a hard anodized door assembly, with access ports available for patient interface (such as patient monitoring, delivery of intravenous fluids, etc.) locking mechanism and interlocking safety device.
The cylinder, end cover and the door assembly are assembled together with stainless steel tie rods/nuts and hinge assembly. Each Hyperbaric Chamber Model comes in two configurations designated with a suffix of "H" and "HR". The suffix "H" is the Standard configuration where the control panel is located on the left of the chamber when facing the door and the door opens from left to right. The suffix "HR" is the Reverse configuration where the control panel is located on the right side of the chamber and the door opens from right to left.
The functional performance system and the safety features incorporated in the Models 3300H/HR and 3600H/HR Hyperbaric Chambers are the same that have been incorporated into the design of the Model 4100H/HR predicate device.
A pressurization cycle counter is provided to maintain a record of the number of pressurization cycles the chamber experienced.
A two-way intercommunication system is used to maintain contact between patient and attendant.
{2}------------------------------------------------
#### 5. Statement of Intended Use
The intended use of the Sechrist Model 3300H/HR and Model 3600H/HR Hyperbaric Chambers are to administer 100% oxygen at pressure greater than ambient, up to 3 atmospheres absolute (30 psi) of pressure.
Specific indication for use of the hyperbaric chambers have been established by the Committee on Hyperbaric Oxygen Therapy of the Undersea and Hyperbaric Medical Society (founded in 1967 to foster exchange of data on the physiology and medicine of commercial and military diving). The committee is comprised of practitioners and scientific investigations in the fields of internal medicine, infectious diseases, pharmacology, emergency medicine, general surgery, orthopedic surgery and aerospace medicine. The committee is responsible for continually reviewing research and clinical data in determining the safety and efficacy of hyperbaric oxygen. Currently, there are thirteen indications that are approved by the committee; these thirteen indications were accepted based on sound physiologic rationale, in vitro studies that demonstrate effectiveness, controlled animal studies, prospective controlled clinical studies and extensive clinical experience from multiple hyperbaric medicine centers. These thirteen indications have been recommended for third-party reimbursements and most insurance carriers have established reimbursement policy based on these recommendations.
The thirteen indications are:
- Air or Gas embolism 1.
- 2. Carbon Monoxide Poisoning
Carbon Monoxide Poisoning Complicated by Cyanide Poisoning
- 3. Clostridial Myonecrosis and Myonecrosis (Gas Gangrene)
- Crush Injury, Compartment Syndrome and Other Acute Traumatic Ischemia's 4.
- ನ. Decompression Sickness
#### 6. Arterial Insufficiencies
Central Retinal Artery Occlusion
Enhancement of Healing In Selected Problem Wounds
- 7. Severe Anemia
- Intracranial Abscess 8.
- ರು Necrotizing Soft Tissue Infections
- 10. Osteomyelitis (Refractory)
- 11. Delayed Radiation Injury (Soft Tissue and Bony Necrosis)
- 12. Comprised Grafts and Flaps
- 13. Thermal Burns
{3}------------------------------------------------
#### SUMMARY OF TECHNOLOGICAL CHARACTERISTICS 6.
#### CHAMBER CONTROLS 6.1.
The Models 3300H/HR and 3600H/HR Hyperbaric Chambers both use the same control panel knobs and switches as used in the predicate device 4100H/HR.
Following is description of the chamber controls.
### 6.1.1. Master Valve
This is a rotary pneumatic valve with three positions. The available positions allow the operator to select the following modes.
- a) Vent
- p) Off
- c) On
### Specification
- a) Valve Type: 3-position pneumatic valve rotary selector
- b} Function Positions: ON, OFF, Emergency Vent
- Operating Pressure: 50 70 psig c)
### 6.1.2. OFF Mode
When the Master Valve is in OFF mode, the hyperbaric chamber will not be capable of developing internal pressure, and the door can be opened and closed.
### 6.1.3. ON Mode
The ON mode shall not be capable of operation unless the chamber door is closed and in the locked position. When the Master Valve is in ON mode, the hyperbaric chamber shall automatically start developing internal pressure (compressing) up to the Pressure Set point and at a rate established by the Rate Set control provided the door is closed and the Door Safety Interlock is engaged. The chamber door cannot be unlocked if there is more than approximately 1 psig pressure within the hyperbaric chamber.
### 6.1.4. Emergency Vent Mode
The Emergency Vent Mode allows the operator to depressurize the hyperbaric chamber rapidly to gain rapid access to the patient in an emergency situation. To activate the Emergency Vent, the Emergency Vent Mode is first selected via the Master Valve and the pneumatic push button manually depressed to actuate pneumatic valves to accomplish a decompression at a maximum rate of 30 psig/min.
