K260564 · Terrats Medical SL · NHA · Jun 9, 2026 · Dental
Device Facts
Record ID
K260564
Device Name
DESS Dental Smart Solutions
Applicant
Terrats Medical SL
Product Code
NHA · Dental
Decision Date
Jun 9, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations. All digitally designed custom abutments for use with DESS Bases or Blanks are to be sent to a Terrats Medical validated milling center for manufacture. The PEEK Abutments are premanufactured prosthetic component directly connected to the endosseous dental implants and are intended for provisional use up to 180 days as an aid in prosthetic rehabilitation.
Device Story
DESS Dental Smart Solutions system provides prosthetic components for endosseous dental implants. Components include PEEK healing and temporary abutments, titanium alloy (Ti-6Al-4V) base abutments (Ti Base, AURUM, C-Base, ELLIPTIBase), multi-unit abutments, and pre-milled blanks. System supports CAD/CAM fabrication of zirconia superstructures at validated milling centers. Components connect to specific OEM implant systems (NobelActive, NobelReplace, NobelParallel, Nobel Brånemark, Keystone Paltop, Genesis Active). Clinicians use these components to support single or multi-unit restorations. Ti-based components feature SelectGrip surface or gold anodization to enhance bonding. Device is provided non-sterile; end-user performs steam sterilization. Output is a stable prosthetic foundation for dental crowns or bridges, facilitating functional and esthetic rehabilitation.
Clinical Evidence
No clinical data provided. Evidence consists of bench testing, including sterilization validation (ISO 17665), biocompatibility testing (ISO 10993-5), and mechanical fatigue testing of worst-case angled abutment/implant assemblies. MRI safety evaluated via scientific rationale and literature review.
Technological Characteristics
Materials: PEEK Classix (Invibio), Ti-6Al-4V ELI (ASTM F136), Zirconia (ISO 6872). Components: Healing/temporary abutments, Ti-base, AURUM, C-Base, ELLIPTIBase, multi-unit abutments, pre-milled blanks. Surface treatments: SelectGrip, gold anodization. Sterilization: Steam sterilization by end-user. Connectivity: Mechanical interface to specific OEM implant systems. CAD/CAM workflow for custom superstructures.
Indications for Use
Indicated for patients requiring prosthetic rehabilitation in the maxillary or mandibular arch using endosseous dental implants. PEEK abutments are indicated for provisional use up to 180 days.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{1}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
Terrats Medical SL
% Melissa Burbage
Principal Regulatory Consultant
Enerxen Consulting
1155 Metcalfe Street
Suite 1572
Montreal, Quebec H3B2V6
CANADA
June 9, 2026
Re: K260564
Trade/Device Name: DESS Dental Smart Solutions
Regulation Number: 21 CFR 872.3630
Regulation Name: Endosseous Dental Implant Abutment
Regulatory Class: Class II
Product Code: NHA
Dated: February 20, 2026
Received: May 6, 2026
Dear Melissa Burbage:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K260564 - Melissa Burbage
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled 'Deciding When to Submit a 510(k) for a Change to an Existing Device' (https://www.fda.gov/media/99812/download) and 'Deciding When to Submit a 510(k) for a Software Change to an Existing Device' (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ('UDI Rule'). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, 'Misbranding by reference to premarket notification' (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{3}
K260564 - Melissa Burbage
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
ANDREW I. STEEN -S
Andrew I. Steen
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (*if known*)
K260564
Device Name
# **DESS Dental Smart Solutions**
Indications for Use (*Describe*)
DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.
All digitally designed custom abutments for use with DESS Bases or Blanks are to be sent to a Terrats Medical validated milling center for manufacture.
The PEEK Abutments are premanufactured prosthetic component directly connected to the endosseous dental implants and are intended for provisional use up to 180 days as an aid in prosthetic rehabilitation.
