MODEL 3300E/ER MONOPLACE HYPERBARIC OXYGEN CHAMBER
K052713 · Sechrist Industries, Inc. · CBF · Oct 28, 2005 · Anesthesiology
Device Facts
Record ID
K052713
Device Name
MODEL 3300E/ER MONOPLACE HYPERBARIC OXYGEN CHAMBER
Applicant
Sechrist Industries, Inc.
Product Code
CBF · Anesthesiology
Decision Date
Oct 28, 2005
Decision
SESE
Submission Type
Special
Regulation
21 CFR 868.5470
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use of the Sechrist Model 3300E/ER Hyperbaric Chamber is to administer 100% oxygen at pressure greater than ambient, up to 3 atmospheres absolute (30 psi) of pressure. Specific indications for use of hyperbaric chambers have been established by the Committee on Hyperbaric Oxygen Therapy of the Undersea and Hyperbaric Medical Society. The current specific thirteen indications are: Air or Gas Embolism; Carbon Monoxide Poisoning and Smoke Inhalation, Carbon Monoxide Complicated by Cyanide Poisoning; Clostridial Myonecrosis (Gas Gangrene); Crush Injury, Compartment Syndrome and Other Acute Traumatic Ischemias; Decompression Sickness; Enhancement of Healing in Selected Problem Wounds; Exceptional Blood Loss (Anemia); Intracranial Abscess; Necrotizing Soft Tissue Infections (Subcutaneous Tissue, Muscle, Fascia); Osteomyelitis (Refractory); Radiation Tissue Damage (Osteoradionecrosis); Skin Grafts and Flaps (Compromised); Thermal Burns.
Device Story
Monoplace hyperbaric chamber; administers 100% oxygen at pressures up to 3 atmospheres absolute (30 psi). Construction: horizontal seamless acrylic cylinder; hard anodized aluminum cover and door assembly; includes access ports for patient monitoring and IV fluid delivery. Control system: pneumatic/electronic; includes pressure display gauges and intercom. Used in clinical settings; operated by trained personnel. Provides hyperbaric oxygen therapy to treat specific medical conditions; supports patient healing and physiological recovery.
Clinical Evidence
Bench testing only. No clinical data presented. Device validation included requirements specification reviews, performance testing, environmental testing, electrical safety/EMC testing, and manufacturing process controls.
Technological Characteristics
Monoplace hyperbaric chamber; acrylic cylinder; hard anodized aluminum door/cover. Pneumatic/electronic control system. Pressure range: up to 3 ATA (30 psi). Standards: ASME Boiler and Pressure Vessel Code Section VIII, Div 1; ASME PVHO-1; NFPA 99. Connectivity: includes intercom and patient monitoring ports.
A hyperbaric chamber is a device that is intended to increase the environmental oxygen pressure to promote the movement of oxygen from the environment to a patient's tissue by means of pressurization that is greater than atmospheric pressure. This device does not include topical oxygen chambers for extremities (§ 878.5650).
{0}------------------------------------------------
### OCT 2 8 2005
510(k) Summary of Safety and Effectiveness Sechrist Industries, Inc., 3300E/ER Hyperbaric Chamber ..
# 510 (k) Summary of Safety and Effectiveness
Hyperbaric Oxygen Chamber Product:
Model: 3300E/ER
510(k) No .:
General Information
Classification: Class II, CBF AN (73), Anesthesiology Panel (Title 21, CFR 868.5470)
Trade Name: Chamber, Hyperbaric
- Sechrist Industries, Inc. Submitter: 4225 E. La Palma Ave. Anaheim, CA 92807
- Greg Godfrey Contact: Vice President, Quality Assurance & Regulatory Affairs
#### Performance Standards:
A performance standard regulation under Section 514 of the Food, Drug and Cosmetic Act has not been promulgated for this device type.
However, the device is designed and manufactured in accordance with the following industry standards:
Section VIII, Division 1 of the American Society of Mechanical Engineers (ASME) Boiler and Pressure Vessel Code
ASME PVHO-1 Pressure Vessels for Human Occupancy Standard
NFPA 99 Safety Standard for Health Care Facilities, Chapter 20
The following quality assurance measures were applied to the development of the 3300E/ER Hyperbaric Chamber - Requirements specification reviews, performance testing and validation, environmental testing, electrical safety/EMC testing.
{1}------------------------------------------------
510(k) Summary of Safety and Effectiveness
Sechrist Industries, Inc., 3300E/ER Hyperbaric Chamber
manufacturing process control procedures, process validation, and design change controls.
### Substantially Equivalent Devices
Manufacturer Sechrist Industries, Inc. Anaheim, CA
Product Model 3200P/PR K950386
#### Intended Use
The intended use of the Sechrist Model 3300E/ER Hyperbaric Chamber is to administer 100% oxygen at pressure greater than ambient, up to 3 atmospheres absolute (30 psi) of pressure.
Specific indications for use of hyperbaric chambers have been established by the Committee on Hyperbaric Oxygen Therapy of the Undersea and Hyperbaric Medical Society.
The current specific thirteen indications are:
- Air or Gas Embolism 1 .
- Carbon Monoxide Poisoning and Smoke Inhalation, Carbon 2. Monoxide Complicated by Cyanide Poisoning
- Clostridial Myonecrosis (Gas Gangrene) 3.
- Crush Injury, Compartment Syndrome and Other Acute Traumatic 4. Ischemias
- 5. Decompression Sickness
- Enhancement of Healing in Selected Problem Wounds 6.
