K260799 · Nipro Technical Solutions, Incoporated · KPO · Aug 19, 2026 · Gastroenterology, Urology
Device Facts
Record ID
K260799
Device Name
Nipro Dry Acid Concentrate Mixer
Applicant
Nipro Technical Solutions, Incoporated
Product Code
KPO · Gastroenterology, Urology
Decision Date
Aug 19, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5820
Device Class
Class 2
Indications for Use
The Nipro Dry Acid Concentrate Mixer is intended for the mixing of dry acid concentrate with treated water that meets the requirements of ISO 23500-3, "Preparation and quality management of fluids for hemodialysis and related therapies—Part 3: Water for hemodialysis and related therapies" to produce an acid concentrate solution intended for use in hemodialysis treatment and related therapies, using a 3-stream dialysate (acid concentrate, bicarbonate concentrate, and water).
Device Story
Nipro Dry Acid Concentrate Mixer (Model NACM 50) prepares liquid acid concentrate for hemodialysis; inputs include dry acid concentrate and hemodialysis-grade reverse osmosis (RO) water. System utilizes 50-gallon HDPE tank, high-flow recirculation pump, and dual opposing Venturi eductors to create high-turbulence flow for dissolution and batch homogeneity. Control console manages fill, mix, rinse, and transfer modes. Device used in clinical settings by facility staff. Output is liquid acid concentrate; facility personnel must verify final solution against manufacturer specifications prior to clinical use. Device ensures repeatable mixing performance; does not perform analytical measurement of solution quality. Benefits include standardized, efficient preparation of dialysate components.
Clinical Evidence
Bench testing only. No clinical data. Performance testing evaluated specific gravity, temperature, electrolyte composition, and endotoxin levels across multiple batches, lots, and operators to ensure consistency with ISO 23500-4:2024 and manufacturer specifications. Human factors/usability engineering validated safe operation.
Technological Characteristics
Materials: HDPE tank. Principle: Mechanical dissolution via high-flow recirculation pump and dual opposing Venturi eductors. Energy: Electric. Connectivity: Standalone. Standards: ANSI/AAMI/ISO 23500-4:2024. Software: Control console for operational modes (Fill, Mix, Rinse, Transfer).
Indications for Use
Indicated for use in hemodialysis facilities to prepare liquid acid concentrate from dry acid concentrate and treated water for use in 3-stream hemodialysis systems.
Regulatory Classification
Identification
A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
Predicate Devices
Dri-Sate Mixer for Preparation of Acidified Dialysate Concentrate (K983618)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
August 19, 2026
Nipro Technical Solutions, Incorporated
Jessica Oswald-Mcleod
Director, Regulatory Affairs
3110 N Oakland, Suite 101
Mesa, AZ 85215
Re: K260799
Trade/Device Name: Nipro Dry Acid Concentrate Mixer
Regulation Number: 21 CFR§ 876.5820
Regulation Name: Hemodialysis System and Accessories
Regulatory Class: II
Product Code: KPO
Dated: July 21, 2026
Received: July 21, 2026
Dear Jessica Oswald-Mcleod:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K260799 - Jessica Oswald-Mcleod
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K260799 - Jessica Oswald-Mcleod
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# MAURA ROONEY -S
Maura Rooney
Assistant Director
DHT3A: Division of Renal, Gastrointestinal, Obesity, and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260799 | ? |
| Please provide the device trade name(s). | | ? |
| Nipro Dry Acid Concentrate Mixer | | |
| Please provide your Indications for Use below. | | ? |
| The Nipro Dry Acid Concentrate Mixer is intended for the mixing of dry acid concentrate with treated water that meets the requirements of ISO 23500-3, "Preparation and quality management of fluids for hemodialysis and related therapies—Part 3: Water for hemodialysis and related therapies" to produce an acid concentrate solution intended for use in hemodialysis treatment and related therapies, using a 3-stream dialysate (acid concentrate, bicarbonate concentrate, and water). | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
510(k) #: K260799
# 510(k) Summary
Prepared on: 2026-07-21
Contact Details
21 CFR 807.92(a)(1)
| Applicant Name | Nipro Technical Solutions, Incorporated |
| --- | --- |
| Applicant Address | 3110 N. Oakland, Suite 101 Mesa AZ 85215 United States |
| Applicant Contact Telephone | 305-432-6699 |
| Applicant Contact | Mrs. Jessica Oswald-McLeod |
| Applicant Contact Email | JessicaO@nipromed.com |
Device Name
21 CFR 807.92(a)(2)
| Device Trade Name | Nipro Dry Acid Concentrate Mixer |
| --- | --- |
| Regulation Name | Hemodialysis system and accessories |
| Classification Name | Dialysate Concentrate For Hemodialysis (Liquid Or Powder) |
| Regulation Number | 876.5820 |
| Product Code(s) | KPO |
Legally Marketed Predicate Devices
21 CFR 807.92(a)(3)
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| --- | --- | --- |
| K983618 | Dri-Sate Mixer for Preparation of Acidified Dialysate Concentrate | KPO |
| K191093 | Isopure Dry Acid Dissolution System | KPO |
| K223431 | Nipro Citric Complete Liquid Citric Acid Concentrate | KPO |
Device Description Summary
21 CFR 807.92(a)(4)
The Nipro Dry Acid Concentrate Mixer (Model NACM 50) is designed to combine dry acid concentrate with hemodialysis grade reverse osmosis (RO) water to produce an acid concentrate solution for use in hemodialysis and related renal replacement therapies. The system integrates mechanical, hydraulic, and electronic components to ensure repeatable and reproducible mixing performance. A 50 gallon HDPE mixing tank, paired with a high flow recirculation pump and dual opposing Venturi eductors, creates high turbulence flow that enhances dissolution efficiency and batch homogeneity.