### Specification
| (a) Control Type: | Pneumatic push-button |
|-------------------|----------------------------------|
| (b) Time: | 30 psig to 0 psig in 119 sec max |
Page 4 of 20
{4}------------------------------------------------
### 6.1.5. Ventilation (Purge) Flow Rate Control
The Ventilation Flow Rate Control allows the operator to adjust the purge flow rate at which gas exhausts from the hyperbaric chamber during treatment. During treatment, once the chamber reaches desired pressure and oxygen concentration, the high ventilation rate may not be necessary and so to conserve oxygen, flow rate can be reduced. This range is clearly marked near the control valve.
### Specification
| (a) Control Type: | Single Turn Gate Valve |
|------------------------------------|----------------------------------------------|
| (b) Ventilation (Purge) Flow Rate: | 80 to 400 lpm min @ 15 psig Chamber Pressure |
| (c) O2 Conservation Range: | 80 to 200 lpm |
| (d) Tolerance: | $\[ \pm \]$ 50 lpm of setting |
### 6.1.6. Pressure Set Control
The desired hyperbaric chamber pressure is adjusted by the rotary Set Pressure Control knob.
### Specification
| (a) Control Type: | Multi-turn, rotary regulating valve |
|----------------------------|--------------------------------------|
| (b) Range: | 1.5 to 30.0 psig or equivalent units |
| (c) Idle/Default Pressure: | 1.5 psig or equivalent units |
| (d) Resolution: | 0.2 psig or equivalent units |
| (e) Tolerance: | ± 0.5 psig of setting |
### 6.1.7. Rate Set Control
The desired pressurization/depressurization rate of the hyperbaric chamber is adjusted by a rotary Rate Set Control knob. The Rate Set is the rate required to reach the desired hyperbaric chamber pressure. The pneumatic system is capable of developing the Set Rate of pressure change within the chamber on a 1:1 ratio.
### Specification
| (a) Control Type: | Single-turn, rotary valve |
|-------------------|--------------------------------------------------------|
| (b) Range: | 1.0 to 5.0 psig/min |
| (c) Profile: | Linear |
| (d) Resolution: | 1.0 psig/min |
| (e) Tolerance: | 20% or 0.5 psig/min of setting at range 1.0 to 3.9 |
| | psig/min & 40% of setting at range 4.0 to 5.0 psig/min |
### 6.1.8. Intercom
The intercom sub-system is an intercommunication system for two-way private conversation capability between the patient inside the hyperbaric chamber and an attendant outside. It is capable of delivering a single channel (mono) or dual channel (stereo) audio input to the speaker inside the hyperbaric chamber for patient entertainment from an auxiliary source such as but not limited to tape
{5}------------------------------------------------
recorder, cd/mp3 player, DVD player and television. If the handset is picked up for conversation, the auxiliary audio signal shall be temporarily disabled.
### Specification
| (a) Supply: | 12 VDC nominal |
|---------------------------|------------------------------------------------------------------------------------------------------------|
| (b) Audio Power: | $1\frac{1}{2}$ watt max inside chamber |
| (c) Speakers: | Dual 4 ohms or 8 ohms, internal and external |
| (d) Internal Microphone: | Electret Condenser type or equivalent |
| (e) Volume Controls: | Rotary potentiometers for Volume In & Out CW to increase; CCW to decrease |
| (f) Aux. Input Connector: | (2) RCA Phone Jacks or equivalent |
| (g) Operator Handset: | Telephone Style or equivalent |
| (h) Indicators: | (2) LEDs<br>1st LED: Green - Main Power ON<br>Black - Main Power OFF<br>2nd LED: Amber – Low Battery Power |
#### MEASURED DISPLAYS 6.2.
### 6.2.1. Chamber Set Pressure Display
A pneumatic pressure gauge having multiple scales is installed to display the set pressure.
### Specification
| (a) | Display Type: | Pneumatic Pressure Gauge |
|-----|----------------|---------------------------------|
| (b) | Display Range: | 0 to 30 psig |
| | | Nominal, 1.5, 2.0, 2.5, 3.0 ATA |
| | | 0 to 200 kpa (optional) |
| (c) | Tolerance: | ±5% of reading |
### 6.2.2. Chamber Actual Pressure Display
A pneumatic pressure gauge having multiple scales is installed to display actual , hyperbaric chamber pressure.
### Specification
| | (a) Display Type: | Pneumatic Pressure Gauge |
|--|---------------------|---------------------------------|
| | (b) Display Range: | 0 to 30 psig |
| | | Nominal, 1.5, 2.0, 2.5, 3.0 ATA |
| | | 0 to 200 kpa (optional) |
| | (c) Tolerance: | ±5% of reading |
{6}------------------------------------------------
#### Chamber Pressure Cycle Counter 6.2.3.
A pressure cycle counter is installed to record the total number of hyperbaric chamber pressurization cycles. The pressure cycle counter can be used to keep track of periodic maintenance requirements of the chamber.
### Specification
| (a) Display Type: | Pneumatic Mechanical Counter |
|-------------------|------------------------------|
| (b) Range: | 0 to 99,999 cycles min |
- (c) Reset: None
#### 6.3. INDICATORS
### 6.3.1. Supply Pressure Indicator
An indicator is provided to indicate the existence of the supply pressure.