Compatible Implant Systems
| Implant Compatibility (Connection) | Implant Body Diameter, mm | Implant Platform, mm |
| --- | --- | --- |
| NobelActive® NobelReplace/ NobelParallel Conical | 3.0 | 3.0 |
| | 3.5 | NP (3.5) |
| | 4.3, 5.0 | RP (3.9) |
| | 5.5 | WP (5.1) |
| NobelReplace® Trilobe | 3.5 | NP (3.5) |
| | 4.3 | RP (4.3) |
| | 5.0 | WP |
| Nobel Brånemark System® | 3.3 | NP |
| | 3.75, 4.0 | RP |
| | 5.0 | WP |
| Keystone Paltop Dynamic (Internal Hex) | 3.0, 3.25 | NP (3.25) |
| PALTOP ADVANCED CLASSIC (Internal Hex) | 3.25 | NP (3.25) |
| PALTOP ADVANCED PLUS (Internal Hex) | 3.0, 3.25 | NP (3.25) |
| Keystone Paltop Dynamic Conical (Internal conical) | 3.25, 3.75, 4.2, 5.0 | CC (3.25/3.75/4.2/5.0) |
| Genesis Active (Internal conical) | 3.5, 3.8, 4.5, 5.5 | CC (3.5/3.8/4.5/5.5) |
| PALTOP Implants (Internal conical) | 3.25 | NP (3.25) |
| Keystone Prima Plus Conical (Internal conical) | 3.5, 4.1, 5.0, 6.0 | CC (3.5, 4.1, 5.0, 6.0) |
Type of Use (*Select one or both, as applicable*)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
{5}
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*“An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.”*
---**FORM FDA 3881 (8/23)**
Page 1 of 2
PSC Publishing Services (301) 443-6740 EF
{6}
Traditional 510(k) Premarket Notification
510(k) Summary
DESS Dental Smart Solutions
K260564
# **510(k) Summary**
**Terrats Medical SL**
**DESS® Dental Smart Solutions**
**K260564**
**June 9, 2026**
ADMINISTRATIVE INFORMATION
| Manufacturer Name | Terrats Medical SL Carrer Mogoda 75-99 Barberà del Vallès 08210 Barcelona, Spain Telephone: +34 935 646 006 Fax: +34 935 647 317 |
| --- | --- |
| Official Contact | Roger Terrats, CEO |
| Representative/Consultant | Melissa Burbage Enerxen Consulting, Inc. 1155 Metcalfe Street, Suite 1572 Montreal, Quebec H3B 2V6 Telephone: +1 619-480-7733 Email: melissa.burbage@enerxen.com |
| | Sergio Muñoz, Terrats Medical SL Barberà del Vallès 08210 Barcelona, Spain Telephone +34 935 646 006 Email sergio.munoz@dessdental.com |
DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | DESS Dental Smart Solutions |
| --- | --- |
| Common Name | Dental implant abutment |
| Regulation Number | 21 CFR 872.3630 |
| Regulation Name | Endosseous dental implant abutment |
| Regulatory Class | Class II |
| Primary Product Code | NHA |
| Classification Panel | Dental Products Panel |
| Reviewing Division | DHT1B: Division of Dental and ENT Devices |
PREDICATE DEVICE INFORMATION
Primary Predicate Device
K170588, DESS Dental Smart Solutions, Terrats Medical SL
Reference Devices
K222269, DESS Dental Smart Solutions, Terrats Medical SL
Page 1 of 10
{7}
Traditional 510(k) Premarket Notification
510(k) Summary
DESS Dental Smart Solutions
K260564
K240208, DESS Dental Smart Solutions, Terrats Medical SL
K120954, Nobel Procera PEEK Abutments, Nobel Biocare AB
Reference Devices for OEM implant body clearances
K210117, Paltop Narrow Implant, Paltop Advanced Dental Solutions, Ltd.
K223814, Genesis Implant System, Keystone Dental, Inc.