The Control Console enables user operation of all functional modes. The device's performance related functions include Fill (Pre-Fill and Final-Fill) and Pump (Mix, Clean/Rinse, and Transfer).
The Nipro Dry Acid Concentrate Mixer performs a controlled mixing process but does not perform any analytical measurement or acceptance determination of the resulting solution; the dialysis facility is responsible for verifying that the incoming RO water meets applicable quality requirements and for confirming the final mixed solution meets the dry acid manufacturer's acceptance criteria prior to clinical use.
Intended Use/Indications for Use
21 CFR 807.92(a)(5)
The Nipro Dry Acid Concentrate Mixer is intended for the mixing of dry acid concentrate with treated water that meets the requirements of ISO 23500-3, "Preparation and quality management of fluids for hemodialysis and related therapies—Part 3: Water for hemodialysis and related therapies" to produce an acid concentrate solution intended for use in hemodialysis treatment and related therapies, using a
{5}
3-stream dialysate (acid concentrate, bicarbonate concentrate, and water)
## Indications for Use Comparison
21 CFR 807.92(a)(5)
The subject device, predicate device Dri-Sate Mixer (K983618), and reference device Isopure (K191093) are all dissolution systems intended to produce liquid acid concentrate, one component in a three-stream dialysate used in hemodialysis treatment. No significant differences were identified in the intended use that would introduce new risks.
Nipro Citric Complete (K223431) is included as a reference device for biocompatibility and performance test alignment. This device is appropriate for comparison due to its formulation, handling, and clinical application context directly relevant to the biological endpoints and acid solution composition associated with the output of the subject device, an acid concentrate solution.
## Technological Comparison
21 CFR 807.92(a)(6)
The subject device exhibits technological characteristics that are substantially equivalent to those of the predicate device, including intended use, use environment, system architecture, materials of construction, operational principles, and energy sources. The Nipro Dry Acid Concentrate Mixer employs the same fundamental dissolution and mixing technology used to prepare liquid acid concentrate for three-stream dialysate. Performance testing verifies that the device consistently produces liquid acid concentrate meeting the specifications of ANSI/AAMI/ISO 23500-4:2024, demonstrating that its technological features function as intended.
## Non-Clinical and/or Clinical Tests Summary & Conclusions
21 CFR 807.92(b)
Bench performance testing was conducted by operating the Acid Mixer according to its intended use and labeling, followed by analytical evaluation of the resulting liquid acid solution. The testing strategy was designed to demonstrate repeatable and reproducible performance of the device across representative concentrate formulations, multiple lots, multiple batches, and multiple operators. Specific gravity and temperature were selected as clinically relevant acceptance parameters, as routine dialysis practice relies on measurement and documentation of these parameters to determine batch acceptability in accordance with the concentrate manufacturer specifications.
No clinical testing was performed.
System validation, including human factors/usability engineering, confirmed that the device is safe and effective when considering its intended use, users and use environment. The primary testing objectives for the Acid Mixer system are to verify that the device is correctly installed, calibrated, and capable of consistently producing liquid acid concentrate that meets all required performance, safety, and regulatory specifications. This includes confirming that installation conditions, utilities, safety systems, and device controls function as intended under both normal and stressed operating scenarios, ensuring the system can safely power, fill, mix, pump, drain, and respond to fault conditions such as dry pump operation and sensor disconnections. The data further aims to demonstrate that the Acid Mixer can repeatedly achieve the required critical quality attributes of the final liquid acid concentrate (specific gravity, electrolyte composition, and endotoxin levels) across multiple units and batch repetitions, using acceptance criteria derived from the acid manufacturer and ISO 23500 4:2024.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.