### Specification
| (a) Indicator Type: | Pneumatically operated<br>Green to indicate supply ON<br>Black to indicate supply OFF |
|---------------------|---------------------------------------------------------------------------------------|
|---------------------|---------------------------------------------------------------------------------------|
### 6.3.2. Chamber Pressure Indicator
An indicator is provided to indicate the existence of the hyperbaric chamber pressure.
### Specification
| (a) Indicator Type: | Pneumatically operated |
|---------------------|----------------------------------------------------------|
| | Red to indicate chamber pressure of ~ 1.5 psig or higher |
| | Black to indicate chamber pressure less than 0.5 psig |
#### SAFETY SYSTEMS 6.4.
### 6.4.1. Door Safety Interlock System
The door Safety Interlock System prevents the hyperbaric chamber door from being accidentally opened while the hyperbaric chamber is pressurized. It also prevents the pressurization of the hyperbaric chamber if the door is not in the fully locked position.
### Specification
- (a) Activation pressure: approximately 1.0 psig
- (b) Deactivation pressure: approximately 0.5 psig
### 6.4.2. Over-Pressure Relief System
{7}------------------------------------------------
Two over-pressure relief valves are installed to prevent hyperbaric chamber from exceeding its maximum allowable working pressure. If the relief valve malfunctions and causes rapid decompression of the hyperbaric chamber pressure, the Manual Shut-Off Valve is also be available to prevent the rapid decompression.
### 3300H/HR:
| (a) Relief Valve Set Pressure: | 35 psig maximum |
|--------------------------------|-----------------------------------|
| (b) Tolerance: | ±2 psig |
| (c) Manual Shut-Off Valve: | Sealed open with a soft wire seal |
### 3600H/HR:
| (a) Relief Valve Set Pressure: | 36 psig maximum |
|--------------------------------|-----------------------------------|
| (b) Tolerance: | $\pm$ 2 psig |
| (c) Manual Shut-Off Valve: | Sealed open with a soft wire seal |
### 6.4.3. Emergency Shutoff and Automatic Vent
The Emergency Shutoff system allows the operator to automatically shut off the gas supply and depressurizes the hyperbaric chamber. This system is activated using a toggle switch protected by a cover to prevent accidental activation.
### Specification
| (a) Control Type: | Pneumatic Toggle Switch |
|-------------------------|----------------------------------|
| (b) Decompression Time: | 30 psig to 0 psig in 119 sec max |
{8}------------------------------------------------
#### TABLE OF COMPARISON 7.
There are no changes in the intended use and there are no significant changes in performance specifications or principal of operation of the Models 3300H/HR and 3600H/HR Hyperbaric Chambers, as compared to the predicate device Sechrist Model 4100H/HR Hyperbaric Chamber. Existing technology, consistent with that used in comparative chambers, has been incorporated inito the design of this device.
Modifications made to the device do not affect the intended use or alter the fundamental scientific technology of the device. Therefore, the modified device is substantially equivalent to the cleared device.
| Description | Model 4100H/HR<br>(Predicate) | Model 3600H | Model 3300H |
|-------------------------------------------|----------------------------------------------------------------------|-----------------------------------|---------------------------------------------------------------|
| Operational Control | | | |
| | Pneumatic Only | No Change | No Change |
| Modes | | | |
| Emergency Vent | Included | No Change | No Change |
| Off | Included | No Change | No Change |
| ON | Included | No Change | No Change |
| Emergency Shutoff and<br>Automatic Vent | Included | No Change | No Change |
| Displays | | | |
| Chamber Pressure Gauge | Mechanical Gauge | No Change | No Change |
| Pressure Set Gauge | Mechanical Gauge | No Change | No Change |
| Pressure Cycle Counter | Mechanical/Pneumatic<br>type | No Change | No Change |
| Alarms/Indicators | | | |
| Supply Pressure Indicator | Pneumatic | No Change | No Change |
| Chamber Pressure Indicator | Pneumatic | No Change | No Change |
| AC Power Indicator | LED | No Change | No Change |
| Battery Charging Indicator | LED | No Change | No Change |
| Low Battery Warning/Caution | LED | No Change | No Change |
| Accessories | | | |
| Intercom System | Included | No Change | No Change |