K243983, Paltop Dental Implant System, Paltop Advance Dental Solutions, Ltd
K232740, Paltop Short Implants, Paltop Advanced Dental Solutions, Ltd
K220200, Paltop Concial Implant System, Paltop Advance Dental Solutions, Ltd
K240966, Prima Plus Conical Implant System, Keystone Dental, Inc.
K142260, NobelActive®, Nobel Biocare AB
K102436, NobelActive® 3.0, Nobel Biocare AB
K173418, NobelParallel™ Conical Connection, Nobel Biocare AB
K050705, TiUnite Implants®, Nobel Biocare AB
K050406, NOBELSPEEDY™ Implants, Nobel Biocare USA LLC
K022562, Various Brånemark System Implants–Immediate Function Indication, Nobel Biocare
INDICATIONS FOR USE STATEMENT
DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.
All digitally designed custom abutments for use with DESS Bases or Blanks are to be sent to a Terrats Medical validated milling center for manufacture.
The PEEK Abutments are premanufactured prosthetic component directly connected to the endosseous dental implants and are intended for provisional use up to 180 days as an aid in prosthetic rehabilitation
| Compatible Implant Systems | | |
| --- | --- | --- |
| Implant Compatibility (Connection) | Implant Body Diameter, mm | Implant Platform, mm |
| NobelActive® NobelReplace/ NobelParallel Conical | 3.0 | 3.0 |
| | 3.5 | NP (3.5) |
| | 4.3, 5.0 | RP (3.9) |
| | 5.5 | WP (5.1) |
| NobelReplace® Trilobe | 3.5 | NP (3.5) |
| | 4.3 | RP (4.3) |
| | 5.0 | WP |
| Nobel Brånemark System® | 3.3 | NP |
| | 3.75, 4.0 | RP |
| | 5.0 | WP |
| Keystone Paltop Dynamic (Internal Hex) | 3.0, 3.25 | NP (3.25) |
| PALTOP ADVANCED CLASSIC (Internal Hex) | 3.25 | NP (3.25) |
| PALTOP ADVANCED PLUS (Internal Hex) | 3.0, 3.25 | NP (3.25) |
| Keystone Paltop Dynamic Conical (Internal conical) | 3.25, 3.75, 4.2, 5.0 | CC (3.25/3.75/4.2/5.0) |
| Genesis Active (Internal conical) | 3.5, 3.8, 4.5, 5.5 | CC (3.5/3.8/4.5/5.5) |
| PALTOP Implants (Internal conical) | 3.25 | NP (3.25) |
| Keystone Prima Plus Conical (Internal conical) | 3.5, 4.1, 5.0, 6.0 | CC (3.5, 4.1, 5.0, 6.0) |
Page 2 of 10
{8}
Traditional 510(k) Premarket Notification
510(k) Summary
DESS Dental Smart Solutions
K260564
## SUBJECT DEVICE DESCRIPTION
The purpose of this submission is to add components to the DESS Dental Smart Solutions system, which includes abutments previously cleared under K170588, K222269, and K240208 to include a healing abutment design using PEEK for three existing Noble connections, a temporary abutment design using PEEK for existing DESS MultiUnite abutments, and include bases and Pre-Milled Blanck for Keystone connections.
| OEM Implant lines | Subject Device Components | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Peek healing | Peek Temporary | Ti base | Aurum | C-Base | Elliptical | Premilled | MUA | Screw |
| NobelActive® NobelReplace Conical NobelParallel Conical | X | | | | | | | | |
| NobelReplace ® Trilobe | X | | | | | | | | |
| Nobel Brånemark System® | X | | | | | | | | |
| Keystone Paltop (Internal Hex) | | | | | | X | | | X |
| Genesis Active (Internal conical) Keystone Paltop Dynamic Conical | | | X | X | X | X | X | X | X |
| Multi unite abutments | | X | | | | | | | |
## Subject Device Designs
**Healing Abutments** are provided in gingival heights from 3 mm to 5 mm to aid in contouring the gingiva during healing. Healing Abutments are made of PEEK.
