DRI-SATE MIXER FOR PREPARATION OF ACIDIFIED DIALYSATE CONCENTRATE
K983618 · Rockwell Medical Technologies, Inc. · KPO · Aug 5, 1999 · Gastroenterology, Urology
Device Facts
Record ID
K983618
Device Name
DRI-SATE MIXER FOR PREPARATION OF ACIDIFIED DIALYSATE CONCENTRATE
Applicant
Rockwell Medical Technologies, Inc.
Product Code
KPO · Gastroenterology, Urology
Decision Date
Aug 5, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5820
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Rockwell Medical Technologies' Dri-Sate™Mixer is designed to mix the Rockwell Medical Technologies Dri-Sate Acid Concentrate Mixes with purified water to produce an acid concentrate solution for hemodialysis for use in 3-stream (acid concentrate, bicarbonate concentrate, and water) hemodialysis machines / monitors. The Rockwell Medical Supply LLC. Dri-Sate Acid Concentrate Solutions which are mixed in the Dri-Sate Mixer™ are indicated for use in acute and chronic hemodialysis and to be used with the appropriate hemodialysis machine/ monitor and bicarbonate concentrate
Device Story
Dri-Sate Mixer prepares acidified dialysate concentrate for hemodialysis; mixes Dri-Sate Acid Concentrate powder/solutions with purified water. Used in clinical settings with 3-stream hemodialysis machines. Output is an acid concentrate solution proportioned into the hemodialysis machine alongside bicarbonate concentrate and water to create final dialysate. Final dialysate circulates through hemodialyzer to remove waste products (urea, creatinine, potassium) via diffusion and ultrafiltration across semipermeable membrane. Facilitates treatment of acute and end-stage renal failure.
Clinical Evidence
No clinical data. No in vitro testing performed.
Technological Characteristics
Mixing device for hemodialysis concentrate solutions. Utilizes standard mixing techniques for concentrate powders/solutions with purified water. Designed for use with 3-stream hemodialysis machines. No specific materials or software algorithms described.
Indications for Use
Indicated for use in acute and chronic hemodialysis patients requiring hemodialysis therapy using 3-stream hemodialysis machines/monitors.
Regulatory Classification
Identification
A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
Predicate Devices
Fresenius USA, Inc. Hemodialysis Concentration Dissolution Unit P/N 89-290-09 (K944493)
Reference Devices
Rockwell Medical Technologies Dri-Sate Acid Concentrate Hemodialysis Mixes (K981003)
Submission Summary (Full Text)
{0}------------------------------------------------
Rockwell Medical Technologies
KID SETS
Rockwell Medical Technologies Dri-Sate™ Mixe
Section 16. 510(k) Summary
# 510K(k) SUMMARY
| SUBMITTER: | Rockwell Medical Technologies<br>28025 Oakland Oaks<br>Wixom, MI 48393<br>Phone: 248-546-0040 |
|-----------------------|-----------------------------------------------------------------------------------------------|
| DATE PREPARED: | October 14th, 1998 |
| DEVICE NAME: | Dri-Sate Mixer for Preparation of Acidified<br>Dialysate Concentrate |
| CLASSIFICATION NAMES: | Accessory to Hemodialysis - Hemodialysis Bath<br>Concentrate Mixing Device |
| PREDICATE DEVICE: | Fresenius USA, Inc. Hemodialysis Concentration<br>Dissolution Unit P/N 89-290-09 |
#### Device Description:
The Rockwell Medical Technologies' Dri-Sate™Mixer is designed to mix the Rockwell Medical Technologies Dri-Sate Acid Concentrate Mixes with purified water to produce an acid concentrate solution for hemodialysis for use in 3-stream (acid concentrate, bicarbonate concentrate, and water) hemodialysis machines / monitors.
Hemodialysis therapy removes blood wastes by diffusion through a dialyzer membrane into a dialysate solution flowing on the opposite side of the dialyzer membrane. Hemodialysis also involves using a differential transmembrane pressure to ultrafilter water from the blood usually resulting in a net patient weight loss during the treatment.