| Purge Rate Adjustment | Included | No Change | No Change |
| Materials | | | |
| | Aluminum | No Change | No Change |
| | Stainless Steel | No Change | No Change |
| Pressure bearing Materials | Alloy Steel | No Change | No Change |
| | Acrylic | No Change | No Change |
| Description | Model 4100H/HR<br>(Predicate) | Model 3600H | Model 3300H |
| Specifications | | | |
| Maximum Operating Pressure | 30.0 psig<br>3.0 ATA<br>2.1 kg/cm²<br>206.8 kPa at ambient | No Change | No Change |
| Operating Temperature Range | 50° - 100° F<br>(10°- 38° C) | No Change | No Change |
| Operating Humidity Range | 30% to 90% at 77° F | No Change | No Change |
| Supply Pressure | 50.0 to 70.0 psig<br>(3.5 to 4.9 kg/cm²)<br>(344.8 to 482.7 kPa) | No Change | No Change |
| Supply Pressure Flow Rate | 42 SCFM | 34 SCFM | 32 SCFM |
| Rate Set | 1- 5.0 psig/min | No Change | No Change |
| Pressure Set | 1.5 - 30 psig | No Change | No Change |
| Purge Rate | 80 to 400 LPM with the chamber pressure set at 15 psig | No Change | No Change |
| Emergency Vent Rate | 30 psig to 0 psig in 119 seconds maximum<br>(0.25 - 0.5 psig/second) | No Change | No Change |
| Relief valves | Quantity (2)<br>Set at 36.00 psig | Quantity (2)<br>Set at 36.00 psig | Quantity (2)<br>Set at 35.00 psig |
| Electrical Power | | | |
| A.C. Input range | 90 to 264 VAC | No Change | No Change |
| A.C. Input Frequency | 47 to 33 Hz | No Change | No Change |
| D.C Output Voltage | +12 VDC Nominal | No Change | No Change |
| D.C. Output Current | 3.3 A Maximum | No Change | No Change |
| Cylinder Dimensions (reference Drawing A) | | | |
| Cylinder Internal Diameter | 41.00 Inches<br>(104 cm) | 35.50 Inches<br>(83 cm) | 32.50 Inches<br>(90 cm) |
| Internal Length | 91.00 Inches<br>(231 cm) | 91.00 Inches<br>(231 cm) | 91.00 Inches<br>(231 cm) |
| External Length(A) | 106.00 Inches<br>(269 cm) | 106.00 Inches<br>(269 cm) | 106.00 Inches<br>(269 cm) |
| External Height (B) | 69.75 Inches<br>(177 cm) | 63.00 Inches<br>(160 cm) | 59.00 Inches<br>(150 cm) |
| External Width (C) | 46.75 Inches<br>(119 cm) | 44.50 Inches<br>(113 cm) | 44.50 Inches<br>(113 cm) |
| Chamber Weight | 3,500 lbs<br>(1591 kg) | 2,220 lbs<br>(1009 kg) | 2,030 lbs<br>(923 kg) |
| Supported Patient Weight | 700 lbs standard<br>(318 kg) | 700 lbs standard<br>(318 kg) | 500 lbs standard<br>(227 kg);<br>700 lbs optional<br>(318 kg) |
The chart below compares the modifications listed above to the predicate device.
Page 9 of 20
{9}------------------------------------------------
# 510(K) SUMMARY
SPECIAL 510(K) PREMARKET NOTIFICATION
3300H/HR and 3600H/HR Hyperbaric Chamber
・
Page 10 of 20
{10}------------------------------------------------
510(K) SUMMARY
SPECIAL 510(K) PREMARKET NOTIFICATION
3300H/HR and 3600H/HR Hyperbaric Chamber
Image /page/10/Figure/1 description: The image shows a technical drawing of a machine with dimensions labeled. The overall length of the machine is labeled as 193.0 inches or 4.9 meters. The drawing also includes labels A, B, and C, which likely represent other key dimensions of the machine, but the specific values for these dimensions are not provided in the image.
Drawing A
{11}------------------------------------------------
#### DETERMINATION OF SUBSTANTIAL EQUIVALENCE 8.
### Summary of Technological Changes
### Similarities
The modifications made to the existing Model 4100H/HR Hyperbaric Oxygen Chamber do not affect the indications for use or theory of operation, and do not change the hyperbaric pressure technology used in the design of the chamber. There has been no change in the science involved with providing hyperbaric oxygen therapy in a monoplace chamber. The weight and dimensions of the two models created by the change are decreased due to the reduction in the size of the acrylic viewport window and pressure retaining metal components. The pressure bearing components and materials have not been changed and comply with the ASME Boiler & Pressure Vessels Code Section VIII Division 1 and ASME PVHO-1 Safety Standard for Pressure Vessels for Human Occupancy. The control panel used for each model is the same as found on the Model 4100H/HR Hyperbaric Chamber with the operational controls remaining the same.
The pressure related safety features of the two Models 3300H/HR and 3600H/HR Hyperbaric Chambers, including the door interlock system, and the door safety switch have not been changed from the Model 4100H/HR Hyperbaric Chamber.