**Temporary Abutments** are provided for multiple-unit restorations, the former with engaging connections to the implants and the latter with non-engaging connections. All Temporary Abutments have a gingival height from 1.5 mm and are made of PEEK, and with a prosthetic platform diameter of 4.5 mm to 6.0 mm.
**Ti Base Abutments** are provided in engaging and non-engaging designs for custom abutment fabrication of a CAD-CAM zirconia superstructure on which a crown may be placed. They also may support a ceramic hybrid abutment in which the crown is included in the design of the zirconia abutment ceramic superstructure.. The two-pieces of the Ti Base Abutment which compose the final abutment consists of the pre-manufactured titanium base component composed of titanium alloy and the CAD-CAM patient matched superstructure or hybrid abutment crown composed of zirconia. The cement required for bonding of superstructures is Multi-Link cement by Ivoclar Vivadent (K130436). Ti Base Abutments are manufactured from titanium alloy (Ti-6Al-4V) with the SelectGrip® surface.
Ti Base Abutments also may support a ceramic hybrid abutment in which the crown is included in the design of the zirconia abutment ceramic superstructure. When used for a direct crown, Ti Base abutments may be used with POM burn out sleeve, an exempt laboratory component not a subject of this submission, that is available for laboratory fabrication of the prosthesis. Ti Bases are not intended for angulation correction when used with a POM burn-out sleeve.
**AURUM Base Abutments** are provided in engaging and non-engaging designs. The two-pieces of the AURUM Base Abutment which compose the final abutment consists of the pre-manufactured titanium base component composed of titanium alloy and the CAD-CAM patient matched superstructure
Page 3 of 10
{9}
Traditional 510(k) Premarket Notification
510(k) Summary
DESS Dental Smart Solutions
K260564
composed of zirconia. The design of the AURUM Base Abutment allows for easier instrument access to the abutment screw and allows for placement of the screw channel out of the esthetic region of the restoration. AURUM Base Abutments are provided with a prosthetic platform diameter of 5.5 mm, with gingival heights (in the base) of 1 mm.
AURUM Base Abutments are manufactured from titanium alloy (Ti-6Al-4V) and anodized a gold color, and the SelectGrip® surface to aid in bonding retention. The design of the AURUM Base Abutments, including the titanium alloy, anodization treatment, and SelectGrip® surface is similar to that of DESS AURUM Base Abutments cleared in K240208.
**C-Base Abutments** are provided in engaging and non-engaging designs. C-Base Abutments are two-piece abutments designed to support a custom CAD-CAM zirconia superstructure on which a single-unit or multi-unit restoration may be placed. The ceramic superstructure produced through CAD-CAM is the second part of the two-piece abutment. The C-Base Abutment also may support a ceramic hybrid abutment in which the crown is included in the design of the zirconia abutment ceramic superstructure.
The C-Base Abutment prosthetic post is 4.68 mm, and the gingival height is 1 mm to 3 mm. All patient-specific custom abutment fabrication is by prescription on the order of the clinician. C-Base Abutments are made of titanium alloy (Ti-6Al-4V) with anodization and a SelectGrip® surface.
C-Base Abutments also may support a ceramic hybrid abutment in which the crown is included in the design of the zirconia abutment ceramic superstructure. When used for a direct crown, C-Base abutments may be used with POMburn out sleeve an exempt laboratory component not a subject of this submission, that is available for laboratory fabrication of the prosthesis. C-Bases are not intended for angulation correction when used with a POM burn-out sleeve.
**ELLIPTIBase Abutments** are available in an engaging design. ELLIPTIBase Abutments are two-piece abutments designed to support a custom CAD-CAM zirconia superstructure on which a single-unit or multi-unit restoration may be placed. The ceramic superstructure produced through CAD-CAM is the second part of the two-piece abutment. The ELLIPTIBase Abutment also may support a ceramic hybrid abutment in which the crown is included in the design of the zirconia abutment ceramic superstructure.