The Rockwell Medical Technologies' Dri-Sate™ Mixer is designed to be used with the Rockwell Medical Supply, LLC. Dri-Sate™ Acid Concentrate Hemodialysis Mixes which contain salt, sugar, and non-sugar and water containing solutions and powders. These are formulated and intended for use in hemodialysis when mixed or proportioned with the appropriate volume of purified water and bicarbonate concentrate / powder. These solutions and powders when proportioned/ mixed with pre-treated or purified water meeting or exceeding AAMI Standards, may be used in conventional and commercially available hemodialysis machines or monitors as a hemodialysis solution. The Dri-Sate™ Acid hemodialysis Mixes which have been presented in a previously cleared 510K Notification (K981003) are intended to be used in three stream hemodialysis machines in which an acidified concentrate is proportioned into one stream, a bicarbonate, chloride and sodium concentrate solution which has been prepared by mixing the powder with a specified volume of water and glacial acetic acid is
{1}------------------------------------------------
# Rockwell Medical Technologies Dri-Sate™ Mixer
proportioned into the second stream of the hemodialysis machine, and water is proportioned into the third stream. These three streams are then mixed to prepare a final proportioned hemodialysis solution. These types of a final hemodialysis solutions are commonly referred to as "Bicarbonate Hemodialysis Solutions." These proportioned hemodialysis solutions are then heated to body temperature and then perfused through the dialysis fluid compartment of artificial kidneys or hemodialyzers. These bicarbonate hemodialysis solutions are separated from the patient's blood by means of a semipermeable cellulosic or non-cellulosic membrane which serves as a molecular weight selective barrier to the passage of molecules beyond a certain molecular weight. The molecular weigh cut-off of each type of membrane may vary depending on the membrane type, manufacturing process, etc. The semi-permeable membrane in a hemodialyzer permits the passage of smaller molecular weight (less than 5,000 daltons for conventional cellulosic membranes), ionized and non-ionized molecules, waste products and toxins (i.e. blood urea nitrogen, creatinine, potassium, etc.) contained in the patient's blood circulating through the dialyzer, to pass through the semi-permeable membrane into the bicarbonate hemodialysis solutions, exit the hemodialyzer, enter the hemodialysis monitor and exit the monitor and are ultimately discarded. The ionic and molecular composition of the hemodialysis solution establishes the concentration gradient between the blood and the hemodialysis solution passing through the hemodialyzer which permits the effective removal of waste products and toxins found in excess in the patient's blood during acute and end-stage renal failure.
Since different patients have different requirements for the removal rates and quantities of various molecules and toxins (i.e. blood urea nitrogen, creatinine, potassium, phosphate, magnesium, chloride, sodium calcium water, etc.) in acute and chronic renal failure, it necessitates having a variety of different bicarbonate containing hemodialysis solutions to satisfy the needs of all acute and end-stage renal failure patients. In addition, there a number of different types of hemodialysis machines which have different proportioning rates.
## TABLE I
| Stream 1 | Stream 2 Acidified Concentrate Proportioning Ratios | Stream 3: Bicarbonate Concentrate Proportioning Ratios |
|----------|-----------------------------------------------------|--------------------------------------------------------|
| Water | 1:35.83 or 1:44.00 or 1:34.00 | 1:19.13 or 1:27.57 or 1:25.16 |
The Rockwell Medical Technologies Dri-Sate™ Acid Concentrate Mixes for hemodialysis have been previously cleared for marketing in the US under a 510(k) Notification on June 16th, 1998 (510(k) Number K981003).
{2}------------------------------------------------
#### Predicate Devices:
The Rockwell Medical Technologies Dri-Sate™ Mixer for Dri-Sate™ acid concentrate solutions is substantially equivalent to the Fresenius USA, Inc. Hemodialysis Concentration Dissolution Unit P/N 89-290-09 (510(k) Number K944493).
Examination of the information pertaining to the Rockwell Medical Technologies Dri-Sate™ Mixer demonstrates that this device is substantially equivalent in composition. intended use, packaging and labeling to other mixing devices for hemodialysis concentrate solutions currently approved for commercial distribution in the United States by the FDA. There are no significant differences between these marketed products and our proposed device.
### Intended Use:
The Rockwell Medical Technologies ' Dri-Sate™Mixer is designed to mix the Rockwell Medical Technologies Dri-Sate Acid Concentrate Mixes with purified water to produce an acid concentrate solution for hemodialysis for use in 3-stream (acid concentrate, bicarbonate concentrate, and water) hemodialysis machines / monitors. The Rockwell Medical Supply LLC. Dri-Sate Acid Concentrate Solutions which are mixed in the Dri-Sate Mixer™ are indicated for use in acute and chronic hemodialysis and to be used with the appropriate hemodialysis machine/ monitor and bicarbonate concentrate
This indication statement is essentially the same as the indication statement for the predicate device.
### Technological Characteristics:
Comparing the proposed device to the predicate devices utilize the same methods and technique for preparing hemodialysis concentrate solutions. There are no significant differences.
### Summary of Non-Clinical Tests:
In vitro testing was not performed was not included in this 510(k) Notification.
### Clinical Test Results:
Clinical testing was not performed
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure, represented by three overlapping profiles, suggesting a sense of community and care. The profiles are positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA", which is arranged in a circular fashion around the figure.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG - 5 1999
Jeffrey R. Shideman, Ph.D. Rockwell Medical Technologies, Inc. 28025 Oakland Oaks Court Wixom. MI 48393
Re: K983618 Dri-Sate™ Mixer for Preparation of Acidified Dialysate Dated: June 15, 1999 Received: June 21, 1999 Requlatory Class: Il 21 CFR §876.5820/Procode: 78 KPO
Dear Dr. Shideman:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If vour device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours.
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
4004
**Page** / of /
#### 510 (k) NUMBER (IF KNOWN): K983618
DEVICE NAME: Rockwell Medical Technologies, Inc. Dri-Sate Mixer for Preparation of Acidified Dialysate
#### INDICATIONS FOR USE:
#### Indications for Use Statement
The Rockwell Medical Technologies' Dri-Sate™Mixer is designed to mix the Rockwell Medical Technologies Dri-Sate Acid Concentrate Mixes with purified water to produce an acid concentrate solution for hemodialysis for use in 3-stream (acid concentrate, bicarbonate concentrate, and water) hemodialysis machines / monitors.
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter-Use (Optional Format 1-2-96)
David G. Hanson
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological De 510(k) Number
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.