### Differences (Changes)
The only changes to the Model 4100H/HR resulting in the two Models 3300H/HR and 3600H/HR Hyperbaric Chambers are as follows:
#### A. Acrylic Cylinder Size
# Model 3600H/HR Hyperbaric Chamber
This Hyperbaric Chamber model has a smaller internal diameter, external height and external width when compared to the Model 4100H/HR Hyperbaric Chamber (predicate). These difference in sizes from the Model 4100H/HR Hyperbaric Chamber (predicate device) can be seen in the Engineering Drawings (see Section 12 - Attachment A).
| Cylinder Internal Diameter | 41.00 Inches (104 cm) | 35.50 Inches (83 cm) |
|----------------------------|-----------------------|-----------------------|
| External Height (B) | 69.75 Inches (177 cm) | 63.00 Inches (160 cm) |
| External Width (C) | 46.75 Inches (119 cm) | 44.50 Inches (113 cm) |
| Chamber Weight | 3,500 lbs (1591 kg) | 2.220 Ibs (1009 kg) |
{12}------------------------------------------------
- · The effects of the modification to this model (i.e., smaller cylinder internal diameter, external height and external widths) results in a reduction of the overall chamber weight when compared to that of the predicate device. This reduction in cylinder size does not affect the operation of the device and its ability to function to the same specifications as the predicate device.
# Model 3300H/HR Hyperbaric Chamber
This Hyperbaric Chamber model has a smaller internal diameter, external height and external width when compared to the Model 4100H/HR Hyperbaric Chamber (predicate). These difference in sizes from the Model 4100H/HR Hyperbaric Chamber (predicate device) can be seen in the Engineering Drawings (see Section 12 - Attachment A).
| Description | Model 4100H/HR<br>(Predicate) | Model 3300H |
|----------------------------|-------------------------------|-----------------------|
| Cylinder Internal Diameter | 41.00 Inches (104 cm) | 32.50 Inches (90 cm) |
| External Height (B) | 69.75 Inches (177 cm) | 59.00 Inches (150 cm) |
| External Width (C) | 46.75 Inches (119 cm) | 44.50 Inches (113 cm) |
| Chamber Weight | 3,500 lbs (1591 kg) | 2,030 lbs (923 kg) |
The effects of the modification to this model (i.e., smaller cylinder internal diameter, external height and external widths) results in a reduction of the overall chamber weight when compared to that of the predicate device.
#### B. Relief Valve
This Hyperbaric Chamber model has two over-pressure relief valves installed to prevent the hyperbaric chamber from exceeding its maximum allowable working pressure which is the design pressure. However, the maximum working pressure is still limited to 30 psi or 3 ATA. The Model 3300H/HR has a setting that differs from the predicate device while the Model 3600H/HR has the same setting as the predicate device. The pressure setting of this model is the same as a previous device (Model 3200) for which the 3300H/HR will eventually replace.
| Description | Model 4100H/HR<br>(Predicate) | Model 3600H | Model 3300H |
|---------------|-----------------------------------|-----------------------------------|-----------------------------------|
| Relief valves | Quantity (2)<br>Set at 36.00 psig | Quantity (2)<br>Set at 36.00 psig | Quantity (2)<br>Set at 35.00 psig |
{13}------------------------------------------------
#### Gas Supply Flow rate C.
The Hyperbaric Chamber requires a gas supply source with a flow rate measured in standard cubic feet per minute (SCFM) units capable of supplying a gas supply pressure within a range of 50-70 psig. The required SCFM is based on the volume requirements during peak flow conditions for each model Hyperbaric Chamber.
The typical treatment cycle of a hyperbaric chamber consists of three phases:
- 1. Peak Flow defined as increase from ambient to the prescribed treatment pressure at worst condition.
- 2. Steady Flow defined as pressure hold at treatment level.
- 3. Minimum Flow defined as exhaust from treatment pressure to ambient.
The required Gas Supply Flow Rates for the Models 3600H/HR and 3300H/HR Hyperbaric Chambers as compared to the predicate model is provided in the table below. The table illustrates the flow rates required for the Model 3600H/HR and Model 3300H/HR are lower than the predicate device Model 4100H/HR Hyperbaric Chamber.
| Description | Model 4100H/HR<br>(Predicate) | Model 3600H/HR | Model 3300H/HR |
|----------------------|-------------------------------|----------------|----------------|
| Gas Supply Flow Rate | 42 SCFM | 34 SCFM | 32 SCFM |
{14}------------------------------------------------
#### TESTING & PERFORMANCE EVALUATION ਰੇ
The Table below summarizes the tests required to validate/verify design changes based on the risk analysis. Design Validation & Verification, Electrical Safety/EMC and Performance test reports are provided in Appendix VIII.