Before attachment of the zirconia superstructure or crown, the ELLIPTIBase post height is 3.0 mm. When used for a single-unit restoration the ELLIPTIBase is to be used with a superstructure to create a minimum post height of 4.0 mm. ELLIPTIBase is made of titanium alloy (Ti-6Al-4V ELI) with a gold anodized surface
The design parameters for the CAD/CAM zirconia superstructures are summarized in Table 1 and Figure 1 below.
Page 4 of 10
{10}
Traditional 510(k) Premarket Notification
510(k) Summary
DESS Dental Smart Solutions
K260564
Table 1 Design Parameters for all DESS Base Abutments
| Abutment Base | | | Zirconia Superstructure | | | | Final Abutment | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Type | Prosthetic Platform Ø (Abutment base Ø), mm | Gingival Height^{1}, mm | Minimum wall thickness, mm | Dimension B Minimum Abutment Post Height^{2} for single-unit restorations, mm | Dimension D Gingival height^{3}, Minimum/maximum, mm | Dimension F Maximum Angulation | Dimension A Total height, Minimum/maximum, mm | Dimension E Width, Minimum/maximum, mm |
| Ti base | 5.5 | 1-2 | 0.4 | 4.2 | 0.5 / 6.0 | 30° | 5.7 / 20.2 | Based on minimum wall thickness / 14 |
| | 5.5 | 3 | 0.4 | 4.2 | 0.5 / 6.0 | 0° | 7.7 / 20.2 | Based on minimum wall thickness / 14 |
| Aurum Base | 5.5 | 1 | 0.4 | 4.0 | 0.5 / 6.0 | 30° | 5.5 / 18 | Based on minimum wall thickness / 14 |
| C-Base | 4.3-5.5 | 1-2 | 0.4 | 4.7 | 0.5 / 6.0 | 30° | 6.2 / 17 | Based on minimum wall thickness / 14 |
| | 4.3-5.5 | 3 | 0.4 | 4.7 | 0.5 / 6.0 | 0° | 8.2 / 17 | Based on minimum wall thickness / 14 |
| ELLIPTIBase (internal conical) | 4.1 | 1 | 0.4 | 4.0 | 0.5/6.0 | 30° | 5.5 / 18 | Based on minimum wall thickness / 14 |
| ELLIPTIBase (internal hex) | 4.1 | 1 | 0.4 | 4.0 | 0.5/6.0 | 0° | 5.5 / 18 | Based on minimum wall thickness / 14 |
$^{1}$ Gingival collar height in the abutment base; $^{2}$ Abutment post height is the portion above the gingival height of the final patient-matched design; $^{3}$ Gingival height in the zirconia superstructure

Figure 1 Design Parameters
All patient-specific custom abutment fabrication for Ti Base Abutments, AURUM Base Abutments, and C-Base Abutments (described below) is by prescription on the order of the clinician. All zirconia superstructures for use with the subject device Ti Base, AURUM Base, and C-Base Abutments will be made at a Terrats Medical validated milling center under FDA quality system regulations, and the zirconia material will conform to ISO 13356.
Multi-Unit Abutments: Straight and Angled are designed for attachment of multi-unit screw-retained restorations and are provided in three (3) designs, straight, angled 17°, and angled 30°. The designs of the
Page 5 of 10
{11}
Traditional 510(k) Premarket Notification
510(k) Summary
DESS Dental Smart Solutions
K260564
subject Multi-Unit Abutments are similar to the designs of Multi-Unit Abutments cleared in K240208 and K222288. All Multi-Unit Abutments are manufactured from titanium alloy (Ti-6Al-4V).
The Straight Multi-Unit Abutments have a non-engaging, threaded design that attaches directly to the implant. Straight Multi-Unit Abutments are provided with a prosthetic platform diameter of 4.8 mm, and with a gingival height of 3 mm and 4 mm.