- 7. Exceptional Blood Loss (Anemia)
- 8. Intracranial Abscess
- 9. Necrotizing Soft Tissue Infections (Subcutaneous Tissue, Muscle, Fascia)
- Osteomyelitis (Refractory) 10.
- Radiation Tissue Damage (Osteoradionecrosis) 11.
- 12. Skin Grafts and Flaps (Compromised)
- 13. Thermal Burns
{2}------------------------------------------------
510(k) Summary of Safety and Effectiveness Sechrist Industries, Inc., 3300E/ER Hyperbaric Chamber
#### Device Description
The Model 3300E/ER Hyperbaric Chamber is a monoplace (one patient) pressure chamber designed to administer 100% oxygen at pressure greater than ambient, up to 3 atmospheres absolute (30 psi) of pressure.
The Model 3300E/ER Hyperbaric Chamber is constructed with a horizontal seamless acrylic inner cylinder. One end is equipped with a hard anodized aluminum cover and the other with a hard anodized aluminum door assembly, with access ports available for patient support (such as patient monitoring and delivery of intravenous fluids). The control system is pneumatic/electronic containing the controls, indicators, chamber pressure display gauges, and intercom assembly.
#### Summary of Substantial Equivalence
The design features and materials used in the manufacture of the Sechrist Industries. Inc. 3300E/ER Hyperbaric Chamber are substantially equivalent several pre-amendment products. Additionally the 3300E Hyperbaric Chamber is of similar shape and functionality to the predicate device.
Therefore, due to the similarity of design features, materials, test results and the similarity of the indicated use to other pre-amendment devices and the predicate device+, Sechrist Industries, Inc. believes this product does not have any safety or efficacy issues.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its wings or body. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
0C1 2 8 2005
Mr. Greg P. Godfrey Vice President, Quality Assurance & Regulatory Affairs Sechrist Industries, Incorporated 4225 East La Palma Avenue Anaheim. California 92807
Re: K052713
Trade/Device Name: MODEL 3300E/ER MONOPLACE HYPERBARIC OXYGEN CHAMBER Regulation Number: 21 CFR 868.5470 Regulation Name: Hyperbaric chamber Regulatory Class: II Product Code: CBF Dated: September 28, 2005 Received: September 30, 2005
Dear Mr. Godfrey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 - Mr. Godfrey
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
Additions Documents 10-27.
### 510(k) Number (if known): K052713
Device Name:
## Sechrist Model 3300E/ER Hyperbaric Chamber
Indications For Use:
The intended use of the Sechrist Model 3300E/ER Hyperbaric Chamber is to administer 100% oxygen at pressure greater than ambient, up to 3 atmospheres absolute (30 psi) of pressure.
Specific indications for use of hyperbaric chambers have been established by the Committee on Hyperbaric Oxygen Therapy of the Undersea and Hyperbaric Medical Society (founded in 1967 to foster exchange of data on the physiology and medicine of commercial and military diving). The committee is comprised of practitioners and scientific investigators in the fields of internal medicine, infectious diseases, pharmacology, emergency medicine, general surgery, orthopedic surgery, trauma surgery, thoracic surgery, otolaryngology, oral and maxillofacial surgery and aerospace medicine. The committee is responsible for continually reviewing research and clinical data in determining the safety and efficacy of hyperbaric oxygen.
Prescription Use X (Part 21 CFR 801 Subpart D)
n (0(t) Number -
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart G)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Clem br AAG
ാന Sign-Off) ിന് Sign-Oif)
Sign-Oiff
Sion of Anesthesiology, General Hospital, this of of of of the Dontal Devices
Page 1 of 3
{6}------------------------------------------------
# Indications for Use
### 510(k) K052713 Additional Documents 10-27-45
### 510(k) Number (if known): K052713
Device Name:
### Sechrist Model 3300E/ER Hyperbaric Chamber
Currently, there are thirteen indications that are Indications For Use: approved by the committee; these thirteen indications were accepted based on sound physiologic rationale, in vivo or in vitro studies that demonstrate effectiveness, controlled animal studies, prospective controlled clinical studies and extensive clinical experience from multiple hyperbaric medicine centers. These thirteen indications have been recommended for third-party reimbursement and most insurance carriers have established reimbursement policy based on these recommendations. The thirteen indications are:
> Air or Gas Embolism 1 -
- Carbon Monoxide Poisoning and Smoke Inhalation, Carbon 2. Monoxide Complicated by Cyanide Poisoning
- 3. Clostridial Myonecrosis (Gas Gangrene)
- Crush Injury, Compartment Syndrome and Other Acute Traumatic 4. Ischemias
- Decompression Sickness 5.
AND/OR Prescription Use X Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Qur
Coff . In esthesiology, General Hospital, on Chirol, Dental Devices
:(k) Number. K052713
Page 2 of 3
{7}------------------------------------------------
# Indications for U
Additional Documents 10-27-0
510(k) Number (if known): K052713 `
Device Name:
# Sechrist Model 3300E/ER Hyperbaric Chamber
Indications For Use:
- Enhancement of Healing in Selected Problem Wounds 6.
- Exceptional Blood Loss (Anemia) 7.
- Intracranial Abscess 8.
- Necrotizing Soft Tissue Infections (Subcutaneous Tissue, Muscle, 9. Fascia)
- 10. Osteomyelitis (Refractory)
- Radiation Tissue Damage (Osteoradionecrosis) 11.
- 12. Skin Grafts and Flaps (Compromised)
- 13. Thermal Burns
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Charles
(Furision Sign-Off) sion of anesthesiology, General Hospital, .an Control, Dental Devices
510(k) Number: K052913
Page 3 of _3
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.