| Device Change | Device Requirement<br>Reference | Testing Performed | Acceptability<br>Criteria | Test<br>Reference |
|----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|-----------------------------|---------------------------------------------|
| | Device specification | System Design Validation | Performs to<br>requirements | 2009-0010 |
| Decreased<br>Acrylic Window<br>Size, Length and<br>Width | ASME Authorized Pressure<br>Vessel Facility | ASME Boiler & Pressure<br>Vessels Code Section VIII<br>Division 1 | ASME<br>Authorization | ASME "U"<br>Certificate of<br>Authorization |
| | PVHO-1 Pressure Bearing<br>Testing | PVHO-1 Safety Standard for<br>Pressure Vessels for Human<br>Occupancy | Tested at 1.1 x<br>MAWP | PVHO-1<br>Appendix I,<br>Enclosure 4 |
| Over Pressure<br>Relief valve<br>pressure<br>change | 1. Device Specification<br>2. Final Test and<br>Calibration Procedures<br>Model 3300H/HR,<br>3600H/HR, 4100H/HR<br>Series | 1. System Design Validation<br>2. Activation pressure of<br>valves | Performs to<br>requirements | Document<br>No. 150156,<br>Section 5 |
| Gas supply<br>flow rate | 1. Device Specification<br>2. Final Test and<br>Calibration Procedures<br>Model 3300H/HR,<br>3600H/HR, 4100H/HR<br>Series | 1. System Design Validation<br>2. Device performs to<br>specification after<br>calibration. | Performs to<br>requirements | 1. 2009-0010<br>2. Doc. No.<br>150156 |
### SYSTEM VALIDATION MATRIX
#### DETAILED DESCRIPTION OF MODIFICATIONS 10.
The modifications resulting in the Model 3300H/HR Hyperbaric Chamber and Model 3600H/HR Hyperbaric Chamber as compared with the Model 4100H/HR Hyperbaric Chamber (predicate) are provided in the following sections and associated tables.
### 10.1. Smaller Acrylic Cylinder Size
### Design Requirement:
The internal diameter of the Model 3300H/HR acrylic cylinder was decreased to 32.50 inches and the internal diameter for the Model 3600H/HR acrylic cylinder was decreased to 35.50 inches from the internal diameter of 41.0 inches of the Model 4100H/HR (predicate device) to provide the most commonly used size of chamber for the most common patient population size on the market and reduce the space necessary to house the hyperbaric chambers.
### Design Modification:
The decrease in the internal diameter of the cylinder from 41.0 inches (Model 4100H/HR) to 32.50 inches (Model 3300H/HR) was made to meet customer requests for a Hyperbaric
{15}------------------------------------------------
Chamber that would accommodate the most common patient size and weight seeking clinical treatment in the market and reduce the space necessary to house the hyperbaric chambers. The reduction in the internal diameter also resulted in reduction of external height and width of the cylinder.
### Hazard / Risk:
Change does not pose a risk to the operator or patient.
### Validation:
The decrease in acrylic viewport size was validated by testing the Model 3300H/HR and Model 3600H/HR Hyperbaric Chambers ability to operate to their approve product specifications. The data indicates the change does not affect the devices ability to perform to its product specifications and performs equivalently to the Model 4100H/HR Hyperbaric Chamber (predicate device).
| Item | Description | Model 4100H/HR<br>(Predicate) | Model 3600H | Model 3300H |
|------|----------------------------|-------------------------------|--------------------------|--------------------------|
| 1 | Cylinder Internal Diameter | 41.00 Inches<br>(104 cm) | 35.50 Inches<br>(83 cm) | 32.50 Inches<br>(90 cm) |
| 2 | External Height (B) | 69.75 Inches<br>(177 cm) | 63.00 Inches<br>(160 cm) | 59.00 Inches<br>(150 cm) |
| 3 | External Width (C) | 46.75 Inches<br>(119 cm) | 44.50 Inches<br>(113 cm) | 44.50 Inches<br>(113 cm) |
# 10.2. Relief Valve Pressure Change
# Design Requirement:
The Relief Valve Pressure for the Model 3300H/HR Hyperbaric Chamber was changed to 35 psig from the pressure of 36 psig offered on the predicate device Model 4100H/HR Hyperbaric Chamber (predicate device). This change was made to offer the same pressure relief option as the Sechrist Model 3200 released by 510(k) K950386.
The Relief Valve Pressure for the Model 3600H/HR Hyperbaric Chamber remains unchanged from the Model 4100H/HR Hyperbaric Chamber (predicate device).
# Design Modification: .
The Relief Valves relief pressure was adjusted by the supplier to trigger at 35 psig ± 2 psig.
# Hazard / Risk:
The identified hazard of excessive chamber pressure could potentially result in patient barotrauma risk. The identified hazard of chamber pressurization exceeding design pressure could potentially result in catastrophic structural failure risk.
{16}------------------------------------------------
### Validation
The activation of the Relief Valve Pressure validated by testing the Model 3300H/HR and Model 3600H/HR Hyperbaric Chambers ability to activate the Relief Valves at their respective preset activation pressures. Each Relief Valve is tested prior to installation during the manufacturing of the Hyperbaric Chamber. . This data is recorded in the "Final Test and Calibration Procedures Model 3300H/HR, 3600H/HR, 4100H/HR Series" (Document 150156).