The Angled Multi-Unit Abutments are provided only in an engaging design that requires an abutment screw. The Multi-Unit Abutments angled 17° are provided with a prosthetic platform diameter of 4.8 mm, and with a gingival height of 4 mm. The Multi-Unit Abutments angled 30° are provided with a prosthetic platform diameter of 4.8 mm, and with a gingival height of 4 mm and 5 mm.
**Pre-Milled Blank Abutments** are designed for custom abutment fabrication by a CAD-CAM process. All patientspecific custom abutment fabrication is by prescription on the order of the clinician. The Pre-Milled Blank Abutments have a maximum (before milling) diameter of 10 mm or 14 mm, and are provided in a solid cylindrical design and with a pre-milled screw-channel. The Pre-Milled Blank Abutments are manufactured from titanium alloy (Ti 6Al-4V).
The design parameters for the Pre-Milled Blank Abutments are:
Minimum wall thickness – 0.45 mm
Minimum abutment post height (length above the abutment collar/gingival height) – 4.0 mm
Minimum gingival height – 0.5 mm
Maximum gingival height – 6.0 mm
Maximum Angulation – 0° - intended for straight abutments only
**DESS Dental Smart Solutions Screws** are designed to attach the abutment to the implant or the prosthesis to the abutment. There are a total of eight (8) subject device screws compatible with the subject device components or previously cleared components. The new screws have designs that are similar to those of screws cleared in K170588, K222269, and K240208. Screws are made of titanium alloy (Ti-6Al-4V).
## MATERIAL COMPOSITION
The subject device healing and temporary abutments are manufactured from PEEK Classix from Invibio Ltd. This is the same material that is used in predicate K120954.
The metallic components of the subject device Ti base Abutments, C-Base Abutments, ELLIPTI Base, AURUM Base, Multi-Unit Abutments, Pre-Milled Blanks, and screws are manufactured from Ti-6Al-4V alloy conforming to ASTM F136 *Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401)*. There have been no changes to the material as cleared in K170588, K240208, and K222269.
The material required for zirconia superstructures on Ti Base, DESS Aurum Base and ELLIPTIBase, and C-Base is VITA YZ ST and VITA YZ XT, conforming to ISO 6872 *Dentistry – Ceramic Materials* and cleared in K180703. The cement required in labeling for bonding of superstructures is Multilink Hybrid Abutment Cement from Ivoclar Vivadent AG, cleared under K130436. There have been no changes to the material as cleared in K170588, K240208, and K222269.
Page 6 of 10
{12}
Traditional 510(k) Premarket Notification
510(k) Summary
DESS Dental Smart Solutions
K260564
# PERFORMANCE DATA
Non-clinical testing data submitted to demonstrate substantial equivalence included:
- Leveraged sterilization validation according to ISO 17665-1 *Sterilization of health care products – Moist heat – Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices* and ISO 17665-2 *Sterilization of health care products — Moist heat — Part 2: Guidance on the application of ANSI/AAMI/ISO 17665-1* for the subject device composed of titanium and zirconia. Conducted validation for the PEEK subject device abutments.
- Biocompatibility testing according to ISO 10993-5 *Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity* for PEEK subject device abutment.
- Leveraged biocompatibility testing according to ISO 10993-5 *Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity* for subject devices composed of titanium and zirconia.
- Non-clinical worst-case MRI review to evaluate the metallic devices in the MRI environment using scientific rationale and published literature (e.g., Woods, Terry O., Jana G. Delfino, and Sunder Rajan. "Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices." *Journal of Testing and Evaluation* 49.2 (2019): 783-795), based on the subject device components and material composition. Rationale addressed parameters per the FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment," including magnetically induced displacement force and torque.
- Leveraged reverse engineering analysis and contractual agreement with OEM manufacture from K170588, K222269, and K240208 of OEM implant bodies, OEM abutments, and OEM abutment screws to confirm compatibility for OEM connections.
- Fatigue testing of OEM implant bodies with patient specific abutments made at worst case angled conditions.