The data indicates that the Model 3300H/HR and Model 3600H/HR Hyperbaric Chambers activates the Relief Valves at the activation pressures as outlined in the product specifications and performs equivalently to the Model 4100H/HR Hyperbaric Chamber (predicate device).
| .Item | Description | Model 4100H/HR<br>(Predicate) | Model 3600H | Model 3300H |
|-------|------------------------------------------------|-------------------------------|------------------|------------------|
| 1 | Relief Valve Pressure<br>(activation pressure) | 36 psig ± 2 psig | 36 psig ± 2 psig | 35 psig ± 2 psig |
### 10.3. Gas Supply Flow Rate Change
The Hyperbaric Chamber requires a gas supply source with a flow rate measured in standard cubic feet per minute (SCFM) units capable of supplying a gas supply pressure within a range of 50-70 psig. The required SCFM is based on the volume requirements during peak flow conditions for each model Hyperbaric Chamber.
The typical treatment cvcle of a hyperbaric chamber consists of three phases:
- 1. Peak Flow defined as increase from ambient to the prescribed treatment pressure.
- 2. Steady Flow defined as pressure hold at treatment level.
- 3. Minimum Flow defined as exhaust from treatment pressure to ambient.
The required Gas Supply Flow Rates for the Models 3600H/HR and 3300H/HR Hyperbaric Chambers as compared to the predicate model is provided in the table below.
| Description | Model 4100H/HR<br>(Predicate) | Model 3600H | Model 3300H |
|----------------------|-------------------------------|-------------|-------------|
| Gas Supply Flow Rate | 42 SCFM | 34 SCFM | 32 SCFM |
### Design Requirement:
The Gas Supply Flow Rate for the Model 3300H/HR Hyperbaric Chamber was reduced to 32 SCFM from the predicate device of 42 SCFM. The Model 3600H/HR Hyperbaric Chamber Gas Supply Flow Rate was also reduced to 34 SCFM from the predicate device of 42 SCFM. The SCFM required to operate each device is based on the volume requirements of each Hyperbaric Chamber model. The Gas Supply Flow Rate required during peak flow conditions is used to establish the minimum SCFM for each model.
{17}------------------------------------------------
### Design Modification:
The Model 3600H/HR and the Model 3300H/HR Hyperbaric Chamber have a lower volume than the Model 4100H/HR Hyperbaric Chamber (predicate device). The Model 3600H/HR requires 34 SCFM to ensure operation to its approved specifications and Model 3300H/HR Hyperbaric Chamber requires 32 SCFM to ensure operation to its approved specifications.
### Hazard / Risk:
The identified potential hazard of fluctuations in gas supplies may lead a risk of the inability to compress Hyperbaric Chamber.
### Validation
The decrease in Gas Supply Flow Rates were validated by testing the Model 3300H/HR and Model 3600H/HR Hyperbaric Chambers ability to operate to their approve product specifications. Each model is tested 100% using the recommended supply pressure settings during manufacturing to ensure it will perform to its approved specification. This data is recorded in the "Final Test and Calibration Procedures Model 3300H/HR, 3600H/HR, 4100H/HR Series" (Document 150156). The data indicates the change does not affect the devices ability to perform to its product specifications and performs equivalently to the Model 4100H/HR Hyperbaric Chamber (predicate device).
| Item | Description | Model 4100H/HR<br>(Predicate) | Model 3600H | Model 3300H |
|------|----------------------|-------------------------------|-------------|-------------|
| 1 | Gas Supply Flow Rate | 42 SCFM | 34 SCFM | 32 SCFM |
{18}------------------------------------------------
### 11. VALIDATION / VERIFICATION MEASURES:
The device modification found in the Model 3300H/HR Hyperbaric Chamber and Model 3600H/HR Hyperbaric Chamber are provided in Attachment A.
The Models 3300H/HR and 3600H/HR Hyperbaric Chambers vary in acrylic viewport size and pressure retaining metal components. The control system, piping and performance specifications did not change nor have an impact on the functionality of the device from the Model 4100H/HR Hyperbaric Chamber. The validation documents tested the worstcase scenario which is the larger Model 4100H/HR Hyperbaric Chamber. Functionality of the Models is tested during production as a condition of release.
The Model 4100H/HR Hyperbaric Chamber is considered the worst-case scenario
| Device Change | Risk Analysis<br>Reference | Device<br>Requirement<br>Reference | Testing<br>Performed | Acceptability<br>Criteria | Test Reference |
|-------------------------------------------------------------|----------------------------|------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|-----------------------------|---------------------------------------------|
| Decreased<br>Acrylic<br>Window Size,<br>Length and<br>Width | Items 1, 2, 3 | Device<br>specification | System Design<br>Validation | Performs to<br>requirements | 2009-0010 |
| | | ASME<br>Authorized<br>Pressure Vessel<br>Facility | ASME Boiler &<br>Pressure<br>Vessels Code<br>Section VIII<br>Division 1 | ASME<br>Authorization | ASME "U"<br>Certificate of<br>Authorization |
| | | PVHO-1<br>Pressure<br>Bearing Testing | PVHO-1 Safety<br>Standard for<br>Pressure<br>Vessels for<br>Human<br>Occupancy | Tested at 1.1 x<br>MAWP | PVHO-1<br>Appendix 1,<br>Enclosure 4 |
| Over Pressure<br>Relief valve<br>pressure<br>change | Item 1 | Device<br>specification | System Design<br>Validation | Performs to<br>requirements | 2009-0010 |
| | | Final Test and<br>Calibration<br>Procedures<br>Model<br>3300H/HR,<br>3600H/HR,<br>4100H/HR<br>Series | Activation of<br>Relief Valve<br>verification and<br>Install | Performs to<br>requirements | Document No.<br>150156 |
| Gas supply<br>flow rate | Item 1 | Device<br>specification | System Design<br>Validation | Performs to<br>requirements | 2009-0010 |
| | | Final Test and<br>Calibration<br>Procedures<br>Model<br>3300H/HR,<br>3600H/HR,<br>4100H/HR<br>Series | Functional<br>performance<br>to specification<br>after<br>calibration. | Performs to<br>requirements | Document No.<br>150156 |
{19}------------------------------------------------
# 12. VALIDATION / VERIFICATION SUMMARY
The validation data indicates the modifications made to the Model 4100H/HR Hyperbaric Chamber (predicate) resulting in the two new Hyperbaric Chamber Models 3300H/HR and 3600H/HR does not affect the Models 3300H/HR and 3600H/HR performance to meet the approved product specifications and these perform as well as and as safe as the predicate device.
{20}------------------------------------------------
Image /page/20/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The text is in all caps and appears to be in a sans-serif font. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 30, 2014
Sechrist Industries, Inc. Mr. Victor Arellano Director Regulatory Affairs, Product Safety & Certification 4225 East La Palma Avenue Anaheim, CA 92807
Re: K140559
Trade/Device Name: Sechrist Models 3300H/HR and 3600H/HR Hyperbaric Chamber Regulation Number: 21 CFR 868.5470 Regulation Name: Hyperbaric Chamber Regulatory Class: II Product Code: CBF Dated: April 30, 2014 Received: May 01, 2014
Dear Mr. Arellano:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave teviewed your becamed the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encrosure) to regary manation poor the Medical Device Amendments, or to Commerce prior to May 20, 1977) are occordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). allo Cosmetic Act (710) that to hot required to the general controls provisions of the Act. The You may, merciole, manel and nevyolyde requirements for annual registration, listing of general controls proficions practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability additeration. Ticaso note: Obter See hat device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (see above) inxisting major regulations affecting your device can be filly be subject to additional ochaons. Existing may of to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{21}------------------------------------------------
Page 2 - Mr. Arellano
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resources/orYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part the regulation entitled, "Misoranams" of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the rou may other baller general mornemational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
(800) 030 2011 01 (30 41 01 (30 40 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit Sheth M.D. Cilnical Deputy Director
DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{22}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K140559
### Device Name
Sechrist Models 3300H/HR Hyperbaric Chamber Sechrist Models 3600H/HR Hyperbaric Chamber
### Indications for Use (Describe)
The intended use of the hyperbaric chambers have been established by the Committee on Hyperbaric Oxygen Therapy of The Intended as of the Irportant in 1967 to foster exchange of data on the physiology and the Onucles of commercial and military diving). The committee is comprised of practitioners and scientific investigation the fields of internal medicine, infectious diseases, pharmacology, emergency medicine, general surgery, orthopedic the freits of mierner modions, incomed essent of the maxillofacial surgery and aerospace medicine. The surely, dauna surgery, and continually reviewing research and clinical data in determining the safety and efficacy of committee is responsible to continually to reville results that are approved by the committee; these thirteen indications were accepted based on sound physiologic rationale, in viro studies that demonstrate effectiveness, controlled were accepted based on some physionals, in 1778 to extensive clinical experience from multiple hyperbaric animal studies, prospective connect states and should be third-party reimbursement and most insurance carriers have established reimbursement policy based on the recommendations.
The thirteen indications are:
- 1. Air or Gas embolism
- 2. Carbon Monoxide Poisoning Carbon Monoxide Poisoning Complicated by Cyanide Poisoning
- 3. Clostridial Myonecrosis and Myonecrosis (Gas Gangrene)
- 3. Crush Injury, Compartment Syndrome and Other Acute Traumatic Ischemias
- 5. Decompression Sickness
- 6. Arterial Insufficiencies
- Central Retinal Artery Occlusion Enhancement of Healing In Selected Problem Wounds
- 7. Severe Anemia
- 8. Intracranial Abscess 9. Necrotizing Soft Tissue Infections
- 10. Osteomyelitis (Refractory)
- 11. Delayed Radiation Injury (Soft Tissue and Bony Necrosis)
- 12. Compromise Grafts and Flaps
- 13. Thermal Burns
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{23}------------------------------------------------
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
THE REAL PART FOR FOR FDAUSE ONLY - FOR FOR FOR FOR FOR FOR FOR FOR FOR SERVEN FOR SERVEN Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Image /page/23/Picture/2 description: The image shows the name "Todd D. Courtney -S" in bold font. Below the name is the date and time "2014.05.30 10:26:31 -04 00" also in bold font. To the right of the date and time is a logo that is difficult to make out.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.