Page 7 of 10
{13}
Traditional 510(k) Premarket Notification
510(k) Summary
DESS Dental Smart Solutions
K260564
# EQUIVALENCE TO MARKETED DEVICES
**Table 2 Table of Substantial Equivalence**
| Comparison | Subject Device | Predicate Devices | Additional Predicate |
| --- | --- | --- | --- |
| | DESS Dental Smart Solutions Terrats Medical SL | K170588, K222269, K240208 DESS Dental Smart Solutions Terrats Medical SL | K120954 Nobel Procera PEEK Abutments |
| Product Code | NHA | NHA | NHA |
| Reason for predicate/reference | n/a | Indications, Abutment design, OEM Connections | PEEK Material |
| Intended Use | Functional and esthetic rehabilitation of the edentulous mandible or maxilla | Functional and esthetic rehabilitation of the edentulous mandible or maxilla | Functional and esthetic rehabilitation of the edentulous mandible or maxilla |
| Indications | DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations. All digitally designed custom abutments for use with DESS Bases or Blanks are to be sent to a Terrats Medical validated milling center for manufacture. The PEEK Abutments are premanufactured prosthetic component directly connected to the endosseous dental implants and are intended for provisional use up to 180 days as an aid in prosthetic rehabilitation | DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations. All digitally designed custom abutments for use with DESS Bases or Blanks are to be sent to a Terrats Medical validated milling center for manufacture. | The Nobel Biocare PEEK Abutments are premanufactured prosthetic components directly connected to endosseous dental implants and are intended for provisional use up to 180 days as an aid in prosthetic rehabilitation. |
| **Design Temporary, Healing** | | K170588 | |
| Abutment Diameter, mm | 4.5 – 6.0 | 3.4 – 5.7 | --- |
| Gingival Height, mm | 1.5 | 0 – 6 | --- |
| Abutment Angulation | Straight | Straight | Straight |
| Abutment/Implant Interface | Internal Thread | Internal Thread | Internal Thread |
| OEM | Nobel Active, Nobel Replace, Nobel Branemark, DESS | Nobel Active, Nobel Replace, Nobel Branemark, DESS | --- |
| Material | PEEK | Ti 6Al-4V ELI | PEEK |
| Duration of PEEK use | 180 days | N/A | 180 days |
| **Design Base Abutment** | | K240208 | |
| Abutment Diameter, mm | 4.1 – 5.5 | 3.85 – 7.0 | --- |
| Gingival Height (of superstructure), mm | 0.5 – 6 | 0.5 – 6 | --- |
| Abutment Angulation | 0° – 30° | 0° – 30° | --- |
Page 8 of 10
{14}
Traditional 510(k) Premarket Notification
510(k) Summary
DESS Dental Smart Solutions
K260564
| Comparison | Subject Device | Predicate Devices | Additional Predicate |
| --- | --- | --- | --- |
| | DESS Dental Smart Solutions Terrats Medical SL | K170588, K222269, K240208 DESS Dental Smart Solutions Terrats Medical SL | K120954 Nobel Procera PEEK Abutments |
| Abutment/Implant Interface | Internal Thread | Internal Thread | --- |
| OEM Connection | Keystone Internal Conical, Keystone Internal Hex | Keystone Internal Conical, Keystone Internal Hex | --- |
| Material | Zirconia and Ti 6Al-4V ELI | Zirconia and Ti 6Al-4V ELI | --- |
| Surface Treatment | SelectGrip or Anodization | SelectGrip or Anodization | --- |
| **Design Premilled Blank** | | K222269 | --- |
| Abutment Diameter, mm | 4.1 – 5.5 | 2.9 – 5.7 | --- |
| Gingival Height, mm | 0.5 – 6 | 0.5 – 6 | --- |
| Abutment Angulation | Straight | Straight | --- |
| Abutment/Implant Interface | Internal Thread | Internal Thread | --- |
| OEM Connection | Keystone Internal Conical | Keystone Internal Conical | --- |
| Material | Ti 6Al-4V ELI | Ti 6Al-4V ELI | --- |
| Surface Treatment | None | None | --- |
| **Design Multi Unit Abutment** | | K222269 | --- |
| Abutment Diameter, mm | 4.8 | 2.9 – 5.7 | --- |
| Gingival Height, mm | 1 – 5 | 0.5 – 6 | --- |
| Abutment Angulation | 0°, 17°, 30° | 0°, 17°, 30° | --- |
| Abutment/Implant Interface | Internal Thread | Internal Thread | --- |
| OEM Connection | Keystone Internal Conical | Keystone Internal Conical | --- |
| Material | Ti 6Al-4V ELI | Ti 6Al-4V ELI | --- |
| Surface Treatment | Anodization | None | --- |
| **Reprocessing** | | | |
| Abutment | End user steam sterilization | End user steam sterilization | End user steam sterilization |
The indications for use of the subject device healing and temporary abutments are identical to the predicate device K120954. Both are premanufactured components that are intended to directly connect to the dental implant for provisional use up to 180 days to aid in prosthetic rehabilitation.
The indications for use of the subject device Ti Base (all types) and Pre-Milled Blank abutments are identical to the predicate device K170588, K222269, and K240208. Both are intended for use with endosseous dental implants to provide functional and esthetic rehabilitation of the edentulous maxilla and mandible. Fabrication of the digitally design abutments are to occur at a DESS validated milling center.
The indications for use of the subject device Multi-Unit abutments are identical to the predicate device K170588, K222269, and K240208. Both are intended for use with endosseous dental implants to provide functional and esthetic rehabilitation of the edentulous maxilla and mandible.
Page 9 of 10
{15}
Traditional 510(k) Premarket Notification
510(k) Summary
DESS Dental Smart Solutions
K260564
All subject device abutments are similar or identical in design, materials and technological characteristics to corresponding abutments of the primary predicate device K170588 and the additional predicate devices K222269, K240208, and K120954.
The design of the Temporary and Healing abutment is the same as predicate devices K170588 and K222269. The PEEK material is the same as additional predicate device K120954.
The design of the Base Abutments (Ti Base, AURUM, C-Base, ELLIPTIBase models) is the same as predicate devices K170588 (abutment design) and K240208 (angulation and OEM connection).
The SelectGrip® surface on the subject device Ti Base Abutments is identical to the Select Grip® surface on equivalent abutments cleared in K170588, K222269, and K240208. The gold anodized surface on the subject device AURUM Base Abutments is identical to the anodized surface on Aurum Abutments of the predicate devices K170588, K222269, and K240208.
The cement required in labeling for bonding of superstructures is Multi-Link cement from Ivoclar Vivadent, cleared under K130436. This is the same cement required in labeling for the predicate devices K170588, K222269, and K240208.
All screws are identical in design, materials, and technological characteristics to those cleared in predicate devices K170588, K222269, and K240208, except for threads and lengths that accommodate the new compatibilities. The Diamond-like carbon (DLC) coating and Anodized applied to all screws is identical to anodizing or DLC surface on screws cleared in predicate device K252384, K240208 and K222269.
The design, including angulation, of the Multi-Unit Abutment is the same as the predicate K222269 with the exception of gold anodization. The gold anodized surface on the subject device Multi-Unit Abutment is identical to the anodized surface on Aurum Abutments of the predicate devices K170588 and K222269.
The risks associated with use of angulated abutments in combination with the compatible implants are mitigated by the mechanical testing.
The subject device is provided non-sterile to the end user for the end user to sterilize prior to use, the same as the primary predicate. The sterilization validation was conducted for the PEEK abutment.
## CONCLUSION
The subject device, the primary predicate device, and the additional predicate devices have the same intended use, have similar technological characteristics, and are made of the same materials. The subject device, the primary predicate, and additional predicate devices encompass the same range of physical dimensions, are packaged in similar materials, and are to be sterilized using similar methods. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
Page 10 of 10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.